K940942 · Medison America, Inc. · IYO · Feb 19, 1997 · Radiology
Device Facts
Record ID
K940942
Device Name
COMBISON 530 DIAGNOSTIC ULTRASOUND SCANNER
Applicant
Medison America, Inc.
Product Code
IYO · Radiology
Decision Date
Feb 19, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
This device is intended for use with the listed transducers for the transcutaneous, transrectal, and transvaginal imaging of organs or structures as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the transcutaneous, or transvaginal imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity.
Device Story
COMBISON 530 is a diagnostic ultrasound scanner providing B-mode, M-mode, Pulsed/Continuous Wave Doppler, and Color Flow imaging. It features the VOLUSON 3D-scan system, which performs automated sweeps of adjacent B-scans to reconstruct and visualize volumes. Operated by clinicians in clinical settings, the device uses various mechanical sector and array transducers to acquire acoustic data. The system processes these signals to display real-time images or 3D volumes. Clinicians use these outputs to detect and assess physical/functional abnormalities and perform fetal measurements. The device aids clinical decision-making by providing diagnostic visualization of internal structures and fetal development.
Clinical Evidence
Bench testing only. No clinical studies were conducted. Performance was validated via nonclinical testing including electrical safety (IEC 601.1) and acoustic output measurements (TRACK III). Imaging performance (gain, depth compensation, magnification, frame rate, measurement accuracy) was verified using phantoms under the Kretztechnik AG Transducer-Software Check Protocol. Acoustic output values were documented to be within established limits and comparable to predicate devices.
Technological Characteristics
Diagnostic ultrasound system using mechanical sector and array transducers. Materials include UDEL polysulphone, TPX, LUCALEN, ULTRADUR, and ABS. Sensing via annular, curved linear, and linear arrays. Energy source: electrical/ultrasonic. Connectivity: standalone. Software: includes 3D volume reconstruction (VOLUSON). Electrical safety: IEC 601.1 (Class 1, Type BF).
Indications for Use
Indicated for transcutaneous, transrectal, and transvaginal imaging of organs, structures, and fetal/uterine structures in obstetric patients to assess physical/functional abnormalities, fetal growth, and maturity.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
1/15
K940942
FEB 19 1997
# 510(k) SUMMARY
## 1. COMPANY INFORMATION.
Name : Kretztechnik AG
Address : Tiefenbach 15
A-4871 Zipf, Austria
Telephone : 011-43-7682-2261-0
Contact Person (U.S.) : Jay C. Cone, President
Cone Instruments, Inc.
5201 Naiman Parkway
Solon, Ohio 44139
(216) 248-1035
Date Prepared : June 28, 1993 (Revised: February 18, 1994)
## 2. DEVICE NOMENCLATURE.
### A. System.
Trade Name : COMBISON® 530
(VOLUSON® for volume scanning feature)
