UNIVEC NON-REUSABLE SYRINGE

K933469 · Univec, Inc. · MEG · Apr 18, 1996 · General Hospital

Device Facts

Record IDK933469
Device NameUNIVEC NON-REUSABLE SYRINGE
ApplicantUnivec, Inc.
Product CodeMEG · General Hospital
Decision DateApr 18, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This syringe is intended for diabetics who inject insulin and healthcare professionals who inject insulin, tuberculin or immunization serum and other injectible medications.

Device Story

Sterile polypropylene piston syringe; 1 cc or 0.5 cc capacity; features 'umbrella' locking mechanism. Used by diabetics or healthcare professionals for aspiration, mixing, and injection of medications. Locking mechanism activated by depressing plunger past resistance point; prevents syringe reuse. Device functions as standard syringe until locked; provides safety benefit by rendering device non-reusable post-injection.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Polypropylene construction; piston syringe design; manual operation; 1 cc or 0.5 cc capacity; non-reusable locking mechanism; sterile.

Indications for Use

Indicated for diabetics requiring insulin injections and healthcare professionals administering insulin, tuberculin, immunization serum, or other injectable medications.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Submission Summary (Full Text)

{0} K933469 # SUMMARY OF SAFETY AND EFFECTIVENESS OF UNIVEC NONREFUSABLE LOCKING SYRINGE (1) Submitter’s Information: UNIVEC, Inc. 999 Franklin Avenue Garden City, New York 11530 (516) 294-1000 (tel) (516) 739-3343 (fax) (2) Device Name: Proprietary Name: Model 884 Umbrella Nonreusable Syringe Common or Usual Name: Piston, Hypodermic Needle Classification Names: Piston Syringe 21 C.F.R. 880.5860 Class II Hypodermic Single Lumen Needle 21 C.F.R. 880.5570 Class II (3) Predicate Devices: UNIVEC Nonreusable Locking Syringe (K943679) FLP Enterprises LOKSHOT Syringe (K891467) (4) Description of Device: The UNIVEC Model 884 Umbrella Nonreusable Syringe is a sterile piston syringe with 1 cc or 0.5 cc capacities fabricated of polypropylene. It has a locking device made of the same polypropylene material. The “umbrella” description describes the locking mechanism which emulates the shape and functioning of an umbrella. Functioning in the traditional manner, this syringe is capable of aspirating and mixing various serums in its barrel by health practitioners and individuals for subsequent injection. It can be locked or prevented from further use by depressing the plunger toward the forward end of the barrel (and past noticeable resistance) in order to activate the locking device. Further use of the syringe is thus prevented. (5) Statement of Intended Use 57 {1} This syringe is intended for diabetics who inject insulin and healthcare professionals who inject insulin, tuberculin or immunization serum and other injectible medications. ## (6) Statement of Technological Characteristics The UNIVEC Nonreusable Syringe has no significant change in design, material, chemical composition, energy source or other characteristics compared to the predicate devices listed. The locking device, fabricated of polypropylene, the same material as the rest of the syringe, is the only difference with one of the predicates, the UNIVEC Model 888 Nonreusable Locking Syringe (K943679), (its locking device is made from stainless steel 304). There are no differences with the FLP Enterprises LOKSHOT Syringe, # K891467
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...