CEP X SPECTRUMORANGE\Y SPECTRUMGREEN DNA PROBE KIT
K954214 · Vysis · OXP · Jan 21, 1997 · Immunology
Device Facts
Record ID
K954214
Device Name
CEP X SPECTRUMORANGE\Y SPECTRUMGREEN DNA PROBE KIT
Applicant
Vysis
Product Code
OXP · Immunology
Decision Date
Jan 21, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.6010
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K954214 · Jan 21, 1997
CEP X SPECTRUMORANGE\Y SPECTRUMGREEN DNA PROBE KIT
Vysis
Archived bone marrow specimens from clinical patients; Historical standard cytogenetic analysis results
The device performance was evaluated by comparing FISH results against standard cytogenetic analysis in archived bone marrow specimens from patients who had undergone opposite-sex and like-sex bone marrow transplants.
Archived specimens; Bone marrow transplant; Comparative study; Clinical specimens
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Multi-center, blinded, controlled, comparative study; Retrospective comparative study using archived clinical specimens
Patients who were recipients of opposite-sex (n=143) and like-sex (n=153) bone marrow transplants; Sample Size: 296 total specimens (143 opposite-sex, 153 like-sex); Number of Sites: 3
Standard cytogenetic analysis (karyotyping)
Relative sensitivity and specificity of FISH in detecting donor cells compared to standard cytogenetic analysis
Indications for Use
The CEP X/Y probe is a combination of CEP X SpectrumOrange and CEP Y SpectrumGreen fluorescently labeled DNA probes for the alpha DNA on the centromeric region of chromosome X and the satellite III DNA at the Yq12 region of chromosome Y. This assay is designed to provide a reliable method for the simultaneous detection and enumeration of chromosomes X and Y in both interphase nuclei and metaphase spreads in bone marrow by fluorescence in situ hybridization (FISH).
Device Story
Device uses fluorescently labeled DNA probes (CEP X SpectrumOrange, CEP Y SpectrumGreen) to target centromeric/satellite DNA regions. Input: bone marrow specimens (unstimulated, cultured). Process: Fluorescence in situ hybridization (FISH) allows visualization of X and Y chromosomes. Output: visual signal enumeration in interphase nuclei and metaphase spreads. Used in clinical laboratories by cytogeneticists/technicians. Healthcare providers use results to assess donor cell engraftment post-bone marrow transplant. Interpretation requires integration with standard cytogenetic analysis and clinical context. Benefits: provides reliable method for detecting sex chromosome chimerism.
Clinical Evidence
Multi-center, blinded, controlled study of 143 opposite-sex bone marrow transplant (BMT) patients (69 CML, 30 AML/ANLL, 7 MDS, 21 ALL, 16 HDNOS). FISH interphase analysis showed 100% relative sensitivity (143/143) for donor cell detection. Specificity assessed in 153 like-sex BMT patients; interphase FISH correctly identified 149/153 (97.4%) as negative. False positives occurred in male recipients, particularly those with abnormal karyotypes (e.g., 46,XY,-Y,+X).
Technological Characteristics
Fluorescently labeled DNA probes targeting chromosome X centromere and Yq12 region. Principle: Fluorescence in situ hybridization (FISH). Form factor: DNA probe kit. Analysis: Visual enumeration of signals in interphase nuclei and metaphase spreads. No software or automated algorithm described.
Indications for Use
Indicated for the detection and enumeration of X and Y chromosomes in bone marrow specimens via FISH, specifically for monitoring donor cell engraftment in patients who have undergone opposite-sex bone marrow transplantation.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
Submission Summary (Full Text)
{0}
K954214
JAN 21 1997
# SUMMARY: SAFETY AND EFFECTIVENESS INFORMATION FOR CEP X SpectrumOrange/CEP Y SpectrumGreen DNA Probe Kit
The CEP X/Y probe is a combination of CEP X SpectrumOrange and CEP Y SpectrumGreen fluorescently labeled DNA probes for the alpha DNA on the centromeric region of chromosome X and the satellite III DNA at the Yq12 region of chromosome Y. This assay is designed to provide a reliable method for the simultaneous detection and enumeration of chromosomes X and Y in both interphase nuclei and metaphase spreads in bone marrow by fluorescence *in situ* hybridization (FISH).