Common Name : Diagnostic Ultrasound Scanner
Classification Names : Ultrasonic Pulsed Echo Imaging System
Ultrasonic Pulsed Doppler Imaging System
Ultrasonic Obstetric-Gynecologic Imager
### B. Transducers.
Trade Names : Sector Transducer Model No. AWP 3.5
Sector Transducer Model No. SWP 3.5
Sector Transducer Model No. VSW 3.5
Sector Transducer Model No. VSW 5.0
Sector Transducer Model No. VWP 3.5 B
Sector Transducer Model No. VWP 5.0
Sector Transducer Model No. VW 7/10
Transrectal Transducer Model No. VRW 77 AK
Transvaginal Transducer Model No. EW 5/7 K
Transvaginal Transducer Model No. VEW 7.5 R
Common Name : Mechanical Sector Ultrasound Imaging Transducer
Classification Name : Diagnostic Ultrasonic Transducer
Trade Names : Curved Linear Array Transducer Model ACA 3.5
Curved Linear Array Transducer Model ACA 5.0
Linear Array Transducer Model PLA 7.5
Common Name : Linear Array Ultrasound Imaging Transducer
Classification Name : Diagnostic Ultrasound Transducer
{1}
2/15
# 3. PREDICATE DEVICES.
a. For Diagnostic Ultrasound Scanner COMBISON® 530:
Kretztechnik Diagnostic Ultrasound Scanner COMBISON® 330
Toshiba Diagnostic Ultrasound Scanner SSH-140A
Siemens Diagnostic Ultrasound Scanner SONOLINE SI-450
b. For Sector Transducer AWP 3.5:
Previously cleared with COMBISON® 330 as Model AWP 3.5/A
c. For Sector Transducer SWP 3.5:
Kretztechnik Sector Transducer AWP 3.5/A
Kretztechnik Sector Transducer AW 4/5 B/A
d. For Sector Transducer VSW 3.5:
Kretztechnik Sector Transducer AWP 3.5/A
e. For Sector Transducer VSW 5.0:
Kretztechnik Sector Transducer AWP 3.5/A
f. For Sector Transducer VWP 3.5 B:
Kretztechnik Sector Transducer AWP 3.5/A
g. For Sector Transducer VWP 5.0:
Kretztechnik Sector Transducer AWP 3.5/A
h. For Sector Transducer VW 7/10:
Kretztechnik Sector Transducer NW 17.5 B/A
i. For Transrectal Transducer VRW 77 AK:
Kretztechnik Transrectal Transducer VRW 177 AK/A
Kretztechnik Transrectal Transducer IRW 177 AK/A
j. For Transvaginal Transducer EW 5/7 K:
Kretztechnik Transvaginal/Urologic Transducer WIW 17.5 AP/A
Kretztechnik Transvaginal Transducer IR 15/7 BG/A
k. For Transvaginal Transducer VEW 7.5 R:
Kretztechnik Transvaginal/Urologic Transducer WIW 17.5 AP/A
Kretztechnik Transvaginal Transducer IR 15/7 BG/A
l. For Curved Linear Array Transducer ACA 3.5:
Acoustic Imaging Convex Sector Transducer CLA 3.5
m. For Curved Linear Array Transducer ACA 3.5:
Acoustic Imaging Convex Sector Transducer CLA 5.0
n. For Linear Array Transducer PLA 7.5:
Acoustic Imaging Linear Array Transducer MLA 7.5
{2}
3/15
# 4. DEVICE DESCRIPTION.
a. **Scanner COMBISON® 530:**
This device is intended to obtain and display B-scan and M-mode images as well as Pulsed Wave and Continuous Wave Doppler spectra, or Color Flow images. Additionally, the COMBISON® 530 offers the VOLUSON® 3D-scan system, which performs a number of B-scans, one adjacent to the other (only with special probes) and gives the user the possibility to visualize planes within the "volume" without the patient.
b. **Transducer AWP 3.5:**
This device is a hand-held, annular array, mechanical sector probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 3.5 MHz. In B-mode the transducer operates as an end-firing sector probe over either of two selectable angle ranges (80°/60°). In M-mode and Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle ranges and operates as a stationary probe.
c. **Transducer SWP 3.5:**
This device is a hand-held, annular array, mechanical sector probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 3.5 MHz. In B-mode the transducer operates as an end-firing sector probe over either of four selectable angle ranges (70°/60°/50°/40°). In M-mode and Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle ranges and operates as a stationary probe.
d. **Transducer VSW 3.5:**
This device is a hand-held, annular array, mechanical sector probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 3.5 MHz. In B-mode the transducer operates as an end-firing sector probe over either of four selectable angle ranges (70°/60°/50°/40°). In M-mode and Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle ranges and operates as a stationary probe. With the COMBISON® 530 this transducer offers the possibility of an automated ±30° sweep of the scan plane. The transducer performs a number of B-scans, one adjacent to the other, to have a "volume" scanned.
e. **Transducer VSW 5.0:**
This device is a hand-held, annular array, mechanical sector probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 5.0 MHz. In B-mode the transducer operates as an end-firing sector probe over either of four selectable angle ranges (70°/60°/50°/40°). In M-mode and Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle ranges and operates as a stationary probe. With the COMBISON® 530 this transducer offers the possibility of an automated ±30° sweep of the scan plane. The transducer performs a number of B-scans, one adjacent to the other, to have a "volume" scanned.
