← Product Code [FFX](/productcode/FFX) · K262621

# Rapid Barostat Bag (RBB) Pump and Catheter System (K262621)

_Mui Scientific, Inc. · FFX · Sep 30, 2026 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K262621

## Device Facts

- **Applicant:** Mui Scientific, Inc.
- **Product Code:** [FFX](/productcode/FFX.md)
- **Decision Date:** Sep 30, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1725
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Rapid Barostat Bag Pump and Catheter are used together for the inflation of a barostat balloon to measure rectal capacity, as well as rectal volume at points of sensation, urge, and discomfort, to determine hypersensitivity, hyposensitivity, or normal rectal sensations. These determinations aid in the diagnosis of anorectal disorders, such as fecal incontinence and constipation.

## Device Story

System consists of air compressor pump unit and disposable PVC catheter with non-compliant balloon; used in clinical settings by healthcare professionals. Pump inflates balloon at constant 2mL/sec rate; patient uses wireless remote to signal sensation, urge, and discomfort points. System records inflation volume and pressure at these points; automatically deflates at 3mL/sec upon discomfort signal. Data stored in internal memory to assist clinicians in diagnosing anorectal disorders. Modifications include updated LCD, touch controller, firmware, and ESD protection to improve reliability and safety.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including usability evaluation, electrical safety testing (IEC 60601-1-2, IEC 61000-4-2), software validation, and design verification.

## Technological Characteristics

System includes air compressor, valves, pressure/flow sensors, and touch-screen display. Catheter: single-lumen PVC tubing with non-compliant plastic balloon. Power: rechargeable batteries. Connectivity: wireless remote control. Software: firmware-based control. Compliance: IEC 60601-1-2, IEC 61000-4-2. Sterilization: not specified.

## Regulatory Identification

A gastrointestinal motility monitoring system is a device used to measure peristalic activity or pressure in the stomach or esophagus by means of a probe with transducers that is introduced through the mouth into the gastrointestinal tract. The device may include signal conditioning, amplifying, and recording equipment. This generic type of device includes the esophageal motility monitor and tube, the gastrointestinal motility (electrical) system, and certain accessories, such as a pressure transducer, amplifier, and external recorder.

## Predicate Devices

- Rapid Barostat Bag (RBB) Pump and Catheter ([K160287](/device/K160287.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** U.S. FOOD & DRUG
ADMINISTRATION

September 30, 2026

Mui Scientific, Inc.
Fatemeh Rahimi
Quality and Regulatory Affairs Manager
145 Traders Blvd. E.,
Unit #34
Mississauga, Ontario L4Z 3L3
Canada

Re: K262621

Trade/Device Name: Rapid Barostat Bag (RBB) Pump and Catheter System
Regulation Number: 21 CFR 876.1725
Regulation Name: Gastrointestinal motility monitoring system
Regulatory Class: Class II
Product Code: FFX
Dated: September 4, 2026
Received: September 4, 2026

Dear Fatemeh Rahimi:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K262621 - Fatemeh Rahimi

Page 2

Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{2}

K262621 - Fatemeh Rahimi

Page 3

Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

MAURA ROONEY -S

for

Anthony C. Lee, Ph.D., M.B.A.

Assistant Director

DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices

OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices

Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262621 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Rapid Barostat Bag (RBB) Pump and Catheter System  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The Rapid Barostat Bag Pump and Catheter are used together for the inflation of a barostat balloon to measure rectal capacity, as well as rectal volume at points of sensation, urge, and discomfort, to determine hypersensitivity, hyposensitivity, or normal rectal sensations. These determinations aid in the diagnosis of anorectal disorders, such as fecal incontinence and constipation.  |   |   |
|  Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |
|  Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ?  |

{4}

K262621
Page 1 of 7

Mui Scientific

Division of
H&A Mui
Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3
Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523
Website: www.muiscientific.com Email: mail@muiscientific.com

August 25, 2026

## Special 510(k) Summary

RE: Rapid Barostat Bag (RBB) Pump and Catheter System

Summary prepared by:

