K262788 · Aura Wellness, LLC · SGT · Aug 31, 2026 · Physical Medicine
Device Facts
Record ID
K262788
Device Name
Nova HD+
Applicant
Aura Wellness, LLC
Product Code
SGT · Physical Medicine
Decision Date
Aug 31, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nova HD+ is indicated: • To temporarily increase local blood circulation in healthy leg muscles • To stimulate healthy muscles in order to improve or facilitate muscle performance
Device Story
Noninvasive physical medicine device; uses electrically generated magnetic fields to stimulate muscles; improves/facilitates muscle performance; increases local blood circulation. System comprises main control unit and five accessory coils (Large Loop, Butterfly Loop, Wings, Half Body Mat, Paddle) for targeted anatomical treatment. Used in home care and clinical settings; operated by patients or professionals. Operator uses 5-button interface to control magnetic pulse generator intensity and treatment duration (1-90 minutes). Microprocessor-controlled firmware manages pulse delivery. Output affects muscle tissue via magnetic field induction; benefits include muscle conditioning and circulation support.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included signal waveform output (volts) and magnetic flux output (mT) measurements across all intensity levels (1-10). Biocompatibility testing performed per ISO 10993-1, -5, and -10.
Technological Characteristics
Microprocessor-controlled powered muscle stimulator; noninvasive magnetic field induction. Main unit: 9.7" x 10.64" x 3.57", 5.6 lbs. Accessories: 5 coils (loops, mat, paddle). Power: 120 Vac, 60 Hz. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 62366-1, IEC 62304. Biocompatibility: ISO 10993 series. Sterilization: Not applicable (reusable, non-disposable).
Indications for Use
Indicated for adults (22+ years) to temporarily increase local blood circulation in healthy leg muscles and to stimulate healthy muscles to improve or facilitate muscle performance.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 31, 2026
Aura Wellness, LLC
Scott Blomberg
Director of Regulatory
2208 Plantside Dr.
Louisville, Kentucky 40299
Re: K262788
Trade/Device Name: Nova HD+
Regulation Number: 21 CFR 890.5850
Regulation Name: Powered Muscle Stimulator
Regulatory Class: Class II
Product Code: SGT
Dated: August 6, 2026
Received: August 6, 2026
Dear Scott Blomberg:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262788 - Scott Blomberg
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K262788 - Scott Blomberg
Page 3
Sincerely,
# Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262788 | ? |
| Please provide the device trade name(s). | | ? |
| Nova HD+ | | |
| Please provide your Indications for Use below. | | ? |
| The Nova HD+ is indicated: • To temporarily increase local blood circulation in healthy leg muscles • To stimulate healthy muscles in order to improve or facilitate muscle performance | | |
| Please select the types of uses. | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
# 510(k) SUMMARY
(as required by 21CFR 807.92)
# I Submitter
Aura Wellness, LLC
2208 Plantside Drive
Louisville, KY 40299
Phone: 502-714-1993
Fax: 502-369-5226
Submitter Contact:
Scott Blomberg
Submission Correspondent:
scott@aurawell.com
Establishment Registration:
10081462
Submission Date:
August 6, 2026
# II Device
Proprietary or Trade Name:
Nova HD+
Common/Usual Name:
Powered Muscle Stimulator
Classification Name:
Stimulator, Muscle, Powered, For Muscle Conditioning
Regulation:
890.5850
Regulatory Class:
2
Product Code:
SGT
Device Panel:
Neurological and Physical Medicine Devices
# III Primary Predicate Device
Nova HD+ (K220938)
Secondary Predicate
Aura Wave (K250896)
# IV Device Description
The Nova HD+ system is a noninvasive physical medicine device that uses electrically generated magnetic fields to stimulate muscles in order to improve and facilitate muscle performance.
The Nova HD+ is noninvasive, fully reusable (no disposable components such as electrodes) and have configurations allowing both home care and clinical professional use.
In addition to the main control unit, the Nova HD+ has five accessory attachments (coils). The attachments vary in size and form which allows the operator to choose which attachment administers optimal treatment to the desired anatomical part of the body.
The device contains firmware that controls the user interface. Through this user interface, the operator controls the magnetic pulse generator intensity and duration of the treatment.
