← Product Code [DEH](/productcode/DEH) · K951965

# DAKO LABELED STREPTAVIDIN-BIOTIN 2 KIT, AND ENVISION (K951965)

_Dako Corp. · DEH · Jul 5, 1996 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K951965

## Device Facts

- **Applicant:** Dako Corp.
- **Product Code:** [DEH](/productcode/DEH.md)
- **Decision Date:** Jul 5, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5550
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

Immunoenzymatic/immunohistochemical (IHC) staining techniques allow for the qualitative identification of tissue antigens. Antigens are visualized via the sequential application of a specific antibody to the antigen (primary antibody) and a detection system. Immunohistochemical detection systems usually consist of a secondary antibody to the primary antibody (link antibody), an enzyme complex and a chromogenic substrate. The enzymatic activation of the chromogen yields a visible reaction product at the antigen site. Results aid in the diagnosis of pathophysiological processes which may or may not be associated with a particular antigen.

## Device Story

DAKO LSAB® 2 Kit and DAKO Envision™ Systems are immunohistochemical (IHC) detection kits; used by laboratory professionals to visualize tissue antigens. LSAB® 2 utilizes modified labeled avidin-biotin (LAB) technique; biotinylated secondary antibody binds primary antibody; peroxidase-conjugated streptavidin binds biotin; chromogen substrate produces visible reaction product. Envision™ System utilizes polymer-based detection; goat anti-mouse/rabbit immunoglobulins conjugated to horseradish peroxidase (HRP) bind primary antibody directly; eliminates sequential link/streptavidin steps. Systems used in clinical pathology labs; results interpreted by pathologists to aid diagnosis of disease states. Envision™ offers two protocols; Protocol #2 increases staining intensity via longer incubation times. Primary antibodies purchased separately. Benefits include standardized antigen visualization for diagnostic assessment.

## Clinical Evidence

Bench testing only. Comparative studies performed using formalin-fixed, paraffin-embedded tissues, frozen tissues, and blood smears. DAKO LSAB® 2 Kit compared to DAKO LSAB® Kit showed equivalent interpretive results. DAKO Envision™ System compared to DAKO LSAB® 2 Kit using both liquid DAB and ready-to-use AEC substrates showed equivalent interpretive results. Protocol #2 for Envision™ demonstrated increased staining intensity compared to Protocol #1. Comparison of 2-component AEC vs. ready-to-use AEC showed equivalent or improved staining performance. No clinical diagnostic trials were required.

## Technological Characteristics

IHC detection kits. LSAB® 2: biotinylated secondary antibody, peroxidase-conjugated streptavidin, chromogenic substrate. Envision™: polymer-based, goat anti-mouse/rabbit immunoglobulins conjugated to HRP. Substrates: DAB or AEC. Form factor: liquid reagents. Connectivity: N/A (manual laboratory procedure). Sterilization: N/A (reagents).

## Regulatory Identification

An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.

## Predicate Devices

- DAKO LSAB® Kit, HRP ([K924726](/device/K924726.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K951965
JUL - 5 1990
Summary of Safety and Effectiveness
DAKO LSA8® 2 Kit, HRP
DAKO ENVISION® SYSTEMS, HRP

# I. Introduction

Immunoenzymatic/immunohistochemical (IHC) staining techniques allow for the qualitative identification of tissue antigens. Antigens are visualized via the sequential application of a specific antibody to the antigen (primary antibody) and a detection system. Immunohistochemical detection systems usually consist of a secondary antibody to the primary antibody (link antibody), an enzyme complex and a chromogenic substrate. The enzymatic activation of the chromogen yields a visible reaction product at the antigen site. Results aid in the diagnosis of pathophysiological processes which may or may not be associated with a particular antigen.

# II. Device Descriptions

The DAKO LSAB® 2 Kit, HRP (Code No. K0677) is based on a modified labeled avidin-biotin (LAB) technique in which a biotinylated secondary antibody forms a complex with peroxidase-conjugated streptavidin molecules. This kit is very similar to the original DAKO LSAB® Kit, HRP (K0680) previously cleared in submission K924726. Refinements to the LSAB® 2 Kit include the elimination of a separate protein blocking reagent incubation and reconfigured substrate-chromogen reagents.

The DAKO Envision™ Systems, HRP, contains a Peroxidase Blocking Reagent similar to the Hydrogen Peroxide provided with the LSAB® and LSAB® 2 Kits. Like the LSAB® 2 Kit, the Envision™ System does not require a protein blocking reagent incubation. The Envision™ System also eliminates the sequential applications of link antibody and streptavidin common to the LSAB® and LSAB® 2 Kits. A polymer, labeled with goat anti-mouse and goat anti-rabbit immunoglobulins conjugated to horseradish peroxidase eliminates these steps. The DAKO Envision™ System offers the user a choice of two protocols. Protocol #1 consecutively incubates the primary antibody and Peroxidase Labeled Polymer for ten (10) minutes. Protocol #2 increases the staining intensity of the Envision™ System by lengthening the incubations of the primary antibody and the labeled polymer to thirty (30) minutes. The DAKO Envision™ System, HRP, is available with DAB (Code No. K1390) or Ready-to-Use AEC substrate-chromogen (Code No. K1391). Both systems use the same Peroxidase Blocking Reagent and Peroxidase Labeled Polymer.

