Biolinq Shine Autonomous Time-in-Range Microsensor

K262599 · Biolinq Incorporated · SFU · Sep 18, 2026 · Clinical Chemistry

Device Facts

Record IDK262599
Device NameBiolinq Shine Autonomous Time-in-Range Microsensor
ApplicantBiolinq Incorporated
Product CodeSFU · Clinical Chemistry
Decision DateSep 18, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1359
Device ClassClass 2

Indications for Use

Biolinq Shine Autonomous Time-in-Range Microsensor consists of a continuous intradermal glucose sensor with a color-changing indicator light. It is intended to measure, record, and analyze physiologic data using an array of microsensors combined with an accelerometer and ambient light sensor. The qualitative display communicates real-time changes in glucose levels in persons 22 years or older not on insulin. The device may be used to aid in the management of a disease or condition related to glycemic control. An app provides more granular glucose information. It analyzes and correlates this glucose information with meals, rest, and activity levels. Use of the glucose and activity information is designed to facilitate acute behavior and lifestyle modifications. Interpretation of device readings should be based on glucose and activity trends and patterns. The user is not intended to take medical action based on the device output without consultation with a qualified healthcare professional.

Device Story

Wearable intradermal glucose sensor; array of seven microelectrodes in papillary dermis; measures interstitial fluid glucose (70-400 mg/dL); includes accelerometer and ambient light sensor. Operates autonomously; autocalibration; no fingerstick required. Provides real-time qualitative feedback via LED lights on device (blue: 70-180 mg/dL; yellow: 181-400 mg/dL; red: <70 mg/dL; purple: end-of-life; white: error). Optional app provides granular data, correlates glucose with meals/rest/activity, and supports lifestyle modifications. Worn on volar forearm. Used by patients for glycemic awareness; not for acute medical action without professional consultation. Benefits include behavior/lifestyle modification through trend/pattern analysis.

Clinical Evidence

No new clinical studies submitted. Clinical evidence relies on data previously reviewed under DEN240080. OTC transition supported by human factors evaluations and labeling assessments.

Technological Characteristics

Intradermal microarray sensor with seven microelectrodes; measures interstitial glucose (70-400 mg/dL). Includes accelerometer and ambient light sensor. Connectivity: Bluetooth (to optional app). Power: Internal battery. Form factor: Wearable for volar forearm. Sterilization: Not specified. Software: Embedded firmware with LED interface and optional mobile app. Standards: ISO 14971, IEC 62304, AAMI TIR57, AAMI TIR97, IEC 62366-1.

Indications for Use

Indicated for persons 22 years or older not on insulin to aid in the management of a disease or condition related to glycemic control.

Regulatory Classification

Identification

The Biolinq Shine Autonomous Time-in-Range Microsensor consists of a continuous intradermal glucose sensor with a color-changing indicator light. It is intended to measure, record, and analyze physiologic data using an array of microsensors combined with an accelerometer and ambient light sensor. The qualitative display communicates real-time changes in glucose levels to aid in the management of a disease or condition related to glycemic control in persons 22 years or older not on insulin. An app provides more granular glucose information, analyzing and correlating glucose information with meals, rest, and activity levels to facilitate acute behavior and lifestyle modifications.

