RDI CARDIAC INSULATION PAD

K960312 · R D Intl. · DWP · Apr 19, 1996 · Cardiovascular

Device Facts

Record IDK960312
Device NameRDI CARDIAC INSULATION PAD
ApplicantR D Intl.
Product CodeDWP · Cardiovascular
Decision DateApr 19, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.4475
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The RDI Cardiac Insulation Pad is intended to be used during open heart surgery to insulate the myocardium from the warmer organs and tissues in the body and thus reduce undesired rewarming of the heart.

Device Story

Pear-shaped, 1/8” thick, closed-cell polyethylene foam pad with 4 mil thick, 1/2” wide, 20” long polyvinyl chloride 'tail'. Used during open heart surgery by surgeons to insulate myocardium from warmer adjacent tissues/organs. Prevents undesired rewarming of heart during procedure. Device provides thermal barrier; performance verified via thermal conductivity, water absorption, and mechanical pull testing compared to predicate.

Clinical Evidence

Bench testing only. Functional testing demonstrated similarity to predicate in thermal conductivity, tail pull strength, and water absorption. Package integrity validated per ASTM F1140-88; shipping/distribution validated per NSTA Project 1A. Biocompatibility compliant with Tripartite Guidance for short-term, direct blood path contact. Non-pyrogenic. Two-year shelf life established via accelerated aging.

Technological Characteristics

Materials: Closed-cell polyethylene foam, polyvinyl chloride. Biocompatibility: Tripartite Guidance (external, blood path direct, short-term). Sterilization: Gamma radiation (AAMI/ISO Method 1). Mechanical: 1/8” thick pad, 20” tail. Standards: ASTM F1140-88 (package integrity), NSTA Project 1A (shipping).

Indications for Use

Indicated for patients undergoing open heart surgery to provide myocardial insulation against surrounding body heat.

Regulatory Classification

Identification

A surgical vessel dilator is a device used to enlarge or calibrate a vessel.

Predicate Devices

Submission Summary (Full Text)

{0} K960312 APR 19 1996 # SUMMARY OF SAFETY AND EFFECTIVENESS ## RDI CARDIAC INSULATION PAD ### I. General Information A. Generic Name: Cardiac Insulation Pad B. Trade Name of Device: RDI Cardiac Insulation Pad C. Applicants Name and Address: RDI Richardson, Texas D. Pre-market Notification Number: Not assigned ### II. Indications for Use The RDI Cardiac Insulation Pad is intended to be used during open heart surgery to insulate the myocardium from the warmer organs and tissues in the body and thus reduce undesired rewarming of the heart. ### III. Device Description The RDI Cardiac Insulation Pad consists of a pear shaped, 1/8” thick, closed cell polyethylene foam material. Attached to the foam is a 1/2” wide, 20” long, 4 mil thick, polyvinyl chloride strip referred to as a “tail”. ### IV. Device Classification: Class II ### V. Safety and Effectiveness: Substantial Equivalence: This device has been shown to be substantially equivalent to the Sorin/Shiley Inc. Cardiac Insulation Pad # ### VI. Other Safety and Effectiveness Data: **Materials:** Fluid contact materials of construction comply with Tripartite Biocompatibility Guidance for external devices, blood path direct, short term use. **Sterilization:** Validated AAMI/ISO Method 1 validation for Gamma radiation sterilization. **Pyrogenicity:** Non-Pyrogenic **Functional Testing:** | Thermal Conductivity | Found to be similar to Sorin/Shiley | | --- | --- | | Tail Pull Test | Found to be similar to Sorin/Shiley | | Water Absorption | Found to be similar to Sorin/Shiley | | Package Integrity | Passed in accordance with ASTM F1140-88 | | Shipping & Distribution | Passed in accordance with NSTA Project 1A vibration/drop tests | | Accelerated Aging | Successful two year shelf life |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...