{3}
4/15
f. Transducer VWP 3.5 B:
This device is a hand-held, annular array, mechanical sector probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 3.5 MHz. In B-mode the transducer operates as an end-firing sector probe over either of two selectable angle ranges (90°/60°). In M-mode and Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle ranges and operates as a stationary probe. With the COMBISON® 530 this transducer offers the possibility of an automated ±15° sweep of the scan plane. The transducer performs a number of B-scans, one adjacent to the other, to have a "volume" scanned.
g. Transducer VWP 5.0:
This device is a hand-held, annular array, mechanical sector probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 5.0 MHz. In B-mode the transducer operates as an end-firing sector probe over either of two selectable angle ranges (90°/60°). In M-mode and Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle ranges and operates as a stationary probe. With the COMBISON® 530 this transducer offers the possibility of an automated ±15° sweep of the scan plane. The transducer performs a number of B-scans, one adjacent to the other, to have a "volume" scanned.
h. Transducer VW 7/10:
This device is a hand-held, dual-element, mechanical sector probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequencies are 7 and 10 MHz. In B-mode the transducer operates as an end-firing sector probe over either of two selectable angle ranges (90°/60°). In M-mode and Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle ranges and operates as a stationary probe. With the COMBISON® 530 this transducer offers the possibility of an automated ±15° sweep of the scan plane. The transducer performs a number of B-scans, one adjacent to the other, to have a "volume" scanned.
i. Transducer VRW 77 AK:
This device is a hand-held, dual-element, mechanical sector probe intended for transrectal use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 7.5 MHz. In B-mode the transducer operates as an end-firing sector probe over either of two selectable scan planes with different angle ranges (longitudinal angle range 100°; transversal angle range 350°). In Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle ranges and operates as a stationary probe. With the COMBISON® 530 this transducer offers the possibility of an automated ±140° sweep of the scan plane. The transducer performs a number of B-scans, one adjacent to the other, to have a "volume" scanned.
j. Transducer EW 5/7 K:
This device is a hand-held, dual-element, mechanical sector probe intended for transvaginal use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating
{4}
5/15
frequencies are 5 and 7.5 MHz. In B-mode the transducer operates as an end-firing sector probe over either of two selectable angle ranges (130°/200°). In M-mode and Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle ranges and operates as a stationary probe.
k. Transducer VEW 7.5 R:
This device is a hand-held, single-element, mechanical sector probe intended for transvaginal use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 7.5 MHz. In B-mode the transducer operates as an end-firing sector probe over an angle range of 100°. In M-mode and Pulsed Wave Doppler-mode the transducer is locked in one selectable position within the available sector angle range and operates as a stationary probe. With the COMBISON® 530 this transducer offers the possibility of an automated 180° rotation of the scan plane. The transducer performs a number of B-scans, one adjacent to the other, to have a "volume" scanned.
l. Transducer ACA 3.5:
This device is a hand-held, 128 element, curved linear array probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 3.5 MHz. In B-mode the transducer operates as an end-firing sector probe over an angle range of 58°. In M-mode and Pulsed Wave Doppler-mode specific element groups can be activated within the array and operate as a stationary probe.
m. Transducer ACA 5.0:
This device is a hand-held, 128 element, curved linear array probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 5.0 MHz. In B-mode the transducer operates as an end-firing sector probe over an angle range of 64°. In M-mode and Pulsed Wave Doppler-mode specific element groups can be activated within the array and operate as a stationary probe.
n. Transducer PLA 7.5:
This device is a hand-held, 128 element, linear array probe intended for transcutaneous use with the COMBISON® 530 Diagnostic Ultrasound Scanner. The nominal operating frequency is 7.5 MHz. In B-mode the transducer operates as an end-firing probe over a width of 55 mm. In M-mode and Pulsed Wave Doppler-mode specific element groups can be activated within the array and operate as a stationary probe.