Contact Person: Fatemeh Rahimi

Title: Quality and Regulatory Affairs Manager

Manufacturer: Mui Scientific Division of H&A Mui Enterprises Inc

Address: 145 Traders Blvd. East, Unit #33-34, Mississauga, Ontario, Canada L4Z 3L3

Phone: (905) 890-5525

Fax: (905) 890-3523

Email: fatemeh.rahimi@muiscientific.com

Trade name: Rapid Barostat Bag (RBB) Pump and Catheter System

Common name: Barostat pump and catheters

Classification name: Gastrointestinal motility system

This Special 510(k) Summary is for the Rapid Barostat Bag Pump and Catheter System

Submission Number: K262621

Regulation Number: 21 CFR § 876.1725

Regulation Name: Gastrointestinal motility monitoring system.

Product code: FFX

Device Class: II

Predicate Submission Number: K160287

Predicate Trade Name: Rapid Barostat Bag (RBB) Pump and Catheter

Indications for Use:

The Rapid Barostat Bag Pump and Catheter are used together for the inflation of a barostat balloon to measure rectal capacity, as well as rectal volume at points of sensation, urge, and discomfort, to determine hypersensitivity, hyposensitivity, or normal rectal sensations. These determinations aid in the diagnosis of anorectal disorders, such as fecal incontinence and constipation.

Device Description:

The Rapid Barostat Bag (RBB) Pump and Catheter System consists of a pump unit and a disposable catheter. The pump is made up of an air compressor pump, valves, pressure and flow sensors, and circuit boards, all contained within a flame retardant plastic enclosure with a touch-screen display. The pump unit is electrically powered by rechargeable batteries. The pump is wirelessly connected to a peripheral remote control for the patient to input different levels of sensation during the study. The catheter is made up of single-lumen PVC tubing with a plastic non-compliant barostat balloon at the distal end, and a female luer at the proximal end. The balloon end of the catheter is inserted into a patient's rectum, and the luer end is connected to the pump.

6-1

{5}

K262621
Page 2 of 7

Mui Scientific

Division of
H&A Mui
Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3
Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523
Website: www.muiscientific.com Email: mail@muiscientific.com

The study consists of two or more rounds of inflation and deflation of the catheter. The first round will inflate the balloon to the set internal pressure of 40 mmHg to determine the maximum volume of the rectum. The second round of inflation will be used to determine the volumetric points of sensation, urge, and discomfort associated with the patient's rectum. During the study, the pump will push air through the catheter tubing into the barostat balloon at a constant rate of 2mL/sec, while displaying the fluctuating pressure within the balloon onto the pump's touch screen display. When the patient begins to feel the sensation of the inflated balloon, they would push the "sensation" button on the remote control to transmit a signal to the pump. The inflation volume and pressure will be recorded at that point of reference. The pump will continue to inflate the balloon for the patient to report the point of the feeling of urge, and then discomfort, and the data points are collected and stored in the system's internal memory. Upon pressing the discomfort indicator button, the system will automatically stop the inflow of air into the balloon and begin to rapidly deflate the balloon at 3 mL/sec. By extrapolating the volume and pressure measurements recorded within the barostat balloon, the pump will be able to provide the healthcare professional with clearer data for a more accurate diagnosis of anorectal functions and rectum disorders.

### 1. Description of modification to RBB Pump:

In 2026, Mui Scientific launched a new model of the RBB pump with minor modifications. Several changes have been made to the previous model to improve the performance of our existing pump, including:

- Certain components were replaced to address minor issues caused by wear and tear.
- Some suppliers were replaced with components of the same specifications that are more efficient and present a lower risk.
- Minor design updates were implemented to the LCD panel, touch controller, and firmware to enhance performance and resolve wear-related issues.
- ESD improvements were made to the Charging Dock PCB to reduce risks and improve safety and ensure compliance with the updated requirements of IEC 60601.
- The user interface has been updated to improve usability and interaction with the device without altering the established test protocol.

