← Product Code [SCA](/productcode/SCA) · K262432

# CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test; CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test (K262432)

_CorDx, Inc. · SCA · Oct 14, 2026 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K262432

## Device Facts

- **Applicant:** CorDx, Inc.
- **Product Code:** [SCA](/productcode/SCA.md)
- **Decision Date:** Oct 14, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.3987
- **Device Class:** Class 2
- **Review Panel:** Microbiology
- **Attributes:** Pediatric

## Indications for Use

The CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider. Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up. The CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider. Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.

## Device Story

Lateral flow immunoassay for qualitative detection of SARS-CoV-2 nucleocapsid, Flu A, and Flu B proteins; uses anterior nasal swab samples. Sample is placed in processing solution to lyse and solubilize viral proteins; extract is loaded onto test cassette. Conjugates on reagent pad bind viral antigens; complex captured by specific monoclonal antibodies on reaction membrane. Visual readout: red line at test region indicates positive; red line at control region indicates valid test. Used at home or point-of-care by lay users or healthcare providers. Results provide rapid screening; negative results are presumptive and require molecular confirmation. Helps guide clinical follow-up and management of respiratory infections.

## Clinical Evidence

Prospective clinical study (N=606) compared self-collected/lay-user-collected anterior nasal swabs to FDA-cleared RT-PCR. SARS-CoV-2 PPA 92.2% (95% CI: 81.5-96.9%), NPA 100%. Flu A PPA 86.5% (95% CI: 74.7-93.3%), NPA 99.8%. Flu B PPA 86.1% (95% CI: 71.3-93.9%), NPA 100%. Usability study (N=55) showed 99.5% correct performance of critical tasks and 98.2% result interpretation accuracy.

## Technological Characteristics

Lateral flow immunochromatographic assay. Components: sample pad, reagent pad (latex particles with monoclonal antibodies), reaction membrane (secondary antibodies), absorbent pad. Cassette housing. Visual readout. Storage 2-30°C. Sterilization via ethylene oxide (ISO 11135, ISO 10993-7).

## Regulatory Identification

A multi-analyte respiratory virus antigen detection test is an in vitro diagnostic device intended for the detection and/or differentiation of respiratory viruses directly from respiratory clinical specimens. The device is intended to be performed at the site of sample collection, does not involve sample storage and/or transport.

## Predicate Devices

- WELLlife COVID-19 / Influenza A&B Home Test ([K243256](/device/K243256.md))
- WELLlife COVID-19 / Influenza A&B Antigen Test ([K243256](/device/K243256.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY

# I Background Information:

A 510(k) Number

K262432

B Applicant

CorDx, Inc.

C Proprietary and Established Names

CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test; CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  SCA | Class II | 21 CFR 866.3987 - Multi-Analyte Respiratory Virus Antigen Detection Test | Microbiology  |

# II Submission/Device Overview:

A Purpose for Submission:

To obtain substantial equivalence determination for the CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test and CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test (hereafter referred to as the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test).

B Measurand:

Protein antigens from SARS-CoV-2, Influenza A and Influenza B

C Type of Test:

Qualitative Lateral Flow Immunoassay

K262432 Page 1 of 22

{1}

### III Intended Use/Indications for Use:

#### A Intended Use(s):

See Indications for Use below.

#### B Indication(s) for Use:

##### CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test:

The CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.

All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens.

Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.

Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.

##### CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test:

The CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.

All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens.

Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider.

Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.

#### C Special Conditions for Use Statement(s):

OTC – Over the counter Use

K262432 Page 2 of 22

{2}

### D Special Instrument Requirements:

Not Applicable

## II Device/System Characteristics:

### A Device Description:

The CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test is a rapid, immunochromatographic membrane assay that uses highly sensitive monoclonal antibodies to detect SARS-CoV-2 nucleocapsid protein and influenza (Flu) A and B proteins in anterior nasal swab specimens. The test package components for each of the devices are as follows:

- CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test
  - 1/2/4/5/8/10/20/25 Test cassettes
  - 1/2/4/5/8/10/20/25 Swabs
  - 1/2/4/5/8/10/20/25 Tubes with sample processing solution
  - 1 Quick Reference Instructions (QRI)
- CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test
  - 2/10/25 Test cassettes
  - 2/10/25 Swabs
  - 2/10/25 Tubes with sample processing solution
  - 1 Quick Reference Guide (QRG)

The test device is composed of a test strip and cassette housing. The test strip, which is enclosed in the cassette housing, is comprised of the following components: sample pad, reagent pad, reaction membrane, and absorbent pad. The reagent pad contains latex particles conjugated with monoclonal antibodies against the proteins of Flu A, Flu B and SARS-CoV-2; the reaction membrane contains the secondary antibodies for the proteins of Flu A, Flu B and SARS-CoV-2.

![img-0.jpeg](img-0.jpeg)

Fig. 1 Diagram of Test Device

### B Principle of Operation:

The CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test employs membrane based lateral flow technology in a sandwich design to detect proteins from SARS-CoV-2, influenza A, and influenza B. Following sample collection, the individual's anterior nasal swab sample is placed

K262432 Page 3 of 22

{3}

in the tube with sample processing solution, to lyse the sample and solubilize viral proteins. After lysis, the sample is loaded onto the test cassette sample well.

When the sample extract is added into the sample well, conjugates dried onto the reagent pad are dissolved and migrate along with the sample. If Flu A, Flu B proteins and/or SARS-CoV-2 nucleocapsid antigen is present in the sample, a complex forms between the anti-Flu A/Flu B/SARS-CoV-2 conjugate and the viral antigen and will be captured by the specific anti-Flu A/Flu B/SARS-CoV-2 monoclonal antibody coated on the test line region (Flu A/Flu B/COV line). Absence of the test line (Flu A/Flu B/COV line) suggests a negative result. To serve as a procedural control, a red line will always appear in the control line region (C) indicating that proper volume of sample has been added and membrane wicking has occurred.

If the control line does not appear, the sample result is invalid. Test results are read visually without the need for any instrument.

