K953116 · Intl. Remote Imaging Systems · GKZ · May 13, 1996 · Hematology
Device Facts
Record ID
K953116
Device Name
WHITE IRIS LEUKOCYTE DIFFERENTIAL ANALYZER
Applicant
Intl. Remote Imaging Systems
Product Code
GKZ · Hematology
Decision Date
May 13, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
The White IRIS is a leukocyte differential analyzer intended for in vitro diagnostic use in determining the proportional leukocyte count (WBC differential) on peripheral blood specimens that have been flagged by an automated hematology analyzer performing differential counts as well as for peripheral blood specimens for which no automated differential has been performed.
Device Story
The White IRIS is an automated leukocyte differential analyzer for clinical laboratory use. It processes whole blood specimens by gravitationally separating leukocytes from erythrocytes in a flow cell. Leukocytes are stained supravitally with 2-Methylpolymethine (IRISpectracolor) to induce metachromasia. A microscope captures images of individual cells, which are processed by color image analysis software. The system sorts cell images into montages on a 20-inch touch-screen monitor based on size and color features. A skilled technologist reviews the 500-cell differential, confirming or modifying the presumptive classifications. This process provides flow cytometry-like precision with microscopic resolution, reducing manual labor and biohazard exposure compared to traditional Wright-stain slide microscopy. The output is a proportional WBC differential count used by physicians as an adjunct to clinical observations for diagnosis and monitoring.
Clinical Evidence
Clinical evaluation included 1,202 patient specimens compared against manual Wright-stain/light microscope reference method (200-cell count). Accuracy studies showed high correlation for major cell types (e.g., Neutrophils r=0.98, Lymphocytes r=0.98). Precision was established per NCCLS H20-A, demonstrating standard deviations comparable to or better than the reference method. Clinical sensitivity analysis confirmed the device's ability to identify abnormal specimens, with an overall agreement of 91.89% (after removing cases within 95% confidence limits). Statistical comparisons (ANOVA, paired t-test) confirmed inter-instrument agreement and consistency.
Technological Characteristics
System utilizes flow microscopy with a cytoprobe for rapid hemacyte fractionation and metachromatic staining (2-Methylpolymethine). Features a 20-inch touch-screen monitor for image display. Connectivity includes automated sampling and barcode reading. Software performs color image analysis and cell sorting. System is designed for closed sampling to minimize biohazard. Operates via automated intelligent microscopy principles.
Indications for Use
Indicated for use in clinical laboratories to perform WBC differential counts on peripheral blood specimens. Intended for use by trained, competent operators to review and confirm/modify presumptive cell classifications. Not for use in determining the presence or absence of nucleated RBCs.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
Wright-stain/Light Microscope
The Yellow IRIS urinalysis workstation and body fluids cell counting system
Reference Devices
Coulter STKS
Submission Summary (Full Text)
{0}
MAY 13 1996
ID:
PAGE 26
# INTERNATIONAL REMOTE IMAGING SYSTEMS, INC.
9162 Eton Avenue
Chatsworth, CA 91311
Phone 818-709-1244
FAX 818-700-9661
K953116
MAY 13 1996
## 510 (k) SUMMARY
| Name of Contact Person: | Jimmie R. Kyle
Vice President, Manufacturing and Product Reliability |
| --- | --- |
| Date of Submission: | June 29, 1995 |
| Name of Device: | • Trade Name - The White IRIS Leukocyte Differential Analyzer
• Common Name - Instrument used in determining WBC Differential
• Classification Name - None Established |
| Predicate Method and Legally Marketed Device to which Substantially Equivalent is claimed: | Wright-stain/Light Microscope and
The Yellow IRIS® urinalysis workstation and body fluids cell counting system |
| Intended Use of Device: | The White IRIS is a leukocyte differential analyzer intended for in vitro diagnostic use in determining the proportional leukocyte count (WBC differential) on peripheral blood specimens that have been flagged by an automated hematology analyzer performing differential counts as well as for peripheral blood specimens for which no automated differential has been performed. |
| Operator competence: | The White IRIS is designed and intended for use by trained, competent operators well skilled in both the use of the instrument and in the recognition of leukocyte classes based on IRISpectracolor stained cell expressions displayed on the ViewStation video monitor. Although leukocyte images are classified presumptively by cell type, a competent technologist is required to confirm or modify the classification of each cell. Laboratories using The White IRIS should maintain adequate operator training and proficiency testing programs for instrument users. |
510 (k) SUMMARY
1
{1}
ID:
PAGE 27
## Limitations:
1. Competent human observer band neutrophil identification is generally limited to those cells with a distinct band-form nucleus. Band neutrophils with lobes connected by thin chromatin strands will generally not be distinguished from segmented neutrophils. Consequently, fewer band neutrophils and more polymorphonuclear neutrophils will be reported. However, the combined neutrophil (band plus segmented) proportional count obtained by a competent human observer using The White IRIS will closely match that obtained from a Wright-stain smear.
