Pen Needle (Ordinary Type I, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Safety Type VII, Ordinary Type IB, Safety type VIII, Safety Type IX, Safety Type X, Ordinary Type XI, Safety Type XII)
K262427 · Hh Global Technology, Inc. · FMI · Aug 10, 2026 · General Hospital
Device Facts
Record ID
K262427
Device Name
Pen Needle (Ordinary Type I, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Safety Type VII, Ordinary Type IB, Safety type VIII, Safety Type IX, Safety Type X, Ordinary Type XI, Safety Type XII)
Applicant
Hh Global Technology, Inc.
Product Code
FMI · General Hospital
Decision Date
Aug 10, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Pen Needles are sterile, single use needles intended for using with pen injector devices for the injection of drug.
Device Story
Sterile, single-use pen needle; consists of stainless-steel cannula, plastic hub, and protective shield; used with compatible pen injectors for subcutaneous medication delivery. Operated by healthcare personnel or adult patients/lay persons in clinical or home settings. Device provides sterile fluid pathway; needle penetrates skin to deliver drug from injector. Benefits include reliable, sterile medication administration. Modification adds 12.7 mm needle length to Ordinary Type I models.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation included mechanical performance testing per ISO 7864 and ISO 11608-2, and accelerated aging studies to confirm performance over 5-year shelf life.
Technological Characteristics
Stainless steel needle; Polypropylene (PP) hub/container; PP/ABS shield/sleeve; medical silicone lubricant; UV glue jointing. Sterilized via irradiation (SAL 10^-6). Complies with ISO 9626, ISO 7864, ISO 11608-1, ISO 11608-2, and ISO 10993 biocompatibility standards.
Indications for Use
Indicated for use by healthcare personnel, adult patients, and adult lay persons for the subcutaneous injection of medication using pen injector devices.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 10, 2026
HH Global Technology, Inc.
% Ningning Wang
Registered Engineer
Tianjin Huahong Technology Co., Ltd.
A01, Plant B, # 278, Hangkong Rd., Tianjin Pilot Free Trade Zone(Air Port Industrial Park)
Tianjin, 300308, China
Re: K262427
Trade/Device Name: Pen Needle (Ordinary Type I, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Safety Type VII, Ordinary Type IB, Safety type VIII, Safety Type IX, Safety Type X, Ordinary Type XI, Safety Type XII)
Regulation Number: 21 CFR 880.5570
Regulation Name: Hypodermic Single Lumen Needle
Regulatory Class: Class II
Product Code: FMI
Dated: July 15, 2026
Received: July 16, 2026
Dear Ningning Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262427 - Ningning Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262427 - Ningning Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
KYRAN R. GIBSON -S
for
Shruti Mistry
Assistant Director
DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K262427 ? |
| Please provide the device trade name(s). | | ? |
| Pen Needle (Ordinary Type I, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Safety Type VII, Ordinary Type IB, Safety type VIII, Safety Type IX, Safety Type X, Ordinary Type XI, Safety Type XII) | | |
| Please provide your Indications for Use below. | | ? |
| Pen Needles are sterile, single use needles intended for using with pen injector devices for the injection of drug. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
510(k) Summary
# 510(k) summary
# I Submitter
HH Global Technology, Inc.
6303 Woodlake Drive
Buford, GA 30518
Contact person: Grace Hammond
President
Tel.: 770-309-8237
E-mail: grace.hammond@hhglobaltech.com
# Correspondent
Tianjin Huahong Technology Co., Ltd.
A01, Plant B, # 278, Hangkong Rd., Tianjin Pilot Free Trade Zone (Air Port Industrial Park),
Tianjin, 300308 CHN
Contact Person: Ningning Wang
Tel.: +86-13021381776
E-mail: ningning.wang@hh-technology.com
Preparation date: July 13, 2026
# II Proposed Device
Trade Name of Device: Pen Needle (Ordinary Type I, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Ordinary Type IB, Ordinary Type XI, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Safety Type VII, Safety Type VIII, Safety Type IX, Safety Type X, and Safety Type XII)
Regulation description: Hypodermic single lumen needle
Regulation Number: 21 CFR 880.5570
Regulatory Class: Class II
Product code: FMI
Review Panel: General Hospital
# III Predicate Device
Predicate Device:
510(k) Number: K252886
Product Code: FMI
Classification: 21 CFR 880.5570
Trade Name: Pen Needle (Ordinary Type I, Ordinary Type VIA, Ordinary Type VIB, Ordinary
1
{5}
510(k) Summary
Type VIC, Ordinary Type IB, Ordinary Type XI, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Safety Type VII, Safety Type VIII, Safety Type IX, Safety Type X, and Safety Type XII)
Manufacturer: Tianjin Huahong Technology Co., Ltd.