### Result Interpretation:

The CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test may give three possible results, positive, negative, or invalid:

|  POSITIVE | ![img-1.jpeg](img-1.jpeg)If the control line (C) is visible and one or more lines appear(s) for any of the viruses, the test is positive for that or those viruses.  |
| --- | --- |
|  NEGATIVE | ![img-2.jpeg](img-2.jpeg)If the control line (C) is visible but no other lines appear, the test is negative.  |
|  INVALID | ![img-3.jpeg](img-3.jpeg)If the control line (C) is not visible the test is invalid, even if any test line is visible.  |

## III Substantial Equivalence Information:

K262432 Page 4 of 22

{4}

# A Predicate Device Name(s):

WELLlife COVID-19 / Influenza A&B Home Test;

WELLlife COVID-19 / Influenza A&B Antigen Test

# B Predicate 510(k) Number(s):

K243256

# C Comparison with Predicate(s):

|  Device & Predicate Device(s): | Subject Device (K262432) | Predicate (K243256)  |
| --- | --- | --- |
|  Device Trade Name | CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test; CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test | WELLlife COVID-19 / Influenza A&B Home Test; WELLlife COVID-19 / Influenza A&B Antigen Test  |
|  Intended Use/Indications For Use | CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test: The CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of | WELLlife COVID-19 / Influenza A&B Home Test: WELLlife COVID-19 / Influenza A&B Home Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection. Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for non-prescription home use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older. All negative results are presumptive and should be confirmed with an FDA cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens. Individuals who test negative and experience continued or worsening  |

K262432 Page 5 of 22

{5}

breath, should seek follow-up care from their healthcare provider. Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.

# CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test:

The CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test is a lateral flow immunoassay intended for the qualitative detection and differentiation of SARS-CoV-2, influenza A, and influenza B protein antigens directly in anterior nasal swab specimens from individuals with signs and symptoms of respiratory tract infection. Clinical symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 years or older testing themselves, or adults testing individuals aged 2 years or older.

All negative results are presumptive and should be confirmed with an FDA-cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2 or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare provider. Positive results do not rule out co-infection with other respiratory pathogens and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.

respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare providers. Positive results do not rule out coinfection with other respiratory pathogens, and therefore do not substitute for a visit to a healthcare provider or appropriate follow-up.

# WELLlife COVID-19/ Influenza A&B Antigen Test:

The WELLlife COVID-19/ Influenza A&B Antigen Test is a lateral flow immunochromatographic assay intended for the qualitative detection and differentiation of influenza A, and influenza B nucleoprotein antigens and SARS-CoV-2 nucleocapsid antigen directly in anterior nasal swab samples from individuals with signs and symptoms of respiratory tract infection.

Symptoms of respiratory infections due to SARS-CoV-2 and influenza can be similar. This test is for use by individuals aged 14 years or older testing themselves, or adults testing aged 2 years or older.

All negative results are presumptive and should be confirmed with an FDA cleared molecular assay when determined to be appropriate by a healthcare provider. Negative results do not rule out infection with influenza, SARS-CoV-2, or other pathogens. Individuals who test negative and experience continued or worsening respiratory symptoms, such as fever, cough and/or shortness of breath, should seek follow-up care from their healthcare providers. Positive results do not rule out coinfection with other respiratory pathogens.

Test results should not be used as the sole basis for treatment or

K262432 Page 6 of 22

{6}

|   |  | other patient management decisions.  |
| --- | --- | --- |
|  General Device Characteristic Similarities  |   |   |
|  Regulation Number | 21 CFR 866.3987 | Same  |
|  Assay Technique | Lateral Flow immunochromatographic assay, visually read | Same  |
|  Analytes | SARS-CoV-2, influenza A, and influenza B, protein antigens | Same  |
|  Intended Specimen | Anterior nasal swab specimen | Same  |
|  Sample Collection Method | Nasal swab supplied in the kit | Same  |
|  Usage | Single use test | Same  |
|  Assay Result | Qualitative (positive, negative or invalid) | Same  |
|  Storage Temperature | Store at 36-86°F /2-30°C | Same  |
|  General Device Characteristic Differences  |   |   |
|  Time to Result | 10-30 min | 10-20 min  |

# IV Standards/Guidance Documents Referenced:

|  Document # | Title | Publisher | Applicable Study  |
| --- | --- | --- | --- |
|  ISO 11135:2014 | Sterilization of health-care products - Ethylene oxide - Requirements for development, validation and routine control of a sterilization process for medical devices | ISO | Sterility  |
|  ISO 10993-7 | Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals | ISO | Sterility  |
|  ISO 10993-1 | Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process | ISO | Biocompatibility  |
|  EP05-A3 | Evaluation of Precision of Quantitative Measurement Procedures | CLSI | Precision  |
|  EP12-A2 | User Protocol for Evaluation of Qualitative Test Performance | CLSI | Precision  |
|  EP25-A | Evaluation of Stability of In Vitro Diagnostic Reagents | CLSI | Reagent stability  |

K262432 Page 7 of 22

{7}

|  Document # | Title | Publisher | Applicable Study  |
| --- | --- | --- | --- |
|  EP37 | Supplemental Tables for Interference Testing in Clinical Chemistry | CLSI | Interference  |

# V Performance Characteristics (if/when applicable):

# A Analytical Performance:

# 1. Precision/Reproducibility:

The multi-lot precision for the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test was evaluated to assess variability across multiple sources of analytical variability, including different testing days, operating runs, operators, and test kit lots. A panel of samples, including one negative sample and seven positive sample combinations: single positive samples (SARS-CoV-2, Flu A, and Flu B separately), doubly co-spiked samples (SARS-CoV-2 + Flu A, SARS-CoV-2 + Flu B, and Flu A + Flu B), and a triply co-spiked sample (SARS-CoV-2 + Flu A + Flu B) were tested. Positive samples were prepared at three concentration levels, including approximately 0.8x LoD, 2x LoD, and 5x LoD. All samples were prepared in pooled negative swab matrix (PNSM).

Testing was performed using three reagent lots and two operators over 10 non-consecutive day. Six panel sets were tested per day, with panel assignments alternated between operators throughout the study to ensure balanced operator representation across concentration levels and reagent lots. A total of 96 tests were performed per study day.