{5}
6/15
# 5. INTENDED USES.
## a. Scanner COMBISON® 530:
This device is intended for use with the listed transducers for the transcutaneous, transrectal, and transvaginal imaging of organs or structures as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the transcutaneous, or transvaginal imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. The same clinical uses were recommended for predicate devices, COMBISON® 330, SSH-140A, and SONOLINE SI-450.
## b. Transducer AWP 3.5:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of intraabdominal organs and structures including the gastrointestinal tract, kidney, bladder, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the transcutaneous imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. The same clinical uses were recommended when this device was previously cleared with the COMBISON® 330.
## c. Transducer SWP 3.5:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of intraabdominal organs and structures including the gastrointestinal tract, kidney, bladder, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the transcutaneous imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. The same clinical uses were recommended for predicate devices AWP 3.5/A and AW 14/5 B/A.
## d. Transducer VSW 3.5:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of intraabdominal organs and structures including the gastrointestinal tract, kidney, bladder, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the transcutaneous imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. The same clinical uses were recommended for predicate device AWP 3.5/A.
## e. Transducer VSW 5.0:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of intraabdominal organs and structures including the gastrointestinal tract, kidney, bladder, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the
{6}
7/15
transcutaneous imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. The same clinical uses were recommended for predicate device AWP 3.5/A.
f. Transducer VWP 3.5 B:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of intraabdominal organs and structures including the gastrointestinal tract, kidney, bladder, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the transcutaneous imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. The same clinical uses were recommended for predicate device AWP 3.5/A.
g. Transducer VWP 5.0:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of intraabdominal organs and structures including the gastrointestinal tract, kidney, bladder, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the transcutaneous imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. The same clinical uses were recommended for predicate device AWP 3.5/A.
h. Transducer VW 7/10:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of peripheral vessels and small organs including breast, thyroid, testes, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria. The same clinical uses were recommended for predicate device NW 17.5 BG/A.
i. Transducer VRW 77 AK:
This device is intended for use with the COMBISON® 530 system for the transrectal imaging of the rectum and surrounding organs and structures including the rectum wall, prostate gland, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria. An approved biopsy attachment is available for use with this transducer. The same clinical uses were recommended for predicate device VRW 177 AK/A and IRW 177 AK/A.
j. Transducer EW 5/7 K:
This device is intended for use with the COMBISON® 530 system in the imaging of the fetus and uterine structures utilizing the transvaginal approach as an aid in the diagnosis of obstetrical and gynecological conditions including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. An approved biopsy attachment is available for use with this transducer. The same clinical uses were recommended for the predicate devices
{7}
8/15
WIW 17.5 AP/A and IR 15/7 BG/A.
k. Transducer VEW 7.5 R:
This device is intended for use with the COMBISON® 530 system in the imaging of the fetus and uterine structures utilizing the transvaginal approach as an aid in the diagnosis of obstetrical and gynecological conditions including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. An approved biopsy attachment is available for use with this transducer. The same clinical uses were recommended for the predicate devices WIW 17.5 AP/A and IR 15/7 BG/A.
l. Transducer ACA 3.5:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of intraabdominal organs and structures including the gastrointestinal tract, kidney, bladder, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the transcutaneous imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. The same clinical uses were recommended for predicate device CLA 3.5.
m. Transducer ACA 5.0:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of intraabdominal organs and structures including the gastrointestinal tract, kidney, bladder, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria; and in obstetric patients for the transcutaneous imaging of the fetus and uterine structures including the performance of fetal measurements for the purpose of assessing fetal growth and maturity. The same clinical uses were recommended for predicate device CLA 5.0.
n. Transducer PLA 7.5:
This device is intended for use with the COMBISON® 530 system for the transcutaneous imaging of peripheral vessels and small organs including breast, thyroid, testes, etc. as an aid in the detection and assessment of physical and functional abnormalities using established diagnostic criteria. The same clinical uses were recommended for predicate device MLA 7.5.