Refer to Table 1: Summary of RBB Pump Modification between the two versions for more details. The details of these changes are outlined below:

Table 1: Summary of RBB Pump Modification

|  Area of Upgrade | Description of Change | Reason/Benefit of Change | Impact Assessment  |
| --- | --- | --- | --- |
|  LCD panel | LCD panel updated to an improved revision with enhanced optical performance and mechanical stability. | - Improves display clarity and long-term reliability - Reduces flickering and ghost-touches - Supports optically bonded touch panel computability - Lower display-related failures and returns - Previous LCD panel was discontinued. | - Supplier change - Modification does not affect the device's intended use, device performance, or device safety, display function remains unchanged. - Changes are limited to the display component only. - No modifications were made to the board hardware design. - Proves to be more efficient and presents a lower risk. No new risks have been introduced.  |

6-2

{6}

K262621

Page 3 of 7

Mui Scientific

Division of

H&A Mui

Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3

Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523

Website: www.muiscientific.com Email: mail@muiscientific.com

|   |  |  | - Verification testing confirmed performance, safety, and reliability remain consistent with original hardware.  |
| --- | --- | --- | --- |
|  Touch panel | The touch panel has been upgraded with enhanced-thickness glass and OCA bonding for improved durability and clarity. | - Increased impact resistance and long-term assembly durability - Enhanced display readability, reduced glass breakage and internal reflections. - Improved perceived product quality. | - Supplier change - Modification does not affect the device's intended use, device performance, or device safety. - No new risks have been introduced. Existing risk controls remain effective. - Verification testing confirmed performance, safety, and reliability remain consistent with original hardware.  |
|  Touch Controller IC | Touch controller IC upgraded to a newer, more robust revision with improved noise immunity and stability. | - Enhanced Noise Immunity: Provides high immunity to inductive power noise, improving reliability. - Enhanced stable I²C communication, auto calibration support. - Improved touch stability, reduced false touches, improved ESD robustness. | - Supplier change - Modification does not affect the device's intended use, device performance, or device safety. - Functional behavior unchanged; improved robustness - Proves to be more efficient and presents a lower risk. No new risks have been introduced. - Verification testing confirmed performance, safety, and reliability remain consistent with original hardware.  |
|  ESD | Added TVS diode on DC input, ferrite bead on power line, and bonded DC barrel jack metal sleeve to system ground. All exposed metal contacts on the dock are now dielectric-coated. | - Improves ESD immunity based on EMC testing observations - Protects sensitive ICs from contact and air discharge events - Reduces transient overvoltage, noise, and latent failures - Improves grounding, power-line protection, and signal stability - Increases long-term reliability and robustness in high-static environments | - Modification does not affect the device's intended use, device performance, or device safety. - No change to electrical ratings or user interaction. - No new hazards; electrical overstress risk reduced. - Proves to be more efficient and presents a lower risk. No new risks have been introduced. - Verification testing confirmed performance, safety, and reliability remain consistent with original hardware.  |
|  Assembly Durability | Improved mechanical reinforcement, connector strain relief, and internal mounting stability. | - Improved mechanical strength and reliability. - Improve long-term durability and reduce wear-related failures. | - Modification does not affect the device's intended use, device performance, or device safety. - No new risks have been introduced. - Verification testing confirmed performance, safety, and reliability remain consistent with original hardware.  |

6-3

{7}

K262621
Page 4 of 7

# Mui Scientific

Division of
H&A Mui
Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3
Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523
Website: www.muiscientific.com Email: mail@muiscientific.com

|  Firmware update | Firmware updated to improve boot stability, power handling, and system monitoring. Firmware structure refactored for reliability | - Improves startup reliability and eliminates intermittent boot failures and unexpected resets - Strengthens power-up sequencing and handling of slow or noisy peripherals - Improves battery efficiency, runtime, and reporting accuracy - Enhances sensor sampling, filtering, and low-flow measurement stability - Improves UI responsiveness and reduces operational latency - Supports future features, diagnostics, and easier firmware maintenance | - Modification does not affect the device's intended use, device performance, or device safety. - Firmware changes does not affect use of device by end user. - All functional tests confirm performance is consistent with prior specifications. - Improves essential performance - Software maintenance update only. - Proves to be more efficient and presents a lower risk. No new risks have been introduced. - Verification testing confirmed performance, safety, and reliability remain consistent with original hardware.  |
| --- | --- | --- | --- |
|  User Interface Software | Updated UI visuals and workflow layouts Faster UI transitions and reduced latency Improved status indicators and enhanced navigation. | - Improves usability and responsiveness, reducing user interaction errors - Reduces operational ambiguity and improves workflow efficiency - Enhances operator confidence during system operation | - Modification does not affect the device's intended use, device performance, or device safety. - Verification testing confirmed performance, safety, and reliability remain consistent with original hardware. - No change to use conditions. - No functional impact; cosmetic/usability only. - Use-related risks reduced. No new risks have been introduced.  |