Results for this study are shown in Table 1 below. The results demonstrated 100% concordance with the expected result across sample types, device test lines and lots with negative samples, 2x LoD and 5x LoD samples. The expected lot-to-lot imprecision was observed with 0.8x LoD samples.

Table 1: Precision Study Results

|  Analyte Concentration | Analyte | Test Lines | # Positives / # Total |   |   | # Positives / # Total (% Positive Rate)  |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Lot 1 | Lot 2 | Lot 3  |   |
|  0 | Negative | COV | 0/40 | 0/40 | 0/40 | 0/120 (0%)  |
|   |   |  Flu A | 0/40 | 0/40 | 0/40 | 0/120 (0%)  |
|   |   |  Flu B | 0/40 | 0/40 | 0/40 | 0/120 (0%)  |
|  0.8x LoD | SARS-CoV-2 | COV | 7/10 | 6/10 | 5/10 | 18/30 (60%)  |
|   |   |  Flu A | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu B | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |  Flu A | COV | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu A | 7/10 | 6/10 | 5/10 | 18/30 (60%)  |
|   |   |  Flu B | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |  Flu B | COV | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu A | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu B | 7/10 | 6/10 | 5/10 | 18/30 (60%)  |
|   |  SARS-CoV-2 + Flu A | COV | 5/10 | 6/10 | 6/10 | 17/30 (56.7%)  |
|   |   |  Flu A | 7/10 | 4/10 | 5/10 | 16/30 (53.3%)  |
|   |   |  Flu B | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |

K262432 Page 8 of 22

{8}

|  Analyte Concentration | Analyte | Test Lines | # Positives / # Total |   |   | # Positives / # Total (% Positive Rate)  |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Lot 1 | Lot 2 | Lot 3  |   |
|   | SARS-CoV-2 + Flu B | COV | 6/10 | 6/10 | 5/10 | 17/30 (56.7%)  |
|   |   |  Flu A | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu B | 4/10 | 4/10 | 6/10 | 14/30 (46.7%)  |
|   |  Flu A + Flu B | COV | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu A | 5/10 | 6/10 | 5/10 | 16/30 (53.3%)  |
|   |   |  Flu B | 4/10 | 6/10 | 4/10 | 14/30 (46.7%)  |
|   |  SARS-CoV-2 + Flu A + Flu B | COV | 5/10 | 5/10 | 6/10 | 13/30 (53.3%)  |
|   |   |  Flu A | 5/10 | 4/10 | 5/10 | 14/30 (46.7%)  |
|   |   |  Flu B | 6/10 | 5/10 | 6/10 | 17/30 (56.7%)  |
|  2.0x LoD | SARS-CoV-2 | COV | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |   |  Flu A | 0/20 | 0/20 | 0/20 | 0/60 (0%)  |
|   |   |  Flu B | 0/20 | 0/20 | 0/20 | 0/60 (0%)  |
|   |  Flu A | COV | 0/20 | 0/20 | 0/20 | 0/60 (0%)  |
|   |   |  Flu A | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |   |  Flu B | 0/20 | 0/20 | 0/20 | 0/60 (0%)  |
|   |  Flu B | COV | 0/20 | 0/20 | 0/20 | 0/60 (0%)  |
|   |   |  Flu A | 0/20 | 0/20 | 0/20 | 0/60 (0%)  |
|   |   |  Flu B | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |  SARS-CoV-2 + Flu A | COV | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |   |  Flu A | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |   |  Flu B | 0/20 | 0/20 | 0/20 | 0/60 (0%)  |
|   |  SARS-CoV-2 + Flu B | COV | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |   |  Flu A | 0/20 | 0/20 | 0/20 | 0/60 (0%)  |
|   |   |  Flu B | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |  Flu A + Flu B | COV | 0/20 | 0/20 | 0/20 | 0/60 (0%)  |
|   |   |  Flu A | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |   |  Flu B | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |  SARS-CoV-2 + Flu A + Flu B | COV | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |   |  Flu A | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|   |   |  Flu B | 20/20 | 20/20 | 20/20 | 60/60 (100%)  |
|  5.0x LoD | SARS-CoV-2 | COV | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |   |  Flu A | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu B | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |  Flu A | COV | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu A | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |   |  Flu B | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |  Flu B | COV | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu A | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu B | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |  SARS-CoV-2 + Flu A | COV | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |   |  Flu A | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |   |  Flu B | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |  SARS-CoV-2 + Flu B | COV | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |   |  Flu A | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu B | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |  Flu A + Flu B | COV | 0/10 | 0/10 | 0/10 | 0/30 (0%)  |
|   |   |  Flu A | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |   |  Flu B | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |

K262432 Page 9 of 22

{9}

|  Analyte Concentration | Analyte | Test Lines | # Positives / # Total |   |   | # Positives / # Total (% Positive Rate)  |
| --- | --- | --- | --- | --- | --- | --- |
|   |   |   |  Lot 1 | Lot 2 | Lot 3  |   |
|   | SARS-CoV-2 + Flu A + Flu B | COV | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |   |  Flu A | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |
|   |   |  Flu B | 10/10 | 10/10 | 10/10 | 30/30 (100%)  |

# 2. Linearity:

This is a qualitative assay and therefore, linearity is not applicable.

# 3. Analytical Specificity/Interference:

# a. Cross-reactivity and Microbial Interference:

Cross-reactivity and microbial interference studies were conducted to determine if pathogens likely to be present in direct nasal swab samples would cause false results by cross reacting or interfering with the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test. A panel of viruses, bacteria, and fungi were evaluated in these studies. Dilutions for testing were made in PNSM. Each organism spiked sample was tested in three replicates in the absence (for cross-reactivity) or presence (for microbial interference) of 3x co-spike LoD for SARS-CoV-2, Flu A virus, and Flu B virus. Cross-reactivity with SARS-CoV-2 was observed with SARS-CoV when tested ≥ 1.60 x 10⁴ TCID₅₀/mL but was not observed at 1.60 x 10³ TCID₅₀/mL.

There was no cross-reactivity or microbial interference observed for any other organisms. The results of cross-reactivity and microbial interference studies are summarized in Table 2.

