KSEA THERMOFLATOR (MODEL 26 4320 20)

K955073 · KARL STORZ Endoscopy-America, Inc. · HIF · Dec 6, 1996 · Obstetrics/Gynecology

Device Facts

Record IDK955073
Device NameKSEA THERMOFLATOR (MODEL 26 4320 20)
ApplicantKARL STORZ Endoscopy-America, Inc.
Product CodeHIF · Obstetrics/Gynecology
Decision DateDec 6, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.1730
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is designed to carefully deliver large flow rates for rapid insufflation, and monitor the amounts of CO₂ gas needed to establish and maintain proper distention of the peritoneal cavity during Ob/Gyn laparoscopic surgical and diagnostic procedures

Device Story

KSEA Thermoflator (model 26 4320 20) is a microprocessor-based laparoscopic insufflator; delivers CO₂ gas to peritoneal cavity; user-adjustable flow rates (0-30 L/min) and pressure (0-30 mmHg); used in clinical settings by surgeons; provides rapid insufflation and maintenance of distention; includes pressure and flow-sensing technology; incorporates safety features to monitor gas delivery; assists surgeons in maintaining surgical field visibility.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Microprocessor-based system; CO₂ gas delivery; pressure and flow-sensing technology; user-adjustable flow (0-30 L/min) and pressure (0-30 mmHg) controls.

Indications for Use

Indicated for patients undergoing Ob/Gyn laparoscopic surgical and diagnostic procedures requiring peritoneal cavity distention via CO₂ insufflation.

Regulatory Classification

Identification

A laparoscopic insufflator is a device used to facilitate the use of the laparoscope by filling the peritoneal cavity with gas to distend it.

Submission Summary (Full Text)

{0} Karl Storz Endoscopy-America, Inc. 600 Corporate Pointe Culver City, California 90230-7600 Phone 310 558 1500 Toll Free 800 421 0837 Fax 310 410 5527 # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CRF 807.92. All data included in this documents are accurate and complete to the best of KSEA's knowledge. **Applicant:** Karl Storz Endoscopy - America, Inc. 600 Corporate Pointe Culver City, CA 90230 (310) 558-1500 DEC - 6 1996 **Contact:** Betty M. Johnson Manager, Regulatory Affairs **Device Identification:** Common Name Laparoscopic Insufflator Trade Name (optional) KSEA Thermoflator (model 26 4320 20) **Indication:** This device is designed to carefully deliver large flow rates for rapid insufflation, and monitor the amounts of CO₂ gas needed to establish and maintain proper distention of the peritoneal cavity during Ob/Gyn laparoscopic surgical and diagnostic procedures **Device Description:** The KSEA Thermoflator (model 26 4320 20) is a microprocessor-based system, with gas output that is user adjustable from 0-30 liters/minute, and pressure that is user adjustable from 0-30 mmHg. The unit uses contemporary technology in pressure and flow-sensing, and incorporates a wide variety of safety features. **Substantial Equivalence:** The KSEA Thermoflator (model 26 4320 20) is substantially equivalent to the predicate devices, since the basic features, design and intended uses are the same. The minor differences between the KSEA Thermoflator (model 26 4320 20) and the predicate device raise no new issues of safety and effectiveness, as these design differences have no effect on the performance, function or intended use of the devices. Signed: Betty M. Johnson Manager, Regulatory Affairs 00173
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...