SOLOPICC 24 (60CM) 20GA CATHETER

K954428 · Douglas Medical Products Corp. · FOZ · Mar 28, 1996 · General Hospital

Device Facts

Record IDK954428
Device NameSOLOPICC 24 (60CM) 20GA CATHETER
ApplicantDouglas Medical Products Corp.
Product CodeFOZ · General Hospital
Decision DateMar 28, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5200
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SoloPICC and SoloMidline Peripheral Intravenous catheters are intended to be used for the administration of medications, fluids and nutritional therapy of thirty days or less duration.

Device Story

SoloPICC and SoloMidline are sterile, single-lumen, radiopaque, flexible intravenous catheters; include PVC female Luer lock adapter, stiffening stylet, and TFX T-Peel™ peelable introducer catheter. Used for short-term (≤30 days) administration of medications, fluids, and nutritional therapy. Operated by clinicians in clinical settings. Catheters facilitate venous access; design ensures compatibility with standard infusion practices. Benefit: provides reliable, temporary vascular access for therapeutic delivery.

Clinical Evidence

Bench testing only; no clinical data provided. Sterility assurance level 10^-6 per AAMI guidelines; LAL testing per USP guidelines.

Technological Characteristics

Radiopaque catheter; PVC female Luer lock adapter; stiffening stylet; polyethylene protector; Tyvek/Mylar packaging. Sterilized per AAMI guidelines (10^-6 SAL).

Indications for Use

Indicated for patients requiring administration of medications, fluids, and nutritional therapy via peripheral intravenous access for durations of thirty days or less.

Regulatory Classification

Identification

An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.

Predicate Devices

Submission Summary (Full Text)

{0} Douglas Medical Products A SoloPak® Company K 954428 September 16, 1995 MAR 28 1996 To Whom it may concern: This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act and CFR 807.92: Trade Name - SoloPICC and SoloMidline Peripheral Intravenous Catheter Common Name - PICC, Peripherally Inserted Central Catheter, Flexible Intravenous Catheter Classification Name - Percutaneous Intravascular Catheter The SoloPICC and SoloMidline Peripheral Intravenous catheters are intended to be used for the administration of medications, fluids and nutritional therapy of thirty days or less duration. The SoloPICC and SoloMidline catheters are sterile, non-pyrogenic, single lumen radiopaque catheters with a PVC female Luer lock adapter. A stiffening stylet is inserted into the female adapter and runs the length of the catheter. The catheter is shrouded in a polyethylene protector and packaged in a printed tyvek\mylar peel pouch. A TFX T-Peel™ peelable introducer catheter is included in the tyvek\mylar package for catheter insertion purposes. The processes and materials used to manufacture the SoloPICC and SoloMidline catheters are the same or the equivalent of the predicate devices named in this submission. The named predicate devices in this submission were the Luther Medical Products, Inc. L-Cath® Intravenous Placement Unit (K801575) and the Medical Profiles, Inc. Peripheral Intravenous Catheter (K914378). The SoloPICC and SoloMidline catheters are sterilized per AAMI guidelines to a 10 -s sterility assurance level. Each production lot is LAL tested per USP guidelines. Based on the fact that the SoloPICC and SoloMidline catheters utilize similar and equivalent designs, materials, and components as currently legally marketed products, these products are safe and effective when used as intended. Sincerely, Ron Haselhorst Director of RA\QA Douglas Medical Products 1840 Industrial Drive, Suite 200, Libertyville, IL 60048 • Telephone: 708-949-7387 • Telefax: 708-949-7388
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...