LINQ II Insertable Cardiac Monitor (ICM) (LNQ22)

K262735 · Medtronic, Inc. · MXD · Aug 28, 2026 · Cardiovascular

Device Facts

Record IDK262735
Device NameLINQ II Insertable Cardiac Monitor (ICM) (LNQ22)
ApplicantMedtronic, Inc.
Product CodeMXD · Cardiovascular
Decision DateAug 28, 2026
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1025
Device ClassClass 2
AttributesPediatric

Indications for Use

The LINQ II ICM is an insertable automatically activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in adult patients, and in pediatric patients who are at least 2 years old, in the following cases: -patients with clinical syndromes or situations at increased risk of cardiac arrhythmias -patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia -patients who have been previously diagnosed with or treated for atrial or ventricular arrhythmias

Device Story

Insertable cardiac monitor (ICM) continuously records subcutaneous ECG; detects tachyarrhythmia, bradyarrhythmia, pause, atrial tachyarrhythmia, and atrial fibrillation. Patient-initiated activation via MyCareLink Heart App or Patient Assistant allows manual symptom recording. Device implanted subcutaneously; communicates via telemetry to mobile manager app or CareLink Network. Healthcare providers review recorded cardiac data to diagnose arrhythmias and guide clinical management. Benefits include long-term monitoring for transient symptoms and arrhythmia management.

Clinical Evidence

Bench testing only. Verification activities confirmed that the new via fill material and associated manufacturing changes do not adversely impact device safety, performance, or reliability.

Technological Characteristics

Implantable hermetic enclosure: sapphire substrate, titanium battery cover. Sensing: titanium foil electrodes with titanium nitride coating. Battery: Lithium anode, silver vanadium oxide/carbon monofluoride cathode (167 mAh). Connectivity: telemetry to mobile app/CareLink. Software: automated arrhythmia detection and command library logic.

Indications for Use

Indicated for adult and pediatric patients (≥2 years) with increased risk of cardiac arrhythmias, transient symptoms (dizziness, palpitation, syncope, chest pain) suggestive of arrhythmia, or prior diagnosis/treatment for atrial or ventricular arrhythmias.

