Retrospective clinical practice records; Off-label clinical use data
A retrospective study of 240 patients was used to evaluate the effectiveness of the device as an adjunct to ACDF surgery. The study compared device-treated patients to untreated controls in a real-world clinical setting to support the expansion of indications.
Patients undergoing 2-level or 1-level (with risk factors) anterior cervical discectomy and fusion (ACDF); Sample Size: 240 patients (110 evaluable at 12 months); Number of Sites: 2 US sites
Untreated controls
Radiographic fusion at 12 months
Indications for Use
The Spinalogic is a portable, battery powered, micro-controlled, noninvasive bone growth stimulator indicated for skeletally mature patients as an adjunct electrical treatment to primary lumbar spinal fusion surgery for one or two levels and adjunct use following primary anterior cervical discectomy and fusion (ACDF) surgery of 2 motion segments (3 vertebrae) from C4 - C7.
Device Story
Portable, battery-powered, micro-controlled, non-invasive bone growth stimulator; uses Combined Magnetic Field (CMF) technology to promote osteogenesis as adjunct to spinal fusion. Device consists of control unit/waveform generator, treatment coil/applicator, and support components. Operated by patient at home under physician prescription; applicator positioned externally over fusion site. Delivers programmed CMF signal (sinusoidal magnetic field, 400mG peak-to-peak, 200mG DC offset, 76.6 Hz) for 30-minute daily sessions. Firmware manages treatment timing and records. Intended to support bone healing; clinical evidence suggests improved radiographic fusion rates in specific populations.
Clinical Evidence
Lumbar: Prospective, randomized, double-masked, placebo-controlled trial (n=243); 64% fusion success (device) vs 43% (placebo) at 9 months. Cervical: Retrospective real-world study (n=240); 12-month radiographic fusion 87.5% (device) vs 73.9% (control). Primary analysis (unadjusted) p=0.077; post-hoc linear regression (controlling for age, sex, BMI, risk factors) showed statistically significant improvement (p<0.05, OR 2.35). No serious adverse events reported.
Technological Characteristics
Non-invasive CMF technology; sinusoidal magnetic field (400mG peak-to-peak, 200mG DC offset, 76.6 Hz). Battery-powered (9V). Dimensions: 28x31x4 cm; Weight: 0.9 kg. IP22 compliant. Firmware-controlled. Non-sterile. Standards: EN 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 62304.
Indications for Use
Indicated for skeletally mature adults as adjunct electrical treatment to primary lumbar spinal fusion (1-2 levels) or primary ACDF surgery (2 motion segments, C4-C7).
Regulatory Classification
Identification
A non-invasive bone growth stimulator is a prescription device that provides stimulation through electrical, magnetic, or ultrasonic fields. The device is intended to be used externally to promote osteogenesis as an adjunct to primary treatments for fracture fixation and spinal fusion or as a treatment for established nonunions or failed fusions. Please reference: www.federalregister.gov/documents/2026/04/16/2026-07366/physical-medicine-devices-reclassification-of-non-invasive-bone-growth-stimulators.
{0}

September 29, 2026
DJO, LLC
% Kristin Duggan
Partner
Hogan Lovells Cadwalader US LLP
555 Thirteenth Street North West
Washington, District of Columbia 20004
Re: K262721
Trade/Device Name: Spinalogic Bone Growth Stimulator
Regulation Number: 21 CFR 890.5870
Regulation Name: Non-Invasive Bone Growth Stimulator
Regulatory Class: Class II
Product Code: LOF
Dated: July 31, 2026
Received: July 31, 2026
Dear Kristin Duggan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262721 - Kristin Duggan
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262721 - Kristin Duggan
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262721 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Spinalogic Bone Growth Stimulator | | |
| Please provide your Indications for Use below. | | ? |
| The Spinalogic is a portable, battery powered, micro-controlled, noninvasive bone growth stimulator indicated for skeletally mature patients as an adjunct electrical treatment to primary lumbar spinal fusion surgery for one or two levels and adjunct use following primary anterior cervical discectomy and fusion (ACDF) surgery of 2 motion segments (3 vertebrae) from C4 - C7. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
# Spinalogic® Bone Growth Stimulator
## 510(k) Summary
Prepared in accordance with 21 CFR 807.92
| 510(k) Owner / Submitter | DJO, LLC |
| --- | --- |
| Address | 3151 Scott St. Vista, CA 92081 |
| Telephone | 1-945-334-7958 |
| Contact Person | Mitchell A. Dhority |
| Email | Mitchell.Dhority@enovis.com |
| Date Summary Prepared | September 29, 2026 |
### 1. Device Name and Classification [21 CFR 807.92(a)(2)]
| Trade / Proprietary Name | Spinalogic Bone Growth Stimulator |
| --- | --- |
| Common Name | Non-invasive bone growth stimulator |
| Classification Name | Stimulator, bone growth, non-invasive |
| 510(k) Number | K262721 |
| Regulation Number | 21 CFR 890.5870 |
| Product Code / Device Class | LOF / Class II (Special Controls) |
### 2. Legally Marketed Predicate Device [21 CFR 807.92(a)(3)]
The subject device is compared to the following legally marketed predicate device:
| Predicate Device | Spinalogic Bone Growth Stimulator |
| --- | --- |
| Predicate Manufacturer | DJO, LLC |
| Predicate Submission / Authorization | P910066/S011 (reclassified into Class II via Final Order 91 FR 20352, effective May 18, 2026) |
| Classification Name | Stimulator, bone growth, non-invasive |
| Regulation / Product Code | 21 CFR 890.5870 / LOF |
### 3. Device Description [21 CFR 807.92(a)(4)]
Spinalogic is a portable, battery-powered, micro-controlled, non-invasive bone growth stimulator intended to provide adjunctive electromagnetic stimulation to support bone healing following spinal fusion surgery. The device uses Combined Magnetic Field (CMF) technology to generate a low-energy electromagnetic field at the treatment site. The stimulation is delivered externally, without implanted components, and is intended for home use under prescription and physician direction.