2. As with Wright-stain smears, distinctions between band and segmented neutrophil, metamyelocyte and myelocyte, myelocyte and promyelocyte will be subject to variation among individual observers.
3. IRISpectracolor stain is not designed for discrimination of nuclear details in leukocytes.
4. Nucleated RBC (NRBC) are generally removed from the leukocyte-rich plasma fraction analyzed by The White IRIS along with RBC. The White IRIS should not be used to determine presence or absence of NRBC.
## Summary of Technological Characteristics:
Combining unique cytoprobe, rapid hemacyte fractionation and novel color image analysis, The White IRIS extends automated intelligent microscopy to leukocyte differentiation. It provides flow cytometry precision and microscopical resolution to review specimens flagged by hematology analyzers with differential capabilities or to complement other analyzers without these capabilities. The system includes compartments for closed sampling, rapid leukocyte-rich plasma preparations, cytoprobe-induced metachromasia, and collection and color analysis of leukocyte images, and presents the results as a single-view 500-cell differential on a 20-in touch-screen monitor for examination by a skilled competent observer.
510 (k) SUMMARY
2
{2}
MAY-09-96 16:12 FROM:IRIS
ID:
PAGE 20
Brief Discussion of Non-Clinical Factors supporting a Determination of Substantial Equivalence:
Brief Discussion of Clinical Tests Supporting a Determination of Substantial Equivalence:
Conclusions Drawn from Clinical Tests:
Safety of Device:
Involves a competent human observer to examine microscopic images as does the predicate method and other similar predicate devices. See attached Substantial Equivalence comparisons.
Conformance to National Committee for Clinical Laboratory Standards Reference leukocyte differential count (proportional) and evaluation of instrumental method: Approved Standard, NCCLS document H20-A and other statistical comparisons to Wright-stain/light microscope reference method.
Use of The White IRIS provides results equal to or better than the reference method with less time expenditure and less biohazard exposure. See attached Efficacy Demonstration.
See attached Hazard Analysis.
510 (k) SUMMARY
3
{3}
MAY-09-96 16:12 FROM:IRIS
ID:
PAGE 29
# Substantial Equivalence.
Table 1- Similarities and differences between The White IRIS and the peripheral blood smear stained with Wright-stain and viewed with a light microscope. (part 1 of 2)
| Substantial Equivalence | The White IRIS | Peripheral Blood Smear Stained with Wright Stain/Light Microscope |
| --- | --- | --- |
| Intended use is the same. | Production of images of cellular components of peripheral blood preparations to permit a skilled, competent observer to characterize these cells, enabling the generation of a white blood cell differential. | Production of images of cellular components of peripheral blood preparations to permit a skilled, competent observer to characterize these cells, enabling the generation of a white blood cell differential. |
| Specimen analyzed is the same. | Whole blood collected in blood specimen tube. | Whole blood collected in blood specimen tube. |
| Stain used differentially colors different cell types in a similar way. | Leukocytes stained supravitally in suspension by 2-Methylpolymethine which produces unique metachromas as among different types of leukocytes. | Leukocytes stained after cells dried and fixed on slide by Wright-stain (consisting of Methylene Blue, Azures, and Eosin Y) produces the Romanowsky effect, differential coloration of cell components. |
| Separation of WBC from RBC background is similar. | Leukocytes separated gravitationally from erythrocytes to allow mono-disperse presentation for automatic randomized selection. | Leukocytes separated by spreading in the feather edge portion of blood smear to allow manual search and selection from among monodispersed cells. |
| Mechanisms for presenting individual WBC images to microscope objective are comparable. | Non-overlapping individual cells, in liquid suspension, in a flow cell are automatically presented to and imaged by a microscope objective. | Non-overlapping individual cells, smeared and fixed on a glass slide are manually presented to and imaged by a microscope objective. |
| Optical means for magnifying images of WBCs for observation and interpretation is the same or similar. | A microscope images individual leukocytes on a camera sensor which produces a magnified video image for view and interpretation. | A microscope directly images and magnifies individual leukocytes for view and interpretation through its oculars. |
510 (k) SUMMARY
4
{4}
M:Y:09-96 16:12 FROM:IRIS
ID:
PAGE 30
Table 1 - continued (part 2 of 2).
