IMPACT AIR 45 4-HOLE

K952927 · Palisades Pharmaceutics, Inc. · EFB · Aug 6, 1996 · Dental

Device Facts

Record IDK952927
Device NameIMPACT AIR 45 4-HOLE
ApplicantPalisades Pharmaceutics, Inc.
Product CodeEFB · Dental
Decision DateAug 6, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4200
Device ClassClass 1

Indications for Use

The Impact Air 45 Handpiece is an air-powered dental handpiece. The handpiece has an angled head shank with a back angle of 45 degrees. It is intended for use with all friction-grip bur that conform to I.S.O. and A.D.A. recommended operating air pressure is between 32 and 40 pounds per square inch ("psi"), which results in high-speed bur rotation (approximately 400,000 to 500,000 revolutions per minute ("rpm")). The device includes a water line that directs water onto the spinning bur, which forms a cooling mist and also helps irrigate the working area. One model of the device also includes a fiberoptic bundle that directs light onto the working area.

Device Story

Air-powered dental handpiece; 45-degree angled head shank; utilizes friction-grip burs. Operates via compressed air (32-40 psi) driving a turbine in the head; achieves 400,000-500,000 rpm. Includes integrated water line for cooling mist/irrigation; optional fiberoptic bundle for illumination. Used by dentists in clinical settings; connects to standard dental unit air supply. Provides high-speed rotation for dental cutting/preparation; cooling mist minimizes thermal damage to tooth structure; fiberoptics improve visibility. Sterilizable via autoclave.

Clinical Evidence

Bench testing only. Data indicates device functions appropriately after 250 autoclave sterilization cycles.

Technological Characteristics

Stainless steel construction; air-powered turbine; 45-degree head angle; 32-40 psi operating pressure; 400,000-500,000 rpm; water irrigation system; optional fiberoptic illumination; compatible with standard dental unit air connections; autoclavable (250 degrees for 35 minutes).

Indications for Use

Indicated for use in dental procedures requiring high-speed bur rotation for cutting or removal of tooth structure.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Submission Summary (Full Text)

{0} FROM : GLENWOOD-PALISADES PHONE NO. : 908 981 9790 Aug. 02 1996 11:14AM P2 AUG 6 1996 K952927 PALISADES PHARMACEUTICALS, INC. PALISADES DENTAL, Division of Palisades Pharmaceuticals, Inc. P.O. BOX 579 • 64 NORTH SUMMIT STREET • TENAFLY, NEW JERSEY 07670 • (201) 569-8502 • 1-(800) 237-9083 • FAX (201) 569-7416 510(K) SUMMARY I. Submitter: Palisades Dental A Division of Palisades Pharmaceuticals, Inc. 64 N. Summit Street Tenafly, NJ 07670 Contact Person: Cynthia Romanoff Phone: (201) 569-8502 Fax: (201) 569-7416 II. Date 510(k) Summary Prepared: August 1, 1996 III. Trade Name: Impact Air 45 Handpiece IV. Classification Name: Dental Handpiece (21 C.F.R. §872.4200) V. Predicate Devices: Midwest Quiet Air Handpiece VI. Device Description: The Impact Air 45 Handpiece is an air-powered dental handpiece. The handpiece has an angled head shank with a back angle of 45 degrees. It is intended for use with all friction-grip bur that conform to I.S.O. and A.D.A. recommended operating air pressure is between 32 and 40 pounds per square inch ("psi"), which results in high-speed bur rotation (approximately 400,000 to 500,000 revolutions per minute ("rpm")). The device includes a water line that directs water onto the spinning bur, which forms a cooling mist and also helps irrigate the working area. One model of the device also includes a fiberoptic bundle that directs light onto the working area. The device may be connected to a standard compressed air source found in virtually all dental units. The use of filtered air is recommended. The air supply enters the device through a tube in the handle and turns the turbine, which is in the head of the device, thus spinning the bur. The Impact air 45 can be sterilized by autoclave when processed for 35 minutes at 250 degrees. Data indicate that after 250 sterilization cycles the handpiece continues to function appropriately. {1} F-3M : GLENWOOD-PALISADES PHONE NO. : 908 981 9790 Aug. 02 1996 11:14AM P3 VII. **Substantial Equivalence**: The Impact Air as compared with the predicate device has the same intended use, does not raise different questions of safety and effectiveness, and is fundamentally the same in operating principle, design, materials, energy source, and other characteristics. Like the predicate device, the Impact air 45 has principally stainless steel construction; is air-powered, with a high-speed turbine in the head of the device; attaches to a standard connection for compressed air found in most dental units; and is slim, lightweight and easy to manipulate.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...