Standard cytogenetic analysis detects the presence of the X and Y chromosomes by karyotyping metaphase spreads after staining the chromosomes with a dye in cultured tissue cells.
Safety and effectiveness issues relevant to FISH assays such as the CEP X/Y assay may include cross-reactivity, poor sensitivity, poor specificity, or poor reproducibility.
# Analytical Sensitivity and Specificity
## Hybridization Efficiency
In a pivotal study, the average percentage of cells with only one hybridization signal was 0.012% (S.D.=0.15%) on 143 bone marrow specimens. Thus, <2% cells with only one signal is a realistic standard of acceptance.
## Analytical Sensitivity
The analytical sensitivity of the CEP X/Y probe was tested in the reproducibility study described below. In that study, the 0% XY specimen was estimated with a mean of 0.00% (s.d.=0.00%) XY nuclei and the 1% XY specimen, 0.94% (s.d.=0.32%). The 0% XX specimen was estimated with a mean of 0.00% (s.d.=0.00%) XX nuclei and the 1% XX specimen, 0.95% (s.d.=0.34%). There was little overlap between the 0% and 1% specimens; the lower 95% confidence limit for the 1% specimen was 0.31% and 0.28% for XY and XX, respectively. Thus, the limit of detection for CEP X/Y is estimated to be 1.0%.
## Analytical Specificity
Locus specificity studies were performed with metaphase spreads according to standard Vysis QC protocols. A total of 65 metaphase spreads were examined sequentially by G-banding to identify chromosomes X and Y, followed by FISH. No cross-hybridization to other chromosome loci was observed in any of the 65 cells examined; hybridization was limited to the centromere of chromosome X and the Yq12 region of chromosome Y.
Safety and Effectiveness Statement
Page 1 of 4
SISAN 870212802
{1}
# Reproducibility
To assess the reproducibility of the CEP X/Y interphase analysis for the percentage of cells with XX and XY signals, bone marrow specimens with approximately 0%/100%, 1%/99%, 5%/95%, 95%/5% XY/XX, 99%/1% and 100%/0% XY/XX were prepared. Inter-site, inter-lot, inter-day, and inter-observer reproducibility were assessed in a pivotal study with two of these bone marrow specimen mixtures (approximately 99%/1% and 100%/0% XY/XX,) and two mixtures of hematologically derived human cells with approximately 0%/100% and 1%/99% XY/XX. The percentage of cells with XX and XY signals were evaluated according to the instructions for signal enumeration in the package insert. Using ANOVA, significant, site-to-site and observer-to-observer variations were observed, reflecting the subjectivity of the visual enumeration process. In addition to the pivotal study, four bone marrow specimens with approximately 0%/100%, 1%/99%, 5%/95% and 95%/5% XY/XX were prepared and analyzed at one site. The mean, standard deviation, and percent CV of the observed percentage of XX and XY nuclei for the bone marrow specimens are shown in Table 1.
Table 1 Precision of the Observed % XY/XX Signaled Nuclei Detection
| Specimen Level of XY/XX: | n | Mean (%) | | Standard Deviation (%) | | Coefficient of Variation (%) | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | XY | XX | XY | XX | XY | XX | |
| 0% | 100% | 10 | 0.00 | 97.4 | 0.00 | 1.18 | — | 1.21 |
| 1% | 99% | 20 | 0.88 | 97.2 | 0.48 | 2.00 | 54.8 | 2.06 |
| 5% | 95% | 20 | 4.90 | 94.9 | 0.99 | 0.99 | 20.2 | 1.04 |
| 95% | 5% | 10 | 95.0 | 4.96 | 1.60 | 1.60 | 1.68 | 32.3 |
| 99% | 1% | 24 | 98.3 | 0.95 | 0.41 | 0.34 | 0.41 | 36.3 |
| 100% | 0% | 24 | 99.0 | 0.00 | 0.47 | 0.00 | 0.48 | — |
Safety and Effectiveness Statement
Page 2 of 4
20:01 26.91 N# 90d 22L
SISN 8p0212802
{2}
# Methods Comparison: Clinical Specimens
A multi-center, blinded, controlled, comparative study was conducted to characterize the performance of the CEP X/Y DNA probe kit in identifying the proportion of XX and XY cells, relative to standard cytogenetic analysis, in recipients of opposite-sex bone marrow transplants (BMT). Archived bone marrow specimens, which were previously evaluated by standard cytogenetic analysis, were selected from a total of 143 patients 72 males and 71 females), who were the recipients of opposite-sex BMTs. Consecutive specimens were selected and evaluated at three sites; site 1 provided and analyzed 40 specimens; site 2, 52 specimens; and site 3, 51 specimens. These specimens were derived from patients with one of the following diagnoses.
