← Product Code [FFK](/productcode/FFK) · K920410

# EMILITH ELECTROMECHANICALIMPACTOR LITHOTRIPSY PROB (K920410)

_Psi Medical Products, Inc. · FFK · Feb 14, 1994 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K920410

## Device Facts

- **Applicant:** Psi Medical Products, Inc.
- **Product Code:** [FFK](/productcode/FFK.md)
- **Decision Date:** Feb 14, 1994
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4480
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Regulatory Identification

An electrohydraulic lithotriptor is an AC-powered device used to fragment urinary bladder stones. It consists of a high voltage source connected by a cable to a bipolar electrode that is introduced into the urinary bladder through a cystoscope. The electrode is held against the stone in a water-filled bladder and repeated electrical discharges between the two poles of the electrode cause electrohydraulic shock waves which disintegrate the stone.

## Special Controls

The special control for this device is FDA's “Guidance for the Content of Premarket Notifications for Intracorporeal Lithotripters.”

---

**Source:** [https://fda-staging.innolitics.com/device/K920410](https://fda-staging.innolitics.com/device/K920410)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).