Table 2: Cross-Reactivity and Microbial Interference Study Results

|  Pathogens | Concentration Tested | # Positive / # Total*  |   |
| --- | --- | --- | --- |
|   |   |  Cross Reactivity | Microbial Interference  |
|  SARS-CoV-1 | 8.00 x 10⁵ TCID₅₀/mL | 3/3** | N/A  |
|   |  1.60 x 10⁵ TCID₅₀/mL | 3/3** | N/A  |
|   |  1.60 x 10⁴ TCID₅₀/mL | 1/3** | N/A  |
|   |  1.60 x 10³ TCID₅₀/mL | 0/3 | 3/3  |
|  MERS-coronavirus | 1.51 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |
|  Human coronavirus OC43 | 1.50 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |
|  Human coronavirus 229E | 1.50 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |
|  Human coronavirus NL63 | 1.50 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |
|  Adenovirus (AV71) | 1.50 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |
|  Adenovirus Type 7 | 1.50 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |
|  Cytomegalovirus | 1.50 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |
|  Epstein Barr Virus Human gammaherpesvirus 4 | > 2.5 x 10³ CEID₅₀/mL | 0/3 | 3/3  |
|  Human Metapneumovirus (hMPV) | 1.50 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |
|  Human Parainfluenza virus 1 | 1.50 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |
|  Human Parainfluenza virus 2 | 1.50 x 10⁵ TCID₅₀/mL | 0/3 | 3/3  |

K262432 Page 10 of 22

{10}

|  Pathogens | Concentration Tested | # Positive / # Total*  |   |
| --- | --- | --- | --- |
|   |   |  Cross Reactivity | Microbial Interference  |
|  Parainfluenza virus 3 | 1.50 x 10^{5} TCID_{50}/mL | 0/3 | 3/3  |
|  Parainfluenza virus 4 | 1.50 x 10^{5} TCID_{50}/mL | 0/3 | 3/3  |
|  Enterovirus Type 68 | 1.50 x 10^{5} TCID_{50}/mL | 0/3 | 3/3  |
|  Rhinovirus 1A | 1.00 x 10^{5} PFU/mL | 0/3 | 3/3  |
|  *Bordetella pertussis* | > 1 x 10^{4} CFU/mL | 0/3 | 3/3  |
|  *Candida albicans* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Chlamydia pneumoniae* | 1.10 x 10^{6} IFU/mL | 0/3 | 3/3  |
|  *Corynebacterium sp.* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Escherichia coli* | > 2.5 x 10^{4} CFU/mL | 0/3 | 3/3  |
|  *Hemophilus influenzae* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Lactobacillus sp.* | > 1 x 10^{4} CFU/mL | 0/3 | 3/3  |
|  *Legionella pneumophila* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Moraxella catarrhalis* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Mycoplasma pneumoniae* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Mycobacterium tuberculosis* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Neisseria meningitidis* | 1.05 x 10^{5} CFU/mL | 0/3 | 3/3  |
|  *Neisseria sp.* | > 1 x 10^{4} CFU/mL | 0/3 | 3/3  |
|  *Pneumocystis jirovecii* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Pseudomonas aeruginosa* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Staphylococcus aureus* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Staphylococcus epidermidis* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Streptococcus salivarius* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Streptococcus pneumoniae* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  *Streptococcus pyogenes* | 1.10 x 10^{6} CFU/mL | 0/3 | 3/3  |
|  Measles | 1.50 x 10^{5} TCID_{50}/mL | 0/3 | 3/3  |
|  Mumps | 1.50 x 10^{5} TCID_{50}/mL | 0/3 | 3/3  |
|  Live RSV strain A | 1.50 x 10^{5} TCID_{50}/mL | 0/3 | 3/3  |
|  Live RSV strain B | 1.50 x 10^{5} TCID_{50}/mL | 0/3 | 3/3  |
|  Human coronavirus HKU1 | 2.61 x 10^{6} GE/mL | 0/3 | 3/3  |
|  Pooled Negative Swab Matrix (PNSM) | NA | 0/3 | 0/3  |
|  3x Positive in PNSM | NA | NA | 3/3  |

*Represents the same results for each of the three analytes. 0/3 and 3/3 represent 0 positive and 3 positive results respectively for all the three analytes tested.

**Results represent only for SARS-CoV-2 test line; Flu A and B test lines showed 0 positives for both Flu A and B.

### b. Endogenous and Exogenous Interfering Substances Study:

Interfering substances studies were conducted to assess the effect on the performance of the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test in the presence of potentially interfering substances that might be present in anterior nasal swab specimens. Each potentially interfering substance was tested in PNSM at the targeted initial concentrations in the absence (negative) and presence (positive) of co-spiked analytes (SARS-CoV-2, Flu A, and Flu B) at 3x LoD. Three (3) replicates were tested for each sample prepared using one lot of the candidate test kit. Interference (false positive results) was observed for FluMist

K262432 Page 11 of 22

{11}

Quadrivalent Live Intranasal Influenza Virus Vaccine for Flu A and Flu B. No interference was observed for the remaining substances at the concentrations tested. The results of the endogenous and exogenous interference are summarized in Table 3.

Table 3: Potential Interfering Substances Study Results

|  Potential Interfering Substance | Concentration | Cross-Reactivity (without analyte) (# Positive / # Total) |   |   | Interference (with analyte) (# Positive / # Total)  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  SARS-CoV-2 | Flu A | Flu B | SARS-CoV-2 | Flu A | Flu B  |
|  Human Whole Blood (EDTA tube) | 4% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Mucin | 2.5 mg/mL | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Chloraseptic (Menthol/Benzocaine) | 3.0 mg/mL | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Naso GEL (NeilMed) | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Nasal Drops (Phenylephrine) | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Nasal Spray (Oxymetazoline) | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Nasal Spray (Cromolyn) | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Nasal Corticosteroid (Dexamethasone) | 1 mg/mL | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Nasal Corticosteroid (Triamcinolone) | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Nasal gel (Galphimia glauca, Histanium hydrochloricum, Luffa operculata, Sulfur) | 1.25 % v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Homeopathic allergy relief (Histaminum hydrochloricum) | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Zicam Nasal Spray (Galphimia glauca, Luffa operculata) | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Homeopathic Nasal Spray (Alkalol) | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Sore Throat Phenol Spray | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Tobramycin | 4 μg/mL | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Mupirocin | 10 mg/mL | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Anti-viral drug (Remdesivir) | 10 mg/mL | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |

K262432 Page 12 of 22

{12}

|  Potential Interfering Substance | Concentration | Cross-Reactivity (without analyte) (# Positive / # Total) |   |   | Interference (with analyte) (# Positive / # Total)  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  SARS-CoV-2 | Flu A | Flu B | SARS-CoV-2 | Flu A | Flu B  |
|  Fluticasone Propionate | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Tamiflu (Oseltamivir Phosphate) | 5 mg/mL | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  FluMist/ FluMist Live intranasal influenza virus vaccine 15% | 15% v/v | 0/3 | 3/3 | 3/3 | NT* | NT* | NT*  |
|  FluMist/ FluMist Live intranasal influenza virus vaccine 1.5% | 1.5% v/v | 0/3 | 3/3 | 3/3 | NT* | NT* | NT*  |
|  FluMist/ FluMist Live intranasal influenza virus vaccine 0.5% | 0.5% v/v | 0/3 | 3/3 | 0/3 | NT* | NT* | NT*  |
|  FluMist/ FluMist Live intranasal influenza virus vaccine 0.25% | 0.25% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Zanamivir | 282 ng/mL | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Zinc (Therazinc throat spray) | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Body & Hand Lotion | 0.5% w/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Body Lotion, with 1.2% dimethicone | 0.5% w/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Hand Lotion | 5% w/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Hand Sanitizer with Aloe, 62% ethyl | 5% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Hand Sanitizer cream lotion | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Hand Sanitizer, 80% ethanol, fast drying | 15% v/v | 0/3 | 0/3 | 0/3 | 3/3 | 3/3 | 3/3  |
|  Hand soap liquid gel | 10% w/v | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3  |
|  PNSM | / | 0/3 | 0/3 | 0/3 | 0/3 | 0/3 | 0/3  |

*NT: Not Tested

### c. Competitive Interference Study:

A competitive inhibition study was conducted to evaluate the potential for a high concentration of one target analyte to interfere with the detection of another target analyte at low concentration. Testing was performed with different combinations of low (3x LoD) and high concentrations (1000x LoD or the highest concentration achievable) of co-spiked virus (SARS-CoV-2, Flu A, and Flu B) solutions. All testing conditions were tested in 3 replicates using a single lot of the candidate test. No competitive interference was observed between SARS-CoV-2, Flu A, and Flu B as listed below in Table 4.

K262432 Page 13 of 22

{13}

**Table 4: Competitive Interference Study Results:**

|  Test Conditions |   |   | Test Results (# Positive / # Total)  |   |   |
| --- | --- | --- | --- | --- | --- |
|  Flu A | Flu B | SARS-CoV-2 | Flu A | Flu B | SARS-CoV-2  |
|  High | Low | Neg | 3/3 | 3/3 | 0/3  |
|  High | Neg | Low | 3/3 | 0/3 | 3/3  |
|  High | Low | Low | 3/3 | 3/3 | 3/3  |
|  Low | High | Neg | 3/3 | 3/3 | 0/3  |
|  Neg | High | Low | 0/3 | 3/3 | 3/3  |
|  Low | High | Low | 3/3 | 3/3 | 3/3  |
|  Low | Neg | High | 3/3 | 0/3 | 3/3  |
|  Neg | Low | High | 0/3 | 3/3 | 3/3  |
|  Low | Low | High | 3/3 | 3/3 | 3/3  |
|  PNSM |   |   | 0/3 | 0/3 | 0/3  |

# 4. Detection Limit and Assay Reportable Range:

# **a. Limit of Detection (LoD):**

# *i. Single Analyte LoD:*

Analytical sensitivity studies were conducted to determine the limit of detection (LoD) for the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test. The LoD is defined as the lowest concentration at which 95% of the replicates tested are positive. A preliminary LoD was established independently for SARS-CoV-2, Flu A, and Flu B by testing serial ten-fold or two-fold dilutions of the viral stock diluted into negative clinical matrix (PNSM) and spiked onto sterile swabs. Three replicates were tested at each dilution. The LoD was confirmed by testing an additional 20 replicates with concentrations at the preliminary LoD or at concentrations below the preliminary LoD for each virus. The results of the LoD study are summarized in Table 5.

**Table 5: Limit of Detection Study Results**

|  Virus | LoD in PNSM | LoD per Swab | # Positive / # Total | % Detected  |
| --- | --- | --- | --- | --- |
|  SARS-CoV-2 (USA-WA1/2020) | 2.55x10^{2} TCID_{50}/mL | 1.28x10^{1} TCID_{50}/swab | 57/60 | 95.0%  |
|  Influenza A H1N1 (A/Indiana/02/2020) | 3.64x10^{6} CEID_{50}/mL | 1.82x10^{5} CEID_{50}/swab | 57/60 | 95.0%  |
|  Influenza A H3N2 (A/Montana/08/2023) | 4.88x10^{5} FFU/mL | 2.44x10^{4} FFU/swab | 57/60 | 95.0%  |
|  Influenza B Victoria (B/Washington/02/2019) | 1.05x10^{6} CEID_{50}/mL | 5.25x10^{4} CEID_{50}/swab | 57/60 | 95.0%  |
|  Influenza B Yamagata (B/Phuket/3073/2013) | 4.13x10^{5} CEID_{50}/mL | 2.07x10^{4} CEID_{50}/swab | 57/60 | 95.0%  |

K262432 Page 14 of 22

{14}

### ii. Co-Spike LoD:

A co-spike equivalency study was conducted to assess whether the co-spiked LoDs of the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test are equivalent to the LoDs of the individually tested analytes. Co-spiked samples were prepared in PNSM and spiked with SARS-CoV-2, Flu A, and Flu B viruses either at 1x LoD or 0.3x LoD for all three assay analytes. For each concentration level, 20 replicates were tested. All replicates produced 100% positive results for each analyte at 1x LoD. As expected, the positive rate dropped at 0.3x LoD. Thus, there was no impact on the sensitivity when all analytes were co-spiked supporting the use of co-spiked samples in subsequent analytical studies.