| Substantial Equivalence | The White IRIS | Peripheral Blood Smear Stained with Wright Stain/Light Microscope |
| --- | --- | --- |
| Viewing of images is similar. | Individual cells are observed on a video monitor in machine-ordered and counted groups of like size and color features. | Individual cells are observed through the oculars one-at-a-time while they are classified and counted. |
| Overview scan analysis is similar and more easily accomplished. | Scanning is accomplished at a glance of 500-cells organized in a montage on a video monitor. | Scanning is accomplished by moving the slide and manually characterizing the kinds of cells observed. |
| Features used to distinguish among the various WBC types are the same or similar. | Cytoplasm of different leukocyte types stained characteristically.
While nucleus is not stained, nuclear shape may often be discerned.
Size of nucleus (relative to cell size) and cell size. | Cytoplasm of different leukocyte types stained characteristically.
Nuclear stain and shape used to distinguish among different leukocyte types.
Size of nucleus (relative to cell size) and cell size. |
| Image interpretation requirements are the same. | Differentiation of cells requires a skilled competent observer. | Differentiation of cells requires a skilled competent observer. |
| Result format is the same and quality is better. | The differential proportional count is based on 500-cells. | The differential proportional count is usually based on only 100-cells. |
510 (k) SUMMARY
5
{5}
MAY-09 10:12:24
ID:
PAGE 31
Table 2 - Similarities and Differences between The White IRIS Flow Microscope and the Conventional Light Microscope.
| Substantial Equivalence | The White IRIS/Flow Microscope | Conventional Slide Microscope |
| --- | --- | --- |
| Intended use is the same. | To enable a skilled, competent observer to examine, characterize and differentiate WBC compositions. | To enable a skilled, competent observer to examine, characterize and differentiate WBC compositions. |
| Means for presenting specimen to device is the same principle. | Flow microscopy of leukocytes monodispersed in suspension. (Used in conjunction with video camera, and monitor). | Slide microscopy of leukocytes monodispersed in dried, stained peripheral blood smear. (In some instances, cells may be suspended on the slide in a wet mount). |
| Microscopic image interpretation requirements are similar. | Differentiation of cells requires a skilled competent observer. | Differentiation of cells requires a skilled competent observer. |
| Results rely on counting individual cells and are based on the same arithmetic and quality is better. | The differential proportional count is based on observing 500 cells. | The differential proportional count is usually based on observing only 100 cells. |
510 (k) SUMMARY
{6}
MAY-09-96 16:13 FROM:IRIS
ID:
PAGE 32
Table 3 - Similarities and Differences between The White IRIS and the Generic Flow Microscope and The Yellow IRIS (part 1 of 2).
| Substantial Equivalence | The White IRIS for WBC Differentiation | Flow Microscope/The Yellow IRIS for UA and Body Fluid Blood Cell Examination and Counting |
| --- | --- | --- |
| Intended use is parallel. | Production of images of cellular components of peripheral blood preparations to permit a skilled, competent observer to characterize and count these cells, enabling the generation of a white blood cell differential. | Production of images of formed elements in urine including blood and other cells, and blood cells in other body fluids, to permit a skilled, competent observer to characterize and count these components to compose a white cell differential. |
| Stain used allows microscopic objects to be identified by differential coloration in a similar way. | Leukocytes stained with 2-Methylpolymethine while in suspension. | Formed elements stained with IRIStain, containing primarily Crystal Violet while in suspension. |
| Mechanisms for presenting individual cell and other formed element images to microscope objective are comparable. | Individual cells, in liquid suspension, in a flow cell are automatically selected and imaged by a microscope objective. | Individual cells and other formed elements, in liquid suspension, in a flow cell are automatically selected and imaged by a microscope objective. |
| Presentation of images for viewing are the same. | Individual cells are presented on a video monitor in machine-ordered and counted groups of like size and color features. | Individual cells and other formed elements are presented on a video monitor in machine-ordered and counted groups of like size and color features. |
| Features used to distinguish among the various leukocyte types are the same or similar. | Cytoplasm of different leukocyte types stained characteristically. While nucleus is not stained, nuclear shape may often be discerned. Size of nucleus (relative to cell size) and cell size. | Cytoplasm of different cells stained characteristically. Nuclear stain and shape used to distinguish among different cell types. Size of nucleus (relative to cell size) and cell size. |
510 (k) SUMMARY
7
{7}
MAY-01-16 16:13 FROM: IRIS
ID:
PAGE 33
Table 3 - continued (part 2 of 2).
