← Product Code [LLZ](/productcode/LLZ) · K262449

# AdvaView (K262449)

_ADVAHEALTH SOLUTIONS, LLC · LLZ · Aug 14, 2026 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K262449

## Device Facts

- **Applicant:** ADVAHEALTH SOLUTIONS, LLC
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Aug 14, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

AdvaView is a software application used for reference and diagnostic viewing and analysis of multi-modality medical imaging and non-imaging data (e.g., video information) with associated reports and information. AdvaView enables qualified healthcare professionals, including (but not limited to) physicians, surgeons, nurses, and administrators to receive and view patient images, documents, and data. AdvaView allows qualified users to perform simple and complex image manipulations (including window/level, markups, 3D visualization) and measurements. Not intended for primary diagnosis of mammographic images. Not intended for diagnostic use on mobile devices.

## Device Story

AdvaView is a zero-footprint, web browser-based medical image viewer; enables access, manipulation, and measurement of multi-modality DICOM images (US, CT, MRI, X-Ray). Operates as a client-side application integrating with external PACS via RESTful APIs; delegates storage, authentication, and encryption to system integrators. Features include window/level adjustment, 3D visualization (MIP, MPR), annotation tools (pencil, ROI, spline, livewire), and clinical measurements (Cobb angle, spine-labeling, cardiothoracic ratio). Used by radiologists, technologists, physicians, and nurses in clinical settings to support diagnostic workflows. Output displayed on web browser; assists clinicians in image interpretation and clinical decision-making.

## Clinical Evidence

Bench testing only. No clinical data was required. Verification and validation included manual testing of measurement tool accuracy using calibrated and un-calibrated phantom DICOM images. Statistical analysis (mean, standard deviation, standard error, one-tailed t-test) confirmed measurement accuracy within +/- 1 unit tolerance across zoom levels.

## Technological Characteristics

Software-only, zero-footprint, web browser-based DICOM viewer. Architecture: client-side application using RESTful APIs for integration with external PACS/infrastructure. Features: multi-modality support (US, CT, MRI, X-Ray), 3D visualization (MIP, MPR), and measurement tools. Connectivity: networked via secured browser interface. Deployment: client-server integration model.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- FlexView Diagnostic ([K233226](/device/K233226.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** U.S. FOOD & DRUG
ADMINISTRATION

August 14, 2026

Advahealth Solutions, LLC
% Meritxell Martinez
Senior Regulatory Affairs Specialist
Innolitics, LLC
1101 W. 34th St.
#550
AUSTIN TX 78705

Re: K262449

Trade/Device Name: AdvaView

Regulation Number: 21 CFR 892.2050

Regulation Name: Medical Image Management And Processing System

Regulatory Class: Class II

Product Code: LLZ

Dated: July 16, 2026

Received: July 16, 2026

Dear Meritxell Martinez:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K262449 - Meritxell Martinez

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K262449 - Meritxell Martinez

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

, for

Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. |   | K262449 ?  |
|  Please provide the device trade name(s). |   | ?  |
|  AdvaView  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  AdvaView is a software application used for reference and diagnostic viewing and analysis of multi-modality medical imaging and non-imaging data (e.g., video information) with associated reports and information. AdvaView enables qualified healthcare professionals, including (but not limited to) physicians, surgeons, nurses, and administrators to receive and view patient images, documents, and data. AdvaView allows qualified users to perform simple and complex image manipulations (including window/level, markups, 3D visualization) and measurements.  |   |   |
|  Not intended for primary diagnosis of mammographic images.  |   |   |
|  Not intended for diagnostic use on mobile devices.  |   |   |
|  Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |
|  Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ?  |

{4}

**ADVA**^{}[] Health

K262449

**510(k) Summary**

## 1. SUBMITTER INFORMATION

|  Company Name | AdvaHealth Solutions  |
| --- | --- |
|  Address | 331 2nd Avenue Suite 400, Minneapolis, MN 55401 US  |
|  Company Contact | Imad Nijim, CEO  |
|  Email | imad.nijim@advahealthsolutions.com  |
|  Primary Correspondent | Meritxell Martinez, Innolitics  |
|  Email | fda@innolitics.com  |
|  Date Prepared | July 16th, 2026  |

