DIOPEXY PROBE

K960522 · Iriderm Div. · HQF · Jul 19, 1996 · Ophthalmic

Device Facts

Record IDK960522
Device NameDIOPEXY PROBE
ApplicantIriderm Div.
Product CodeHQF · Ophthalmic
Decision DateJul 19, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4390
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of this device is transscleral retinal photocoagulation (TSRPC) for retinopexy.

Device Story

The DioPexy Probe is a fiber optic laser delivery device used by ophthalmologists during retinal detachment surgery. It connects to an IRIS OcuLight SL or SLx 810 nm infrared photocoagulator. The device features a contact prism tip that acts as a scleral depressor and deflects laser energy 90 degrees. During surgery, the physician places the probe on the scleral surface, uses a visible pilot beam to target the retina under indirect ophthalmoscopy, and delivers infrared laser energy to create chorioretinal adhesions. The physician monitors the burn intensity and duration via footswitch control. This process provides a seal around retinal breaks, preventing vitreous fluid leakage and facilitating retinal reattachment, offering an alternative to traditional diathermy or cryotherapy.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on the established clinical principles of retinopexy and the known safety/efficacy profile of 810 nm diode laser energy for transscleral photocoagulation as demonstrated by predicate devices.

Technological Characteristics

Multiple-use fiber optic laser delivery device; 3-meter fiber optic cable; contact prism tip for 90-degree laser deflection and scleral depression; compatible with 810 nm infrared photocoagulators; sterilization via Ethylene Oxide (EtO) gas.

Indications for Use

Indicated for patients requiring retinal detachment repair via retinopexy to create chorioretinal adhesions between the sensory retina and the RPE/choroid.

Regulatory Classification

Identification

An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.

Predicate Devices

Submission Summary (Full Text)

{0} K960522 510(k) Summary JUL 19 1996 IRIS Medical Instruments, Inc. DioPexy Probe ## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Theodore A. Boutacoff IRIS Medical Instruments, Inc. 340 Pioneer Way Mountain View, CA 94041 415-962-8100 Contact Person: (same as above) Date Prepared: January 15, 1996 ## Name of Device and Name/Address of Sponsor DioPexy Probe IRIS Medical Instruments, Inc. 340 Pioneer Way Mountain View, CA 94041 ## Classification Name Ophthalmic Laser (Accessory) CFR Section: 886.4390 Product Code: 86 HQF ## Predicate Devices The IRIS DioPexy Probe for transscleral retinal photocoagulation (TSRPC) is substantially equivalent to four predicate devices currently marketed: (1) the MIRA MD-1000 Microdiathermy System (pre-amendment), (2) the CooperSurgical Frigitronics CE-2000 Ophthalmic Cryosurgical System (K872362, K891282), (3) the IRIS Medical Laser Indirect Ophthalmoscope delivery device (K903288), and (4) the IRIS Medical G-Probe (K915236). Page 1 {1} Page 2 # Intended Use The intended use of this device is transscleral retinal photocoagulation (TSRPC) for retinopexy. # Technological Characteristics and Substantial Equivalence The DioPexy Probe enables the ophthalmologist to produce chorioretinal adhesions during retinal detachment surgery using delivery of transscleral infrared laser for retinopexy. Retinal detachments develop when a full thickness break in the retina allows vitreous fluid to dissect beneath the sensory retina causing it to separate from the retinal pigment epithelium (RPE)/choroid. To avoid progression to severe visual loss, a surgical procedure is necessary to reattach the retina. The principles of retinal detachment repair include retinopexy. Retinopexy is used to provide a seal around all retinal breaks to prevent further leakage of vitreous fluid into the subretinal space. Retinopexy creates an adhesion between the sensory retina and the RPE/choroid. This is stimulated by the post treatment inflammatory reaction at the level of the RPE/choroid and results in a chorioretinal scar. Current sources of energy used for retinopexy are heat (diathermy and laser) or freezing (cryotherapy). The MIRA utilizes diathermy for retinopexy. Diathermy has been used for transscleral retinopexy for over 60 years. When used, a diathermy probe is placed on the scleral surface of the eye and heat generated at the tip of the probe is delivered transsclerally to the inside of the eye to the retina/choroid. Diathermy usually requires scleral dissection to minimize scleral damage caused by the heat conduction. The physician determines when sufficient energy is delivered by either observing the burn develop on the retina with a diagnostic indirect ophthalmoscope or by observing an external scleral reaction. The Frigitronics utilizes cryotherapy for retinopexy. Cryotherapy is delivered transsclerally and was introduced in 1964. When used, a cryoprobe is placed on the surface of the sclera, the sclera is depressed, and cold generated at the tip of the probe creates an ice ball which conducts across the sclera. The physician observes the extent of ice ball progression inside the eye with a diagnostic indirect ophthalmoscope and when sufficient terminates energy delivery. The IRIS Medical Laser Indirect Ophthalmoscope utilizes 810 nm diode laser photocoagulation for retinopexy. Diode laser {2} Page 3 photocoagulation delivered transpupillary has been used since 1989 for retinopexy. When used, the LIO is placed on the physician’s head and with the aid of a hand-held indirect lens is used to observe the treatment site. Only breaks which can be viewed directly, with or without the aid of scleral depression, can be treated. When the target is illuminated with the pilot/aiming beam, the physician activates the footswitch and delivers laser energy. Transscleral application of diode laser has been shown to reduce intraocular pressure in the treatment of glaucoma using the IRIS Medical G-Probe (K915236) through transscleral cyclophotocoagulation (TSCPC). Transscleral photocoagulation of the sub-surface structures can be accomplished by placing a diode laser probe on the surface of the eye and allowing the infrared laser energy to be absorbed by the deep pigmented tissues. Typical laser parameters for TSCPC are 2000 milliseconds duration and 1750 milliwatts of power. The DioPexy Probe is intended to be used to perform transscleral retinal photocoagulation (TSRPC) in retinopexy procedures to accomplish retinal detachment repair. When using the DioPexy Probe the retinal treatment site is always under direct observation by the treating physician using an indirect ophthalmoscope. The prism tip of the DioPexy Probe is placed on the outer surface of the sclera and the sclera is depressed. A low power visible pilot/aiming beam is projected transsclerally onto the retina. This visible aiming beam can be observed by the physician on the retina during treatment. Size and intensity of the burn can be monitored and titrated with the laser footswitch. The extent of the laser lesion depends upon the energy delivered to the treatment site. Typical laser parameters for TSRPC are 2000 milliseconds duration and 1100 milliwatts of power. When longer duration burns and lower powers are used, burn development is slower. Chorioretinal adhesion also occurs without visible retinal uptake, as with cryotherapy in bullous retinal detachments. ## Device Description The IRIS DioPexy Probe is a multiple-use fiber optic laser delivery device with integral optics, which terminates with a contact prism tip. It must be directly connected to a compatible IRIS OcuLight SL or SLx (810 nm-infrared) Photocoagulator. The prism tip is designed to deflect laser energy 90° from the axis of the shaft and be simultaneously act as a scleral depressor. The fiber optic cable is 3 {3} meters (10 feet). The DioPexy Probe is shipped from the factory non-sterile in a hard plastic container. The device can be sterilized between each use in this container using Ethylene Oxide (EtO) gas in accordance with the user instructions. Page 4
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...