**NOTE:** All modifications have been implemented to the RBB pump itself, and the updated pump continues to use the same catheter as the previous model. Therefore, there are no changes are made to the RBB catheter.

All of these changes have been validated and verified as outlined below:

- A comprehensive validation was conducted to assess and evaluate all implemented changes.
- Software validation was completed for the new software version to verify compatibility with the overall system.
- New electrical safety and EMC testing were performed. Updated Electrical Safety testing report from Third party)
- An assessment was conducted of the Risk Management File to ensure that the risk profile had not changed.
- A new set of usability and human factors evaluations was conducted.
- A new set of Literature search and clinical Evaluation was conducted.
- A new set of post market assessment was conducted.

The changes are minor and were implemented to reduce risks and improve the device. They do not affect the device's intended use, fundamental technology, or performance specifications, and they do not introduce new risks or significantly modify existing risks. Comprehensive verification, validation, electrical safety/EMC, and usability testing confirmed that the device remains substantially equivalent to the previous version.

6-4

{8}

K262621
Page 5 of 7

Mui Scientific

Division of
H&A Mui
Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3
Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523
Website: www.muiscientific.com Email: mail@muiscientific.com

We have a robust work instruction in our QMS for managing substantial changes (201W1-Substantial Change). All relevant regulatory bodies were notified of the changes and asked to review the device modifications. After a detailed review of the changes in the second version, the modification was determined to be minor.

Refer to Table 2: Parameter Comparison, which presents a comparison between the two versions.

Table 2 Parameter Comparison

|  Parameter | Previous Model to New Model  |
| --- | --- |
|  Device Name / Model | Identical  |
|  Indications for Use | Identical  |
|  Target Population | Identical  |
|  Principle of Operation | Identical  |
|  Design (Geometry, Dimensions) | Identical  |
|  Materials (Patient-contacting) | Identical  |
|  Materials (Non-contacting) | Identical  |
|  Shelf Life | Identical  |
|  Packaging | Identical  |
|  Software | V1.3.121.85_260317  |
|  Firmware | V3.0.1_20260113  |
|  Accessories / Components | Identical  |
|  Usability / User Interface | Minor updates have been made to the user interface, but they were validated through extensive validation and usability testing.  |
|  Performance Specifications | Identical  |
|  Biocompatibility | Both of them uses the same catheter  |
|  Electrical Safety (if applicable) | Revised ESD (Electrostatic Discharge) to align with new IEC standards (IEC 60601-1-2, IEC 61000-4-2)  |
|  Mechanical Properties | Identical  |
|  Risk Profile | The risk management file has been updated for the new version of the device to include additional identified hazards. These risks have been evaluated and appropriately controlled in accordance with ISO 14971. The overall risk profile remains acceptable, and there is no change to the device classification.  |

6-5

{9}

K262621
Page 6 of 7

Mui Scientific

Division of
H&A Mui
Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3
Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523
Website: www.muiscientific.com Email: mail@muiscientific.com

|  Clinical Data | The Clinical Evaluation Report has been updated with additional details and requirements; however, no significant changes were made to the clinical data.  |
| --- | --- |
|  Labeling / IFU | Laser engraved label  |
|  Manufacturing Process | Improved mechanical reinforcement, connector strain relief, and internal mounting stability  |
|  Standards Compliance | Identical  |

## 2. Reason for changes in RBB pump:

The RBB pump was modified to improve reliability, performance, and regulatory compliance. Changes included replacing certain components and suppliers with equivalent alternatives to reduce risk and improve efficiency, implementing minor design updates to the LCD panel, touch controller, and firmware to address wear-related issues and enhance performance, and improving ESD protection on the Charging Dock PCB to increase safety and ensure compliance with the latest IEC 60601 requirements.