## 2. DEVICE INFORMATION

|  **Trade Name** | AdvaView  |
| --- | --- |
|  **Common Name** | DICOM Viewer  |
|  **Product Code** | LLZ  |
|  **Regulation Number** | 892.2050  |
|  **Classification Name** | Picture archiving and communications system  |
|  **Class** | Class II  |
|  **Panel** | Radiology  |

## 3. PREDICATE DEVICE INFORMATION

|  **Predicate Device Name** | FlexView Diagnostic  |
| --- | --- |
|  **Manufacturer** | AdvaHealth Solutions  |

Page 1 of 6

{5}

ADVA Health

510(k) Summary

|  **510(k) Number** | K233226  |
| --- | --- |
|  **Product Code** | LLZ  |
|  **Regulation Number** | 892.2050  |
|  **Classification Name** | Picture archiving and communications system  |
|  **Regulatory Class** | Class II  |
|  **Review Panel** | Radiology  |

## 4. DEVICE DESCRIPTION

AdvaView is a medical image viewer which allows, reviewing, manipulating, and visualizing medical multi-modality image data in DICOM format and other data. AdvaView is a server-based solution that may integrate with healthcare facility software systems and display DICOM images. AdvaView enables healthcare professionals to access, manipulate, and measure DICOM images and collaborate using diagnostic-quality medical images without installing client software.

The AdvaView application has the following primary features and functions:

- Zero-footprint web browser-based
- Allows access to images for all participants in the healthcare process, including radiologists, technologists, physicians, nurses, and other patient care practitioners
- Contains functionality to view, annotate, and measure images
- Contains industry-standard tools for image manipulation, annotation, and measurements
- Advanced image manipulation functions like view synchronization across series, 3D visualization such as MIP and MPR
- Encrypted transmission of medical images through secured networks
- HIPAA-compliant data management, including centralized storage of user activities via audit trails

Page 2 of 6

{6}

**ADVA**^{}[] Health

**510(k) Summary**

AdvaView is a configurable, software-only, web browser-based medical image viewer that displays and processes digital medical images and associated medical information to support clinical workflows.

AdvaView functions as a client application, enabling users to access, retrieve, and interact with imaging data through interfaces and APIs.

## **4.1. Key Marketing Claims**

AdvaView is a zero-footprint web viewer for medical images. AdvaView allows easy access to image data through systems that implement the prescribed API Specifications.

## **5. INDICATIONS FOR USE**

AdvaView is a software application used for reference and diagnostic viewing and analysis of multi-modality medical imaging and non-imaging data (e.g., video information) with associated reports and information. AdvaView enables qualified healthcare professionals, including (but not limited to) physicians, surgeons, nurses, and administrators to receive and view patient images, documents, and data. AdvaView allows qualified users to perform simple and complex image manipulations (including window/level, markups, 3D visualization) and measurements.

**Not intended for primary diagnosis of mammographic images.**

**Not intended for diagnostic use on mobile devices.**

## **6. PREDICATE DEVICE COMPARISON**

### **6.1. Summary of Technological Characteristics Comparison**

AdvaView is a modification of the predicate device, FlexView Diagnostic (K233226), submitted under the Special 510(k) pathway. The subject and predicate devices have the same intended use and the same fundamental scientific technology: both are web browser-based, multi-modality medical image viewers that load, display, manipulate, annotate, and measure DICOM images and associated data, and both rely on an existing PACS as the source of imaging studies. Both provide the same core review capabilities, including window/level adjustment,

Page 3 of 6

{7}

ADVA Health

510(k) Summary

image view controls (rotate, pan, and zoom), measurement functions, text annotations, report generation, and DICOM windowing, and both support the same imaging modalities (US, CT, MRI, and X-Ray).