| Substantial Equivalence | The White IRIS for WBC Differentiation | Flow Microscope/The Yellow IRIS for UA and Body Fluid Blood Cell Examination and Counting |
| --- | --- | --- |
| Presentation of images to skilled, competent observer is the same | Leukocytes in flow cell are viewed through a microscope by a video camera. Images are “captured” and electronically presented to a skilled, competent observer for interpretation on a video monitor. | Analytes in flow cell are viewed through a microscope by a video camera. Images are “captured” and electronically presented to a skilled, competent observer for interpretation on a video monitor. |
| Image montage organization is similar. | Cells are sorted into ranks of comparable size and color. | Cells and other formed elements are sorted into ranks of comparable size. |
| Results are observed and expressed in the same way. | Scanning is accomplished at a glance of 500-cells organized in a montage on a video monitor. The differential proportional count is based on competent observer acknowledged and assigned classification of 500 cell images. | Scanning is accomplished by observing a sequences of montages on a video monitor. Counts are based on competent observer assigned classification of cell images. |
510 (k) SUMMARY
8
{8}
MAY-09-96 16:13 FROM:IRIS
ID:
PAGE 34
# Efficacy Demonstration.
Differential leukocyte composition determined using The White IRIS™ was compared to results obtained by visual interpretation of blood smears stained with Wright-stain as the reference method. Count accuracy, clinical sensitivity and precision were compared according to the NCCLS Approved Standard H20-A "Reference leukocyte differential count (proportional) and evaluation of instrumental methods," National Committee for Clinical Laboratory Standards, Villanova, PA, March 1992, hereinafter referred to as H20-A. Other suitable, well accepted statistical comparisons not prescribed in H20-A were also performed to illustrate the effectiveness of The White IRIS even further. In addition, comparisons to proportional count and flagging results obtained with a Coulter STKS for the same group of specimens demonstrated the need for post primary analyzer review and the efficacy of The White IRIS in fulfilling this need as well as provided perspective with regards to current state-of-the-art hematology analyzers.
The following general conclusions can be made:
1. Inaccuracy, clinical sensitivity and imprecision studies according to H20-A have demonstrated substantial equivalence of leukocyte differential counts (proportional) derived from competent human interpretation of leukocyte images produced by The White IRIS to those from Wright-stained blood smears.
2. Additional statistical comparison (oneway ANOVA and paired t-test) of the results from the two instruments used in the study further demonstrated skilled competent observer inter-instrument agreement as good or better than agreement between the two Wright-stain readers performing the predicate method. Thus both agreement of the proposed new method with the established reference method and consistency within-run, instrument-to-instrument and over a few month period of intense use have been demonstrated in a way even more rigorous than specified by H20-A.
3. Actual comparative performance measure in both review and primary procedural modes demonstrate how well The White IRIS can either complement or replace a 5-part WBC differential capability on the Coulter STKS, the most popular such contemporary product in the market today.
4. Besides its demonstrated Wright-stain equivalency, and demonstrated superiority to the leading contemporary product in every leukocyte measure it makes, use of The White IRIS is also accompanied by significant safety and substantial labor saving benefits for the clinical laboratory.
510 (k) SUMMARY
9
{9}
MAY-09-16 15:13 FROM IRIS
ID:
PAGE 35
# Performance Characteristics on which Substantial Equivalence is Based
## 1 Accuracy
Differential leukocyte counts for 1,202 normal and abnormal patient specimens were obtained using both manually prepared and observed Wright stained smears (200 cell differential), read by two skilled technologists, and automated analysis on The White IRIS (2 instruments x 500 cell differential), reviewed by the same two skilled technologists.
Table 4 - Leukocyte Differential count correlation between methods. Neutrophils are the sum of polymorphonuclear neutrophils and bands, lymphocytes are the sum of normal lymphocytes and variant immature lymphocytes.