### IV Device description
The proposed device is a sterile, single-use Pen Needle intended for use with pen injectors for subcutaneous injection of medication. The Pen Needle consists of a stainless-steel needle cannula attached to a plastic hub and is designed to provide a sterile pathway for medication delivery when used with compatible pen injectors.
The proposed device family includes the following models: Ordinary Type I, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Ordinary Type IB, Ordinary Type XI, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Safety Type VII, Safety Type VIII, Safety Type IX, Safety Type X, and Safety Type XII.
The proposed devices have the same intended use, fundamental design features, materials, manufacturing process, sterilization method, and packaging configurations as the predicate device cleared under K252886.
For this Special 510(k) submission, a new needle length specification of 12.7 mm is added to the Ordinary Type I Pen Needle models. The existing needle length options remain unchanged, and no modifications are made to the other Pen Needle models, including the Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Ordinary Type IB, Ordinary Type XI and all Safety Type models. Except for the addition of the 12.7 mm needle length specification for Ordinary Type I models, the proposed device remains identical to the predicate device in terms of design characteristics and technological features.
### V Indications for use
Pen Needles are sterile, single use needles intended for using with pen injector devices for the injection of drug.
### VI Comparison of technological characteristics with the predicate device
The comparison and discussion between the proposed device and the predicate device are provided in Table 1.
2
{6}
510(k) Summary
Table 1 Substantial Equivalence Comparison
| Item | Predicate Device K252886 | Proposed device | Comments |
| --- | --- | --- | --- |
| Trade Name | Pen Needle (Ordinary Type I, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Ordinary Type IB, Ordinary Type XI, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Safety Type VII, Safety Type VIII, Safety Type IX, Safety Type X, and Safety Type XII) | Pen Needle (Ordinary Type I, Ordinary Type VIA, Ordinary Type VIB, Ordinary Type VIC, Ordinary Type IB, Ordinary Type XI, Safety Type IIA, Safety Type IIB, Safety Type IIIA, Safety Type IIIB, Safety Type IV, Safety Type V, Safety Type VII, Safety Type VIII, Safety Type IX, Safety Type X, and Safety Type XII) | Same |
| Product Code | FMI | FMI | Same |
| Regulation number | 880.5570 | 880.5570 | Same |
| Classification Name | Hypodermic Single Lumen Needle | Hypodermic Single Lumen Needle | Same |
| Device Class | Class II | Class II | Same |
| Intended Use | Intended for use with pen injector devices for injection of drugs | Intended for use with pen injector devices for injection of drugs | Same |
| Intended Patient Population | Healthcare personnel, adult patients and adult lay persons | Healthcare personnel, adult patients and adult lay persons | Same |
3
{7}
510(k) Summary
| Item | Predicate Device K252886 | Proposed device | Comments |
| --- | --- | --- | --- |
| Indications for Use | Pen Needles are sterile, single use needles intended for using with pen injector devices for the injection of drug. | Pen Needles are sterile, single use needles intended for using with pen injector devices for the injection of drug. | Same |
| Prescription/OTC | OTC | OTC | Same |
| Supplied sterile | Yes | Yes | Same |
| Sterility | SAL:10^{-6} | SAL:10^{-6} | Same |
| Sterilization | Irradiation Sterilized | Irradiation Sterilized | Same |
| Single use | Yes | Yes | Same |
| Shelf Life | 5 years | 5 years | Same |
| Needle Gauge | 29G, 30G, 31G, 32G, 33G, 34G | 29G, 30G, 31G, 32G, 33G, 34G | Same |
| Needle Length | 3.5mm, 4mm, 5mm, 6mm, 8mm, 10mm, 12mm | 3.5mm, 4mm, 5mm, 6mm, 8mm, 10mm, 12mm, 12.7mm (12.7 mm added for Ordinary Type I models only) | Different |
| Materials | Needle Tube: Stainless Steel Needle Hub: Polypropylene (PP) Needle Container: PP Needle Shield: PP/ABS (ABS is only available for safety type) | Needle Tube: Stainless Steel Needle Hub: Polypropylene (PP) Needle Container: PP Needle Shield: PP/ABS (ABS is only available for safety type) Adjusting Sleeve: PP/ABS Lubricant: Medical silicone-based lubricant | Same |
4
{8}
510(k) Summary
| Item | Predicate Device K252886 | Proposed device | Comments |
| --- | --- | --- | --- |
| | Adjusting Sleeve: PP/ABS Lubricant: Medical silicone-based lubricant Jointing medium: UV glue Spring: Stainless steel (Safety type only) | Jointing medium: UV glue Spring: Stainless steel (Safety type only) | |
| Biocompatibility | Complied with ISO10993 series standards, and the following tests are performed - Cytotoxicity: No cytotoxicity - Skin Irritation: No evidence of skin irritation - Skin Sensitization: No evidence of sensitization -Acute and Subacute Systemic Toxicity: No systemic toxicity -Hemolysis: No evidence of hemolysis -Pyrogen: Non-pyrogenic | Complied with ISO10993 series standards, and the following tests are performed - Cytotoxicity: No cytotoxicity - Skin Irritation: No evidence of skin irritation - Skin Sensitization: No evidence of sensitization -Acute and Subacute Systemic Toxicity: No systemic toxicity -Hemolysis: No evidence of hemolysis -Pyrogen: Non-pyrogenic | Same |
