K960032 · Coherent Medical Group · GEX · Apr 1, 1996 · General, Plastic Surgery
Device Facts
Record ID
K960032
Device Name
COHERENT VERSAPULSE AESTHETIC SURGICAL LASERS
Applicant
Coherent Medical Group
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 1, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Coherent VersaPulse Aesthetic Surgical Lasers and the Delivery Accessories that are used with them to deliver 532 nm, 1064 nm and 755 nm laser energy that are indicated for use in surgical applications requiring photothermolysis of soft tissue, and 2940 nm laser energy that is indicted for use in surgical applications requiring excision, incision, ablation, vaporization, and coagulation of soft tissue in medical specialties including: general and plastic surgery; and dermatology.
Device Story
VersaPulse Aesthetic Surgical Lasers deliver laser energy (532 nm, 1064 nm, 755 nm, 2940 nm) for soft tissue surgery. System includes laser console/tower, articulated arm, fiber port, control/display panels, footswitch/handswitch, and various delivery accessories (handpieces, scanners, epidermal cooling). Operated by clinicians in dermatology, plastic, or general surgery. Device transforms electrical energy into specific laser wavelengths to achieve photothermolysis, excision, incision, ablation, vaporization, or coagulation. Clinician uses control panel to set parameters; output delivered via handpiece to target tissue. Benefits include precise tissue interaction for aesthetic and surgical procedures.
Clinical Evidence
In-vivo testing performed to demonstrate laser energy equivalence to currently marketed surgical lasers. Biocompatibility testing conducted for new delivery device materials contacting patient tissue. No clinical human trial data reported.
Indicated for surgical applications requiring photothermolysis of soft tissue (532 nm, 1064 nm, 755 nm) and excision, incision, ablation, vaporization, and coagulation of soft tissue (2940 nm) in general surgery, plastic surgery, and dermatology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Candela 10 mm Circular Spot Laser Handpiece (K950661)
Coherent Family of Scanner Handpieces (K951812 & K946304)
Cyrus Chess Cool Laser Optics (CLO) Unit (K934486)
Submission Summary (Full Text)
{0}
K960032
APR - 1 1996
Attachment 27
510(k) Summary Statement for the
Coherent VersaPulse Aesthetic Surgical Lasers
# I. General Information
Submitter: Coherent Medical Group
3270 West Bayshore Road
Palo Alto, CA 94303
Contact Person: Anne C. Worden
Summary Preparation Date: December 29, 1995
# II. Names
Device Names: Coherent VersaPulse Aesthetic Surgical Lasers and Delivery Accessories.
Primary Classification Name: Laser Powered Surgical Instrument (and Accessories).
# III. Predicate Devices
- Laserscope Orion SL Series & KTP/532, KTP:YAG Surgical Lasers (K941841, K933880 & K911830)
- Continuum Biomedical Medlite 532 & 1064 nm Surgical Lasers (K952636, K922935 & K922577)
- Continuum Biomedical MedAlex 755 nm Surgical Lasers (K935631)
- Candela TL-1 Alexandrite Laser (K944090 & K940173)
- Pfizer/Premier Centauri & Pegasus 2940 nm Surgical Lasers (K905802 & K904627)
- Continuum Multilite Er:YAG (755 nm) Surgical Laser (K933057)
- Laserscope SmartScan Fiber Delivery Device for use with Orion SL Series & KTP/532, KTP:YAG Surgical Lasers (K941841, K933880 & K911830)
- Candela 10 mm Circular Spot Laser Handpiece (K950661)
- Coherent Family of Scanner Handpieces for use with Coherent ULTRAPULSE Carbon Dioxide Surgical Lasers (K951812 & K946304)
- Cyrus Chess Cool Laser Optics (CLO) Unit for use with Argon (488/514 nm), Argon/Dye (577 nm and 585 nm), and KTP (532 nm) Surgical Lasers (K934486)
# IV. Product Description
The Coherent VersaPulse Aesthetic Surgical Lasers (and its delivery accessories) are intended to be used to deliver: frequency doubled, Q-switched frequency doubled, and Q-switched Nd:YAG; Q-switched Alexandrite:YAG laser energy for use in surgical applications requiring the photothermolysis (photocoagulation) of soft tissue; and Er:YAG laser energy for use in surgical applications requiring the excision, incision, ablation, vaporization, and coagulation of soft tissue in medical specialties including: general and plastic surgery; and dermatology.