1. Chronic myelogenous leukemia (CML): 69 specimens
2. Acute myeloid leukemia (AML) or Acute nonlymphocytic leukemia (ANLL): 30 specimens
3. Myelodysplastic syndrome (MDS): 7 specimens
4. Acute lymphoid leukemia (ALL): 21 specimens
5. Hematological disorder not otherwise specified, but in which cytogenetics are commonly requested (HDNOS): 16 specimens
All sites utilized unstimulated, cultured specimens for both standard cytogenetic and FISH analyses. Each site followed its own in-house protocol for standard cytogenetic analysis; FISH analyses were performed according to the instructions in the CEP X/Y DNA probe kit package insert. The number of donor and recipient cells were enumerated by FISH in a minimum of 20 metaphase and 500 interphase cells.
As expected for specimens with presumed sex chromosome chimerism after opposite-sex BMT, donor cells were detected in each of the 143 specimens by standard cytogenetic analysis. Interphase FISH analysis designated 143/143 specimens as positive for the presence of donor cells (100% relative sensitivity). FISH metaphase analysis detected donor cells in 141/141* specimens (100% relative sensitivity).
In addition to assessing the performance of FISH in the target population of patients with opposite-sex BMT, the ability of interphase and metaphase FISH to correctly designate specimens with like-sex BMT as negative was assessed in 153 patients with like-sex BMTs; the distribution of diagnoses for these patients was similar to those with opposite-sex BMTs. FISH interphase analysis correctly designated 149/153 (97.4%) as negative. All of the four false positive cases occurred in male recipients of like-sex BMT. One case had a 46,XY,-Y,+X karyotype, which led to a FISH result of 37.4% of cells with XX signals; the FISH results of the other three cases showed low levels of XX cells (4.6%, 1.6%, and 0.8%). FISH metaphase analyses designated 151/153 (98.7%) as negative. Both false positive cases were the same patients as those with discrepant FISH interphase analysis. One case had a 46,XY,-Y,+X karyotype, which led to a FISH result of 20% of cells with XX signals; the FISH results of the other case showed 7.1% XX cells.
Safety and Effectiveness Statement
Page 3 of 4
80:16.91 N51 20d 23L
SISAN 8p021L2802
{3}
The misclassification of a like-sex BMT recipient with an abnormal acquired karyotype demonstrates the importance of performing pre-BMT cytogenetic analysis in conjunction with FISH. The other 3 false positive cases by FISH had low levels of XX cells; both recipient and donor cells showed a 46,XY karyotype. Although no females with like-sex BMT cases were misclassified by FISH, low levels of donor/recipient cells by FISH should be interpreted with caution. All FISH results should be interpreted in conjunction with standard cytogenetic analysis and within the context of other relevant clinical information.
## Conclusions
The performance of CEP X/Y is supported by the Vysis Quality Control Procedures and is demonstrated in clinical studies. When the CEP X SpectrumOrange / CEP Y SpectrumGreen DNA Probe is used as instructed in the package insert, the above statements describe its performance.
Safety and Effectiveness Statement
80:16,91 Nf 80d 2εL
SISN 8p0212802
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
Vysis
c/o Ms. Vicki Anastasi
Manager, Regulatory Affairs
3100 Woodcreek Dr.
Downers Grove, IL 60515
Re: k954214
Trade/Device Name: CEP X Spectrum Orange/Y Spectrum Green DNA Probe Kit
Regulation Number: 21 CFR §866.6010
Regulation Name: Tumor-associated antigen immunological test system
Regulatory Class: II
Product Code: OXP, KIR
Dated: November 19, 1996
Received: November 20, 1996
Dear Ms. Anastasi:
This letter corrects our substantially equivalent letter of January 21, 1997.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and
{5}
Page 2 – Ms. Vicki Anastasi
809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Maria M. Chan, Ph.D.
Director
Division of Immunology and Hematology Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
Center for Devices and Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.