U.S. ENDOSCOPY PULL PEG KIT & U.S. ENDOSCOPY GUIDEWIRE PEG KIT
K955248 · United States Endoscopy Group, Inc. · KNT · May 21, 1996 · Gastroenterology, Urology
Device Facts
Record ID
K955248
Device Name
U.S. ENDOSCOPY PULL PEG KIT & U.S. ENDOSCOPY GUIDEWIRE PEG KIT
Applicant
United States Endoscopy Group, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
May 21, 1996
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Percutaneous endoscopic gastrostomy (PEG) has become the method of choice for patients, both adults and children, unable to take nutrition through normal means, yet, have a functional gastrointestinal tract. Percutaneous endoscopic gastrostomy is a less invasive and traumatic method of placement than a surgical procedure.
Device Story
U.S.E. PULL PEG and GUIDEWIRE PEG Trays provide necessary components for percutaneous endoscopic gastrostomy. Trays include silicone feeding tubes, bolsters, feeding adapters, and accessories (lubricating jelly, disposable hemostat, grasping snare). Used by clinicians to establish enteral access in patients with functional GI tracts but impaired oral intake. Procedure involves endoscopic placement of feeding tube through abdominal wall into stomach. Benefits include less invasive/traumatic alternative to surgical gastrostomy. Device performance verified against predicate specifications.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by literature review of established percutaneous endoscopic gastrostomy procedures and adherence to performance specifications.
Technological Characteristics
Components include silicone feeding tubes, bolsters, and adapters. Manufacturing utilizes quality assurance procedures based on MIL-STD-9858. Materials are certified to standards appropriate for medical use.
Indications for Use
Indicated for adults and children unable to take nutrition orally but possessing a functional gastrointestinal tract. Contraindications: massive ascites, sepsis, esophageal/gastric obstructions, esophageal/gastric varices, morbid obesity, inability to transluminate the stomach, or combative patients who cannot be sedated.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Submission Summary (Full Text)
{0}
K955248
MAY 21 1996
# SECTION II
## SUMMARY AND CERTIFICATION
### SAFETY AND EFFECTIVENESS
#### PULL PEG TRAY SUMMARY
The Summary of Safety and Effectiveness on percutaneous endoscopic gastrostomy and the Pull PEG tray used reflects data available and presented at the time the submission was prepared, but, caution should be exercised in interpreting the data. The results of future studies may require alterations of the conclusions or recommendations set forth.
## Procedure/Product Overview
Percutaneous endoscopic gastrostomy (PEG) has become the method of choice for patients, both adults and children, unable to take nutrition through normal means, yet, have a functional gastrointestinal tract. Percutaneous endoscopic gastrostomy is a less invasive and traumatic method of placement than a surgical procedure.
The placement of PEGs, via endoscopy, has been clinical acceptable for more than 10 years. Several manufacturers provide PEG units to the market.
A U.S.E. PULL PEG Tray includes all necessary products (eg. silicone feeding tube, bolster, feeding adapter, etc.) and accessory products (eg. lubricating jelly, disposable hemostat, grasping snare) to complete the endoscopic gastrostomy.
## Contraindications For Utilizing a PEG
The following are contraindications for percutaneous endoscopic gastrostomy.
1. Patients with massive ascites, sepsis, esophageal or gastric obstructions, esophageal or gastric varices and morbid obesity.
2. If the patient’s stomach can not be transluminated.
3. If the patient is combative and can not be sedated.
## Manufacturing Overview
U.S.E. manufactures and tests the product to performance specifications based on predicate and/or substantially equivalent devices.
U.S.E. manufacturing processes and procedures are based on good manufacturing practices. Quality assurance methods and procedures based on MIL-STD-9858 are utilized to assure conformance to design
{1}
specifications.
Materials used in the manufacturing process are certified to standards appropriate for their use.
## Bibliography
Mamel, J.J.: Percutaneous Endoscopic Gastrostomy: A Review. *American Society of Parenteral & Enteral Nutrition* 87: 65, 1987.
Mellinger, J.D. and Ponsky, J.L.: Percutaneous Endoscopic Gastrostomy. *Endoscopy* 24: 64-67 Jan-Feb 1992.
Ponsky, J.L. and Gauderer, M.: Percutaneous Endoscopic Gastrostomy: A Nonoperative Technique for Feeding Gastrostomy. *Gastrointestinal Endoscopy* 27: 9-11, 1981.
