K960755 · Medtronic Cardiac Surgery, Medtronic, Inc. · DQX · May 23, 1996 · Cardiovascular
Device Facts
Record ID
K960755
Device Name
CARDIOTHERM BLOOD CARDIOPLEGIA SYSTEMS
Applicant
Medtronic Cardiac Surgery, Medtronic, Inc.
Product Code
DQX · Cardiovascular
Decision Date
May 23, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medtronic CardioTherm™ Blood Cardioplegia System is intended for mixing, cooling, warming, and delivery of oxygenated blood and/or asanguineous cardioplegia solution.
Device Story
Single-use disposable device; integrates into extracorporeal circuit. Inputs: arterial blood from oxygenator; asanguineous cardioplegia solution. Mechanism: polycarbonate housing with nonporous polypropylene hollow fibers; blood/cardioplegia flows through fibers; temperature-regulated water flows around fibers for heat exchange. Output: temperature-controlled blood/cardioplegia mixture delivered to patient. Clinician-operated; used in surgical settings. Tubing sets with specific inner diameters determine blood-to-cardioplegia mixing ratios (e.g., 1:1, 2:1, 4:1, 9:1). Device includes bubble chamber, vent port, temperature probe port, and pressure monitoring line. Facilitates myocardial protection during cardiac surgery by providing precise temperature and composition control of cardioplegic solution.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, design analysis, and comparison of technological characteristics and nominal specifications against legally marketed predicate devices.
Technological Characteristics
Heat exchanger with polycarbonate housing; nonporous polypropylene hollow fibers; polyethylene screen support. Tubing sets: polyvinyl chloride (PVC) with polycarbonate/PVC connectors. Dimensions: tubing inner diameters 3/32" to 17/64". Connectivity: standalone extracorporeal circuit component. Sterilization: not specified. Sensing: temperature probe port, pressure monitoring line. Energy: external water-based cooling/warming source.
Indications for Use
Indicated for patients undergoing procedures requiring extracorporeal circulation where mixing, cooling, warming, and delivery of oxygenated blood and/or asanguineous cardioplegia solution is required.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K896807/ K944219)
Avecor Cardiovascular - MYOtherm™ Cardioplegia Delivery Systems (K904171)
Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)
Medtronic InterSept™ Crystalloid Cardioplegia Delivery System (K885154)
Submission Summary (Full Text)
{0}
MAY 23 1996
K960755
# SUMMARY OF SAFETY AND EFFECTIVENESS
## COMPANY AND CONTACT PERSON
Medtronic, Inc.
Cardiopulmonary Division
4633 E. La Palma Avenue
Anaheim, CA 92807
714-779-3700
Debra J. Kridner, Manager
Regulatory Affairs
## DEVICE NAME
CardioTherm™ Blood Cardioplegia Delivery Systems
(CTXXX Series and CTXXXBR Series)
## NAME OF PREDICATE OR LEGALLY MARKETED DEVICE
- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K896807/ K944219)
- Avecor Cardiovascular - MYOtherm™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)
## DESCRIPTION OF DEVICE
Each Medtronic CardioTherm™ Blood Cardioplegia System is a single use, disposable device designed to mix arterial blood from an oxygenator with asanguineous cardioplegia solution in specific ratios depending on the tubing set configurations. The blood/cardioplegia solution is then cooled/warmed and delivered to the patient.
The Medtronic CardioTherm™ Cardioplegia Heat Exchanger consists of a polycarbonate housing which incorporates a bubble chamber. A stopcock is attached to the upper luer port of the housing lid allowing for air venting through a purge line during priming and pressure monitoring through a vented or nonvented pressure monitoring line throughout the procedure. A temperature probe port is located adjacent to the blood outlet port at the bottom of the device, allowing for temperature monitoring of the blood/cardioplegia solution prior to patient delivery.
The heat exchanger housing contains nonporous polypropylene hollow fibers, which are supported on their exterior by a polyethylene screen. These polypropylene fibers run longitudinally through the polycarbonate chamber. Blood/cardioplegia solution flows through the polypropylene hollow
510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 144
{1}
fibers. The walls of these fibers provide a barrier between the blood/cardioplegia solution and the cooling/warming water.
The flow of solution through the Medtronic CardioTherm™ Blood Cardioplegia Heat Exchanger is as follows:
## Blood/Cardioplegia
The blood and/or cardioplegia solution enters the bottom of the device, flows up through the polypropylene hollow fibers, flows back down a center collecting tube, and exits the from the bottom of the device.
## Water
The cooling/warming water enters a side port, passes around the outside walls of the polypropylene fibers, then exits through a parallel side port.
