ULTRACUT 2000

K960374 · Ultralase Technologies, Inc. · LFL · Jun 28, 1996 · SU

Device Facts

Record IDK960374
Device NameULTRACUT 2000
ApplicantUltralase Technologies, Inc.
Product CodeLFL · SU
Decision DateJun 28, 1996
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

The Ultracut 2000 is an ultrasonic surgical instrument intended for use to cut or fragment tissue, bone, or organs in orthopedic surgery.

Device Story

Ultrasonic surgical instrument; utilizes piezoelectric transducer to generate 40kHz vibrations; variable amplitude (0-500μ) and power (0-100 watts); features integrated aspiration and irrigation; operated via footswitch; used in orthopedic surgery for tissue/bone/organ fragmentation; provides surgeon control over cutting parameters; benefits include precise tissue removal with concurrent aspiration/irrigation.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Piezoelectric transducer; 40kHz operating frequency; variable power (0-100W) and amplitude (0-500μ); titanium or stainless steel surgical tips; autoclave sterilization; console dimensions 4.5 x 12.25 x 11.125 inches; 115VAC power source.

Indications for Use

Indicated for use in orthopedic surgery to cut or fragment tissue, bone, or organs.

Predicate Devices

Submission Summary (Full Text)

{0} K960374 # APPENDIX C ## SUMMARY OF SAFETY AND EFFECTIVENESS JUN 28 1996 The Ultralase Ultracut 2000 is equivalent to the predicate device Cavitron Corporation’s Cavitron Ultrasonic Surgical Aspirator (CUSA). The Ultracut 2000 has the same general intended use, similar principles of operation, and similar technological characteristics as the predicate device. The Ultracut 2000 and CUSA are both ultrasonic surgical instruments intended for use to cut or fragment tissue, bone, or organs in orthopedic surgery. ## Comparison Chart | | Ultracut 2000 | CUSA (Model 200H) | | --- | --- | --- | | 510(k) | Applied For | K801623 | | Console Dimensions (in) | 4.5 x 12.25 x 11.125 | 76 x 23 x 26 | | Power Specifications | 115VAC; 50/60 Hz | 115VAC; 50/60 Hz | | Footswitch Control | Yes | Yes | | Intended Use | Ultrasonic Surgery | Ultrasonic Surgery | | Frequency | 40kHz | 37kHz | | Vibration Amplitude | Variable, 0-500μ | Variable, 0-355μ | | Power | Variable, 0-100 watts | Variable, 0-70 watts | | Timer | Yes (1 sec - 100 min) | Yes (0 - 99 min) | | Aspiration | Yes | Yes | | Irrigation | Yes | Yes | | Modes of Operation | Continuous | Continuous | | | Pulsed On/Off | Pulsed On/Off | | Handpiece Frequency | 40kHz | 37kHz | | Dimensions (inches) | L 5.75 x D 1.25 | L 6.75 | | Removable Surgical Tips | Yes | Yes | | Tip Material | Titanium or Stainless Steel | Titanium | | Transducer | Piezoelectric | Magnetostrictive | | Sterilization | Autoclave | EtO or Steam |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...