# SUNMAX ENTERPRISE SHANGHAI VINYL POWDERED PATIENT EXAM. GLOVES (K960746)

_Sunmax Enterprise Shanghai Co. , Ltd. · LYZ · Jun 27, 1996 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K960746

## Device Facts

- **Applicant:** Sunmax Enterprise Shanghai Co. , Ltd.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Jun 27, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

## Device Story

Vinyl patient examination gloves; powdered with absorbable dusting powder (USP). Worn on hands by healthcare personnel; acts as protective barrier to prevent cross-contamination between wearer and patient/environment. Device used in clinical settings; provides physical barrier against fluids and waste.

## Clinical Evidence

No clinical data. Bench testing only: physical dimensions, water leak test (1000 ml, AQL 2.5), and biocompatibility (primary skin irritation and sensitization) performed per ASTM D5250-92.

## Technological Characteristics

Vinyl material; powdered with absorbable dusting powder (USP). Dimensions and physical properties conform to ASTM D5250-92. Biocompatibility verified via skin irritation/sensitization testing. FDA 1000 ml water fill test (AQL 2.5, Inspection Level S-4).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Honyee Vinyl Patient Examination Gloves ([K893822](/device/K893822.md))
- Cheer & Merit Vinyl Patient Examination Gloves ([K941809](/device/K941809.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

ATTACHMENT A
Page 1 of 2
JUN 27 1996
K960746

# 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: K960746

1. **Submitter's Identification:**
Mr. Ernie Liu
Sunmax Enterprise Shanghai Co., Ltd.
#2 Zhu Hang New Industry Zone Jinshan
County, Shanghai, China

**Date Summary Prepared:** February 15, 1996

2. **Name of the Device:**
Sunmax Enterprise Shanghai Co., Ltd. Vinyl Patient Examination Gloves - Powdered

3. **Predicate Device Information:**
Honyee Vinyl Patient Examination Gloves, K#893822
Cheer &amp; Merit Vinyl Patient Examination Gloves, K#941809

4. **Device Description:**
Classified by FDA's General and Plastic Surgery Device Panel as Class I, 21 CFR 880.6250, Vinyl Patient Examination Glove, 80LYZ, Powdered with Absorbable Dusting Powder, USP, Class III and meets all requirements of ASTM Standard D5250-92.

5. **Intended Use:**
A glove is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

6. **Comparison to Predicate Devices:**
Sunmax Enterprise Shanghai Co., Ltd. Vinyl Patient Examination Gloves-Powdered, is substantially equivalent in safety and effectiveness to the Honyee Vinyl Patient Examination Glove and the Cheer &amp; Merit Patient Examination Glove.

{1}

ATTACHMENT A
Page 2 of 2

- 2 -

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:

The standards used for Sunmax Enterprise Shanghai Co., Ltd. glove production are based on ASTM-D-5250-92. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, Inspection Level S-4, meeting these requirements. Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.

Sunmax Enterprise Shanghai Co., Ltd. operates in compliance with FDA's GMPs.

8. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic claim.

9. Conclusions:

Sunmax Enterprise Shanghai Co., Ltd. Vinyl Patient Examination Gloves conform fully to ASTM-D-5250-92 standards as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

[Signature]

ERNE LIU

FEB. 16, 1996

DATE:

---

**Source:** [https://fda-staging.innolitics.com/device/K960746](https://fda-staging.innolitics.com/device/K960746)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).