| Cell Type | Correlation | Slope | Intercept | Wright-Stain Mean (%) | TWI Mean (%) |
| --- | --- | --- | --- | --- | --- |
| Neutrophils | 0.98 | 0.96 | 4.16 | 56.13 | 57.90 |
| Lymphocytes | 0.98 | 0.93 | -0.97 | 31.06 | 27.85 |
| Monocytes | 0.91 | 1.03 | 1.19 | 7.69 | 9.12 |
| Eosinophils | 0.98 | 0.97 | 0.20 | 3.53 | 3.64 |
| Basophils | 0.87 | 0.78 | 0.18 | 0.70 | 0.72 |
| | | | | | |
| Segmented Neutrophils | 0.93 | 0.94 | 7.18 | 49.15 | 53.56 |
| Bands | 0.54 | 0.25 | 2.58 | 6.99 | 4.34 |
| Metamyelocytes | 0.84 | 0.64 | 0.08 | 0.40 | 0.33 |
| Myelocytes | 0.74 | 0.38 | 0.00 | 0.12 | 0.05 |
| Promyelocytes | 0.97 | 1.00 | 0.00 | 0.08 | 0.08 |
| Blasts | 0.96 | 1.13 | 0.03 | 0.25 | 0.32 |
| | | | | | |
| Normal Lymphocytes | 0.98 | 0.93 | -1.05 | 30.52 | 27.33 |
| Variant Lymphocytes | 0.41 | 0.31 | 0.35 | 0.55 | 0.52 |
510 (k) SUMMARY
10
{10}
MAY-09 96 16 14 FROM
ID:
PAGE 36
# 2 Precision
Short term standard deviations calculated according to H20-A are tabulated for The White IRIS and for the Wright stain reference method in Table 5. Short term imprecision is based on the root square of the differences between replicates. For The White IRIS, each replicate is the average of two determinations (specimen volume permitting) or a single determination for each instrument. For the reference method, each replicate is a 200 cell differential by one or the other reader.
Table 5 - Imprecision of The White IRIS compared to that of the Wright-stain reference method represented by standard deviation (SD) calculated according to the method specified by H20-A.
| | The White IRIS
(1014 Specimens) | | Wright-stain
(1277 Specimens) | |
| --- | --- | --- | --- | --- |
| Cell type | Mean (%) | SD | Mean (%) | SD |
| Neutrophils | 58.96 | 2.72 | 56.28 | 3.55 |
| Lymphocytes | 26.00 | 3.17 | 30.91 | 3.68 |
| Monocytes | 9.03 | 1.62 | 7.69 | 2.29 |
| Eosinophils | 3.69 | 0.83 | 3.54 | 1.31 |
| Basophils | 0.73 | 0.35 | 0.71 | 0.62 |
| Segmented Neutrophils | 54.64 | 4.25 | 49.47 | 4.72 |
| Bands | 4.33 | 2.94 | 6.81 | 3.68 |
| Metamyelocytes | 0.33 | 0.66 | 0.39 | 0.56 |
| Myelocytes | 0.05 | 0.17 | 0.12 | 0.27 |
| Promyelocytes | 0.04 | 0.17 | 0.08 | 1.33 |
| Blasts | 0.32 | 1.73 | 0.24 | 1.82 |
| Normal Lymphocytes | 26.38 | 3.19 | 30.38 | 3.67 |
| Variant Lymphocytes | 0.52 | 0.95 | 0.53 | 0.99 |
The following "within" run precision was established using 22 replicates of the same sample.
Table 6 - Precision of the Leukocyte Differential parameters is specified at 95% confidence limits.
| Leukocyte | Mean (%) | 95% Confidence Limits |
| --- | --- | --- |
| Neutrophils | 54.96 | ± 5.0 |
| Lymphocytes | 33.46 | ± 5.0 |
| Monocytes | 7.11 | ± 2.5 |
| Eosinophils | 3.93 | ± 1.5 |
| Basophils | 0.51 | ± 1.0 |
510 (k) SUMMARY
{11}
MAY-09-2018 16:44 FROM
ID:
PAGE 37
# 3 Clinical Sensitivity
The following clinical sensitivity analysis is based on the evaluation of 1202 Leukocyte Differential Summaries performed on The White IRIS.
Table 7 - H20-A defined requirements for abnormal specimens to be included in the clinical sensitivity study.
| H20-A requires at least 5 cases of each of these abnormal conditions | | | This Study |
| --- | --- | --- | --- |
| Characteristic Leukocyte Differential Count Finding | Absolute Cell Count | Proportional Cell Count | Number of Specimens |
| Granulocytosis and/or Left Shift (Bands) | ≥ 9.0 x 10^{9}/L
≥ 0.9 x 10^{9}/L | > 80%
> 6% | 152
197 |
| Monocytosis | ≥ 0.8 x 10^{9}/L | > 10% | 64 |
| Eosinophilia | ≥ 0.5 x 10^{9}/L | > 7% | 89 |
| Lymphocytosis and/or Lymphocytes, Variant Forms | ≥ 3.5 x 10^{9}/L
≥ 0.7 x 10^{9}/L | > 50% | 64
5 |
| Granulocytopenia | ≤ 1.5 x 10^{9}/L | < 10% | 11 |
| Lymphopenia | ≤ 1.0 x 10^{9}/L | < 7% | 70 |
| Immature Cells, including Blasts | ≥ 0.1 x 10^{9}/L | > 2% | 66 |
Table 8 reflects reference ranges established from a group of 154 normal (according to H20A criteria) specimens.
