K960373 · Bioject, Inc. · KZE · Mar 5, 1997 · General Hospital
Device Facts
Record ID
K960373
Device Name
BIOJECTOR 2000 MODEL 1B02000
Applicant
Bioject, Inc.
Product Code
KZE · General Hospital
Decision Date
Mar 5, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5430
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Biojector® 2000 is indicated for delivery of subcutaneous (SC) or intramuscular (IM) injections of vaccines and other injectable drugs. The Biojector® 2000 may be used by physicians, nurses, veterinarians, podiatrists and other practitioners who routinely administer injections. The Biojector® 2000 may also be used by patients authorized by their physicians to self inject, or have other individuals administer injections of prescribed medication.
Device Story
Needle-free injection system; delivers subcutaneous or intramuscular injections via high-velocity fluid jet. Components: single-use sterile medication syringe, injector body, CO2 cartridge (or tank with adapter). Operation: user depresses actuator; CO2 gas release drives plunger; drug forced through syringe orifice at high velocity to penetrate skin. CO2 does not contact drug. Used in clinical settings or home by patients/caregivers. Output: drug delivery into tissue. Benefits: needle-free administration of prescribed medications.
Clinical Evidence
Bench testing only. Product qualification testing and biocompatibility data demonstrate functional equivalence to the predicate device (K920631).
Technological Characteristics
Non-electrically powered jet injector. Materials: plastic and metal components. Power source: CO2 cartridge or tank. Syringe volumes up to 1.0 cc; orifice diameters 0.004 to 0.014 inches. Depth of penetration determined by orifice size. Manual activation.
Indications for Use
Indicated for subcutaneous or intramuscular delivery of vaccines and injectable drugs. Population includes patients authorized by physicians for self-injection or administration by others, and professional practitioners (physicians, nurses, veterinarians, podiatrists).
Regulatory Classification
Identification
A nonelectrically powered fluid injector is a nonelectrically powered device used by a health care provider to give a hypodermic injection by means of a narrow, high velocity jet of fluid which can penetrate the surface of the skin and deliver the fluid to the body. It may be used for mass inoculations.
{0}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
MAR - 5 1997
Mr. Mary K. Norton
Manager, Regulatory Affairs/Quality Assurance
Bioject, Incorporated
7620 S.W. Bridgeport Road
Portland, Oregon 97224
Re: K960373
Trade Name: Biojector 2000 Needle-Free Injection Management System
Regulatory Class: II
Product Code: KZE
Dated: December 5, 1996
Received: December 6, 1996
Dear Ms. Norton:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of
{1}
Page 2 - Ms. Norton
the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Timothy A. Ulatowski
Director
Division of Dental, Infection Control and General Hospital Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}
Page 1 of 1
510(k) Number (if known) K960373
Device Name: **Biojector® Needle-Free Injection Management System, Biojector® 2000**
## Indications For Use:
The Biojector® 2000 is indicated for delivery of subcutaneous (SC) or intramuscular (IM) injections of vaccines and other injectable drugs. The Biojector® 2000 may be used by physicians, nurses, veterinarians, podiatrists and other practitioners who routinely administer injections. The Biojector® 2000 may also be used by patients authorized by their physicians to self inject, or have other individuals administer injections of prescribed medication.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) *Tobacco Occur*
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number *K 960373*
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over The-Counter Use ☐
Optional Format 1-2-96)
011-1 3
{3}
K960373
BIOJECT
January 24, 1996
MAR - 5 1997
BIOJECT INC.
7620 S.W. BRIDGEPORT ROAD
PORTLAND, OREGON 97224
TELEPHONE: (503) 639-7221
FAX: (503) 624-9002
# 510(k) SUMMARY (21CFR Subpart 807.92)
## (a)(1) Submitter Identification:
Bioject, Inc.