Regulatory Classification

Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA** U.S. FOOD & DRUG ADMINISTRATION August 28, 2026 Medtronic, Inc. Shreya Chenni Sr. Regulatory Affairs Specialist 8200 Coral Sea St. Mounds View, Minnesota 55340 Re: K262735 Trade/Device Name: LINQ II Insertable Cardiac Monitor (ICM) (LNQ22) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: MXD Dated: July 31, 2026 Received: July 31, 2026 Dear Shreya Chenni: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K262735 - Shreya Chenni Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K262735 - Shreya Chenni Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # ALEXANDRA K. MANARAS -S Alexandra Manaras Acting Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262735 | ? | | Please provide the device trade name(s). | | ? | | LINQ II Insertable Cardiac Monitor (ICM) (LNQ22) | | | | Please provide your Indications for Use below. | | ? | | The LINQ II ICM is an insertable automatically activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in adult patients, and in pediatric patients who are at least 2 years old, in the following cases: -patients with clinical syndromes or situations at increased risk of cardiac arrhythmias -patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia -patients who have been previously diagnosed with or treated for atrial or ventricular arrhythmias | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | | Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? | {4} K262735 # 510(k) Summary Submitter Information | Date Prepared: | 8/28/2026 | | --- | --- | | Submitter | Medtronic, Inc. | | | Cardiac Rhythm Management 8200 Coral Sea Street NE Mounds View, MN 55112 ERN: 2182208 | | Contact Person | Shreya Chenni Sr. Regulatory Affairs Specialist Email: shreya.chenni@medtronic.com Phone: +12025692535 | | Alternate Contact | Haley Liberty Prin. Regulatory Affairs Specialist Email: haley.liberty@medtronic.com Phone: +17635262359 | General Information Trade name: LINQ II™ Insertable Cardiac Monitor Common name: Insertable cardiac monitor Product code: MXD Classification: Class II Regulation number: 21 CFR 870.1025 Special Controls: Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm Predicate Device: LINQ II™ Insertable Cardiac Monitor (model LNQ22) K240693 Device Description The LINQ II Insertable Cardiac Monitor (ICM) Model LNQ22 is a programmable device that continuously monitors a patient's ECG and other physiological parameters. The device records cardiac information in response to automatically detected arrhythmias and patient-initiated activation or markings. The device is designed to automatically record the occurrence of an episode of arrhythmia in a patient. Note: Arrhythmias are classified as tachyarrhythmia, bradyarrhythmia, pause, atrial tachyarrhythmia, or atrial fibrillation. Patients may also manually record symptoms. To manually record symptoms, the patient will also need either the MyCareLink Heart App (patient app on mobile device) or the Patient Assistant Model PA97000. The patient can use the MyCareLink Heart App or the Patient Assistant to manually record his or her cardiac rhythm while experiencing or immediately after a symptomatic event. Accessories The LINQ II ICM includes the following accessories: - The LINQ Tool Kit Model LNQ22TK includes two implant tools: The Incision Tool is used to make a small incision through the patient's skin; and the Insertion Tool is used to insert the device through the incision and into the patient's body at the desired location. - The Reveal LINQ Mobile Manager (LMM) Model MSW002 is a programmer designed as a mobile app that communicates with the LINQ II ICM devices via the existing model 24967 Page 1 of 3 {5} K262735 telemetry head. - The 2692 Device Command Library (DCL) software is a component of the Data Transformation Services subsystem of the CareLink Network. This software is responsible for understanding command status, implant device state within CareLink, and the initiation of commands for the implanted device. - The 2691 Instrument Command Library (ICL) software is the logic and data files required for remote programming ## Indications for Use The indications for use for the modified LINQ II ICM are: The LINQ II ICM is an insertable automatically activated and patient-activated monitoring system that records subcutaneous ECG and is indicated in adult patients, and in pediatric patients who are at least 2 years old, in the following cases: - patients with clinical syndromes or situations at increased risk of cardiac arrhythmias - patients who experience transient symptoms such as dizziness, palpitation, syncope, and chest pain that may suggest a cardiac arrhythmia - patients who have been previously diagnosed with or treated for atrial or ventricular arrhythmias The modified LINQ II ICM has an indication for use statement that differs from the predicate device. The statement was revised to provide additional clarity regarding the patient population, but it does not introduce a new patient population or a new disease. The modified LINQ II device continues to have the same intended diagnostic use as the predicate device, and this update does not result in any new or increased risks. ## Technological Characteristics The predicate LINQ II ICM consists of a hybrid substrate that is made of sapphire. The sapphire provides part of the implantable hermetic enclosure, integrates the feedthroughs directly into the substrate, and provides a substrate for component attachment/interconnect. The antenna and sense electrodes are titanium foil laser bonded to the outside of the sapphire substrate and connected directly to the embedded feedthroughs. The sense electrodes are coated with sputtered titanium nitride. The sapphire is laser bonded to the titanium battery cover, which provides the complete hermetic enclosure. The battery is Lithium anode, silver vanadium oxide/carbon monofluoride cathode with a capacity of 167 mAh. The subject LINQ II ICM will continue to use the same technology. It is designed to automatically record the occurrence of an arrhythmia in a patient, continuously sense the patient's subcutaneous ECG, and analyze the timing of ventricular events to detect possible episodes of arrhythmia. The LINQ II ICM has a small form factor, and uses Sapphire, Titanium, Parylene, and Titanium Nitride coating on the sensing electrodes as body contacting materials. When compared to the predicate LINQ II ICM (K240693), the modified LINQ II ICM described herein has the same: - Operating principle - Design features - Device functionality Page 2 of 3 {6} K262735 - Biological safety - Packaging materials - Shelf life The modified LINQ II ICM differs from the predicate in that the modified device includes a new via fill material. ## Substantial Equivalence The modified LINQ II ICM is substantially equivalent to its predicate device based on the comparisons of the device functionality, technology characteristics and intended use. The changes to the subject device were verified through design verification activities. Testing demonstrated that the new via fill material and associated manufacturing changes do not adversely impact the safety, performance, or reliability of the LINQ II ICM device. All verification activities were completed successfully and demonstrated there was no adverse impact to the function of the subject device. The verification results did not raise new safety concerns or performance issues. ## Conclusion Based on the intended use, fundamental technological characteristics, and results of verification activities, the subject device has demonstrated to be substantially equivalent to the predicate device. Page 3 of 3
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...