← Product Code [OWN](/productcode/OWN) · K262547

# KARL STORZ TIPCAM1 Rubina (K262547)

_KALR STORZ SE & Co. KG · OWN · Sep 21, 2026 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K262547

## Device Facts

- **Applicant:** KALR STORZ SE & Co. KG
- **Product Code:** [OWN](/productcode/OWN.md)
- **Decision Date:** Sep 21, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Pediatric

## Indications for Use

The TIPCAM1 Rubina videoscope is intended to be used together with the camera control unit during diagnostic and/or surgical procedures when endoscopic video assistance is required. For use in all endoscopy and endoscopic surgery within the peritoneal and thoracic cavities, including gynecological and urological anatomy. Indicated for selected thoracoscopically-assisted cardiac surgical procedures, such as:- Mitral valve repair- IMA mobilization for cardiac revascularization- Patent foramen ovale closure- Atrial septal defect repair- Left atrial appendage closure/occlusion- Atrial myxoma excision- Mitral valve replacement- Tricuspid valve repair- Epicardial pacing lead placement

## Device Story

TIPCAM1 Rubina is an integrated rigid videoendoscope and camera unit; used with compatible light source (TL400 Power LED Rubina) and camera control unit (IMAGE1 S or IMAGE1 CCU). Device transmits visible (VIS) and near-infrared (NIR) light into patient anatomy; captures reflected light via integrated optics; processes signals through CCU for display on monitor. Operated by physicians in clinical/surgical settings for endoscopic visualization. Provides real-time imaging to assist surgeons during diagnostic/surgical procedures; enables visualization of anatomy and fluorescent imaging; supports decision-making during complex cardiac and general endoscopic surgeries.

## Clinical Evidence

No new clinical trials conducted. Published literature provided to support the removal of original contraindications present in the predicate submission (K212695).

## Technological Characteristics

Integrated rigid videoendoscope; utilizes VIS and NIR light sources; connects to IMAGE1 S or IMAGE1 CCU via 4U connector. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, and ISO 8600 series (1, 3, 4, 5). No changes in technological characteristics compared to predicate K212695.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- KARL STORZ ICG Imaging System ([K212695](/device/K212695.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

![LOGO](FDA U.S. FOOD & DRUG ADMINISTRATION)

September 21, 2026

KALR STORZ SE & Co. KG
Frank Sugden
Senior Regulatory Intelligence Manager
Dr.-Karl-Storz-Straße 34
Tuttlingen, Baden-Württemberg 78532
Germany

Re: K262547

Trade/Device Name: KARL STORZ TIPCAM1 Rubina
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and accessories
Regulatory Class: Class II
Product Code: OWN, GWG
Dated: July 23, 2026
Received: July 23, 2026

Dear Frank Sugden:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K262547 - Frank Sugden

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K262547 - Frank Sugden

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# MARCO CANNELLA -S

Marco Cannella Ph.D.

Acting Assistant Director

Division of Cardiac Electrophysiology,

Diagnostics, and Monitoring Devices

Office of Cardiovascular Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262547 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  KARL STORZ TIPCAM1 Rubina  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The TIPCAM1 Rubina videoscope is intended to be used together with the camera control unit during diagnostic and/or surgical procedures when endoscopic video assistance is required. For use in all endoscopy and endoscopic surgery within the peritoneal and thoracic cavities, including gynecological and urological anatomy. Indicated for selected thoracoscopically-assisted cardiac surgical procedures, such as:- Mitral valve repair- IMA mobilization for cardiac revascularization- Patent foramen ovale closure- Atrial septal defect repair- Left atrial appendage closure/occlusion- Atrial myxoma excision- Mitral valve replacement- Tricuspid valve repair- Epicardial pacing lead placement  |   |   |
|  Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |
|  Please select the age group(s) for which the device(s) is to be used. | Neonates/Newborns (Birth to < 29 days old)Infants (29 days old to < 2 years old)Children (2 years old to < 12 years old)Adolescents (12 years old to < 22 years old)Adults (22 years old and greater) | ?  |

