TROJAN POLYURETHANE CONDOM (MODIFICATION)
Device Facts
| Record ID | K955672 |
|---|---|
| Device Name | TROJAN POLYURETHANE CONDOM (MODIFICATION) |
| Applicant | Carter Products |
| Product Code | MOL · Obstetrics/Gynecology |
| Decision Date | Aug 29, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 884.5300 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The subject polyurethane condom has the same intended use as the predicate latex condoms. The condom is used for contraceptive and for prophylactic purposes (helping to prevent transmission of venereal disease). If used properly, condoms will help to reduce the risk of transmission of HIV infection (AIDS) and many other sexually transmitted diseases, including chlamydia, genital herpes, genital warts, gonorrhea, hepatitis B, and syphilis. No contraceptive can guarantee 100% effectiveness. Any use of the condom for other than vaginal intercourse can increase the potential of damage to the condom. For maximum benefits, it is important to follow the instructions for use that accompany the product. Failure to do so may result in the loss of the benefits of a condom. During intimate contact, lesions and various body fluids can transmit sexually transmitted diseases (STDs); therefore, a new condom should be applied each and every time before any such contact occurs. The condom with spermicidal lubricant combines the use of a condom with a spermicidal lubricant. The spermicide, nonoxynol-9, reduces the number of active sperm, thereby decreasing the risk of pregnancy if erection is lost before withdrawal and some semen spills outside the condom. However, the extent of decreased risk has not been established. A condom with spermicidal lubricant should not be used as a substitute for the combined use of a vaginal spermicide and a condom.
Device Story
Sheath-style physical barrier device; covers penis during intercourse; prevents exchange of semen, body fluids, and microorganisms. Composed of aliphatic polyurethane; includes reservoir tip and integral ring. Coated with nonoxynol-9 (8%) spermicidal lubricant. Used by consumers during sexual intercourse. Provides physical barrier; nonoxynol-9 reduces active sperm count. Benefits include contraception and reduction of STD/HIV transmission risk. Polyurethane material avoids latex protein sensitization; offers higher tensile strength, break force, and resistance to oil, UV, and oxidation compared to latex.
Clinical Evidence
Consumer use studies with heterosexual couples demonstrated equivalent performance to predicate latex condoms regarding breakage and slippage. Viral permeability study confirmed equivalent barrier properties for preventing transmission of small viral particles.
Technological Characteristics
Aliphatic polyurethane membrane; 59 mm width, 180 mm length; reservoir tip; integral ring. Nonoxynol-9 (8%) spermicidal lubricant. Higher tensile strength and break force than latex; resistant to oil, UV, and oxidation. No latex proteins.
Indications for Use
Indicated for use by individuals for contraception and the prevention of transmission of HIV and other sexually transmitted diseases (STDs).
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
Predicate Devices
- TROJAN® Latex Condoms