Accessories to the device
| Name | Description | |
| --- | --- | --- |
| Large Loop | For treatment of upper back and shoulders, and lower back | Provided with system |
| Butterfly Loop | For treatment of knee, ankle, foot, hand, arm and shoulder | Optional accessory |
| Wings | For treatment of back, shoulders, and neck | Optional accessory |
| Half Body Mat | For treatment of back, lower back, hips, and hamstring muscles | Optional accessory |
| Paddle | For treatment of hip, thigh, buttocks, hands, and feet | Optional accessory |
{5}
# V Indication for Use
The Nova HD+ is indicated:
- To temporarily increase local blood circulation in healthy leg muscles.
- To stimulate healthy muscles in order to improve or facilitate muscle performance.
Environments of use: Over the Counter (OTC), both home care and clinical professional use
# VI Comparison of Technological Characteristics and Performance with the Predicate Devices
Table 1 below provides the summarized substantial equivalence comparison of general intended uses/actions, specific indications for use, equivalence of key clinical and technical features between subject and predicate device, along with a full listing of technical and conformance specifications.
Table 1: Comparison of Subject vs. Predicate Devices
| | | Nova HD+ Subject Device | Nova HD+ Primary Predicate Device K220938 | Aura Wave Secondary Predicate Device K250896 |
| --- | --- | --- | --- | --- |
| Classification Code | Regulation Identical Subject and Secondary Predicate Identical | Primary: Powered Muscle Stimulator SGT 890.5850 | Primary: Powered Muscle Stimulator NGX 890.5850 | Primary: Powered Muscle Stimulator SGT 890.5850 |
| Indication for Use | Subject and Predicate Similar Subject and Secondary Predicate Identical | The Nova HD+ is indicated: • To temporarily increase local blood circulation in healthy leg muscles. • To stimulate healthy muscles in order to improve and facilitate muscle performance | The Nova HD+ is indicated: • To stimulate healthy muscles in order to improve and facilitate muscle performance | The Aura Wave is indicated: • To temporarily increase local blood circulation in healthy leg muscles. • To stimulate healthy muscles in order to improve and facilitate muscle performance |
| Primary Mode of Action | Identical | Non-invasive tissue stimulation via magnetic field induction | Non-invasive tissue stimulation via magnetic field induction | Non-invasive tissue stimulation via magnetic field induction |
| Therapy Timer | Subject and Primary Predicate Similar Subject and Secondary Predicate Identical | 1-90 minutes (10 minute default) Recommended treatment time is 10-20 minutes for a single location on the body. | 1-20 minutes (10 minute default) | 1-90 minutes (10 minute default) Recommended treatment time is 10-20 minutes for a single location on the body. |
| Model (System) | Similar | Nova HD+ | Nova HD+ | Aura Wave |
| Available Accessories | Subject and Primary Predicate Similar Subject and Secondary Predicate Identical | Large Loop Butterfly Loop Wings Half Body Mat Paddle | Large Loop | Large Loop Butterfly Loop Wings Half Body Mat Paddle |
{6}
| | | Nova HD+ Subject Device | Nova HD+ Primary Predicate Device K220938 | Aura Wave Secondary Predicate Device K250896 |
| --- | --- | --- | --- | --- |
| Weight | Subject and Primary Predicate Identical | 5.6 pounds (2.5 kg) | 5.6 pounds (2.5 kg) | 8.1 pounds (3.7 kg) |
| Dimensions (Main Unit): | Subject and Primary Predicate Identical | 9.7" x 10.64" x 3.57" | 9.7" x 10.64" x 3.57" | 11.9" x 13.7" x 4.6" |
| Size of Therapy Area | Subject and Primary Predicate Similar Subject and Secondary Predicate Identical | 50 in² (Paddle) to 573 in² (Wings) | 147.2 in² (Large Loop) | 50 in² (Paddle) to 573 in² (Wings) |
| Magnetic Field Output RMS Average | Similar | Up to 3.6 mT | Up to 1.8 mT | Up to 0.19 T |
| Maximum Induced Current* | The maximum induced current for the subject device does not raise safety or effectiveness concerns as it falls below the accepted standards for safe operation. In addition, the maximum induced current is similar to the Secondary Predicate device. | 0.063 mA/cm² | 0.039 mA/cm² | 0.084 mA/cm² |
| Pulse Repetition Rate | Subject and Primary Predicate Identical | 0.4 to 5.4 Hz | 0.4 to 5.4 Hz | 0.4 to 5.8 Hz |
| Pulse Duration | Subject and Primary Predicate Identical | 0.6 to 0.7 mS | 0.6 to 0.7 mS | 0.13 to 0.24 mS |
| Power Consumption | Subject and Primary Predicate Identical | 81.6 Watts maximum | 81.6 Watts maximum | 90 Watts maximum |