The system consists of a control unit/waveform generator, treatment coil or applicator assembly, patient-contacting support/positioning components, user interface indicators/prompts, power source, and labeling/instructions for use. During treatment, the applicator is positioned externally over the spinal fusion site. The device delivers a programmed CMF signal for the prescribed daily treatment duration of 30 minutes, consistent with the operating specifications and labeling.
The scientific concept supporting the device is the use of externally applied electromagnetic fields to promote osteogenesis as an adjunct to the primary surgical treatment.
{5}
Spinalogic was FDA approved on December 17, 1999 (P910066/S011) as a portable, battery-powered, micro-controlled, noninvasive bone growth stimulator indicated as an adjunct electromagnetic treatment to primary lumbar spinal fusion surgery for one or two levels. Spinalogic is part of a larger family of CMF Bone Growth Stimulator devices covered under PMA P910066, approved March 4, 1994. No modifications to the physical components of the device, its software, or manufacturing controls that affect consistent treatment delivery are made as part of this submission. The only changes are to the labeling and indications for use.
### 4. Intended Use / Indications for Use [21 CFR 807.92(a)(5)]
The Spinalogic Bone Growth Stimulator is intended to provide non-invasive electromagnetic stimulation to promote osteogenesis as an adjunct treatment for spinal fusion. The indications for use of the predicate approved Spinalogic device are as follows:
- The Spinalogic is a portable, battery powered, micro-controlled, noninvasive bone growth stimulator indicated as an adjunct electrical treatment to primary lumbar spinal fusion surgery for one or two levels.
This 510(k) notice updates the indications for use to the following:
- The Spinalogic is a portable, battery powered, micro-controlled, noninvasive bone growth stimulator indicated for skeletally mature patients as an adjunct electrical treatment to primary lumbar spinal fusion surgery for one or two levels and adjunct use following primary anterior cervical discectomy and fusion (ACDF) surgery of 2 motion segments (3 vertebrae) from C4 - C7.
### 5. Technological Characteristics Compared with Predicate [21 CFR 807.92(a)(6)]
Both Spinalogic and the predicate device use non-invasive stimulation technology to promote bone growth as an adjunct to spinal fusion. The subject device remains the same portable, battery powered, micro-controlled, non-invasive bone growth stimulator as the predicate. The device components, materials, patient-contacting elements, energy source, output characteristics, accessories, software, manufacturing processes, and fundamental technological characteristics are unchanged.
Because there are no technological modifications, the subject device does not include new components or accessories that are absent from the predicate device. The expanded indication is supported by clinical and labeling information described below. The device continues to perform in accordance with its established specifications and does not raise different types of safety and effectiveness questions from the predicate device.
