K942301 · Boston Scientific Corp · KNS · Nov 2, 1994 · Gastroenterology, Urology
Device Facts
Record ID
K942301
Device Name
MICROVASIVE INJECTION GOLD PROBE
Applicant
Boston Scientific Corp
Product Code
KNS · Gastroenterology, Urology
Decision Date
Nov 2, 1994
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Microvasive Injection Gold Probe is indicated for use in endoscopic injection therapy and electrohemostasis of actual or potential bleeding sites in the gastrointestinal tract, including, but not limited to: the esophagus, stomach, duodenum, and colon. These sites include the following: peptic ulcers; Mallory-Weiss tears; arteriovenous malformations (AVMs); Dieulafoy lesions; bleeding polyp stocks; and angiomata.
Device Story
Endoscopic catheter for injection therapy and bipolar electrohemostasis; used in GI tract (esophagus, stomach, duodenum, colon). Device features bipolar ceramic tip for electrohemostasis and integrated needle for injection of agents. Operated by physician during endoscopy. Input: electrical energy from bipolar electrosurgical generator; injection agents (sclerosing/vasoconstrictor). Output: thermal energy for hemostasis; localized injection. Benefits: dual-modality treatment of bleeding lesions. Used in clinical/endoscopic settings.
Clinical Evidence
Bench testing only. In vitro laboratory testing performed on finished, sterilized catheters to verify clinical performance and product integrity. Full Tripartite Biocompatibility testing performed with satisfactory results.
Technological Characteristics
Bipolar electrosurgical catheter with integrated injection needle. Features bipolar ceramic tip, electrical wires, and irrigation capability. Dimensions: 210-350 cm length, 7-10 F shaft OD, 25-gauge needle with 4-6 mm extension. Connectivity: Bipolar Erbe generator connector. Sterilization: Ethylene oxide gas per AAMI protocol.
Indications for Use
Indicated for endoscopic injection therapy and electrohemostasis of actual or potential GI bleeding sites (esophagus, stomach, duodenum, colon) including peptic ulcers, Mallory-Weiss tears, AVMs, Dieulafoy lesions, bleeding polyp stocks, and angiomata. Contraindicated for non-focal bleeding, varices, diffuse lesions, injection agent allergies, or use with monopolar electrosurgical generators.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}
Boston Scientific Corporation
May 10, 1994
510(k) Premarket Notification
Microvasive® Injection Gold Probe™
K942301
NOV 2 1994
Section V
# **FOI RELEASABLE**
# **510(k) Summary**
Page 41 of 159
{1}
Boston Scientific Corporation
May 10, 1994
510(k) Premarket Notification
Microvasive® Injection Gold Probe™
# FOI RELEASABLE
K942301
## 510(k) Summary
Pursuant to §513(i)(3)(A) of the Food, Drug, and cosmetic Act, Boston Scientific Corporation is required to submit with this Premarket Notification "... adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Boston Scientific Corporation chooses to submit a summary of information respecting safety and effectiveness.
### A. GENERAL INFORMATION
| Owner Operator Submitting this Premarket Notification: | Boston Scientific Corporation 480 Pleasant Street Watertown, MA 02172 (617) 923-1720 |
| --- | --- |
| Contact Person: | Sheri D. O'Brien Senior Regulatory Affairs Specialist |
| --- | --- |
| Device Generic Name: | Electrohemostasis and Injection Therapy Catheter |
| --- | --- |
| Device Trade Name: | Microvasive® Injection Gold Probe™ |
| --- | --- |
| Classification Name: | Boston Scientific Corporation believes that the Injection Gold Probe is best described by the following device classification names: |
| --- | --- |
| Unit, Electrosurgical, Endoscopic (with Accessories): | 78KNS |
| --- | --- |
| Endoscope and/or Accessories: | 78KOG |
Device Classification:
Although the Injection Gold Probe has not been specifically classified under 21 CFR § 876, the status of the classification names presented below supports a Class II device classification.
| Endoscope Electrosurgical Unit and Accessories: | Class II, 21 CFR § 876.4300 |
| --- | --- |
| Endoscope and Accessories: | Class II, 21 CFR § 876.1500 |
### B. INDICATIONS FOR USE
The Microvasive Injection Gold Probe is indicated for use in endoscopic injection therapy and electrohemostasis of actual or potential bleeding sites in the gastrointestinal tract, including, but not limited to: the esophagus, stomach, duodenum, and colon. These sites include the following: peptic ulcers; Mallory-Weiss tears; arteriovenous malformations (AVMs); Dieulafoy lesions; bleeding polyp stocks; and angiomata.
### C. CONTRAINDICATIONS
Contraindications for this device are those specific to injection therapy and bipolar electrohemostasis treatments. These contraindications include, but are not limited to: non-focal bleeding sites; esophageal/gastric varices; diffuse lesions; allergies to injection agents; and use of the Injection Gold Probe with monopolar electrosurgical generators
Page 42 of 159
{2}
Boston Scientific Corporation
May 10, 1994
510(k) Premarket Notification
Microvasive® Injection Gold Probe™
# D. POTENTIAL COMPLICATIONS
Potential complications include, but may not be limited to: perforation; bleeding, post-injection ulceration with delayed bleeding; aspiration pneumonia; pleural effusion; other respiratory difficulties; hepatic failure; septicemia/infection; chest pain; esophageal ulcers; esophageal strictures; and dysphagia.