| Input Power | Subject and Primary Predicate Identical | 120 Vac, 60 Hz, 0.68 A | 120 Vac, 60 Hz, 0.68 A | 120 Vac, 50 or 60 Hz, 0.75 A |
| Number of Output Channels | Identical | 1 | 1 | 1 |
| User Interface | Identical | Simple 5-button interface | Simple 5-button interface | Simple 5-button interface |
| Microprocessor controlled | Identical | Yes | Yes | Yes |
| Therapy Amplitude Control | Subject and Primary Predicate Identical | Power level 1-10 controlled by operator | Power level 1-10 controlled by operator | Power level 1-20 controlled by operator |
| User-controlled therapy start/stop | Identical | Yes | Yes | Yes |
| Compliance with voluntary standards | Identical | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 62366-1 IEC 60601-1-11 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 62366-1 IEC 60601-1-11 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-6 IEC 62366-1 IEC 60601-1-11 |
| Firmware developed per IEC 62304 | Identical | Yes | Yes | Yes |
*Induced current measured according to IEC 60601-2-10.
{7}
## VII Equivalence Rationale
The subject device Nova HD+ and the Primary Predicate device Nova HD+ (K220938) are both noninvasive physical medicine devices that use electrically generated magnetic fields to stimulate muscles and other tissues for a variety of benefits to the patient. Those benefits include a temporary increase in blood circulation in healthy leg muscles, and the stimulation of healthy muscles in order to improve or facilitate muscle performance.
Both systems share the following characteristics:
- A main unit which is a solid-state microprocessor controlled electronic device and is powered from mains.
- Large Loop accessory coil.
- Relatively small (< 10 pounds) and can be easily transported.
- Tested to applicable safety and EMC standards for both home care and professional use environments.
- Operator adjustable therapy timer and adjustable power output.
- Identical user interface.
- Very similar waveform output in terms of waveform shape and pulse rate.
- Similar induced current (estimated for tissue).
- Both offer one operating channel.
The subject device Nova HD+, and the Secondary Predicate device Aura Wave both offer a variety of accessories to give the user the option of treating either a targeted part of the anatomy with the smaller accessories to large areas of the anatomy such as the back with the largest accessories. The accessories for the subject device and the Secondary Predicate device are identical.
### Non-clinical Testing
Performance testing demonstrated that the subject treatment applicator performed equivalent to the Primary Predicate applicator. Testing included:
Performance testing involved multiple measurements of:
- Signal waveform output (volts) generated from Nova HD+ with its applicators.
- Magnetic flux output (mT) generated from the applicator at all signal intensity input levels 1-10.
### Software
- All software features documented and tested in accordance with IEC 62304 for 'Moderate' level of concern.
### Electrical / EMC evaluated per
- IEC 60601-1 - Medical Electrical Equipment - Part 1: General Requirements for Basic Safety And Essential Performance
- IEC 60601-1-2 - Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests
- IEC 60601-1-11 - Medical Electrical Equipment - Part 1-11: General Requirements for Basic Safety And Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in The Home Healthcare Environment
- IEC 60601-1-6 - Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
- IEC 62366-1 - Medical devices — Part 1: Application of usability engineering to medical devices
### Biocompatibility
- ISO 10993-1 - Biological evaluation of medical devices — Part 1: Evaluation and testing within a risk management process
{8}
- ISO 10993-5 - Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10 - Biological Evaluation of Medical Devices - Part 10: Tests for Irritation And Skin Sensitization
### Clinical Testing
Not applicable. No clinical testing was performed.
### CONCLUSIONS
The analysis of the differences between Nova HD+ and the predicate devices does not raise new questions of safety and effectiveness for the subject device. Based on device performance test results, Aura Wellness determines that the Nova HD+ system performs within its design specifications and is substantially equivalent to the predicate devices.
The information in this 510(k) submission demonstrates that the Nova HD+ system is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.