← Product Code [JPA](/productcode/JPA) · K955720

# AMAX CS-190 COAGULATION ANALYZER SYSTEM (K955720)

_Sigma Diagnostics, Inc. · JPA · Mar 13, 1996 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K955720

## Device Facts

- **Applicant:** Sigma Diagnostics, Inc.
- **Product Code:** [JPA](/productcode/JPA.md)
- **Decision Date:** Mar 13, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5425
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The Amelung AMAX CS-190 Coagulation Analyzer is an automated random access multipurpose analyzer. The AMAX CS-190 Coagulation Analyzer can be used for the detection of fibrin formation utilizing either mechanical principles (ball method) or photo-optical principles to perform clot based tests such as prothrombin time(PT), activated partial thromboplastin time (APTT), fibrinogen, factor assays, and other clotting tests. In addition, the AMAX CS-190 Coagulation Analyzer can be used for chromogenic assays such as antithrombin III (AT III).

## Device Story

Automated random access multipurpose coagulation analyzer; utilizes mechanical (ball method) or photo-optical principles to detect fibrin formation. Performs clot-based assays (PT, APTT, fibrinogen, factor assays) and chromogenic assays (AT III). Used in clinical laboratory settings by trained personnel. Provides quantitative results for coagulation parameters; assists clinicians in diagnosing and monitoring coagulation disorders. Benefits patients through rapid, automated, and precise coagulation testing.

## Clinical Evidence

Bench testing only. Comparison studies performed against TOA CA-5000 (mechanical), BBL Fibrometer, and manual spectrophotometric methods. PT (optical) r=0.989; APTT (optical) r=0.907; Factor IX (optical) r=0.954; Factor X (optical) r=0.980; Fibrinogen (optical) r=0.979; AT III r=0.848. Precision studies showed CV <4% for PT/APTT, <9% for factor assays, <10% for fibrinogen, and <6% for AT III.

## Technological Characteristics

Automated random access analyzer; dual-mode detection: mechanical (ball method) and photo-optical. Performs clot-based and chromogenic assays. No specific materials, connectivity, or software architecture details provided.

## Regulatory Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

## Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

## Reference Devices

- TOA Medical Electronics CA-5000
- BBL Fibrometer

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

510(k) NOTIFICATION
Sigma Diagnostics
545 South Ewing Avenue
St. Louis, MO 63103
AMAX CS-190
Coagulation Analyzer
December 15, 1995

K955720

MAR 13 1996

# ATTACHMENT 1

## Summary of Safety and Effectiveness

{1}

510(k) NOTIFICATION
Sigma Diagnostics
545 South Ewing Avenue
St. Louis, MO 63103
AMAX CS-190
Coagulation Analyzer
December 15, 1995

# ATTACHMENT 1

## 510(k) Summary of Safety and Effectiveness

The Amelung AMAX CS-190 Coagulation Analyzer is an automated random access multipurpose analyzer. The AMAX CS-190 Coagulation Analyzer can be used for the detection of fibrin formation utilizing either mechanical principles (ball method) or photo-optical principles to perform clot based tests such as prothrombin time(PT), activated partial thromboplastin time (APTT), fibrinogen, factor assays, and other clotting tests. In addition, the AMAX CS-190 Coagulation Analyzer can be used for chromogenic assays such as antithrombin III (AT III).

In comparison studies of clot based assays between the AMAX CS-190 Coagulation Analyzer (photo-optical or mechanical) and the TOA Medical Electronics CA-5000 (mechanical), the following regression statistics were obtained:

|  PT (optical) | r = 0.989 | y = 1.113x - 0.9  |
| --- | --- | --- |
|  APTT (optical) | r = 0.907 | y = 0.848x + 7.3  |
|  Factor IX (optical) | r = 0.954 | y = 0.950x + 0.7  |
|  Factor IX (mechanical) | r = 0.882 | y = 0.882x + 2.3  |
|  Factor X (optical) | r = 0.980 | y = 1.019x + 0.0  |
|  Factor X (mechanical) | r = 0.976 | y = 1.001x - 0.3  |
|  Fibrinogen (optical) | r = 0.979 | y = 1.211x + 0.9  |

In comparison studies of mechanical clot based assays between the AMAX CS-190 Coagulation Analyzer and the BBL Fibrometer, the following regression statistics were obtained:

|  PT | r = 0.985 | y = 1.067x - 0.3  |
| --- | --- | --- |
|  APTT | r = 0.885 | y = 1.427x + 0.4  |

In comparison studies of chromogenic assays between the AMAX CS-190 Coagulation Analyzer and the manual spectrophotometric method, the following regression statistics were obtained:

|  AT III | r = 0.848 | y = 0.848x + 5.0  |
| --- | --- | --- |

The coefficient of variation for precision studies was less than 4% for the PT and APTT assays, less than 9% for the factor assays, less than 10% for the fibrinogen assays, and less than 6% for AT III.

The safety and effectiveness of the AMAX CS-190 Coagulation Analyzer is demonstrated by its substantial equivalency to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K955720](https://fda-staging.innolitics.com/device/K955720)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).