# E. DESCRIPTIVE AND TECHNOLOGICAL CHARACTERISTICS OF PROPOSED AND PREDICATE DEVICES
Boston Scientific Corporation believes that the proposed Microvasive Injection Gold Probe is substantially equivalent to the currently marketed Microvasive Gold Probe™ and Microvasive Variject™ Clear Injection Therapy Needle catheter. Table 1 compares the descriptive characteristics of these products.
)
)
Page 43 of 159
{3}
Boston Scientific Corporation
May 10, 1994
510(k) Premarket Notification
Microvasive® Injection Gold Probe™
Table 1:
Similarities and Differences between the
Proposed Microvasive Injection Gold Probe and the
Predicate Microvasive Gold Probe and Variject Clear Injection Therapy Needle Catheter
| Device Characteristic | Microvasive® Gold Probe™ (Predicate Device) (K885005) | Microvasive® Variject™ Clear Injection Therapy Needle Catheter (Predicate Device) (K854238) | Microvasive® Injection Gold Probe™ (Proposed Device) (This 510(k)) |
| --- | --- | --- | --- |
| Indications for Use | Endoscopic electrohemostasis of visible bleeding and non-bleeding sites in the gastrointestinal tract. | Used endoscopically to introduce a sclerosing agent or vasoconstrictor into selected sites to control actual or potential bleeding lesions in the digestive system. | Endoscopic injection therapy and electrohemostasis of actual or potential bleeding sites in the gastrointestinal tract. |
| **Catheter Shaft:** | | | |
| Material | Confidential | Confidential | Confidential |
| Usable Length | 290 cm - 350 cm | 200 cm | 210 - 350 cm |
| Shaft OD | 7 - 10 F | 5.4 F (1.8 mm) | 7 - 10 F |
| Coating | Confidential | NA | Confidential |
| **Bipolar Electrohemostasis:** | | | |
| Bipolar Ceramic Tip | Confidential | NA | Confidential |
| Electrical Wires | Confidential | NA | Confidential |
| Generator Connector | Bipolar Erbe Connector | NA | Bipolar Erbe Connector |
| Electrical Bifurcation | Confidential | NA | Confidential |
| **Injection Therapy:** | | | |
| Needle Material | NA | Confidential | Confidential |
| Needle Gauge | NA | 23 & 25 Gauge | 25 Gauge |
| Needle Extension Length | NA | 5 mm | 4 - 6 mm |
| Needle Luer Hub Material | NA | Confidential | Confidential |
| **Irrigation Capability:** | | | |
| Irrigation Capability | Yes | No | Yes |
| Irrigation Luer Hub Material | Confidential | NA | Confidential |
Page 44 of 159
{4}
Boston Scientific Corporation
May 10, 1994
510(k) Premarket Notification
Microvasive® Injection Gold Probe™
# F. PERFORMANCE CHARACTERISTICS
In vitro laboratory testing regarding both clinical performance characteristics and product integrity was performed on finished, sterilized Injection Gold Probe catheters to verify the device's safety and effectiveness. In addition, Full Tripartite Biocompatibility testing was performed on the Injection Gold Probe with satisfactory results.
# G. PACKAGING, STERILIZATION, AND PYROGENICITY
The package is the same as that presently used for the currently marketed Microvasive Gold Probe. The Injection Gold Probe will be sterilized using ethylene oxide gas using the AAMI protocol for ethylene oxide sterilization. Pyrogenicity testing will be performed on a periodic basis to monitor bacterial endotoxin levels.
# H. CONCLUSION
Boston Scientific Corporation believes that the proposed Microvasive Injection Gold Probe is substantially equivalent to the Microvasive Gold Probe and Variject catheters. As demonstrated in Table 1, the Injection Gold Probe combines their indications for use, while combining their designs. In addition, BSC has presented Laboratory and Biocompatibility Testing data to verify the safety and effectiveness of the proposed Injection Gold Probe.
Based on the information presented, Boston Scientific Corporation believes that the proposed Microvasive Injection Gold Probe is safe and effective and it will meet the minimum requirements that are considered acceptable for its intended use.
Injection Gold Probe is a trademark of Boston Scientific Corporation.
Gold Probe is a trademark of Boston Scientific Corporation.
Variject is a trademark of Boston Scientific Corporation.
Microvasive is a registered trademark of Boston Scientific Corporation.
Page 45 of 159
{5}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
NOV 2 1994
Food and Drug Administration
1390 Piccard Drive
Rockville MD 20850
Ms. Sheri D. O'Brien
Senior Regulatory Affairs Specialist
Boston Scientific Corporation
480 Pleasant Street
Watertown, MA 02172-2414
Re: K942301/B
Microvasive Injection Gold Probe
Dated: October 19, 1994
Received: October 20, 1994
Regulatory Class: II
21 CFR 876.4300 Procode: 78 KNS
Dear Ms. O'Brien:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter immediately will allow you to begin marketing your device as described in your 510(k) premarket notification. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market, but it does not mean that FDA approves your device. Therefore, you may not promote or in anyway represent your device or its labeling as being approved by FDA. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), promotion, or advertising please contact the Office of Compliance, Promotion and Advertising Policy Staff (HFZ-300) at (301) 594-4639. Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Lillian Yin
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.