### iii. International Standard Material NIBSC code: 21/368 – Limit of Detection:

The LoD was established using the 1st International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in PNSM. Initially, a preliminary LoD test was performed by spiking \(50~\mu \mathrm{L}\) of each diluted sample onto the sample collection swab head in three replicates. Following this, a confirmatory LoD test with an additional 20 replicates was conducted at the preliminary LoD concentration. To determine LoD concentration, another 20 replicates of three targeted dilutions bracketing the LoD concentration (3x LoD, 1x LoD, and 1/3x LoD) were prepared in PNSM and tested. It was determined that the LoD of the 1st International Standard for SARS-CoV-2 Antigen (NIBSC 21/368) in PNSM was confirmed to be \(2.50\times 10^{2}\) IU/mL (\(1.25\times 10^{1}\) IU/swab).

### b. High Dose Hook Effect:

A High Dose Hook Effect study was conducted to determine if a hook effect (a false negative result) would be observed at high concentrations of the analytes. Undiluted stock of each virus (heat-inactivated SARS-CoV-2, Flu A H1N1, Flu A H3N2, Flu B Victoria lineage, and Flu B Yamagata lineage) was applied separately to the swab and tested in triplicate using a single candidate test lot. No hook effect was observed for any of the tested analytes as shown in Table 6 below.

Table 6: High Dose Hook Effect Study Results

|  Virus/Strain | Concentration | Conc. Units | # Positive / # Total  |   |   |
| --- | --- | --- | --- | --- | --- |
|   |   |   |  SARS-CoV-2 | Flu A | Flu B  |
|  SARS-CoV-2: USA-WA1/2020 | 1.02E+08 | TCID50/mL | 3/3 | 0/3 | 0/3  |
|  Flu A H1N1pdm09: A/Indiana/02/2020 | 9.70E+08 | CEID50/mL | 0/3 | 3/3 | 0/3  |
|  Flu A H3N2: A/Montana/08/2023 | 2.60E+08 | FFU/mL | 0/3 | 3/3 | 0/3  |
|  Flu B Victoria: B/Washington/02/2019 | 2.10E+09 | CEID50/mL | 0/3 | 0/3 | 3/3  |

K262432 Page 15 of 22

{15}

|  Virus/Strain | Concentration | Conc. Units | # Positive / # Total  |   |   |
| --- | --- | --- | --- | --- | --- |
|   |   |   |  SARS-CoV-2 | Flu A | Flu B  |
|  Flu B Yamagata: B/Phuket/3073/2013 | 1.10E+09 | CEID50/mL | 0/3 | 0/3 | 3/3  |

# c. Inclusivity:

The inclusivity study was performed to determine if the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test can detect target analytes across a variety of strains by establishing analytical sensitivity for SARS-CoV-2 strains, Flu A strains, and Flu B strains on the candidate test kit. Studies were done by performing testing using one lot of the candidate device on a total of 1 SARS-CoV-2 strain, 20 Flu A strains, and 6 Flu B strains. Three replicates were tested per dilution for each virus. The lowest concentration in which all replicates tested positive was determined to be the analytical sensitivity. The results of the inclusivity study summarized in Table 7 below demonstrate that the test results can detect analytes across a range of viral strains.

Table 7: Inclusivity Study Results

|  Virus | Strain | Minimum Detectable Concentration  |
| --- | --- | --- |
|  Influenza A (H1N1) | A/Brownsville/39H/2009 | 2.0 x 10^{2} TCID_{50}/mL  |
|   |  A/Massachusetts/15/2013 | 1.6 x 10^{6} CEID_{50}/mL  |
|   |  A/Bangladesh/3002/2015 | 3.3 x 10^{5} CEID_{50}/mL  |
|   |  A/Michigan/45/2015 | 5.0 x 10^{6} CEID_{50}/mL  |
|   |  A/St. Petersburg/61/2015 | 9.3 x 10^{5} CEID_{50}/mL  |
|   |  A/Hawaii/66/2019 | 1.9 x 10^{7} CEID_{50}/mL  |
|   |  A/Wisconsin/588/2019 | 1.4 x 10^{4} FFU/mL  |
|   |  A/Iowa/53/2015 | 2.9 x 10^{6} CEID_{50}/mL  |
|   |  A/Idaho/07/2018 | 3.2 x 10^{2} TCID_{50}/mL  |
|   |  A/Dominican/Republic/7293/2013 | 5.0 x 10^{3} TCID_{50}/mL  |
|   |  A/Ohio/09/2015 (H1N1)v | 1.8 x 10^{6} CEID_{50}/mL  |
|  Influenza (H1N2) | A/Minnesota/19/2011 (H1N2)v | 8.0 x 10^{6} CEID_{50}/mL  |
|  Influenza A (H3N2) | A/Hong Kong/2671/2019 | 3.1 x 10^{6} CEID_{50}/mL  |
|   |  A/Singapore/INFIMH-16-0019/2016 | 1.1 x 10^{5} CEID_{50}/mL  |
|   |  A/New York/21/2020 | 1.3 x 10^{5} FFU/mL  |
|   |  A/Tasmania/503/2020 | 3.3 x 10^{4} FFU/mL  |
|   |  A/Alaska/232/2015 | 6.5 x 10^{4} CEID_{50}/mL  |
|   |  A/Indiana/08/2011 (H3N2)v | 8.1 x 10^{2} TCID_{50}/mL  |
|  Influenza A (H7N3) | A/northern pintail/Illinois/10OS3959/2010 (H7N3) | 3.5 x 10^{5} CEID_{50}/mL  |
|  Influenza A (H5N1) | A/mallard/Wisconsin/2576/2009 (H5N1) | 4.0 x 10^{5} CEID_{50}/mL  |
|  Influenza B (non-Victoria non-Yamagata) | B/Maryland/1/1959 | 2.2 x 10^{1} CEID_{50}/mL  |

