FLUID REISISTANT EYESHIELD MASK
Device Facts
| Record ID | K955557 |
|---|---|
| Device Name | FLUID REISISTANT EYESHIELD MASK |
| Applicant | American Threshold Healthcare |
| Product Code | FXX · General, Plastic Surgery |
| Decision Date | Apr 25, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4040 |
| Device Class | Class 2 |
Indications for Use
This product is intended for use in infection control practices to minimize contamination caused by exhaled microorganisms and reduce the potential exposure of the wearer to blood and body fluids.
Device Story
Fluid Resistant Eyeshield Mask provides barrier protection for healthcare workers. Device consists of three-layer construction: spunbonded polyolefin or cellulosic outer/inner facings with meltblown polyolefin filter media. Layers are sonically welded; includes malleable aluminum nose wire for fit and optical quality eyeshield attached via ultrasonic welding. Worn by clinicians in surgical or clinical settings to reduce exposure to blood, body fluids, and exhaled microorganisms. Device filters particles at >95% efficiency (1.0 and 0.1 micron) and maintains breathability (average pressure drop ≤ 5.0 mm water).
Clinical Evidence
Bench testing only. Testing included: 1) Water Resistance (Impact Penetration) per Nelson Labs SOP#MBG/037; 2) Differential Pressure (Delta P) per Nelson Labs SOP#ARO/006; 3) Latex Particle Challenge per Nelson Labs SOP#ARO/011. Results: Average Delta P 2.19-2.54 mm water; Filtration efficiency >98% for 0.1 and 1.0 micron particles.
Technological Characteristics
Materials: Spunbonded polyolefin, cellulosic nonwoven, meltblown polyolefin filter media, aluminum nose wire. Construction: Three-layer laminate, sonically welded edges. Form factor: Mask with integrated optical quality eyeshield. Performance: >95% particle filtration efficiency, ≤ 5.0 mm water pressure drop.
Indications for Use
Indicated for use by healthcare personnel to minimize contamination from exhaled microorganisms and reduce exposure to blood and body fluids during surgical or clinical procedures.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Reference Devices
- American Threshold Surgical Mask (K801036)