{8}
9/15
# 6. COMPARISON WITH PREDICATE DEVICES(S).
## a. Scanner COMBISON 530:
This scanner is capable of B-, M-, Pulsed Wave and Continuous Wave Doppler, and Color Flow Mapping Modes. The predicate scanner COMBISON 330 is only capable of B- and M-Mode, while the Siemens SONOLINE SI-450 scanner also includes Pulsed Wave and Continuous Wave Doppler, and the Toshiba SSH-140A offers Color Flow Mapping in addition. The COMBISON 530 volume-scanning feature is an extension of the cine-loop storage/ retrieval capability found in all three predicate devices.
## b. Transducer AWP 3.5:
This transducer was already approved for use with the scanner COMBISON 330. There has been no change performed to the original version. Safety and effectiveness are not affected.
## c. Transducer SWP 3.5:
This transducer differs from the predicate device only in minor respects; none of these minor differences adversely effects safety or effectiveness. The nominal frequency of both transducers is 3.5 MHz and both utilize a mechanically sectored annular array; a three-phase motor is used to provide the rocking motion in transducer SWP 3.5, a stepper motor in the AWP 3.5/A. Both have similar sector angle ranges--70°, 60°, 50°, or 40° selectable in the case of the SWP 3.5, 80° fixed in the case of AWP 3.5/A.
Both of the patient contact materials utilized in transducer SWP 3.5 (UDEL polysulphone and TPX) have been employed in the previously marketed transducer AWP 3.5/A.
## d. Transducer VSW 3.5:
This transducer differs from the predicate device only in minor respects; none of these minor differences adversely effects safety or effectiveness. The nominal frequency of both transducers is 3.5 MHz and both utilize a mechanically sectored annular array; a three-phase motor is used to provide the rocking motion in transducer VSW 3.5, a stepper motor in the AWP 3.5/A. Both have similar sector angle ranges--70°, 60°, 50°, or 40° selectable in the case of the SWP 3.5, 80° fixed in the case of AWP 3.5/A.
The VSW 3.5 is also able to move the B-scan plane to obtain a VOL-scan.
Both of the patient contact materials utilized in transducer VSW 3.5 (LUCALEN and ULTRADUR) have been employed in food-, drug-, and cosmetics packaging.
## e. Transducer VSW 5.0:
This transducer differs from the predicate device only in minor respects; none of these minor differences adversely effects safety or effectiveness. The nominal operating frequency of the VSW 5.0 is 5.0 MHz, while, for transducer AWP 3.5/A, the nominal frequency is 3.5 MHz. Both utilize a mechanically sectored annular array; a three-phase motor is used to provide the rocking motion in transducer VSW 5.0, a stepper motor in the AWP 3.5/A. Both have similar sector angle ranges--70°, 60°, 50°, or 40° selectable in the case of the SWP 3.5, 80° fixed in the case of AWP 3.5/A.
{9}
10/15
The VSW 3.5 is also able to move the B-scan plane to obtain a VOL-scan.
Both of the patient contact materials utilized in transducer VSW 5.0 (LUCALENᵃ and ULTRADURᵃ) have been employed in food-, drug-, and cosmetics packaging.
f. Transducer VWP 3.5 B:
This transducer differs from the predicate device only in minor respects; none of these minor differences adversely effects safety or effectiveness. The nominal frequency of both transducers is 3.5 MHz and both utilize a mechanically sectored annular array; a stepper motor is used to provide the rocking motion in both transducers. Both have similar sector angle ranges--90°, 60° selectable in the case of the VWP 3.5 B, 80° fixed in the case of AWP 3.5/A.
The VWP 3.5 B is also able to move the B-scan plane to obtain a VOL-scan.
