NIDEK INC. MODELS YC-2000 & 3000 OPHTHALMIC LASERS

K892204 · Nidek, Inc. · LXS · Jun 22, 1989 · Ophthalmic

Device Facts

Record IDK892204
Device NameNIDEK INC. MODELS YC-2000 & 3000 OPHTHALMIC LASERS
ApplicantNidek, Inc.
Product CodeLXS · Ophthalmic
Decision DateJun 22, 1989
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4392
Device ClassClass 2

Regulatory Classification

Identification

The Nd:YAG laser for posterior capsulotomy and peripheral iridotomy consists of a mode-locked or Q-switched solid state Nd:YAG laser intended for disruption of the posterior capsule or the iris via optical breakdown. The Nd:YAG laser generates short pulse, low energy, high power, coherent optical radiation. When the laser output is combined with focusing optics, the high irradiance at the target causes tissue disruption via optical breakdown. A visible aiming system is utilized to target the invisible Nd:YAG laser radiation on or in close proximity to the target tissue.

Special Controls

*Classification.* Class II (special controls). Design Parameters: Device must emit a laser beam with the following parameters: wavelength = 1064 nanometers; spot size = 50 to 100 micros; pulse width = 3 to 30 nanoseconds; output energy per pulse = 0.5 to 15 millijoules (mJ); repetition rate = 1 to 10 pulses; and total energy = 20 to 120 mJ.

Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...