BRAUN HEMOSTASIS INTRODUCERS

K960406 · B.Braun Medical, Inc. · DYB · Apr 12, 1996 · Cardiovascular

Device Facts

Record IDK960406
Device NameBRAUN HEMOSTASIS INTRODUCERS
ApplicantB.Braun Medical, Inc.
Product CodeDYB · Cardiovascular
Decision DateApr 12, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The Braun Hemostasis Valve Introducer Kits are designed to facilitate percutaneous introduction of catheters into the vasculature while preventing the backflow of blood.

Device Story

Braun Hemostasis Introducer Kits consist of hemostasis valve introducer sheath, dilator, and optional guidewire/obturator. Used by physicians in clinical settings to facilitate percutaneous catheter insertion into vasculature. Sheath design prevents blood backflow during procedure. Available in straight and contralateral curve configurations, 5-9 French sizes, 10-60 cm lengths. Device provides access path for diagnostic or interventional catheters; enables clinician to maintain vascular access while minimizing blood loss.

Clinical Evidence

Bench testing only. Performance metrics include: hub-to-body tensile strength (14.5-18.4 lbs), shape recovery (85-88%), kink resistance (no kinks at 170° for 5F), and leak testing (no leaks at 100 psi for sheath, 6 psi for gasket).

Technological Characteristics

Materials: PEBA sheath. Construction: Braided body. Sizes: 5-9 French. Lengths: 10-60 cm. Configurations: Straight and contralateral curve. Features: Hemostasis valve, dilator, optional guidewire/obturator. Energy source: None (mechanical).

Indications for Use

Indicated for patients requiring percutaneous catheterization of the vasculature, facilitating device introduction while maintaining hemostasis.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} K960406 # II. SUMMARY OF SAFETY AND EFFECTIVENESS Applicant Name & Address: B. Braun Medical Inc., Cardiovascular Division, 13000 Hwy 55 Minneapolis, MN 55441-3813 (612) 559-3370 Contact: Barbara Atzenhoefer, Regulatory Affairs Manager Date Prepared: January 22, 1996 Common or Usual Name: Braun Hemostasis Introducer Kits Device Classification Name: Catheter Introducer ## Substantial Equivalence: K895044* Balkin Up & Over Cook K945396 Braun Fluoro-Tip™ Hemostasis Introducer Kits B. Braun Cardiovascular Device Description: Braun Hemostasis Introducer Kits consist of a hemostasis valve introducer sheath packaged with a corresponding dilator. One or more of the following accessories may also be packaged with the sheath/dilator; guidewire, and/or obturator. The introducer sheaths are available in both straight and contralateral curve versions. They will be available in 5 - 9 French sizes, in lengths from 30 - 60 cm in the curved version and 10 - 60 cm in the straight version. Each will be packaged with an appropriately sized dilator. Optional items to be kitted with the sheath and dilator include a guidewire, and/or obturator. The components, their materials and accessories, are similar to those previously cleared under Braun 510(k)s; K894446, K930057, K931743, K945396, and K950882. Statement of Intended Use: The Braun Hemostasis Valve Introducer Kits are designed to facilitate percutaneous introduction of catheters into the vasculature while preventing the backflow of blood. Summary/Comparison of Technological Characteristics: | Characteristic | Braun Hemostasis Introducer Kits | Cook Up & Over Balkin Contralateral- K895044 | Braun Fluoro-Tip K945396 | | --- | --- | --- | --- | | Available French Sizes | 5 - 9 French | Curved versions 5.5, 6 & 7 fr Straight versions of the Check-Flo introducer sheaths are available in 4-9 french | 7 - 10 fr | | Available Lengths | curved - 30 - 60 cm straight - 10 - 60 cm | curved - 40 cm straight - 7 - 13 cm | 10 - 32 cm | | Body Stock Design | Braided | non-braided | non-braided, double layer sheath | | Sheath Material | peba | Teflon® | FEP (Teflon®) | | Hub to Body Tensile | 5 fr - 14.5 lbs avg 9 fr - 18.4 lbs avg | 7.2 lbs avg | 7 fr - 14 lbs avg | | Shape Recovery | 5 fr 85% recovery @ 5 sec 86% recovery @ 60 sec 9fr 85% recovery @ 5 sec 88% recovery @ 60 sec | 72% recovery @ 5 sec 77% recovery @ 60 sec | Not applicable | | Sheath Leak Test | No leaks at 100 psi | No leaks at 100 psi | Not performed | | Gasket leak Test | No leaks at 6 psi | No leaks at 6 psi | No leaks at 10 psi | | Kink Resistance | 5 fr - No kinks at 170° 9 fr - Kinked at avg of 107° | Kinked at an avg of 76° | Not performed | | Radiopacity | More visible than Cook | Less visible than Braun | Not compared against Cook | 3
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...