Both of the patient contact materials utilized in transducer VWP 3.5 B (LUCALENᵃ and ULTRADURᵃ) have been employed in food-, drug-, and cosmetics packaging.
g. Transducer VWP 5.0:
This transducer differs from the predicate device only in minor respects; none of these minor differences adversely effects safety or effectiveness. The nominal operating frequency of the VWP 5.0 is 5.0 MHz, while, for transducer AWP 3.5/A, the nominal frequency is 3.5 MHz. Both utilize a mechanically sectored annular array; a stepper motor is used to provide the rocking motion in both transducers. Both have similar sector angle ranges--90°, 60° selectable in the case of the VWP 5.0, 80° fixed in the case of AWP 3.5/A.
The VWP 5.0 is also able to move the B-scan plane to obtain a VOL-scan.
Both of the patient contact materials utilized in transducer VWP 5.0 (LUCALENᵃ and ULTRADURᵃ) have been employed in food-, drug-, and cosmetics packaging.
h. Transducer VW 7/10:
This transducer differs from the predicate device only in minor respects; none of these minor differences adversely effects safety or effectiveness. The nominal operating frequencies of the VW 7/10 are 7.5 and 10 MHz, while, for transducer NW 17.5 B/A, the nominal frequency is only 7.5 MHz. Both utilize a mechanically sectored single-element; a stepper motor is used to provide the rocking motion in both transducers. Both have similar sector angle ranges--110°, 80° selectable in the case of the VW 7/10, 80° fixed in the case of NW 17.5 B/A.
The VW 7/10 is also able to move the B-scan plane to obtain a VOL-scan.
Both of the patient contact materials utilized in transducer VW 7/10 (LUCALENᵃ and ULTRADURᵃ) have been employed in food-, drug-, and cosmetics packaging.
i. Transducer VRW 77 AK:
This transducer differs from the predicate device only in minor respects; none of these minor differences adversely
{10}
11/15
effects safety or effectiveness. The only difference between the transducer VRW 77 AK and the predicate transducer VRW 177 AK/A is the longitudinal scan angle range, which is 100° in case of the VRW 77 AK transducer and 90° in case of the predicate transducer VRW 177 AK/A.
Both of the patient contact materials utilized in transducer VRW 77 AK have been employed in the previously marketed transducers AWP 3.5/A and VRW 177 AK/A.
j. Transducer EW 5/7 K:
This transducer differs from the predicate transducers only in minor respects; none of these minor differences adversely affect safety or effectiveness. Transvaginal transducer EW 5/7 K offers selectable nominal operating frequencies of 5.0 and 7.5 MHz. Predicate transducer IR15/7BG/A operated at a fixed nominal frequency of 5.0 MHz, while transducer WIW17.5AP/A operated at a fixed nominal frequency of 7.5 MHz. Mechanical sectoring of the two elements in transducer EW 5/7 K is accomplished by means of a rope drive; mechanical sectoring of both transducer IR15/7BG/A and transducer WIW17.5AP/A is accomplished using a cogwheel gear mechanism. Transducer EW 5/7 K offers two selectable sector angle ranges (130° or 200°). The sector angle range of transducer IR15/7BG/A was fixed at 240°; the sector angle range of WIW17.5AP/A was fixed at 118°.
Both of the patient contact materials utilized in transducer EW 5/7 K (TPX and UDEL® polysulphone) have been employed in previously marketed transducers--TPX in transducer IR15/7BG/A and WIW17.5AP/A, and UDEL® polysulphone in sector transducer AWP3.5/A.
k. Transducer VEW 7.5:
This transducer differs from the predicate transducers only in minor respects; none of these minor differences adversely affect safety or effectiveness. Transvaginal transducer VEW 7.5 R offers a nominal operating frequency of 7.5 MHz. Predicate transducer IR15/7BG/A operated at a fixed nominal frequency of 5.0 MHz, while transducer WIW17.5AP/A operated at a fixed nominal frequency of 7.5 MHz. Mechanical sectoring of the single element in transducer VEW 7.5 R is accomplished by means of a rope drive; mechanical sectoring of both transducer IR15/7BG/A and transducer WIW17.5AP/A is accomplished using a cogwheel gear mechanism. Transducer VEW 7.5 R offers a sector angle range of 100°. The sector angle range of transducer IR15/7BG/A was fixed at 240°; the sector angle range of WIW17.5AP/A was fixed at 118°. The VEW 7.5 R is also able to rotate the B-scan plane 180°. (Combined motion in case of VOL-scan.)