K262432 Page 16 of 22

{16}

|  Virus | Strain | Minimum Detectable Concentration  |
| --- | --- | --- |
|  Influenza B (Victoria Lineage) | B/Colorado/6/2017 | 4.0 x 10^{4} CEID_{50}/mL  |
|   |  B/Florida/78/ 2015 | 4.3 x 10^{5} CEID_{50}/mL  |
|  Influenza B (Yamagata Lineage) | B/Wisconsin/1/2010 | 3.5 x 10^{1} TCID_{50}/mL  |
|   |  B/Florida/4/2006 | 1.8 x 10^{4} CEID_{50}/mL  |
|   |  B/Texas/06/2011 | 1.6 x 10^{6} CEID_{50}/mL  |
|  SARS-CoV-2 (B.1.1.529, Omicron) | USA/MD-HP20874/2021 | 2.5 x 10^{2} TCID_{50}/mL  |

# **d. Assay Reportable Range:**

Not applicable; the device is a binary qualitative assay that is visually read.

# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

# **a. Controls:**

# *i. Internal Controls:*

Each CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test has a built-in “Control” region which serves as an internal procedural control indicated by the presence of a colored line in the control line region (“C line”). The “C line” should always appear if the test has been performed properly.

# *ii. External Controls:*

External quality control materials are not included in the test kit but are available separately for optional use by professional users.

# **b. Stability:**

# *i. Real-time Stability:*

A real-time stability study was conducted to establish the shelf-life of the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test. Three different lots were stored at four conditions 2°C, 15°C, 30°C and RT, and were tested at 3-month intervals from time 0 (within one month after production of the three device lots included in the study) up to 27 months. The testing panel consisted of a negative sample and a positive sample with co-spiked analytes at 2x LoD. The samples were prepared in pooled nasal wash matrix. Each sample was tested in five replicates across all three lots stored at specified storage duration. At the time of clearance, all study data met the protocol defined acceptance criteria and support storage of the test kits from 2-30°C for up to 24 months.

# *ii. Open-kit Stability:*

An open-kit stability study was conducted to determine the stability of the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test kit when left outside of its packaging. A test panel comprised of five (5) PNSM only negative control samples and five (5) low positive (2x LoD co-spiked) contrived samples were tested at zero (0) hours, one (1) hour, and two (2) hours after the device packaging was opened, with the one-hour and two-hour storage taking place at the worst-case condition for room temperature (i.e., 30 ± 3°C). A single lot of the candidate test kit was used in the study. Results were acceptable for all conditions tested.

K262432 Page 17 of 22

{17}

# iii. Shipping Stability:

The purpose of this study was to evaluate the stability of the unopened CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test kit under extreme summer and winter temperature and humidity conditions that may be encountered during the shipping and handling of the test. A single test lot was exposed to simulated extreme summer shipping temperature and humidity condition (60°C and 85% RH) over an 8-day period with testing occurring every 24 hours and winter shipping temperature conditions (-10°C for 8 hrs and 18°C for 4 hrs) with testing occurring at the end of entire 12-hr simulation. Performance was assessed pre- (T0) and post-distribution (Td) across five replicates of PNSM negative sample and co-spiked moderate positive (5x LoD) sample contrived in PNSM. The test kit was stable at all shipping conditions tested.

# 6. Assay Cut-Off:

Not applicable as this is a qualitative visually read assay without numeric data.

# B Comparison Studies:

# 1. Method Comparison with Predicate Device:

See section C (clinical studies) below.

# 2. Matrix Comparison:

Not Applicable, the device only uses one matrix (i.e., anterior nasal swab). However, a matrix equivalency study was conducted between pooled negative nasal swab matrix (PNSM) and surrogate pooled negative nasal wash matrix (PNW) that was used in multiple analytical studies. The data demonstrated the equivalent performance of the test with both matrices.

# C Clinical Studies:

# 1. Clinical Sensitivity and Specificity:

A prospective clinical study to assess the performance of the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test for the detection of SARS-CoV-2, Flu A, and Flu B protein antigens in self-collected anterior nasal swab samples when compared to an FDA cleared RT-PCR assay was conducted from April 2025 to May 2026 at seven (7) sites throughout the United States. Subjects who exhibited symptoms of infection consistent with COVID-19 or influenza at the time of collection were enrolled in the clinical study.

A total of A total of 632 subjects were enrolled in the study, of which 606 subjects were evaluable within 5 days post symptoms onset (DPSO). The subjects enrolled either self-collected a dual anterior nares sample or had a dual anterior nares sample collected from him/her by another individual for testing. A matched anterior nasal swab sample was also taken from each subject by a healthcare professional for testing on a highly sensitive, FDA 510(k) cleared RT PCR test as the comparator.

The subject demographics are shown in Table 8 below.

K262432 Page 18 of 22

{18}

**Table 8: Subject Demographics**

|   | Subjects by lay-user (Tester) collection and testing (N=150) | Self-collecting and testing (N=456) | Overall (N=606)  |
| --- | --- | --- | --- |
|  Age: Mean (SD) | 11.8 (12.9) | 40.8 (17.9) | 33.6 (20.9)  |
|  Age: Median [Min, Max] | 9 [2, 83] | 38 [14, 95] | 30 [2, 95]  |
|  **Age Group**  |   |   |   |
|  2 - <14 years of age | 126 (84.0%) | 0 (0.0%) | 126 (20.8%)  |
|  14-21 years of age | 12 (8.0%) | 64 (14.0%) | 76 (12.5%)  |
|  22-64 years of age | 10 (6.7%) | 338 (74.1%) | 348 (57.4%)  |
|  ≥65 years of age | 2 (1.3%) | 54 (11.8%) | 56 (9.2%)  |
|  **Sex at Birth***  |   |   |   |
|  Male | 86 (57.3%) | 159 (34.9%) | 245 (40.4%)  |
|  Female | 64 (42.7%) | 296 (64.9%) | 360 (59.4%)  |
|  **Ethnicity**  |   |   |   |
|  Hispanic/Latino | 64 (42.7%) | 112 (24.6%) | 176 (29.0%)  |
|  Not Hispanic/Latino | 81 (54.0%) | 315 (69.1%) | 396 (65.3%)  |
|  Unknown/Prefer not to answer | 5 (3.3%) | 29 (6.4%) | 34 (5.6%)  |
|  **Race**  |   |   |   |
|  American Indian or Alaskan Native | 2 (1.3%) | 5 (1.1%) | 7 (1.2%)  |
|  Asian | 19 (12.7%) | 77 (16.9%) | 96 (15.8%)  |
|  Black or African American | 16 (10.7%) | 89 (19.5%) | 105 (17.3%)  |
|  Native Hawaiian/Pacific Islander | 0 (0.0%) | 1 (0.2%) | 1 (0.2%)  |
|  White | 59 (39.3%) | 235 (51.5%) | 294 (48.5%)  |
|  Unknown/Prefer not to answer | 14 (9.3%) | 19 (4.2%) | 35 (5.4%)  |
|  Other (Mixed race/biracial) | 40 (26.7%) | 30 (6.6%) | 70 (11.6%)  |