Heat exchange occurs as the blood/cardioplegia solution passes through the hollow fibers, while the temperature regulated cooling/warming water passes around the outside walls of the same fibers. The blood/cardioplegia solution is cooled/ warmed as it flows through the heat exchanger.
By connecting tubing of specific inner diameters to the heat exchanger, a predetermined approximate ratio of blood to cardioplegia solution may be delivered to the patient. The tubing sets are manufactured with polyvinyl chloride tubing and polycarbonate/polyvinyl chloride connectors which are currently used as components in the commercially distributed Medtronic Extracorporeal InterSept™ Custom Tubing Packs (K800178) and Medtronic InterSept™ Crystalloid Cardioplegia Delivery System (K885154). Four of the various delivery ratios available are:
1. a tubing set for 1:1 ratio of blood to asanguineous cardioplegia.
2. a tubing set for 2:1 ratio of blood to asanguineous cardioplegia.
3. a tubing set for 4:1 ratio of blood to asanguineous cardioplegia.
4. a tubing set for 9:1 ratio of blood to asanguineous cardioplegia.
These tubing ratios are common to commercially available blood cardioplegia systems; including the Medtronic Electromedics, Inc. Blood Cardioplegia Delivery Systems, the Gish Biomedical, Inc. Single Pass Blood Cardioplegia (SPBC) Systems, the Avecor Cardiovascular MYOtherm™ Cardioplegia Delivery Systems and the Sorin Biomedical BCD ADVANCED Blood Cardioplegia System.
Tubing inner diameters for commercially available tubing sets generally range from 3/32" (2.4 mm) to 17/64" (6.7mm).
510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 145
{2}
# STATEMENT OF INTENDED USE
The Medtronic CardioTherm™ Blood Cardioplegia System is intended for mixing, cooling, warming, and delivery of oxygenated blood and/or asanguineous cardioplegia solution.
# STATEMENT OF INTENDED USE OF PREDICATE/MARKETED DEVICES
The Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems are intended to mix, cool, and deliver oxygenated blood/cardioplegia solution in a predetermined ratio.
The Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems are intended to mix, cool/warm, and deliver oxygenated blood/cardioplegia solution.
The Avecor Cardiovascular - MYOtherm™ Cardioplegia Delivery Systems are intended to mix, cool, and deliver oxygenated blood/cardioplegia solution in a predetermined ratio.
The Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System are intended to mix, cool/warm, and deliver oxygenated blood/cardioplegia solution.
# STATEMENT OF TECHNOLOGICAL CHARACTERISTICS COMPARISON
A table comparing the intended use and technological characteristics of the Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems with the four noted substantially equivalent devices is provided in Attachment I.
# DETERMINATION OF SUBSTANTIAL EQUIVALENCE
This premarket notification submission provides substantial equivalence information and rationale which addresses the introduction to commercial distribution of the Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia System.
The Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems are substantially equivalent to other Blood Cardioplegia Systems currently in commercial distribution. These predicate/marketed devices include:
- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K896807/K944219)
- Avecor Cardiovascular - MYOtherm™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)
The Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems have an intended use which is substantially equivalent to other Blood Cardioplegia Systems currently in commercial distribution. These predicate/marketed devices include:
510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 146
{3}
- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K896807/K944219)
- Avecor Cardiovascular - MYOther™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)
The Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems have technological characteristics which are substantially equivalent to other Blood Cardioplegia Systems currently in commercial distribution. These predicate/marketed devices include:
- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K89680/K944219)
- Avecor Cardiovascular - MYOther™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)
The design, construction, materials and nominal specifications of the Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems are either identical or substantially equivalent to other Blood Cardioplegia Systems currently in commercial distribution. These predicate/marketed devices include:
- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K896807/K944219)
- Avecor Cardiovascular - MYOther™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)
510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 147
{4}
DEVICE COMPARISONS -- GENERAL CHARACTERISTICS AND NOMINAL SPECIFICATIONS
| | Medtronic, Inc.
CardioTherm™
Blood Cardioplegia System | Medtronic, Inc./Electromedics Inc.
Blood Cardioplegia.
Delivery Systems | Gish Biomedical, Inc.
