SYNCHRON SYSTEMS CANNABINOID 20 NG REAGENT

K955677 · Beckman Instruments, Inc. · LDJ · Apr 3, 1996 · Clinical Toxicology

Device Facts

Record IDK955677
Device NameSYNCHRON SYSTEMS CANNABINOID 20 NG REAGENT
ApplicantBeckman Instruments, Inc.
Product CodeLDJ · Clinical Toxicology
Decision DateApr 3, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3870
Device ClassClass 2

Indications for Use

The SYNCHRON Systems Cannabinoid 20 ng Reagent, in conjunction with the SYNCHRON Systems THC Urine Calibrators, is intended for use in the qualitative determination of cannabinoids in human urine samples. This assay is designed for use with the family of SYNCHRON Systems, which include analyzers such as the SYNCHRON® CX4, CX®4CE, CX4 DELTA, CX5, CX5CE, CX5 DELTA, CX7, and CX7 DELTA Systems.

Device Story

The SYNCHRON Cannabinoid 20 ng Reagent is an enzyme immunoassay test kit for use on Beckman SYNCHRON chemistry analyzers. It detects cannabinoids in human urine via competitive binding between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH). Enzyme activity is measured by the conversion of NAD to NADH, which is monitored spectrophotometrically. The device is used in clinical laboratory settings by trained personnel to provide preliminary screening results. The output is a qualitative determination of cannabinoid presence, which assists healthcare providers in identifying potential drug use. The liquid-stabilized reagent requires no preparation and is designed for automated processing on SYNCHRON systems.

Clinical Evidence

Bench testing only. Method comparison study of 136 urine samples against the Emit II Cannabinoid Reagent demonstrated 99% overall agreement, 99% relative sensitivity, and 100% relative specificity. Imprecision studies (n=20 per material) showed %CV of 0.3% across negative, low, and high calibrator/control levels. Stability studies confirmed 12-month unopened shelf-life and 60-day on-instrument stability.

Technological Characteristics

Homogeneous enzyme immunoassay; competitive binding principle using G6PDH enzyme label; spectrophotometric measurement of NAD to NADH conversion. Liquid-stabilized reagent; automated pipetting and reaction timing on SYNCHRON chemistry analyzers. Constant reaction temperature. 60-day on-instrument stability.

Indications for Use

Indicated for the qualitative screening of cannabinoids and their metabolites in human urine samples to detect drug presence.

Regulatory Classification

Identification

A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.

Special Controls

*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0} 200 South Kraemer Boulevard, Box 8000 Brea CA 92621 8000 • (714) 993-5321 26 # BECKMAN Appendix C Summary of Safety & Effectiveness SYNCHRON® Systems Cannabinoid 20 ng Reagent APR - 3 1995 K955677 ## 1.0 Submitted By Sheri Hall Manager, Product Submissions Beckman Instruments, Inc. 200 S. Kraemer Blvd. W-337 Brea, California 92621 Telephone: (714) 993-8916 FAX: (714) 961-4457 ## 2.0 Date Submitted 15 December 1995 ## 3.0 Device Name(s) ### 3.1 Proprietary Names SYNCHRON Systems Cannabinoids 20 ng (THC2) Reagent SYNCHRON Systems DAT Negative Urine Calibrator SYNCHRON Systems 20 ng/mL THC Urine Calibrator SYNCHRON Systems 50 ng/mL THC Urine Calibrator SYNCHRON Systems 0 ng/mL THC Urine Control SYNCHRON Systems 50 ng/mL THC Urine Control ### 3.2 Classification Names Cannabinoid test system (21 CFR 862.3870) ## 4.0 Predicate Device(s) Emit II Cannabinoid 20 ng Assay, Syva Company, K940158 Cannabinoid Reagent, Diagnostic Reagents, Inc, K943998 THC Urine Calibrators and Controls, Diagnostic Reagents, Inc., K944701 ## 5.0 Description The SYNCHRON Cannabinoid 20 ng Reagent is an enzyme immunoassay reagent test kit for the SYNCHRON Systems and provides a rapid screening procedure for determining the presence of cannabinoids and it's metabolites in urine. It is designed to complement the current panel of CX Drugs of Abuse Assays, previously cleared by FDA. The kit consists of reagent and calibrators. ## 6.0 Intended Use The SYNCHRON Systems Cannabinoid 20 ng Reagent, in conjunction with the SYNCHRON Systems THC Urine Calibrators, is intended for use in the qualitative determination of cannabinoids in human urine samples. This assay is designed for use with the family of SYNCHRON Systems, which include analyzers such as the SYNCHRON® CX4, CX®4CE, CX4 DELTA, CX5, CX5CE, CX5 DELTA, CX7, and CX7 DELTA Systems. thc2510k.sse Beckman Instruments, Inc. twx: 910-592-1260 • telex 06-78413 {1} Beckman Instruments, Inc. Section 510(k) Notification SYNCHRON® Systems Cannabinoid 20 ng Reagent # 7.0 Comparison to Predicate(s) The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary. | SYNCHRON Cannabinoid Reagent | Emit II Cannabinoid Reagent | | --- | --- | | Similarities | | | Homogeneous enzyme immunoassay intended for the qualitative analysis of cannabinoids in human urine | Same | | Method based on competition between drug in the sample and drug labeled with the enzyme G6PDH; enzyme activity measured as NAD is converted to NADH | Same | | Reagent designed for chemistry analyzers which maintain constant reaction temperature, use automated pipetting, measure enzyme rates, mix, and accurately time reactions | Same | | Reagent provides preliminary analytical test for screening purposes | Same | | Differences | | | SYNCHRON reagent is liquid stabilized and requires no preparation | Emit II reagent is lyophilized and requires reconstitution | | SYNCHRON reagent is intended for qualitative determinations only | Emit II reagent has a specialized application for semi-quantitative determination of drug concentration | | SYNCHRON reagent is stable for 60 days once opened, when stored properly | Emit II reagent is stable for 12 weeks once opened, if handled as directed | # 8.0 Summary of Performance Data The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, and imprecision experiments that relate results obtained from the SYNCHRON Systems Cannabinoid 20 ng Reagent to the Emit II Reagent. Relative Sensitivity and Specificity Study Results SYNCHRON Cannabinoid 20 ng Reagent vs Emit II | Overall Agreement | 99% | | --- | --- | | Relative Sensitivity | 99% | | Relative Specificity | 100% | | Number of Disconcordant w/ Emit II | 1 | | Total Number of Samples | 136 | SYNCHRON Cannabinoid 20 ng Reagent Stability Study Results | Product Claim | 510(k) Summary | | --- | --- | | Shelf-life | 12 months unopened | | On-Instrument Stability | 60 days | | Calibration Frequency | 7 days | thc2510k.sse {2} Beckman Instruments, Inc. Section 510(k) Notification SYNCHRON® Systems Cannabinoid 20 ng Reagent Estimated Within-Run Imprecision SYNCHRON Cannabinoid 20 ng Reagent | Material | Mean (mA/min) | SD (mA/min) | %CV | Number of Results | | --- | --- | --- | --- | --- | | Negative Calibrator | 245 | 0.8 | 0.3 | 20 | | Control 1 (0 ng/mL) | 245 | 0.8 | 0.3 | 20 | | Low Calibrator (20 ng/mL) | 274 | 0.9 | 0.3 | 20 | | Control 2 (50 ng/mL) | 318 | 1.0 | 0.3 | 20 | | High Calibrator (50 ng/mL) | 318 | 1.1 | 0.3 | 20 | This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92. thc2510k.sse
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...