← Product Code [JAK](/productcode/JAK) · K950972

# REAL-TIME RECONSTRUCTION SYSTEM MODEL TSXF-001A (K950972)

_Toshiba America Medical Systems, In.C · JAK · Apr 15, 1996 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K950972

## Device Facts

- **Applicant:** Toshiba America Medical Systems, In.C
- **Product Code:** [JAK](/productcode/JAK.md)
- **Decision Date:** Apr 15, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Computed Tomography X-Ray Scanners generally support three modes of imaging: 1) Multiple contiguous slices: conventional CT where individual slices of axial data are acquired, with the table advancing a set distance until a volume of interest is covered. 2) Helical Scanning: a volume of interest is scanned continuously while the patient table is advanced longitudinally, until a desired volume of data is acquired. 3) Dynamic scanning: a single slice location is scanned repeatedly over a period of time. Using the Real-Time Reconstruction System (RTRS) Option, the scanner effectively operates in the dynamic scan mode with the added flexibility of table and/or gantry movement during scanning. Under the control of the operating physician the RTRS is intended to speed up interventional procedures typically performed on CT. By permitting the operator to view the scanned area in virtual real time the RTRS is intended to support existing CT system clinical uses by offering enhanced operator flexibility coupled with the potential of reduced scan time.

## Device Story

TSXF-001A is a field-installed hardware/software upgrade for Xpress/SX CT system (TSX-011A/50). Input: X-ray projection data. Transformation: Real-time reconstruction of axial slices at 6 images/second; enables dynamic scanning with simultaneous table/gantry movement. Output: Real-time axial image display on in-room monitor. Operation: Physician-controlled via added table-side control panel for X-ray initiation/termination, table motion, and gantry tilt. Context: Interventional CT procedures. Safety features: Automatic attenuating filter insertion; 100-second scan duration limit; elapsed time display. Benefit: Increased operator flexibility and reduced scan time for interventional procedures.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Hardware/software upgrade kit. Features solid-state X-ray detector, larger X-ray generator, and real-time reconstruction engine. Includes automatic attenuating filter mechanism. Field-installed by service engineer.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- TSX-011A (Xpress CT System) (Toshiba Corp.)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K95-0972

K950972

# 510 (k) Summary

APR 15 1996

Date: 16 February, 1995

Submitter: Javad Seyedzadeh, Director of Regulatory Affairs
Toshiba America Medical Systems, Inc.
2441 Michelle Drive; Tustin, CA 92681-2068

Contact: Mark Stauffer, Regulatory Affairs Specialist
(714) 730-5000

Device Name: TSXF-001A

Common Name: Real-Time Reconstruction System Option for Xpress/SX CT System

Classification Name: Computed Tomography X-Ray System

Classification: Class II

Establishment Reg. Number: 2020563

Predicate Device: TSX-011A (Xpress CT System) [Toshiba Corp.]

Description of this device: The TSXF-001A is an optional upgrade kit available for the Xpress/SX CT system (TSX-011A/50). The kit consists of hardware and software that can be field installed by a qualified Field Service Engineer. When equipped with this option, the Xpress/SX system will have the capability to reconstruct and display axial slices at the rate of six images per second, effectively emulating a real-time display presentation. Additionally, an added control panel provides table-side control of x-ray initiation and termination, table longitudinal and vertical motion, and gantry tilt.

Intended use of this device: Computed Tomography X-Ray Scanners generally support three modes of imaging:

1) Multiple contiguous slices: conventional CT where individual slices of axial data are acquired, with the table advancing a set distance until a volume of interest is covered.
2) Helical Scanning: a volume of interest is scanned continuously while the patient table is advanced longitudinally, until a desired volume of data is acquired.
3) Dynamic scanning: a single slice location is scanned repeatedly over a period of time.

Using the Real-Time Reconstruction System (RTRS) Option, the scanner effectively operates in the dynamic scan mode with the added flexibility of table and/or gantry movement during scanning. Under the control of the operating physician the RTRS is intended to speed up interventional procedures typically performed on CT. By permitting the operator to view the scanned area in virtual real time the RTRS is intended to support existing CT system clinical uses by offering enhanced

{1}

operator flexibility coupled with the potential of reduced scan time.

Features included with the RTRS to help control patient radiation dose are:

1) Use of an attenuating filter which is moved automatically into the x-ray beam when the RTRS mode is selected.
2) Automatic interruption of scanning after a maximum period of one hundred seconds.
3) Display of the elapsed scan time on the in-room viewing monitor.

Comparison to predicate device: The Xpress/SX with Real-Time Reconstruction System option is an upgraded version of the predicate, the Xpress CT System. The enhancements of the Xpress/SX with RTRS include a larger x-ray generator, faster reconstruction time, a solid-state x-ray detector, and real-time reconstruction capabilities. Taken together, these features provide user flexibility and improvements to the efficiency with which the system is utilized. There are no perceived or imagined new intended uses for CT introduced by the Xpress/SX with RTRS option and the safety and effectiveness of the system are believed to remain within the accepted boundaries for this modality.

---

**Source:** [https://fda-staging.innolitics.com/device/K950972](https://fda-staging.innolitics.com/device/K950972)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).