## 3. Substantial Equivalence Discussion

Following extensive verification testing, we conclude that the modified pump is substantially equivalent to the previously marketed version in terms of intended use, indications for use, and overall performance characteristics. The modifications were introduced to improve performance based on post-production feedback, particularly to address issues such as ghost touch and screen flickering. Additional design changes were also implemented to meet updated requirements in the latest revision of the electrical safety standards and to ensure compliance with the revised electrical safety testing requirements. These changes do not affect the device's intended use, fundamental performance or safety characteristics. We are claiming substantial equivalence between the modified version and its previously cleared model, referring to Table 5.

Table 3: Existing Model vs Modified Model

|  Product Name | 510(k) Number | Version | Manufacturer  |
| --- | --- | --- | --- |
|  Rapid Barostat Bag Pump and its catheter | K160287 | V1 | Mui Scientific, Division of H & A Mui Enterprises Inc  |
|  Rapid Barostat Bag Pump and its catheter | Waiting for Special 510K clearance | V2 | Mui Scientific, Division of H & A Mui Enterprises Inc  |

The modified version (V2) of the Rapid Barostat Bag Pump is similar to the previous model. Both versions have the same intended use, indications for use, intended patient population, contraindications, warnings, and other key characteristics. There are no modifications or changes to the RBB catheter, and both models use the same catheter.

## 4. Summary of non-clinical performance evaluations

Clinical testing is not required for this Special 510(k) because the modification does not change the device's intended use, fundamental scientific technology, or raise any new questions of safety or effectiveness. The change has been thoroughly evaluated through Mui Scientific's design control process, including risk analysis, design verification, and design

6-6

{10}

K262621^{}[] Page 7 of 7

Mui Scientific

Division of^{}[] H&A Mui^{}[] ^{}[] Enterprises Inc.

145 Traders Blvd. E., Unit #34, Mississauga, ON, Canada, L4Z 3L3^{}[] Tel: (905) 890-5525 Toll Free: 1-800-303-6611 Fax: (905) 890-3523^{}[] ^{}[] Website:^{}[] www.muiscientific.com Email:^{}[] mail@muiscientific.com

validation activities. In addition, numerous non-clinical performance tests have been conducted to assess the impact of the modification and to confirm that the device continues to meet all applicable performance, safety, and functional requirements. The testing demonstrated that the modified device performs as intended, remains substantially equivalent to the previously cleared device, and does not introduce any new hazards or increase existing risks. Based on the comprehensive non-clinical testing results and risk management evaluation, sufficient objective evidence exists to support the safety and effectiveness of the modified device without the need for clinical data.

Refer to Table 6 for a summary of the tests performed:

**Table 4: Non-Clinical Bench Performance Testing**

|  Evaluation Description | Summary  |
| --- | --- |
|  Usability Evaluation | Evaluation of the usability of the Modified device, it's IFU, and labels.  |
|  Electrical Safety Evaluation | Evaluation of the electrical safety testing of the device.  |
|  Software Evaluation including cybersecurity assessment | Evaluation of the software and its validation  |
|  Design Verification and Validation | Evaluation of the design of the device.  |

We have not made any changes to the materials or components that have direct or indirect patient contact; therefore, new biocompatibility testing is not required. The component changes were implemented to improve the long-term durability of the RBB pumps, which is expected to maintain or extend performance over time, not reduce it. As a result, there is no risk of decreased shelf life or service life, and additional shelf-life testing is not necessary. In addition, the packaging remains unchanged, so no further packaging validation or testing is required.

## 5. Conclusion

The modifications to the device do not introduce any significant changes to its intended use, indications for use, or fundamental performance characteristics. The information provided demonstrates that the modifications do not affect the device's intended use or overall safety and effectiveness.

6-7

---

**Source:** [https://fda-staging.innolitics.com/device/K262621](https://fda-staging.innolitics.com/device/K262621)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