Concentrated primary rabbit/mouse antibodies or DAKO® Ready-to-Use N-Series Primary Antibodies and Negative Control Reagents are suitable for use with the DAKO LSAB® 2 Kit and DAKO Envision™ Systems. Primary antibodies are not included with the DAKO LSAB® 2 Kit, HRP or DAKO Envision™ Systems, and must be purchased separately by the user.

# III. Experimental Data

Concurrent testing was done with the DAKO LSAB® Kit, HRP and the DAKO LSAB® 2 Kit, HRP using formalin-fixed, paraffin embedded tissues. Equivalent interpretive results were observed in all test specimens.

Summary of Safety &amp; Effectiveness/ rev. C/ p. 1

{1}

Summary of Safety &amp; Effectiveness/ rev. C/ p. 2

# III. Experimental Data, continued

Similar concurrent testing was done with the DAKO LSAB® 2 Kit, HRP and the DAKO Envision™ System, HRP with Ready-To-Use AEC substrate-chromogen (K1391) using formalin-fixed, paraffin embedded tissues. Equivalent interpretive results were observed in all test specimens.

This test procedure was repeated using the DAKO LSAB® 2 Kit, HRP and the DAKO Envision™ System, HRP with liquid DAB substrate-chromogen (K1390). Equivalent interpretive results were observed in all test specimens.

Additional testing was done with the DAKO LSAB® 2 Kit, HRP and the DAKO Envision™ System, HRP with Ready-To-Use AEC substrate-chromogen (K1391) using acetone-fixed, frozen tissues and blood smears. Results were equivalent.

Results obtained from comparative testing of the DAKO Envision™ System, HRP with Ready-to-Use AEC substrate-chromogen using Protocols #1 and #2 support package insert claims for Protocol #2. Protocol #2 increases the staining intensity of the DAKO Envision™ System. Concentrated Primary antibodies used with protocol #2 may be diluted up to twenty times the optimal dilution used with Protocol #1.

Comparison testing of LSAB2 with 2 component AEC versus ready-to-use AEC was performed using LCA (N1514), UCHL1 (N1520), and HMB45 (N1545). The tissues tested included tonsil (N1514, and N1520) and melanoma (N1545). Results showed that the same level of staining was obtained with both product configurations for N1545 and N1520. Slightly more intense staining (2 component AEC staining = 3-3.5 vs ready-to-use AEC staining = 4+) was noted for ready-to-use AEC with N1514. Background staining of both negative controls and positively stained tissues was evaluated as negative i.e., no staining. Thus, the LSAB2 with the ready-to-use AEC is similar or better in performance to LSAB2 with the 2-component AEC.

# IV. Product Specific Limitations

## DAKO LSAB® 2, Kit (K0677):

Endogenous avidin-binding activity (EABA) has been noted in frozen sections of liver (entire hepatic nodule) and kidney (tubular epithelium), as well as in frozen and formalin-fixed lymphoid tissues (paracortical histiocytes). EABA can be suppressed by sequential 20-minute incubations, first with 0.1% avidin and then with 0.01% biotin in 0.05 M Tris-HCl buffer, pH 7.2-7.6 prior to the application of the primary antibody, or use DAKO® Biotin Blocking System (code no. X0590).

## DAKO Envision™ System, HRP (K1390, K1391):

Some primary antibodies may require proteolytic digestion and target retrieval to achieve optimal staining results. Protocol #2 is recommended for use. Contact DAKO Technical Services at 800/424-0021 for additional information.

Endogenous peroxidase or pseudoperoxidase activity can be found in hemoproteins such as hemoglobin, myoglobin, cytochrome, and catalase as well as in eosinophils. This activity can be inhibited by incubating specimens with Peroxidase Blocking Reagent Bottle #1 of the DAKO Envision™ System, HRP for five (5 ± 1) minutes prior to the application of the primary antibody. Blood and bone marrow smears and frozen tissues can also be treated with this reagent. However, this procedure does not abolish the reddish-brown pigment of hemoproteins. Alternatively, a solution of methanol-hydrogen peroxide can be used. Some antigens may become denatured with this procedure.

---

**Source:** [https://fda-staging.innolitics.com/device/K951965](https://fda-staging.innolitics.com/device/K951965)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).