The principal technological difference between the two devices is a change in deployment architecture. FlexView Diagnostic was deployed as a hosted, multi-tenant platform that operated self-managed backend infrastructure. AdvaView has been re-implemented as a client-side, zero-footprint viewer that runs within the web browser and integrates with external systems for image storage, configuration, authentication, and supporting services through defined RESTful API interfaces. Consistent with this architecture, functions that were previously performed by the self-operated backend (e.g., patient-data storage and confidentiality, service availability, in-transit encryption, and user authentication and access control) are delegated to the system integrator and the third-party PACS. This is a change in how the software is deployed and integrated rather than a change in the methods by which images are rendered, manipulated, or measured. The fundamental scientific technology of the viewer is unchanged.

AdvaView also adds a set of viewer features that extend the functionality of the predicate. These fall into four categories:

- Image display and projection functions, including Maximum, Minimum, and Average Intensity Projection modes, a viewport scale overlay, and scrollbar.
- Region-of-interest and measurement tools, including ellipse, rectangle, circle, and freehand ROI tools; spline and livewire contour tools; and Cobb angle, spine-labeling, and cardiothoracic ratio (CTR) tools.
- Annotation and workspace customization, including a pencil annotation tool, an annotation show/hide toggle, hanging-protocol configuration, multiple-monitor support, and configurable thumbnail location, radial menu, toolbars, viewport labels, and mouse bindings.
- Input-handling robustness, through validation of DICOM input received from external systems.

These differences between the subject and predicate devices do not alter the intended use and do not raise new questions of safety or effectiveness.

Page 4 of 6

{8}

**ADVA**^{}[] Health

**510(k) Summary**

## **7. DISCUSSION OF PERFORMANCE TESTING**

### **7.1. Software Verification and Validation**

Software verification and validation testing were conducted on the AdvaView system and documentation was provided as recommended by FDA’s 2023 Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions.”

The software verification and validation testing verified that the design requirements were successfully met. The intended use and user needs were successfully validated.

#### **7.1.1. Measurement Accuracy**

Verification of AdvaView included manual testing of the accuracy of the measurement tools using phantom (calibrated and un-calibrated) DICOM images.

Measurement accuracy testing involved evaluating the difference (i.e., delta) between the expected length and the measurement entered by the user. Delta results were calibrated to +/- 1 units, which is equivalent to +/- 1 units relative to the acquisition size and display resolution. This calibration ensured consistent accuracy across different zoom levels.

Acceptance criteria for this test included statistical analyses such as mean, standard deviation, standard error and a one tailed t-test. Verification testing for all measurement tools passed the predetermined acceptance criteria.

### **7.2. Conclusion**

AdvaView is a modification of AdvaHealth Solutions’ own legally marketed device, FlexView Diagnostic (K233226), submitted under the Special 510(k) pathway. AdvaView retains the same intended use and the same fundamental scientific technology as the predicate: both are zero-footprint, web browser-based, multi-modality medical image viewers that load, display, manipulate, annotate, and measure DICOM images and associated data, and both rely on an existing PACS as the source of imaging studies.

The modifications comprise a change in deployment architecture together with a set of additional viewer features. Each modification was evaluated under the manufacturer’s design control procedures using well-established methods, and the associated risk analysis,

Page 5 of 6

{9}

![img-0.jpeg](img-0.jpeg)

## 510(k) Summary

verification, and validation activities demonstrate that the changes were adequately verified and validated. No clinical testing was necessary to support the modifications. The results confirm that the modifications do not alter the intended use and do not raise new questions of safety or effectiveness relative to the predicate.

Based on the risk analysis, the verification and validation results, and the technological characteristics comparison presented above, AdvaView is substantially equivalent to FlexView Diagnostic (K233226).

Page 6 of 6

---

**Source:** [https://fda-staging.innolitics.com/device/K262449](https://fda-staging.innolitics.com/device/K262449)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