Special Controls

In combination with the general controls of the FD&C Act, the glucose range monitoring system is subject to the following special controls: (1) Design verification and validation must include the following: (i) Robust clinical data must demonstrate the accuracy of the device for its intended use. (ii) The clinical data must include a comparison between GRM outputs and blood glucose values in specimens collected in parallel that are measured on an FDA-accepted laboratory-based glucose measurement method that is precise and accurate, and that is traceable to a higher order (e.g., an internationally recognized reference material and/or method). Performance must be assessed using appropriate statistical methods for the type of output. (iii) The clinical data must be obtained from a clinical study designed to fully represent the performance of the device throughout the intended use population and throughout the measuring range of the device. (iv) Clinical study results must demonstrate consistent analytical and clinical performance throughout the sensor wear period. (v) Data must demonstrate that throughout the claimed sensor life, the device does not allow clinically significant gaps in sensor data availability that would prevent the GRM from achieving its intended use. (2) Design verification and validation must include adequate controls established during manufacturing and at product release to ensure the released product meets the performance specifications as defined in paragraphs (b)(1) and (b)(2) of this section. (3) The device must demonstrate clinically acceptable performance in the presence of clinically relevant levels of potential interfering substances that are reasonably present in the intended use population, including but not limited to endogenous substances and metabolites, foods, dietary supplements, and medications. (4) The device must include appropriate measures to ensure that a disposable sensor cannot be used beyond its claimed sensor wear period. (5) Design verification and validation must include results obtained through a usability study that demonstrates that the intended user can use the device safely and correctly interpret device outputs, contraindications, warnings, precautions, and limitations, as applicable. (6) The labeling required under § 809.10(b) of this chapter must include a separate description of the following sensor performance data observed in the clinical study performed in conformance with paragraph (b)(1) of this section for each intended use population, in addition to separate sensor performance data for each different GRM insertion or use sites (e.g., abdomen, arm, buttock): (i) A description of all device outputs, including glucose data as well as information generated by the device based on the glucose data, their intended purposes, and limitations. (ii) A description of the accuracy of the glucose data. (iii) A description of the frequency and duration of gaps in sensor data. (iv) A description of the observed duration of GRM life for the device. (v) A statement that the device is not intended for insulin dosing, medication adjustment, or other acute medical decision making.

Predicate Devices

Submission Summary (Full Text)

