BRAUN HEMOSTASIS INTRODUCERS

K955820 · B.Braun Medical, Inc. · DYB · Mar 25, 1996 · Cardiovascular

Device Facts

Record IDK955820
Device NameBRAUN HEMOSTASIS INTRODUCERS
ApplicantB.Braun Medical, Inc.
Product CodeDYB · Cardiovascular
Decision DateMar 25, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

For percutaneous introduction of catheters into the vasculature while preventing the backflow of blood.

Device Story

Device consists of 16 and 18 French catheter introducers and vessel dilators; includes silicone gasket to prevent blood backflow. Used by clinicians for percutaneous vascular access. Introducer sheath (FEP with BaSO4) and dilator (high-density polyethylene with bismuth subcarbonate) facilitate catheter placement. Device functions as a mechanical conduit for vascular procedures.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Sheath: FEP with BaSO4. Dilator: High-density polyethylene with bismuth subcarbonate. Gasket: Silicone. Sizes: 16 and 18 French. Lengths: 10-40 cm. Packaging: Tyvek/Polymylar pouch. Sterilization: Not specified.

Indications for Use

Indicated for percutaneous introduction of catheters into the vasculature in patients requiring vascular access.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Submission Summary (Full Text)

{0} 3 90 # II. SUMMARY OF SAFETY AND EFFECTIVENESS Applicant Name & Address: B. Braun Medical Inc., Cardiovascular Division, 13000 Hwy 55 Minneapolis, MN 55441-3813 (612) 559-3370 Contact: Barbara Atzenhoefer, Regulatory Affairs Manager Date Prepared: December 19, 1995 Common or Usual Name: Catheter introducer and Vessel dilator for percutaneous catheterization Device Classification Name: Introducer: 21 CFR Reference- 870.1340, Class II Panel- 74 DYB Dilator: 21 CFR Reference- 870.1310, Class II, Panel- 74 DRE Substantial Equivalence: | 510(k) # | Product | Manufacturer | | --- | --- | --- | | K894446 | Angeion Hemostasis Valve Introducer Kit | B. Braun Cardiovascular - MN | | K931743 | Burron Cardiovascular Angestat | B. Braun Cardiovascular - MN | | K871153 | ANGETEAR Tear-away Introducer Kits | B. Braun Cardiovascular - MN | | K930057 | Fluoro-tip Hemostasis Introducer Set | B. Braun Cardiovascular - MN | | K945396 | Fluoro-tip Hemostasis Introducer Set | B. Braun Cardiovascular - MN | Device Description: 16 & 18 french versions of the Braun Hemostasis Introducer with silicone gasket, in lengths of 10 - 40 cm. Statement of Intended Use: For percutaneous introduction of catheters into the vasculature while preventing the backflow of blood. Summary/Comparison of Technological Characteristics: | Characteristic | Braun Hemostasis Introducers | K894446 Hemostasis Introducer | K931743 Angestat | K871153 ANGETEAR | | --- | --- | --- | --- | --- | | A) Labeling/Pkg | | | | | | Labels & IFUs | See Attachment B | All Braun Hemostasis Introducer Label formats are identical | All Braun Hemostasis Introducer Label formats are identical | same information provided on pouch and box labels as on hemostasis labels | | Expiration Dating | 2 years | 2 years | 2 years | 4 years | | Pyrogen Free | Yes | Yes | Yes | Yes | | Packaging | Tyvek/Polymylar pouch | Tyvek/Polymylar pouch | Tyvek/Polymylar pouch | Tray w/ tyvek lid w/in a tyvek/polymylar pouch | | B) Intended Use | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood | percutaneous introduction of catheters into the vasculature while minimizing the backflow of blood | percutaneous introduction of catheters into the vasculature | | C) Physical Char. | | | | | | Available french sizes | 4-18 french | 5-14 french | 4-15 french | 4-16 french | | Lengths | 10-40 cm | 12-24 cm | 15-32 cm | 16-30 cm | | gasket material | silicone | Natural rubber | silicone | none | | sheath material | FEP with BaSO4 | FEP teflon with BaSO4 | FEP with BaSO4 | polyethylene with bismuth subcarbonate | | dilator material | high density polyethylene with bismuth subcarbonate | polyethylene with bismuth subcarbonate | polyethylene with bismuth subcarbonate | FEP with BaSO4, or polyethylene with bismuth subcarbonate | | D) Anatomical Sites | vasculature | vasculature | vasculature | vasculature |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...