*One subject preferred not to answer sex at birth

The performance of the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test when compared to the FDA cleared highly sensitive RT-PCR molecular assays are presented in the tables below.

**Table 9: SARS-CoV-2 Clinical Performance**

|   | Comparator Positive | Comparator Negative | Total  |
| --- | --- | --- | --- |
|  **Candidate Positive** | 47 | 0 | 47  |
|  **Candidate Negative** | 4 | 555 | 559  |
|  **Total** | 51 | 555 | 606  |
|  **Positive Percent Agreement (PPA) = (47/51) = 92.2% (95% CI: 81.5% - 96.9%)**  |   |   |   |
|  **Negative Percent Agreement (NPA) = (555/555) = 100.0% (95% CI: 99.3% - 100.0%)**  |   |   |   |

K262432 Page 19 of 22

{19}

Table 10: Influenza A Clinical Performance

|   | Comparator Positive | Comparator Negative | Total  |
| --- | --- | --- | --- |
|  Candidate Positive | 45 | 1 | 46  |
|  Candidate Negative | 7 | 553 | 560  |
|  Total | 52 | 554 | 606  |
|  Positive Percent Agreement (PPA) = (45/52) = 86.5% (95% CI: 74.7% - 93.3%)  |   |   |   |
|  Negative Percent Agreement (NPA) = (553/554) = 99.8% (95% CI: 99.0% - 100.0%)  |   |   |   |

Table 11: Influenza B Clinical Performance

|   | Comparator Positive | Comparator Negative | Total  |
| --- | --- | --- | --- |
|  Candidate Positive | 31 | 0 | 31  |
|  Candidate Negative | 5 | 570 | 575  |
|  Total | 36 | 570 | 606  |
|  Positive Percent Agreement (PPA) = (31/36) = 86.1% (95% CI: 71.3% - 93.9%)  |   |   |   |
|  Negative Percent Agreement (NPA) = (570/570) = 100.0% (95% CI: 99.3% - 100.0%)  |   |   |   |

### 2. Clinical Cut-Off:

Not Applicable. The candidate device is a qualitative assay with a visually read binary result without numeric raw data.

### 3. Usability and Readability Study:

#### a. Usability and User Comprehension Study:

A total of 55 evaluable subjects participated in the human factors assessment portion of the study to evaluate the usability of the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test, and to evaluate the labeling and comprehension of the investigational test QRI, when performed by lay users in a simulated home environment.

Evaluation of the human user experience indicated high usability of the investigational test. Of the 55 subjects who participated in the usability and comprehension assessment, 53 (96.4%) found the instructions clear and easy to follow, 54 (98.2%) found the sample collection easy to perform, and all subjects had no difficulty reading the test results. Overall, 99.5% of all critical tasks associated with sample collection and running the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test kit were performed correctly. Among all non-critical tasks, 98.8% were performed correctly as shown in Table 12 below.

K262432 Page 20 of 22

{20}

**Table 12: Critical vs Non-critical Tasks Correctly Performed**

|  Steps | Tasks performed correctly | Total number of tasks | Percentage of tasks performed correctly  |
| --- | --- | --- | --- |
|  **Critical** | 438 | 440 | 99.5%  |
|  **Non-Critical** | 163 | 165 | 98.8%  |
|  **Total** | 601 | 605 | 99.3%  |

# ***b. Readability Study:***

All 55 subjects who participated in the human factors assessment also interpreted a panel of mock investigational tests with various results that reflected test concentration levels at 1.9x and 5x LoD in a blinded and randomized fashion.

An overall mock result interpretation accuracy of 98.2% across negative, invalid, and positive test results at both 1.9x LoD and 5x LoD was observed, indicating that lay users were able to correctly interpret test results, including low-intensity positive results. Thus, the readability study results are acceptable.

# **D Other Supportive Study/Device Information**

# **1. Flex Studies:**

To assess the robustness and risk for false results of the CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid Test when deviating from the IFU/QRI test steps, flex studies were conducted that assessed all major aspects of the test procedure (sample volume, reading time, other deviations from the procedure [delay in mixing, delay in addition of sample to the well, incubation time] and variability of environmental test conditions that the test may be subjected to when in use (lighting, disturbance during use, temperature and humidity stress conditions). Testing was performed with negative samples and contrived positive samples generated by diluting viruses into PNSM at a co-spiked concentration of 2x LoD. False results are observed with reading at less than 8 minutes, and the use of only 1 drop of extracted sample for testing. Thus, the studies support that the test is robust in the intended use condition with an insignificant risk of erroneous result.

# **E Expected Values/Reference Range:**

A patient sample is expected to be negative for influenza A, influenza B, and SARS-CoV-2.

# **VI Proposed Labeling:**

The labeling supports the finding of substantial equivalence for this device.

K262432 Page 21 of 22

{21}

# **VII Conclusion:**

The submitted information in this premarket notification is complete and supports a substantial equivalence decision to the predicate.

K262432 Page 22 of 22

---

**Source:** [https://fda-staging.innolitics.com/device/K262432](https://fda-staging.innolitics.com/device/K262432)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).