Each of the patient contact materials utilized in transducer VEW 7.5 R (TPX and ULTRADUR®/ABS) have been employed in previously marketed transducers--TPX in transducer IR15/7BG/A and WIW17.5AP/A. ULTRADUR®/ABS is widely used for food-, drug-, and cosmetics packaging.
l. Transducer ACA 3.5:
This transducer differs from the predicate transducers in no respects adversely affecting safety or effectiveness. The ACA 3.5 type transducer is manufactured and also marketed by
{11}
12/15
Acoustic Imaging as model CLA 3.5 (Kretztechnik is an OEM customer of Acoustic Imaging).
m. Transducer ACA 5.0:
This transducer differs from the predicate transducers in no respects adversely affecting safety or effectiveness. The ACA 5.0 type transducer is manufactured and also marketed by Acoustic Imaging as model CLA 5.0 (Kretztechnik is an OEM customer of Acoustic Imaging).
n. Transducer PLA 7.5:
This transducer differs from the predicate transducers in no respects adversely affecting safety or effectiveness. The PLA 7.5 type transducer is manufactured and also marketed by Acoustic Imaging as MLA 7.5 (Kretztechnik is an OEM customer of Acoustic Imaging).
{12}
13/15
# 7. PERFORMANCE DATA.
a. **Nature of Data Submitted:** Appropriate nonclinical tests as described below have been performed to assess pertinent performance variables. Formal clinical studies were not conducted because the scanner COMBISON® 530, as well as the listed transducers, do not raise new questions concerning safety or effectiveness when compared with predicate devices.
b. **Electrical Safety:** Tests utilizing the listed transducers with COMBISON® 530 scanner were performed to measured electrical leakage currents according to IEC Standard 601.1. Guard wire resistance and the following leakage currents: (1) ground, (2) housing, (3) patient device, and (4) mains voltage at patient device were measured and found to be well within the limits for Class 1, Type BF devices established by IEC Standard 601.1.
c. **Imaging Performance:** Performance characteristics including gain, depth compensation, magnification, frame rate, image orientation, measurement accuracy, sector angle, and noise level for each possible operating mode were tested using appropriate phantoms under the Kretztechnik AG Transducer-Software Check Protocol. All performance values were found to be within the limits stated in product specifications.
d. **Acoustic Output:** The maximum acoustic output for each transducer and each operating mode was measured under worst-case conditions and documented according to TRACK III reporting procedures as described in FDA guidelines. Results, expressed as derated values to account for assumed attenuation in tissue, were as follows:
| TRANSDUCER | NOMINAL FREQUENCY | MODE | I_{SFTA} (mW/cm²) | I_{SFPA} (W/cm²) |
| --- | --- | --- | --- | --- |
| AWP 3.5 | 3.5 MHz | B-Mode | 26.40 | 101.65 |
| | | M-Mode | 49.00 | 102.00 |
| | | PW-Mode | 574.00 | 30.00 |