| Principle of Operation | Used with a pen injector for injection of drugs | Used with a pen injector for injection of drugs | Same |
5
{9}
510(k) Summary
| Item | Predicate Device K252886 | Proposed device | Comments |
| --- | --- | --- | --- |
| Performance | Complied with ISO 9626, ISO 7864, ISO11608-1, ISO 11608-2 | Complied with ISO 9626, ISO 7864, ISO11608-1, ISO 11608-2 | Same |
| **Difference Analysis:** The proposed device adds a 12.7 mm needle length option for Ordinary Type I Pen Needle models only. Except for this additional needle length, the proposed device remains identical to the predicate device in terms of intended use, design, materials, manufacturing process, sterilization, packaging, and other technological characteristics. The additional needle length has been verified through applicable performance testing and does not raise new questions regarding safety or effectiveness when compared to the predicate device. | | | |
6
{10}
510(k) Summary
## VII Non-Clinical Testing
### 7.1 Verification and Validation Activities – Design Change
The proposed Pen Needle differs from the predicate device (K252886) by the addition of a new needle length specification of 12.7 mm for the Ordinary Type I Pen Needle models. No changes have been made to the intended use, fundamental operating principle, device design, materials, manufacturing process, sterilization method, packaging configuration, or overall performance specifications of the device.
Design verification and validation activities were conducted to evaluate the impact of the added needle length specification on device safety and performance in accordance with established design control procedures.
### 7.2 Non-Clinical Testing Overview
Non-clinical testing was performed to support the safety and performance of the proposed device with the new 12.7 mm needle length specification. The testing program included general performance testing, needle dimension and mechanical performance evaluation in accordance with ISO 7864, Pen Needle performance testing in accordance with ISO 11608-2, and accelerated aging evaluation to assess performance throughout the claimed shelf life.
Testing was performed on the proposed Ordinary Type I Pen Needle with the 12.7 mm needle length specification. Both time-zero and after accelerated aging testing were conducted to verify that the modified device maintains its established performance requirements after aging.
All testing was conducted in accordance with internal specifications and applicable standards. Acceptance criteria were established based on the predicate device specifications and applicable regulatory and consensus standards.
### 7.3 Test Results Summary
All non-clinical verification and validation activities met their predefined acceptance criteria. The results of performance testing, ISO 7864 testing, and ISO 11608-2 testing demonstrated that the proposed Pen Needle with the 12.7 mm needle length specification meets the applicable requirements for needle dimensions, mechanical performance, and functional performance.
7
{11}
510(k) Summary
The addition of the 12.7 mm needle length specification for the Ordinary Type I Pen Needle models does not adversely affect device performance, sterility assurance, or packaging integrity.
### 7.4 Conclusions
The results of non-clinical testing demonstrate that the proposed Pen Needle is substantially equivalent to the predicate device (K252886). The addition of a new 12.7 mm needle length specification for the Ordinary Type I Pen Needle models does not alter the device's intended use, fundamental scientific technology, or overall device design.
The design modification does not introduce new risks or raise different questions of safety and effectiveness when compared to the predicate device. The proposed device remains as safe and effective as the legally marketed predicate device.
### VIII Clinical Testing
No clinical study is included in this submission.
### IX Conclusion
Based on the design verification and validation activities and non-clinical performance testing, the proposed Pen Needle was demonstrated to meet all predefined acceptance criteria.
The addition of a 12.7 mm needle length specification for the Ordinary Type I Pen Needle models does not change the intended use, fundamental operating principle, materials, manufacturing process, sterilization method, packaging configuration, or other technological characteristics of the device. The verification and validation results confirm that the design modification does not adversely affect device safety, effectiveness, or performance.
In conclusion, the proposed device is substantially equivalent to the predicate device (K252886). The design modification does not raise different questions of safety and effectiveness when compared to the predicate device, and the proposed device is as safe and effective as the legally marketed predicate device.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.