Table 8 - Reference Ranges (%) for Wright-Stain Smear and The White IRIS.
| | Wright-Stain Smear | | The White IRIS | |
| --- | --- | --- | --- | --- |
| Leukocyte Type | Lower Range | Upper Range | Lower Range | Upper Range |
| Segmented Neutrophils | 31.25 | 71.68 | 34.48 | 72.74 |
| Band Neutrophils | 0.00 | 9.25 | 0.00 | 7.68 |
| Lymphocytes | 17.50 | 56.25 | 14.50 | 48.97 |
| Variant Lymphocytes | 0.00 | 1.25 | 0.00 | 1.17 |
| Monocytes | 3.25 | 11.00 | 4.60 | 12.27 |
| Eosinophils | 0.00 | 9.00 | 0.00 | 9.71 |
| Basophils | 0.00 | 2.00 | 0.00 | 1.54 |
| Metamyelocytes | 0.00 | 0.50 | 0.00 | 0.28 |
| Myelocytes | 0.00 | 0.25 | 0.00 | 0.05 |
| Promyelocytes | 0.00 | 0.00 | 0.00 | 0.00 |
| Blasts | 0.00 | 0.00 | 0.00 | 0.00 |
510 (k) SUMMARY
{12}
MAY-09-96 16:14 FROM: K.S
ID:
PAGE 38
Using the reference ranges from Table 8, all specimens were classified as normal (all parameters within normal range) or abnormal (any parameter outside of normal range). The comparison after removing those cases for which parameters causing disagreement were within 95% confidence limits is shown in Table 9.
Table 9 - Crosstabulation of combined abnormal classifications after false positive and false negative cases for which differences that are within 95% confidence limits are removed. Values in parentheses () are after false positive and false negative band decisions are removed.
| | Results of The White IRIS
(Test Method) | | | |
| --- | --- | --- | --- | --- |
| | | Positive
(Distributional &/or
Morphological
Abnormals) | Negative
(Normal) | Total |
| Results of
Wright-Stain/Light
Microscope
(Reference
Method) | Positive
(Distributional &/or
Morphological
Abnormal) | 716 | 55 (22) | 752 (738) |
| | Negative
(Normal) | 53 (48) | 292 | 345 (340) |
| | Total | 769 (764) | 328 (314) | 1097 (1078) |
Summary
Agreement: 91.89% (93.51%)
False Positive: 15.36% (14.12%)
False Negative: 4.79% (2.98%)
510 (k) SUMMARY
13
{13}
MAY-09-96 16:15 FROM:IRIS
ID:
PAGE 39
A summary of discrepant cases is in Table 10. The 5% significance level of McNemar's statistic is 3.84. Thus for those cases that disagree outside the 95% confidence limits, other than excessive false negative bands and excessive false positive metamyelocytes, discrepancies are balanced.
