ADVANTX LCV+

K960575 · GE Medical Systems · IZI · May 7, 1996 · Radiology

Device Facts

Record IDK960575
Device NameADVANTX LCV+
ApplicantGE Medical Systems
Product CodeIZI · Radiology
Decision DateMay 7, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1600
Device ClassClass 2

Indications for Use

The Advantx LVC+ x-ray imaging system is intended to be used for general purpose diagnostic angiographic fluoroscopy and radiographic studies.

Device Story

Advantx LCV+ is an angiographic and radiographic positioner; utilizes 32 cm intensifier; integrates with vascular table and x-ray system. Used in clinical settings by medical professionals for diagnostic imaging. System incorporates hardware and software redundancies to mitigate single-point failures regarding unintended motion. Safety managed via FMEA and adherence to UL 187 standards. Output provides diagnostic x-ray images for clinical assessment.

Clinical Evidence

Bench testing only; system evaluation performed to ensure performance to specifications and Federal Regulations; FMEA conducted to control potential hazards.

Technological Characteristics

Materials compliant with UL 187; 32 cm intensifier; energy source 220 V ac 50/60 Hz; includes hardware and software redundancies for motion control.

Indications for Use

Indicated for patients requiring general purpose diagnostic angiographic fluoroscopy and radiographic studies.

Regulatory Classification

Identification

An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Submission Summary (Full Text)

{0} K960575 MAY - 7 1996 GE Medical Systems P.O. Box 414, W-657 Milwaukee, WI 53201 USA # SUMMARY OF SAFETY & EFFECTIVENESS 8 February, 1996 This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21CFR Part 807.87(h). Contact: Larry A. Kroger, Ph.D. Phone: 414-544-3894 Fax: 414-544-3863 Product: Advantx LCV+ X-ray Imaging System Manufactured by GE Medical Systems - Europe, Buc, France Distributed by GE Medical Systems, Milwaukee, WI ## Indications for Use The Advantx LVC+ x-ray imaging system is intended to be used for general purpose diagnostic angiographic fluoroscopy and radiographic studies. ## Identification of the Product Name: Advantx LCV+ Mfg: GE Medical Systems - Europe 238 Rue de la Miniere 78530 Buc, FRANCE ## Indications for Use The Advantx LCV+ imaging system is intended for general purpose diagnostic angiographic fluoroscopy and radiography studies. ## Device Descriptions The Advantx LCV+ imaging system is an angiographic and radiographic positioner with a 32 cm intensifier, for use with a vascular table and an x-ray system. **Materials:** All construction and materials are compliant with UL 187. **Design:** There are hardware and software redundancies to prevent single point failures that could cause unintended motion. **Energy Source:** 220 V ac 50/60 Hz ## Marketing History In the opinion of GE Medical Systems, the LCA is of a comparable type and substantially equivalent to currently marketed diagnostic x-ray systems that comply with the same or equivalent standards and have the same intended uses. ## Adverse Effects on Health The potential hazards (unintended emission of x-rays, excessive radiation, mechanical and electrical) are identified in a Hazards Analysis and controlled by: - Failure Mode and Effects Analysis (FMEA) to demonstrate the non-existence or extremely low probability of unwanted events. - System evaluation to insure performance to specification and Federal Regulations. - Adherence to Industrial Standards (UL). ## Conclusions Use of the Advantx LCV+ imaging system does not result in any new potential safety risks.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...