MEDI-TECH EDDY INTRAVASCULAR INFUSION CATHETER

K954457 · Medi-Tech, Inc. · KRA · May 15, 1996 · Cardiovascular

Device Facts

Record IDK954457
Device NameMEDI-TECH EDDY INTRAVASCULAR INFUSION CATHETER
ApplicantMedi-Tech, Inc.
Product CodeKRA · Cardiovascular
Decision DateMay 15, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1210
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EDDY™ Intravascular Infusion Catheter is intended for general Intravascular use, including neuro, peripheral, and coronary vasculature. The catheter is intended to be flow tracked in order to access distal tortuous vasculature. Once the subselective region has been accessed, the catheter can be used for the controlled and selective infusion of diagnostic, embolic, or therapeutic materials into vessels. Diagnostic, embolic, or therapeutic agents to be used in accordance with specifications outlined by the manufacturer.

Device Story

Single-lumen intravascular infusion catheter; flow-tracked or used with steerable hydrophilic guidewires (≤.016"). Features: hydrophilic coating, radiopaque markers, steam-shapeable distal tip, standard luer adapter. Used by physicians in clinical settings to access distal vasculature; facilitates targeted delivery of diagnostic/therapeutic agents. Benefits: enables subselective vessel access; allows controlled infusion. Disposable, single-use device.

Clinical Evidence

Bench testing included shaft burst pressure, tensile strength, infusion rate, tip flexibility, torque testing, and coating durability/lubricity. Biocompatibility of blood-contacting materials verified. In vivo studies performed to confirm safety and performance.

Technological Characteristics

Single-lumen catheter; two shaft sizes (2.7F/2.2F and 2.4F/1.7F); lengths 135-165cm. Hydrophilic coating on distal shaft. Radiopaque markers for fluoroscopy. Steam-shapeable distal tip. Standard luer adapter. Sterile, single-use.

Indications for Use

Indicated for patients requiring access to distal tortuous vasculature in neuro, peripheral, and coronary vessels for the controlled infusion of diagnostic, embolic, or therapeutic agents.

Regulatory Classification

Identification

A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.

Predicate Devices

Submission Summary (Full Text)

{0} K954457 MAY 15 1996 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (Pursuant to Section 12, Safe Medical Devices Act of 1990) EDDY™ Intravascular Infusion Catheter 1. The trade or proprietary name of the device is Medi-tech EDDY™ Intravascular Infusion Catheter. The common or usual name is Continuous Flush Catheter. 2. The EDDY™ Intravascular Infusion Catheter is intended for general Intravascular use, including neuro, peripheral, and coronary vasculature. The catheter is intended to be flow tracked in order to access distal tortuous vasculature. Once the subselective region has been accessed, the catheter can be used for the controlled and selective infusion of diagnostic, embolic, or therapeutic materials into vessels. Diagnostic, embolic, or therapeutic agents to be used in accordance with specifications outlined by the manufacturer. The device is a single lumen device which facilitates distal vascular access. The catheter is available in two shaft sizes. The larger shaft size models incorporate a 2.7F proximal shaft with a 2.2F flexible distal shaft and a distal internal diameter of .019". The smaller shaft size models incorporate a 2.4F proximal shaft with a 1.7F flexible distal shaft and a distal internal diameter of .013". The catheter is available in usable shaft lengths from 135 to 165cm, in 5cm increments. The catheter is intended to be flow tracked, but may accommodate steerable hydrophilic guidewires of ≤.016" for the larger shaft size models. The EDDY™ is available in straight and bulb tip configurations. Radiopaque markers at the distal tip and the mid-shaft bond facilitate fluoroscopic visualization of the device. The proximal end of the infusion catheter incorporates a standard luer adapter to facilitate the attachment of accessories. The distal end of the catheter is steam shapeable as are other flow tracked catheters currently on the market. A white ink proximal mark is located 100cm from the tip of the catheter to indicate when the stylet is to be removed during insertion of the infusion catheter into the guide catheter. The outer surface of the EDDY™ catheter is coated with a hydrophilic coating system, except for the proximal 30cm of shaft. The device will be provided sterile, and is intended for one procedure use only (disposable). Testing and evaluation included shaft burst pressure, tensile, infusion rate, tip flexibility, torque testing, coating durability and lubricity, and in vivo studies. Testing was performed to assess the biocompatibility of the device's blood contact materials. {1} 3. Test results verified that the Medi-Tech EDDY™ Intravascular Infusion Catheter meets all of the minimum requirements and is deemed adequate for its intended use. The Medi-Tech EDDY Intravascular Infusion Catheter is considered to be “substantially equivalent” to the ZEPHYR® Infusion Catheter currently marketed by Target Therapeutics, Inc., the Balt MAGIC™ Infusion Catheter marketed by Target Therapeutics, Inc., and the Venture II Intravascular Infusion Catheter currently marketed by Medi-Tech.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...