| SWP 3.5 | 3.5 MHz | B-Mode | 10.90 | 145.88 |
| | | M-Mode | 58.00 | 146.00 |
| | | PW-Mode | 458.00 | 27.00 |
| VSW 3.5 | 3.5 MHz | B-Mode | 20.40 | 159.26 |
| | | M-Mode | 65.00 | 159.00 |
| | | PW-Mode | 676.00 | 21.00 |
| | | VOL-Mode | 2.00 | 159.26 |
| VSW 5.0 | 5.0 MHz | B-Mode | 9.80 | 171.74 |
| | | M-Mode | 60.00 | 172.00 |
| | | PW-Mode | 543.00 | 27.00 |
| | | VOL-Mode | 0.40 | 171.74 |
{13}
14/15
| TRANSDUCER | NOMINAL FREQUENCY | MODE | I_{SPTA} (mW/cm²) | I_{SPFA} (W/cm²) |
| --- | --- | --- | --- | --- |
| VWP 3.5 B | 3.5 MHz | B-Mode | 46.30 | 159.26 |
| | | M-Mode | 65.00 | 159.00 |
| | | PW-Mode | 676.00 | 21.00 |
| | | VOL-Mode | 5.10 | 159.26 |
| VWP 5.0 | 5.0 MHz | B-Mode | 15.50 | 171.74 |
| | | M-Mode | 60.00 | 172.00 |
| | | PW-Mode | 396.00 | 22.00 |
| | | VOL-Mode | 1.10 | 171.74 |
| VW 7/10 | 7.5 MHz | B-Mode | 9.00 | 170.36 |
| | | M-Mode | 43.00 | 170.00 |
| | | PW-Mode | 463.00 | 85.00 |
| | | VOL-Mode | 0.60 | 170.36 |
| | 10 MHz | B-Mode | 6.00 | 163.14 |
| | | M-Mode | 31.00 | 163.00 |
| | | PW-Mode | 463.00 | 85.00 |
| | | VOL-Mode | 0.30 | 163.14 |
| EW 5/7 K | 5.0 MHz | B-Mode | 5.40 | 67.84 |
| | | M-Mode | 31.00 | 68.00 |
| | | PW-Mode | 550.00 | 29.00 |
| | 7.5 MHz | B-Mode | 5.40 | 145.12 |
| | | M-Mode | 32.00 | 145.00 |
| | | PW-Mode | 3.90 | 20.00 |
| VEW 7.5 R | 7.5 MHz | B-Mode | 3.80 | 140.25 |
| | | M-Mode | 20.00 | 140.00 |
| | | PW-Mode | 355.00 | 29.00 |
| | | VOL-Mode | 3.80 | 140.25 |
| VRW 77 AK | 7.5 MHz | B-Mode | 7.80 | 137.51 |
| | | M-Mode | 34.00 | 138.00 |
| | | PW-Mode | 474.00 | 101.00 |
| | | VOL-Mode | 0.50 | 164.97 |
| ACA 3.5 | 3.5 MHz | B-Mode | 3.50 | 50.80 |
| | | M-Mode | 19.00 | 51.00 |
| | | PW-Mode | 431.00 | 15.00 |
| | | CFM-Mode | 564.60 | 61.30 |
{14}
15/15
| TRANSDUCER | NOMINAL FREQUENCY | MODE | I_{SPTA} (mW/cm²) | I_{SPFA} (W/cm²) |
| --- | --- | --- | --- | --- |
| ACA 5.0 | 5.0 MHz | B-Mode | 6.90 | 65.30 |
| | | M-Mode | 24.00 | 102.00 |
| | | PW-Mode | 427.00 | 16.00 |
| | | CFM-Mode | 685.70 | 118.20 |
| PLA 7.5 | 7.5 MHz | B-Mode | 3.50 | 77.40 |
| | | M-Mode | 9.00 | 77.00 |
| | | PW-Mode | 429.00 | 89.00 |
| | | CFM-Mode | 703.30 | 70.00 |
All values obtained for the listed transducers during acoustic output testing were lower than the officially published maxima for previously approved devices for the same intended uses. Since the reported acoustic output values for all transducers listed in Section 3 as predicate devices were also lower than the established limits, the test data indicate substantial equivalence with respect to the acoustic output.
e. **Conclusions:** Nonclinical performance testing of the scanner COMBISON® 530 and the listed transducers has been performed, and the results demonstrate that these devices are safe, effective, and perform as well or better than the legally marketed devices with which this scanner and these transducers have been compared.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.