{0} [LOGO] FDA U.S. FOOD & DRUG ADMINISTRATION ### 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ### I Background Information: #### A 510(k) Number K262599 #### B Applicant Biolinq Incorporated #### C Proprietary and Established Names Biolinq Shine Autonomous Time-in-Range Microsensor #### D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | SFU | Class II | 21 CFR 862.1359 Glucose range monitoring system | Clinical Chemistry | #### E Purpose for Submission: Modification from Prescription Use Only to Over The Counter use ### II Intended Use/Indications for Use: #### A Intended Use(s): See Indications for Use below. #### B Indication(s) for Use: Biolinq Shine Autonomous Time-in-Range Microsensor consists of a continuous intradermal glucose sensor with a color-changing indicator light. It is intended to measure, record, and analyze physiologic data using an array of microsensors combined with an accelerometer and ambient light sensor. The qualitative display communicates real-time changes in glucose levels in persons 22 years or older not on insulin. The device may be used to aid in the management of a disease or condition related to glycemic control. An app provides more granular glucose information. It analyzes and correlates this glucose information with meals, rest, and activity levels. Use of the glucose and activity information is designed to facilitate acute behavior and lifestyle modifications. Interpretation of device readings should be based on glucose and activity Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} trends and patterns. The user is not intended to take medical action based on the device output without consultation with a qualified healthcare professional. # C Special Conditions for Use Statement(s): OTC - Over The Counter - Remove the sensor before Magnetic Resonance Imaging (MRI), Computed Tomography (CT) scan, high-frequency electrical heat (diathermy) treatment, X-ray machines, and Advanced Imaging Technology (AIT) body scanners. The magnetic fields and radiation could damage the device components. - This device is not intended for pregnant women, people on dialysis, critically ill patients, or persons under 22 years of age. - Do not use for hypoglycemia awareness. The sensor must not be used in persons with Type 1 Diabetes, those at significant risk of severe hypoglycemia, those taking any form of insulin, or with Automated Insulin Delivery (AID) Systems. The sensor does not provide glucose information below 70 mg/dL and does not provide alerts or alarms. - Do not use if you have significant color blindness, profound visual impairment that cannot be corrected, monochromacy (complete colorblindness), or inability to see small objects as the device uses small colored lights. - Do not use while taking hydroxyurea (used for cancer and sickle cell anemia treatment) as it causes inaccurate readings. - Do not use during or shortly after medical tests involving drinking xylose (D-Xylose absorption test). - Acetaminophen may falsely raise sensor readings (Acetaminophen can be found in many medications, including Tylenol®, Excedrin®, and cold medicines). Use a fingerstick blood glucose meter if sensor readings don't match symptoms when taking acetaminophen. - Environmental and activity limitations: Do not use for SCUBA diving, above 10,500 feet (3,200 meters) altitude, or outside temperature range of 50-107°F (10-42°C). - Water resistance: Can be submerged up to 6 feet (2 meters) for up to 30 minutes maximum. - For airport security, use hand-wanding, pat-downs, or metal detectors instead of AIT body scanners. - Wear on inner forearm only, avoiding areas too close to elbow or wrist. Do not place on irritated or damaged skin, scarring, cuts, burns, wounds, or other skin damage. - Single use only - cannot be reused or re-applied if it falls off. # III Device Description Biolinq Shine Autonomous Time-in-Range Microsensor is an autonomous intradermal time-in range microarray sensor intended to measure glucose in people not using insulin. The wearable consists of an array of seven microelectrodes ("microsensors") that reside in the papillary dermis to continuously measure interstitial fluid glucose. The device is designed to be worn on the volar forearm (inner surface of the forearm) for placement in the user's field of vision. It continuously measures glucose levels (when concentrations are within 70-400 mg/dL) and stores data for up to five days. The device includes a built-in user interface that provides real-time feedback through LED lights visible directly on the wearable. The lights blink throughout the wear period. A blue LED light K262599 - Page 2 of 4 {2} indicates glucose levels in the range of 70 to 180 mg/dL. A yellow LED light indicates glucose levels 181-400 mg/dL, and a red LED light indicates glucose levels below 70 mg/dL. Additional features on the wearable include concurrent flashing blue in the top and bottom hemisphere of the wearable to reflect when the user's glucose is rising more rapidly and directionally headed to the yellow zone, and the wearable flashing increases in intensity when the user's glucose is above 400 mg/dL. If the algorithm determines glucose levels cannot be provided, the wearable flashes white light in the upper hemisphere as a system error notification. When the device reaches end of use life, it displays a purple light to indicate the sensor should be replaced. After application, the device starts operating automatically without user interaction and begins providing continuous glucose measurements in two hours without requiring external reference fingerstick blood glucose values (autocalibration). Combined with the integrated lights and its visible wear location, it can be used as a standalone device not requiring a separate display device or app. An app is available, but not required, for additional context and behavioral insights about glucose time-in-range related to daily activities. The device is applied using a single-use applicator and includes additional features such as ambient light sensing for LED brightness adjustment and accelerometer-based activity tracking. # IV Substantial Equivalence Information: # A Predicate Device Name(s): Biolinq Shine Autonomous Time-in-Range Microsensor # B Predicate 510(k) Number(s): DEN240080 # C Comparison with Predicate(s): | Device & Predicate Device(s): | K262599 | DEN240080 | | --- | --- | --- | | Device Trade Name | Biolinq Shine Autonomous Time-in-Range Microsensor | Same | | **General Device Characteristic Similarities** | | | | Intended Use/Indications For Use | Automatically measure glucose in bodily fluids continuously for a specified period of time. | Same | | **General Device Characteristic Differences** | | | | Availability to user | Over the counter | Prescription use only | K262599 - Page 3 of 4 {3} # V Standards/Guidance Documents Referenced: ISO 14971 Third Edition 2019-12 Medical devices - Application of risk management to medical devices. Recognition Number: 5-125 ANSI AAMI IEC 62304:2006/A1:2016 Medical device software - Software life cycle processes [Including Amendment 1 (2016)]. Recognition Number: 13-79 AAMI TIR57:2016 Principles for medical device security - Risk management. Recognition Number: 13-83 AAMI TIR97:2019 Principles for medical device security - Postmarket risk management for device manufacturers. Recognition Number: 13-112 ANSI AAMI IEC TIR 80001-2-2:2012 Application of risk management for IT Networks incorporating medical devices - Part 2-2: Guidance for the disclosure and communication of medical device security needs, risks and controls. Recognition Number: 13-42 ANSI AAMI IEC 62366-1:2015+AMD1:2020 (Consolidated Text) Medical devices Part 1: Application of usability engineering to medical devices including Amendment 1. Recognition Number: 5-129 # VI Performance Characteristics: # A. Non-Clinical Performance The proposed OTC indication was supported through human factors evaluations and labeling modifications. The human factors evaluation and labeling modifications were reviewed and found to be adequate in supporting OTC use. # B. Clinical Studies: No new clinical studies were submitted. Clinical data was provided in DEN240080 and is still applicable to this device. # C. Other Supportive Device Performance Characteristics Data None. # VII Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. # VIII Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K262599 - Page 4 of 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...