Table 10 - Analysis of combined abnormality classifications crosstabulation
| Statistical Significance | Disagreements by cell types | Total Number of Cases | Number of Combined Abnormal False Negative | Number of Combined Abnormal False Positive | McNemar's Statistic |
| --- | --- | --- | --- | --- | --- |
| All Disagreements | TOTAL | 194 | 78 | 116 | 7.44 |
| | | | | | |
| Within 95% Confidence Limits | PMN | 10 | 2 | 8 | 3.60 |
| | Lymphocyte | 11 | 5 | 6 | 0.09 |
| | Lymphocyte and PMN | 5 | 2 | 3 | 0.20 |
| | Monocyte | 27 | 10 | 17 | 1.81 |
| | Monocyte and Lymphocyte | 1 | 0 | 1 | 1.00 |
| | Eosinophil | 5 | 3 | 2 | 0.20 |
| | Basophil | 7 | 1 | 6 | 3.57 |
| | Basophil and Lymphocyte | 1 | 0 | 1 | 1.00 |
| | Band neutrophil | 1 | 1 | 0 | 1.00 |
| | Band neutrophil and Basophil | 1 | 0 | 1 | 1.00 |
| | Variant lymphocyte | 10 | 6 | 4 | 0.40 |
| | Metamyelocyte | 7 | 3 | 4 | 0.14 |
| | Metamyelocyte and Monocyte | 1 | 1 | 0 | 1.00 |
| | Metamyelocyte and Variant lymphocyte | 1 | 0 | 1 | 1.00 |
| | Myelocyte | 6 | 2 | 4 | 0.67 |
| | Myelocyte and Lymphocyte | 1 | 1 | 0 | 1.00 |
| | Myelocyte and Metamyelocyte | 1 | 1 | 0 | 1.00 |
| | Promyelocyte | 1 | 1 | 0 | 1.00 |
| | Blast | 2 | 0 | 2 | 2.00 |
| | Blast and Variant lymphocyte | 1 | 0 | 1 | 1.00 |
| | Megakaryocyte | 3 | 1 | 2 | 0.33 |
| | Megakaryocyte, Lymphocyte and Variant lymphocyte | 1 | 1 | 0 | 1.00 |
| | Plasma cell | 1 | 1 | 0 | 1.00 |
| TOTAL | 105 | 42 | 63 | 4.20 | |
| | | | | | |
| Outside 95% Confidence Limits | PMN | 2 | 1 | 1 | 0.00 |
| | Lymphocyte | 5 | 2 | 3 | 0.20 |
| | Lymphocyte and PMN | 6 | 1 | 5 | 2.67 |
| | Monocyte | 11 | 3 | 8 | 2.27 |
| | Monocyte and PMN | 1 | 0 | 1 | 1.00 |
| | Monocyte and Lymphocyte | 1 | 0 | 1 | 1.00 |
| | Eosinophil | 1 | 0 | 1 | 1.00 |
| | Basophil | 3 | 0 | 3 | 3.00 |
| | Band neutrophil | 19 | 14 | 5 | 4.26 |
| | Band neutrophil and PMN | 2 | 2 | 0 | 2.00 |
| | Band neutrophil and Lymphocyte | 2 | 1 | 1 | 0.00 |
| | Variant lymphocyte | 20 | 9 | 11 | 0.20 |
| | Variant lymphocyte and PMN | 2 | 1 | 1 | 0.00 |
| | Variant lymphocyte and Monocyte | 1 | 0 | 1 | 1.00 |
| | Variant lymphocyte and Band neutrophil | 2 | 1 | 1 | 0.00 |
| | Metamyelocyte | 10 | 1 | 8 | 6.40 |
| | Blast | 1 | 0 | 1 | 1.00 |
| | TOTAL | 89 | 36 | 53 | 3.25 |
510 (k) SUMMARY
{14}
MAY-09-96 16:15 FROM:IRIS
ID
PAGE 40
The comparison of normal and abnormal cases prior to removing false positive and false negative cases that are within 95% confidence limits is shown in Table 11.
Table 11 - Crosstabulation of combined abnormality classifications
| | Results of The White IRIS
(Test Method) | | | |
| --- | --- | --- | --- | --- |
| | | Positive
(Distributional
and/or
Morphological
Abnormal) | Negative
(Normal) | Total |
| Results
of
Wright-stain/
Light
Microscope
(Reference
Method) | Positive
(Distributional
and/or
Morphological
Abnormal) | 716 | 76 | 794 |
| | Negative
(Normal) | 116 | 292 | 408 |
| | Total | 832 | 370 | 1202 |
| Summary | Agreement: | 82.95% |
| --- | --- | --- |
| | False Positive: | 28.43% |
| | False Negative: | 9.82% |
510 (k) SUMMARY
15
{15}
MAY 11
ID:
PAGE 41
Table 11 is based on a $4 \times 4$ crosstabulation of the four mutually exclusive outcomes for each sample processed:
1. Normal
2. Distributionally abnormal
3. Morphologically abnormal
4. Both distributionally and morphologically abnormal.
Table 12 - Crosstabulation of specimen classifications by Wright-stain versus those of The White IRIS™
| Wright-stain classification | The White IRIS™ classification | | | | Row Sum |
| --- | --- | --- | --- | --- | --- |
| | Normal | Distributionally abnormal | Morphologically abnormal | Both distributionally and morphologically abnormal | |
| Normal | 292 | 62 | 40 | 14 | 408 |
| Distributionally abnormal | 30 | 230 | 8 | 73 | 341 |
| Morphologically abnormal | 38 | 14 | 32 | 11 | 95 |
| Both distributionally and morphologically abnormal | 10 | 122 | 15 | 211 | 358 |
Table 12 summarizes classifications for both methods. Overall agreement in this table is $63.6\%$. Based on Cohen's kappa$^1$ of 0.49 for this table one can reject the hypothesis that the entries in the table are purely random and there is no agreement between methods.