7620 SW Bridgeport Road
Portland, OR 97224
800-683-7221
503-624-9002 (fax)
**Contact Person:**
Mary K. Norton
Manager, Regulatory Affairs/Quality Assurance
## (a)(2) Device Name:
Biojector® 2000 Needle-Free Injection Management System™
**Common Name:** Needleless Injector, Jet Injector
**Classification Name:** Non-electrically Powered Fluid Injector, 21 CFR 880.5430
## (a)(3) Predicate Device:
Biojector® 2000 Jet Injection System (K920631)
## (a)(4) Device Description:
The Biojector® 2000 is a needle free injection management system designed to deliver drugs both subcutaneously (SC) and intramuscularly (IM). It is a non-electrically powered device which is intended to administer an injection by means of a high velocity jet of fluid that penetrates the surface of the skin and delivers drug to the body. The system is comprised of three major components: The single use sterile medication syringe (medication container), the Biojector® (injector), and the Carbon Dioxide (CO₂) cartridge (power source). The device is also capable of being powered with a CO₂ tank with a specialized adapter.
The syringe holds a variable volume of drug up to a maximum of one cubic centimeter (cc) or milliliter (ml.). Volume increments are marked at 0.10, 0.20, 0.25, 0.30, 0.40, 0.50, 0.60, 0.75, 0.80, 0.90, and 1.0 cc or ml. The syringe is filled using either a fill needle, or a plastic fluid transfer device. Once the syringe is filled, it is loaded in the Biojector®. Syringes are available in five sizes which are numbered 2, 3, 4, 5 and 7. The syringe sizes have the following orifice diameters: 0.004, 0.006, 0.008, 0.010 & 0.014 inches, respectively. The number 2 syringe is used for all subcutaneous injections. The remaining syringes, numbered 3, 4, 5, and 7, are used for intramuscular injections. Depth of penetration of the drug varies with the orifice diameter selected. The larger the orifice diameter, the deeper the penetration of the drug.
022
510(k) Summary
Appendix A
{4}
Activation of the Biojector® is initiated when the actuator is depressed. CO2 gas is released, through the action of a series of valves within the injector. This causes the plunger to push the drug out of the sterile syringe, through the syringe orifice at a high velocity, allowing it to penetrate the skin and be deposited in the tissue. Exhaust CO2 is expelled, through the exhaust and bleed valves, in the rear section of the Biojector®, as well as into the cartridge compartment. The CO2 does not come in contact with the drug.
(a)(5) Intended Use: The Biojector® 2000 Needle-Free Injection Management System™ which is the subject of this submission and the currently marketed (predicate) Biojector® 2000 Jet Injection System (with tank adapter), K920631, are intended to deliver subcutaneous (SC) or intramuscular (IM) injections of vaccines and other injectable medications. Both the Biojector® 2000 (subject of submission) and the predicate device may be used by physicians, nurses, veterinarians, and podiatrists and other practitioners who routinely administer injections. The Biojector® 2000 (subject of submission) may also be used by patients authorized by their physicians to self inject at home, or have other individuals administer injections of prescribed drug. It should be noted that a prior submission made by Bioject (K861687), predicate to K920631, was cleared via the 510(k) process for professional and home use.
(a)(6) This Premarket Notification describes modifications to the Biojector® 2000 device previously described via the 510(k) process (K920631). A summary of these modifications includes:
- Changing from a two piece to a one piece main body design with related minor design changes to ease manufacturing;
- Incorporation of design modifications to the exhaust and bleed valves to improve CO2 efficiency and performance;
- Changes in material involved replacing metal parts with plastic parts for cost reduction purposes and to add alternate vendors for existing materials;
- Labeling revisions including the addition of a Home Use Instruction Manual.
There are no changes in intended use or operational principle. Substantial equivalence is based upon equivalence in intended use, labeling, design, materials and operational principle to the Biojector® Jet Injection System (with tank adapter), K920631 and there are no new safety and efficacy issues as a result of the design modifications.
Product qualification testing and biocompatibility data for the modified device (subject of submission) and the current, predicate device (K920631), demonstrate the new model to be functionally equivalent to the predicate device. The device is as safe and effective as the legally marketed, predicate device, and does not raise questions of safety and efficacy different than that of the predicate.
510 (k) Summary
023
Appendix A
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.