| | **SPINALOGIC** (Predicate: P910066/S011) | **SPINALOGIC** (Subject Device: K262721) |
| --- | --- | --- |
| **Intended Use** | Non-invasive stimulation to promote osteogenesis as an adjunct to spinal fusion/clinical management | No change |
{6}
| | **SPINALOGIC** (Predicate: P910066/S011) | **SPINALOGIC** (Subject Device: K262721) |
| --- | --- | --- |
| **Indications for Use** | The Spinalogic is a portable, battery powered, micro-controlled, noninvasive bone growth stimulator indicated as an adjunct electrical treatment to primary lumbar spinal fusion surgery for one or two levels. | The Spinalogic is a portable, battery powered, micro-controlled, noninvasive bone growth stimulator indicated for skeletally mature patients as an adjunct electrical treatment to primary lumbar spinal fusion surgery for one or two levels and adjunct use following primary anterior cervical discectomy and fusion (ACDF) surgery of 2 motion segments (3 vertebrae) from C4 - C7. |
| **User Population** | Physicians (Prescription Use Only) | No change |
| **Technological Characteristics** | The Bone Growth Stimulator emits a sinusoidal magnetic field with a peak to peak amplitude of 400mG and a 200mG D.C. offset at a frequency of 76.6 Hz. Magnetic energy is considered to be non- ionizing radiation. | No change |
| **Accessories** | Not applicable | No change |
| **Dimensions (l x w x h)** | 11 in x 12 in x 1.5 in (28 cm x 31 cm x 4 cm) | No change |
| **Weight** | 2.0 lbs (0.9 kg) | No change |
| **Power Source** | Battery operated, replaceable, 9 volts direct current. Battery operated device (9V standard battery), not to use lithium batteries. | No change |
| **Safety Features** | Unit is fully enclosed and not serviceable by user (aside from 9V battery installation); IP22 compliant | No change |
| **Software** | Yes, device uses firmware to generate magnetic field treatment, maintain treatment record, daily treatment time countdown | No change |
| **Sterilization** | Non-sterile; no sterilization is required | No change |
{7}
## 6. Nonclinical Performance Data [21 CFR 807.92(b)(1)]
The Spinalogic Bone Growth Stimulator incorporates the same technological features as the predicate and the treatment signal is identical, so no additional non-clinical data was required to support substantial equivalence. A series of toxicological studies on isolated cells, as well as animals, were performed to evaluate the safety of combined static and dynamic magnetic fields. Further, in vitro and in vivo studies were conducted to determine whether the application of these magnetic fields in animal models would stimulate bone healing and other related biological responses. Many of these previous laboratory experiments included whole body exposure to the test animals (and thus included vertebrae and nerve tissue). The device has been tested and found to comply with the electromagnetic compatibility (EMC) limits for medical devices to EN 60601-1:2006+A1:2013+AC:2014+A12:2014 +A2:2020, IEC 60601-1-2:2015+A1:2020, IEC 60601-1-6:2010 + A1:2013, IEC 60601-1-11:2015, and IEC 62304:2006 + A1:2015. These limits are designed to provide reasonable protection against harmful interference in a typical medical installation.
## 7. Clinical Performance Data [21 CFR 807.92(b)(2)]
Lumbar Indications (P910066). The supporting clinical data was derived from a single prospective, randomized, double-masked, placebo-controlled trial where the placebo was a non-functional (sham) Spinalogic device. Spinalogic was studied as an adjunct to surgery in 243 patients undergoing lumbar spine fusion (one or two levels between L3 and S1). Of the 201 subjects evaluable at the 9 months study endpoint, fusion success was determined for 64% of Spinalogic device users and 43% of the placebo controls. This positive treatment effect was statistically significant for the overall population and for women but not for males. Spinalogic was approved as an adjunct treatment to primary lumbar spinal fusion surgery for one or two levels.
Cervical Indications (K262721). The supporting clinical data was derived from a single retrospective, controlled, real-world evidence (off-label clinical practice) study enrolling 240 patients at 2 US sites. Spinalogic was evaluated as an adjunctive post-operative treatment for patients undergoing 2-levels (consecutive or adjacent) or 1-level (with presence of at least one risk factor for fusion) anterior cervical discectomy and fusion (ACDF). Study success was determined at 12 months based on the proportion of subjects that achieved radiographic fusion. Of the 110 12-months completers (subjects who had interpretable radiographic studies at the 12-months visit), 87.5% (56/64) of the Spinalogic treated and 73.9% (34/46) of the untreated controls achieved radiographic fusion. Treatment-related improvement in radiographic fusion was not statistically significant in the unadjusted analysis performed for the 12-months completers (13.6% p = 0.077), diabetic subjects (20/24 versus 15/19 p = 0.716) and active smokers (23/29 versus 19/29 p = 0.234). Female subjects (91.3% versus 73.1% p = 0.009) benefited more from treatment than males (74.2% versus 72.5% p = 0.844).
When variables such as age, sex, BMI, risk factors and 2-level fusions were controlled for in a post-hoc linear regression model, there was a statistically significant improvement in fusion rates for the Spinalogic treatment group versus those in the full control group (p<0.05); individuals who received the Spinalogic treatment were 2.35 times more likely to achieve fusion. The study did not observe treatment-related improvement in secondary non-fusion outcomes (Visual Analogue Scale - pain, Odom Criteria). Study safety was consistent with that reported for Spinalogic's lumbar spine study with no identified serious adverse events related to device use.
In conclusion, statistical significance for the pre-specified primary analysis was not obtained within this real-world evidence study; however, the secondary effectiveness analyses provide evidence for device-related improvement in radiographic fusion at 12 months of follow-up.
## 8. Conclusions [21 CFR 807.92(b)(3)]
The clinical data collected on the cervical indication demonstrate that the subject device is substantially equivalent to the predicate device identified.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.