SPBC
Blood Cardioplegia Systems | Avecor Cardiovascular
MYOtherm™
Cardioplegia Delivery Systems | Sorin Biomedical
BCD ADVANCED
Blood Cardioplegia System |
| --- | --- | --- | --- | --- | --- |
| 510(k) Number: | This submission | K934355 | K896807 | K904171 | K925369 |
| Catalog Number(s): | CT-XXX Series
CT-XXXBR Series | D1081A, D1080A, D1082A | SPBC-XX Series
SPBC-XXBR Series | MYOtherm™ XX Series
MYOtherm™ XX BR Series | BCD ADV X Series
(includes bridge/shunt system) |
| Intended Use: | To mix, cool, warm and deliver
oxygenated blood/cardioplegia
solution. | To mix, cool, and deliver
oxygenated blood/cardioplegia
solution in a predetermined ratio. | To mix, cool/warm, and deliver
oxygenated blood/cardioplegia
solution. | To mix, cool, and deliver
oxygenated blood/cardioplegia
solution in a predetermined ratio. | To mix, cool/warm, and deliver
oxygenated blood/cardioplegia
solution. |
| Performance Characteristics: | | | | | |
| Heat Exchanger | | | | | |
| Priming Volume (mL) | 46 | 55 | 83 | 68 | 55 |
| Maximum Flow Rate (LPM) | 1 | 3 | not rated | not rated | .5 |
| Fluid Path Pressure (maximum) | 500 mmHg | not rated | 300 mmHg | 500 mmHg | not rated |
| Water Path Pressure (maximum) | 45 psi | 45 psi | 65psi | 40 psi | 80 psi |
| Technological Characteristics: | | | | | |
| Configuration | Heat Exchanger | Heat Exchanger | Heat Exchanger | Heat Exchanger | Heat Exchanger |
| Heat Exchanger | | | | | |
| Vent Port | Yes | Yes | Yes | Yes | Yes |
| Bubble Chamber | Yes | Yes | Yes | Yes | Yes |
| Temperature Well | Yes | Yes | Yes | Yes | Yes |
| Pressure Monitoring Line | Yes | Yes | Yes | Yes | Yes |
| Patient Delivery Line | Yes | Yes | Yes | Yes | Yes |
| | Various Tubing Sets | Various Tubing Sets | Various Tubing Sets | Various Tubing Sets | Various Tubing Sets |
510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 148
{5}
DEVICE COMPARISONS -- GENERAL CHARACTERISTICS AND NOMINAL SPECIFICATIONS
| | Medtronic, Inc.
CardioTherm™
Blood Cardioplegia System | Medtronic, Inc./Electromedics Inc.
Blood Cardioplegia.
Delivery Systems | Gish Biomedical, Inc.
SPBC
Blood Cardioplegia Systems | Avecor Cardiovascular
MYOTherm™
Cardioplegia Delivery Systems | Sorin Biomedical
BCD ADVANCED
Blood Cardioplegia System |
| --- | --- | --- | --- | --- | --- |
| Materials | | | | | |
| Heat Exchanger
Housing
Heat Exchanger | Polycarbonate
Polypropylene | Polycarbonate
Stainless steel | Polycarbonate
Aluminum | Polycarbonate
Stainless steel | Polycarbonate
Aluminum |
| Tubing Sets | Polyvinyl chloride
Polycarbonate | Polyvinyl chloride
Polycarbonate | Polyvinyl chloride
Polycarbonate | Polyvinyl chloride
Polycarbonate | Polyvinyl chloride
Polycarbonate |
| Specifications | | | | | |
| Tubing Set Ratios | Separate sets
(1:1, 2:1, 4:1 and 9:1) | Separate sets
(1:1, 2:1 and 4:1) | Separate sets
(1:1, 2:1, 4:1 and 9:1) | Separate sets
(1:1, 2:1, 4:1 and 9:1) | Separate sets
(1:1, 2:1, 4:1 and 8:1) |
| Tubing Inner Diameter | 1/12 inch to 1/4 inch | 1/8 inch to 1/4 inch | 3/32 inch to 17/64 inch | 1/8 inch to 1/4 inch | 3/32 inch to 17/64 inch |
| Ratio Capability | Variable (w/bridge clamps)
(all blood, 1:1, 2:1, 4:1, 9:1
and all crystalloid with
various tubing sets) | Variable
(1:1, 2:1 and 4:1) | Variable (w/bridge clamps)
(all blood, 1:1, 2:1, 4:1, 9:1,
and all crystalloid with
various tubing sets) | Variable (w/bridge clamps)
(all blood, 1:1, 2:1, 4:1, 9:1,
and all crystalloid with
various tubing sets) | Variable (w/bridge clamps)
(all blood, 1:1, 2:1, 4:1, 8:1,
and all crystalloid with
various tubing sets) |
| Mode of Operation | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need. | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need. | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need. | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need. | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need. |
10(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 149
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.