510 (k) SUMMARY
16
1 Cohen, J. 1960. A coefficient of agreement for nominal scales. Educational and Psychological Measurement, 20: 37-46.
{16}
MAY-09-36 16:16 FROM:IRIS
ID:
PAGE 42
# Hazard Analysis.
Laboratory instrument results are generally used by the physician as an adjunct to clinical observations made in the course of establishing a diagnosis and in monitoring the course of disease progress, when appropriate thereafter. Therefore, The White IRIS leukocyte differential analyzer is unlikely to present the patient with a direct hazard. Also, as long as the instrument operator follows the recommended operating procedures in the Operators Manual this risk is minimized.
It is expected that anyone using this instrument is a qualified laboratorian who will follow universal precautions and other "Standard Laboratory Practices" in regards to working with laboratory instruments, reagents, consumables, and patient blood specimens in a safe and precautious manner. Also, the laboratorian is expected to be trained as a skilled competent observer familiar with and experienced in the interpretative qualities of leukocyte images.
The hazard analysis below is intended to identify the potential hardware and software failures that might cause inadvertent erroneous test results and lists the system checks in the design to safeguard against adverse consequences.
Table 13 - List of Hazards, their Causes, and System Responses.
| Potential Hazard | Potential Causes | Software/Hardware Responses |
| --- | --- | --- |
| Bar code reading failure could cause results to be reported to the wrong patient. | Bar Code Reader failure. Unreadable bar code label. No bar code label. | Software checksum analysis with operator alerts. No detectable bar code or bar code error - no sample processing or missing code flagging. |
| Sample wheel positioning failure could cause wrong sample to be aspirated and results reported to the wrong patient ID. | Motor failure. Encoder failure. | Optical sensor coupled with software detection and operator alerts. |
| Insufficient sample aspirated for analysis could cause erroneous results. | Insufficient sample. Sample probe clog. No vacuum or pressure. | Optical and ultrasonic sensors coupled with software detection and operator alerts. |
| Failure of red blood cells to settle out leaving too many rbc's could cause erroneous results. | No IRISETlant reagent present. IRISETlant reagent no good. Settling tubes non functional. | Optical sensor with software detection and operator alerts. |
| Insufficient leukocyte-rich plasma preparation could cause erroneous results. | Insufficient sample. No settling of RBCs. | Optical and ultrasonic sensors coupled with software detection and operator alerts. |
510 (k) SUMMARY
17
{17}
PAGE 43
| Potential Hazard | Potential Causes | Software/Hardware Responses |
| --- | --- | --- |
| Non staining or improper staining of white blood cells could cause erroneous results. | No IRISpectracolor reagent present. IRISpectracolor no good. Specimen too old. | Optical and software detection with operator alerts. |
| Total or partial interruption of flow of white cells through the White IRIS flowcell could cause erroneous results. | Flowcell partial or total clog. No vacuum or pressure. Hardware failure. | Optical and software detection with operator alerts. |
| Failure of the ViewStation sorting algorithms could cause erroneous results. | Power outage with bad reboot. Spike in line voltage. Software lockup. | Reportable results require the optical review, edit and approval of competent operator. |
| Previous patient sample not properly washed from system (carryover) could cause erroneous results. | No IRIScrub reagent present. No vacuum or pressure. Hardware failure. | Optical and ultrasonic sensors with software detection and operator alerts. |
| Image data transfer from Flow Microscope Module to View-Station corrupted during transmission could cause erroneous results. | Power outage with bad reboot. Spike in line voltage. Software lockup. | Software detection using checksums to validate data packets. Reportable results require the optical review, edit and approval of competent operator. |
| Optical System failure could cause erroneous or no results to be reported. | Light source inoperable. Strobe inoperable. Camera inoperable. | Optical and software detection with operator alerts. |
| Editing errors by competent operator could cause erroneous results to be reported. | Wrong tag used to classify cells. Accidental touch stroke. | Edit confirmation summary to be verified by competent operator prior to reporting results. |
| Contact of the operator with the cap piercing unit during operation could cause injury to the operator. | Removal of needle cover during operation. | Hardware detection with software operator alerts and instrument shutdown. |
| Contact with high voltage inside the instrument could cause electrical shock to operator. | Removal of protective covering isolating high voltages. Excessive leakage current. | Warning labels on system and warnings in operating instructions. Grounding per code and all metal parts bonded to ground system. |
510 (k) SUMMARY
18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.