Ponsky, J.L., Gauderer, M. and Stellato, T.A.: Percutaneous Endoscopic Gastrostomy in Children. *Gastrointestinal Endoscopy* 27: 128, May 1981.
Ponsky, J.L., Gauderer, M. and Stellato, T.A.: Percutaneous Endoscopic Gastrostomy: Review of 150 Cases. *Archives of Surgery* 118: 913-914, August 1983
Ponsky, J.L., Gauderer, M., Stellato, T.A. and Aszodi, A.: Percutaneous Approaches to Enteral Alimentation. *The American Journal of Surgery* 149: 102-105, January 1985.
Ponsky, J.L.: Percutaneous Endoscopic Gastrostomy - A Closer Look. *Endoscopy Review* 1: 24-29, December 1984.
Starkey J.F., Jefferson P.A., Kirby, D.F.: Taking Care of Percutaneous Endoscopic Gastrostomy. *American Journal of Nursing* 88: 42, 1988.
{2}
# SAFETY AND EFFECTIVENESS
## GUIDEWIRE PEG TRAY SUMMARY
The Summary of Safety and Effectiveness on percutaneous endoscopic gastrostomy and the Guidewire PEG tray used reflects data available and presented at the time the submission was prepared, but, caution should be exercised in interpreting the data. The results of future studies may require alterations of the conclusions or recommendations set forth.
## Procedure/Product Overview
Percutaneous endoscopic gastrostomy (PEG) has become the method of choice for patients, both adults and children, unable to take nutrition through normal means, yet, have a functional gastrointestinal tract. Percutaneous endoscopic gastrostomy is a less invasive and traumatic method of placement than a surgical procedure.
The placement of PEGs, via endoscopy, has been clinical acceptable for more than 10 years. Several manufacturers provide PEG units to the market.
A U.S.E. GUIDEWIRE PEG Tray includes all necessary products (eg. silicone feeding tube, bolster, feeding adapter, etc.) and accessory products (eg. lubricating jelly, disposable hemostat, grasping snare) to complete the endoscopic gastrostomy.
## Contraindications For Utilizing a PEG
The following are contraindications for percutaneous endoscopic gastrostomy.
1. Patients with massive ascites, sepsis, esophageal or gastric obstructions, esophageal or gastric varices and morbid obesity.
2. If the patient’s stomach can not be transluminated.
3. If the patient is combative and can not be sedated.
## Manufacturing Overview
U.S.E. manufactures and tests the product to performance specifications based on predicate and/or substantially equivalent devices.
U.S.E. manufacturing processes and procedures are based on good manufacturing practices. Quality assurance methods and procedures based on MIL-STD-9858 are utilized to assure conformance to design specifications.
Materials used in the manufacturing process are certified to
{3}
standards appropriate for their use.
**Bibliography**
Mamel, J.J.: *Percutaneous Endoscopic Gastrostomy: A Review*. *American Society of Parenteral & Enteral Nutrition* 87: 65, 1987.
Mellinger, J.D. and Ponsky, J.L.: *Percutaneous Endoscopic Gastrostomy*. *Endoscopy* 24: 64-67, Jan-Feb 1992.
Ponsky, J.L. and Gauderer, M.: *Percutaneous Endoscopic Gastrostomy: A Nonoperative Technique for Feeding Gastrostomy*. *Gastrointestinal Endoscopy* 27: 9-11, 1981.
Ponsky, J.L., Gauderer, M. and Stellato, T.A.: *Percutaneous Endoscopic Gastrostomy in Children*. *Gastrointestinal Endoscopy* 27: 128, May 1981.
Ponsky, J.L., Gauderer, M. and Stellato, T.A.: *Percutaneous Endoscopic Gastrostomy: Review of 150 Cases*. *Archives of Surgery* 118: 913-914, August 1983.
Ponsky, J.L., Gauderer, M., Stellato, T.A. and Aszodi, A.: *Percutaneous Approaches to Enteral Alimentation*. *The American Journal of Surgery* 149: 102-105, January 1985.
Ponsky, J.L.: *Percutaneous Endoscopic Gastrostomy - A Closer Look*. *Endoscopy Review* 1: 24-29, December 1984.
Starkey J.F., Jefferson P.A., Kirby, D.F.: *Taking Care of Percutaneous Endoscopic Gastrostomy*. *American Journal of Nursing* 88: 42, 1988.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.