Coherent VersaPulse Aesthetic Surgical Lasers are comprised of the following main components:
510(k) Coherent VersaPulse Aesthetic Surgical Lasers
Attachment 27 - Page 1
{1}
- a laser console and tower;
- a counterbalanced articulated arm and delivery systems;
- a fiber port and delivery systems;
- control and display panels;
- footswitch and handswitch delivery controls;
- a remote control unit; and
- a variety of delivery device accessories (articulated arm and fiber handpieces, scanners, and epidermal cooling devices).
## V. Indications for Use
The Coherent VersaPulse Aesthetic Surgical Lasers and the Delivery Accessories that are used with them to deliver 532 nm, 1064 nm and 755 nm laser energy that are indicated for use in surgical applications requiring photothermolysis of soft tissue, and 2940 nm laser energy that is indicted for use in surgical applications requiring excision, incision, ablation, vaporization, and coagulation of soft tissue in medical specialties including: general and plastic surgery; and dermatology.
## VI. Rationale for Substantial Equivalence
The Coherent VersaPulse Aesthetic Surgical Lasers and their delivery device accessories share the same indications for use, similar design features (including control system, wavelengths, beam quality, laser tube configuration, active medium, cooling system, and controls and displays), functional features (including power, repetition rate, energy, spot sizes and treatment areas), and therefore are substantially equivalent to the Laserscope Orion SL Series & KTP/532, KTP:YAG Surgical Lasers (K941841, K933880 & K911830), the Continuum Biomedical Medlite 532 nm & 1064 nm Surgical Lasers (K952636, K922935 & K922577), the Continuum Biomedical MedAlex 755 nm Surgical Lasers (K935631), the Candela TL-1 Alexandrite Surgical Lasers (K944090 & K940173), the Pfizer/Premier Centauri & Pegasus 2940 nm Surgical Lasers (K905802 & K904627), and the Continuum Multilite Er:YAG Surgical Laser (K933057), and to the Laserscope SmartScan Fiber Delivery Device for use with Orion SL Series & KTP/532, KTP:YAG Surgical Lasers (K941841), the Candela 10 mm Circular Spot Laser Handpiece (K950661), the Coherent Family of Scanner Handpieces for use with Coherent ULTRAPULSE Carbon Dioxide Surgical Lasers (K951812 & K946304), as well as the Cyrus Chess Cool Laser Optics (CLO) Unit for use with Argon (488/514 nm), Argon/Dye (577 nm and 585 nm), and KTP (532 nm) Surgical Lasers (K934486).
## VII. Safety and Effectiveness Information
In-vivo testing (was performed to demonstrate that the laser energy provided by the Coherent VersaPulse Aesthetic Surgical Lasers is equivalent to that provided by the currently marketed surgical lasers. Testing is performed to establish the biocompatibility of new delivery device materials that could potentially contact patient tissue.
510(k) Coherent VersaPulse Aesthetic Surgical Lasers
Attachment 27 - Page 2
{2}
# VIII. Conclusion
The Coherent VersaPulse Aesthetic Surgical Lasers were found to be substantially equivalent to similar currently marketed surgical laser devices. The Coherent VersaPulse Aesthetic Surgical Lasers share the same indications for use, similar design features, and similar functional features as the currently marketed Surgical Lasers and delivery devices. The physical test results demonstrated that the Coherent VersaPulse Aesthetic Surgical Lasers have acceptable performance characteristics.
510(k) Coherent VersaPulse Aesthetic Surgical Lasers
Attachment 27 - Page 3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.