← Product Code [IPF](/productcode/IPF) · K955666

# EGS4000 (K955666)

_S&P Electrical Industries, Inc. · IPF · Apr 15, 1997 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K955666

## Device Facts

- **Applicant:** S&P Electrical Industries, Inc.
- **Product Code:** [IPF](/productcode/IPF.md)
- **Decision Date:** Apr 15, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5850
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

Powered Muscle Stimulator Device: 1. Relaxation of Muscle Spasm. 2. Prevention or retardation of disuse atrophy. 3. Increasing local blood circulation. 4. Muscle re-education. 5. Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis. 6. Maintaining or increasing range of motion. TENS Device: Symptomatic relief of chronic intractable pain. CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.

## Device Story

EGS4000 is a portable, battery-operated high-voltage pulsed galvanic stimulator (neuromuscular stimulator) used for muscle stimulation and TENS therapy. It delivers monophasic pulsatile current to patients via electrodes. Operated by clinicians or patients under practitioner orders, the device features dual channels, independent intensity potentiometers, a timer, and multiple switching rates. It can be powered by a 9V battery or an optional AC adapter. A data port allows for recording elapsed usage time. The device replicates the waveform, amplitude, duration, and frequency characteristics of its predicate. It is used to manage muscle conditions and chronic pain. The output is controlled by the user to achieve therapeutic stimulation, aiding in muscle rehabilitation and pain management.

## Clinical Evidence

No clinical data.

## Technological Characteristics

High-voltage pulsed galvanic stimulator (monophasic pulsatile current). Features dual-channel output, independent intensity potentiometers, timer, and multiple switching rates. Powered by 9V battery or optional AC adapter. Includes data port for elapsed time tracking. Updated lead wire receptacles to meet safety standards. Dimensions and weight are reduced compared to the predicate.

## Regulatory Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Predicate Devices

- EGS 300 ([K790847](/device/K790847.md), K802107A, [K832371](/device/K832371.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

EXHIBIT #1
Page 1 of 3
K955646
April 15, 1997

# 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: _______________.

## 1. Submitter's Identification:

S&amp;P Electrical Industries, Inc.
D/B/A Electro-Med Health Industries
11601 Biscayne Boulevard Suite 200A
Miami, Florida 33181

## Date Summary Prepared:

December 11, 1995

## 2. Name of the Device:

EGS4000 (High Voltage Pulsed Galvanic Stimulator)

## 3. Predicate Device Information:

EGS 300 (Portable High Voltage Pulsed Galvanic Stimulator)

Specification Developer:

S&amp;P Electrical industries, Inc. D/B/A Electro-Med Health Industries

510(k) Numbers:

- K790847 -- Original submission-
Substantial equivalence established

- K802107A -- Addition of TENS language and treatment of pain-
Substantial equivalence established
December 4, 1981 letter from Office of Compliance permitting reference to pain.

- K832371 -- Dental submission-
Substantial equivalence again established and labeling accepted.

{1}

Exhibit #1
Page 2 of 3

## 4. Device Description:

The EGS4000 is a Powered Muscle Stimulator and TENS device. It is a small battery operated device, as its predecessor, the EGS 300. The additional operational features include a second channel, a timer, additional switching rates, a change in lead wire receptacles -- to comply with the newest proposal on lead wire attachment. Additionally, independent intensity potentiometers are used. To enhance the versatility of the predicate this unit may also operate with an a/c adaptor. For recording time elapsed in usage, a data port allows for accessing the information. These last two features are performed with all the appropriate mechanisms to ensure patient safety.

The waveform, intensity available and pulse rates available are all as in the predicate device. Battery operation of the EGS4000 is performed with a 9 volt battery as opposed to the previous special 14.4 volt battery. This will allow the end user greater convenience while additionally allowing longer operating time with a single battery.

## 5. Intended Use:

### Powered Muscle Stimulator Device:

1. Relaxation of Muscle Spasm.
2. Prevention or retardation of disuse atrophy.
3. Increasing local blood circulation.
4. Muscle re-education.
5. Immediate postsurgical stimulation of calf muscles to prevent venous thrombosis.
6. Maintaining or increasing range of motion.

### TENS Device:

Symptomatic relief of chronic intractable pain.

**CAUTION:** Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.

## 6. Comparison to Predicate Devices:

The predicate device, the EGS 300, is a battery operated "high voltage pulsed galvanic stimulator", also known as a neuromuscular stimulator or monophasic pulsatile current. This device replicates the waveform, in amplitude, duration and frequencies available. The predicate device uses knobs

Page 2 of 3

{2}

EXHIBIT #1
Page 3 of 3

(potentiometers) for patient controls. This device does the same. The predicate device utilizes a switch to change polarity of the active electrode, this device does the same.

The additional features on the EGS4000 are an additional output channel, a timer, a separate power On-Off switch, 2 additional switching rates, a change in lead wire receptacles--to comply with the newest proposal on lead wire attachment, operation on a 9V battery as opposed to a 14.4V dedicated battery, optional a/c adaptor and an elapsed time reading mechanism.

The obvious difference is one-third the weight and less than one-half the size.

7. **Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:**

The development has been with oscilloscope comparisons for waveforms and its characteristics. Timing was verified with timing checks as well as oscilloscope comparisons. Calculations for pulse charge, current density and power density were accomplished.

8. **Discussion of clinical tests performed:**

Non-Applicable

9. **Conclusion:**

The EGS4000 is presented for approval to meet today’s changes in technology as well as newer FDA requirements for cables and lead wires. The changes and upgrades to the device meet both the FDA requirements and some market requirements. Function and output have not changed. Indications, contraindications and precautions are all updated.

86P ELECTRICAL INDUSTRIES, INC.
D/B/A ELECTRO-MED HEALTH INDUSTRIES

By: Phyllis Lehman, President

DATE: December 11, 1995

Page 3 of 3

---

**Source:** [https://fda-staging.innolitics.com/device/K955666](https://fda-staging.innolitics.com/device/K955666)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).