RETROSPECTIVE II-ELITE ECG MEMORY LOOP RECORDER

K954163 · Cardiac Evaluation Center, Inc. · DXH · Jul 11, 1996 · Cardiovascular

Device Facts

Record IDK954163
Device NameRETROSPECTIVE II-ELITE ECG MEMORY LOOP RECORDER
ApplicantCardiac Evaluation Center, Inc.
Product CodeDXH · Cardiovascular
Decision DateJul 11, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2920
Device ClassClass 2

Indications for Use

The sole intended use for the device or its predicate is for the long term monitoring of ECG in occasionally symptomatic patients. The purpose of both devices is to record occasional short-lived arrhythmia. Their usage is usually confined to otherwise healthy individuals on an out-patient basis. These recorders are not meant for use in life support or ICU situations. Their normal function is to record all ECG data and save a particular episode only when activated by the patient. The stored episode contains data before and after the activation so that the actual arrhythmia will be captured. The data is then saved for transmission to a qualified technician for interpretation. The recorders take no action and advises none to the patient other than to call the receiving station. They are totally passive devices and offers no diagnosis or perform any signal analysis.

Device Story

Portable ECG memory loop recorder; inputs ECG signals via 3 or 5 disposable surface electrodes. Microprocessor amplifies and digitizes signals; stores data in RAM. Patient-activated button saves recent data; send button triggers acoustic frequency modulation (FM) transmission over voice-grade telephone lines to receiving hardware. Used by patients in outpatient settings; data interpreted by qualified technicians. Features include simultaneous two-channel transmission, automatic baseline centering, multiple event storage, and digital time/date stamps. Device is passive; provides no diagnosis or signal analysis. Benefits include capturing transient arrhythmias for physician review.

Clinical Evidence

Bench testing only. Verification testing confirmed compliance with AAMI performance requirements for ambulatory ECG monitors.

Technological Characteristics

Solid-state design; no magnetic media or moving parts. Powered by alkaline or carbon-zinc batteries; no AC connection. Acoustic coupling for telephone transmission. 3 or 5 disposable surface electrodes. Conforms to AAMI specifications for ambulatory ECG monitors.

Indications for Use

Indicated for long-term ECG monitoring in occasionally symptomatic, otherwise healthy out-patients to record short-lived arrhythmias. Not for use in life support or ICU settings.

Regulatory Classification

Identification

A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.

Predicate Devices

Submission Summary (Full Text)

{0} K954163 U JUL 11 1995 # 510(K) SUMMARY ## Retrospective II-Elite CEC Medical Systems, Inc. 2063 South 116th Street Milwaukee, WI 53227 Voice: (414) 545-6777 FAX: (414) 545-4126 Date of Summary: August 25, 1995 Contact for this submission: John Ellenz, Director of Engineering | Device Trade Name: | Retrospective II-Elite | | --- | --- | | Device Common Name: | ECG Memory Loop Recorder | | Device Classification: | Transmitters and Receivers, Electrocardiographic, Telephone. | | Classification Number: | 74DXH | Claimed Predicate Device: Retrospective ECG Memory Loop Recorder Manufacturer: CEC Medical Systems 510(K) Document #: K831406 ### Description of Device The Retrospective II is a portable ECG memory loop recorder. It is a non-critical Type II device. The device, as well as its predicate, continuously record ECG and store several minutes of the most recent data. When the patient experiences a symptom he/she presses a button which saves this stored data. The patient then calls a receiving service and transmits the data over the telephone. The service is then able to interpret or forward the data to the patient’s physician. Both the Retrospective and the Retrospective II-Elite have this same operational sequence. Both devices obtained ECG data via 3 or 5 disposable surface electrodes. The ECG signals are amplified and then digitized. A microprocessor controls the sampling of the data and then stores the data in RAM. When the symptom button is pressed the most recent stored data is then saved. When the send button is presses the saved data is recalled by the microprocessor. The data is then frequency modulated and used to drive a speaker. This audio data is transmitted over a voice grade telephone line to the receiving hardware. Both devices accomplish this entire process using solid-state File:c:\usr\john\rem\elite\summary.doc PAGE 1/3 08/25/95 8:27 AM {1} techniques, that is no magnetic media or moving parts are employed. Both devices are powered by alkaline batteries and have no AC power connection. ## Intended Use of Device The sole intended use for the device or its predicate is for the long term monitoring of ECG in occasionally symptomatic patients. The purpose of both devices is to record occasional short-lived arrhythmia. Their usage is usually confined to otherwise healthy individuals on an out-patient basis. These recorders are not meant for use in life support or ICU situations. Their normal function is to record all ECG data and save a particular episode only when activated by the patient. The stored episode contains data before and after the activation so that the actual arrhythmia will be captured. The data is then saved for transmission to a qualified technician for interpretation. The recorders take no action and advises none to the patient other than to call the receiving station. They are totally passive devices and offers no diagnosis or perform any signal analysis. ## Summary of Comparison to Predicate Device The Retrospective II-Elite is an evolution of the original Retrospective. Both were designed with the same intended use. Improvements in technology over the past decade allow the Retrospective II-Elite to be smaller, lighter, and have longer battery life than the original device. However the underlying principal of operation is identical for both devices. Both amplify and digitize ECG data, store that data digitally in static RAM, and transmit the data via acoustic FM. Additional features of the Elite model include the simultaneous transmission of two channels of ECG data, automatic baseline centering, multiple events, and digital time/date stamps. These features do not alter the intended use of the device, rather they allow the device to be competitive with others on the market. The quality of transmitted ECG data is not compromised by the additional features. In fact the Elite is designed to conform with ECG quality specifications suggested by The Association for the Advancement of Medical Instrumentation (AAMI) which are more rigorous than those used in the original Retrospective. Both devices are powered by alkaline (or other standard consumer types such as carbon zinc) batteries. There is no provision for the connection of either device to any external supply. No AC risk currents exist. Data is transmitted to a receiving station over normal telephone lines. No electrical connection is made with the telephone system, as the coupling method is acoustic. Therefore the Retrospective II-Elite presents no additional electrical risks to the patient. File:c:\usr\john\rem\elite\summary.doc 08/25/95 8:27 AM {2} # Test Results The Retrospective II-Elite was designed to meet the specifications suggested by the Association for the Advancement of Medical Instrumentation (AAMI) for Ambulatory ECG Monitors. Verification testing on a sample device has shown that the Retrospective II-Elite complies with applicable AAMI performance requirements. File:c:\usr\john\rem\elite\summary.doc PAGE 3/3 08/25/95 8:27 AM
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...