{4}

**STORZ**  
KARL STORZ — ENDOSKOPE

K262547

## 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

|  Submitter: | KALR STORZ SE & Co. KG Dr.-Karl-Storz-Straße 34 Tuttlingen, Baden-Württemberg 78532 Germany  |
| --- | --- |
|  Contact: | Frank Sugden Senior Manager Regulatory Intelligence Phone: 805-562-3953 Email: Frank.Sugden@karlstorz.com  |
|  Date of Preparation: | September 21, 2026  |
|  Type of 510(k) Submission: | Traditional  |
|  Device Identification: | Trade Name: KARL STORZ TIPCAM1 Rubina Classification Name: Endoscope and accessories (21 CFR 876.1500) 510(k) Number: K262547  |
|  Regulatory Class: | II  |
|  Product Code: | OWN, GWG  |
|  Classification Panel: | Gastroenterology/Urology  |
|  Predicate Device(s): | KARL STORZ ICG Imaging System (K212695)  |
|  Device Description: | The KARL STORZ Imaging System is intended to provide real-time visible (VIS) and near-infrared (NIR) fluorescent imaging and documentation of endoscopic and microscopic procedures in adults and pediatrics. Following the “light in to light out” mental model, the KARL STORZ Imaging System incorporates one of multiple compatible light sources to provide either or both VIS or NIR light into the operating scene within a patient. This light is captured via either a passive telescope or an active videoendoscope (flexible or rigid). In the case of a passive telescope the light is then transferred to a connected camera head or exoscope. All of these active videoendoscopes, camera heads, and exoscopes make of use a connector (multiple options available) to transmit the light to a camera control unit (CCU). Finally, the  |

Page 1 of 3

{5}

**STORZ**  
KARL STORZ—ENDOSKOPE

K262547

|   | CCU transmits the light to a monitor for the medical professional to see. The TIPCAM1 Rubina videoendoscope is an integrated unit comprised of a camera and a rigid endoscope. A connected light source (TL400 Power LED Rubina for NIR) transmits light (VIS or NIR) through the integrated rigid endoscope via a light guide. Either the IMAGE1 S or IMAGE1 CCU systems process the image transmitted from the TIPCAM1 Rubina via the 4U connector.  |
| --- | --- |
|  Intended Use and Indications for Use: | The TIPCAM1 Rubina videoscope is intended to be used together with the camera control unit during diagnostic and/or surgical procedures when endoscopic video assistance is required. For use in all endoscopy and endoscopic surgery within the peritoneal and thoracic cavities, including gynecological and urological anatomy. Indicated for selected thoracoscopically-assisted cardiac surgical procedures, such as: - Mitral valve repair - IMA mobilization for cardiac revascularization - Patent foramen ovale closure - Atrial septal defect repair - Left atrial appendage closure/occlusion - Atrial myxoma excision - Mitral valve replacement - Tricuspid valve repair - Epicardial pacing lead placement  |
|  Technological Characteristics. | TIPCAM1 Rubina remains identical to the cleared KARL STORZ ICG Imaging System in K212695 and reviewed in K233333. There are no changes in technological characteristics.  |
|  Non-Clinical Performance Data: | There are no performance standards or special controls developed under Section 514 of the FD&C Act for endoscopes. However, the KARL STORZ Imaging System is tested in accordance with the following standards and FDA Guidance: **Electrical Safety and EMC** - IEC 60601-1 - IEC 60601-1-2 - IEC 60601-2-18 **ISO Endoscopic Standards** - ISO 8600-1 - ISO 8600-3 - ISO 8600-4 - ISO 8600-5 **Software Verification and Validation Testing** - Guidance for the Content of Premarket Submissions for Software  |

Page 2 of 3

{6}

STORZ
KARL STORZ—ENDOSKOPE

K262547

|   | Contained in Medical Device • Level of concern: Moderate Conformance to these standards was not the basis upon which substantial equivalence was determined.  |
| --- | --- |
|  Clinical Performance Data: | Published literature was provided to support the change of the original contraindication in the predicate submission K212695.  |
|  Conclusion: | The subject device and predicate device have the same intended use and fundamental endoscopic visualization function. Although the subject device expands the indications to include selected thoracoscopically-assisted cardiac procedures, the physical and technological characteristics remain identical to the cleared system. Therefore, the differences in indications do not affect the safety or effectiveness of the device and do not alter its therapeutic, diagnostic, prosthetic, or surgical use. The conclusions drawn from the technological characteristics that the subject device is as safe and effective as the predicate device.  |

Page 3 of 3

---

**Source:** [https://fda-staging.innolitics.com/device/K262547](https://fda-staging.innolitics.com/device/K262547)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).