The Willow 12 Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.
Device Story
Single-use guidewire with shapeable tip; facilitates intravascular access, positioning, and exchange of interventional devices in small-diameter, tortuous neuro and peripheral vasculature. Physician-operated in clinical settings; wire torqued manually to navigate vasculature. Input: manual manipulation by physician. Output: mechanical guidance for catheters. Benefits: enables precise navigation to target locations during diagnostic and interventional procedures. Includes introducer, torque device, and shaping mandrel.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence is supported by extensive non-clinical bench performance testing, including coating lubricity/durability, tensile strength, torqueability, kink resistance, flex fatigue, and simulated use in tortuous models. Biocompatibility is supported by testing on the larger Willow 24 reference device.
Technological Characteristics
Guidewire with nickel-titanium laser-cut tip, platinum-tungsten alloy coil, and core wire with PTFE coating. Dimensions: ≤0.012" distal, <0.013" proximal. Features proprietary hydrophilic coating. Sterilized via 100% Ethylene Oxide (ISO 11135:2014). Accessories: introducer, torque device, shaping mandrel. Mechanical device; no software or electronic components.
Indications for Use
Indicated for adult patients undergoing endovascular treatment in neurovascular and peripheral vasculatures requiring selective introduction and positioning of catheters and interventional devices.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 14, 2026
Arbor Endovascular, LLC
Kathy Tansey
Vice President of Regulatory, Clinical, and Quality
2345 Bering Drive
San Jose, California 95131
Re: K262424
Trade/Device Name: Willow 12 Guidewire
Regulation Number: 21 CFR 870.1330
Regulation Name: Catheter Guide Wire
Regulatory Class: Class II
Product Code: MOF, DQX
Dated: July 15, 2026
Received: July 16, 2026
Dear Kathy Tansey:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262424 - Kathy Tansey
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262424 - Kathy Tansey
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# NAIRA MURADYAN -S
Naira Muradyan, PhD
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262424 | ? |
| Please provide the device trade name(s). | | ? |
| Willow 12 Guidewire | | |
| Please provide your Indications for Use below. | | ? |
| The Willow 12 Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures. | | |
| Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
Arbor Endovascular, LLC
510(k) Premarket Notification
Willow® 12 Guidewire
# 510(k) Summary K262424
# Submitter Name, Address and Contact:
| Submitter: | Arbor Endovascular, LLC 2345 Bering Drive San Jose, CA 95131 |
| --- | --- |
| Sponsor Contact: | Kathy Tansey Vice President of Regulatory, Clinical, and Quality Email: ktansey@arborendo.com Phone: +1 (408) 391-6536 |
| --- | --- |
| Date Prepared: | August 10, 2026 |
| --- | --- |
# Device Name and Classification:
| Trade/Proprietary Name: | Willow® 12 Guidewire |
| --- | --- |
| Common Name: | Wire, Guide, Catheter |
| Classification Name: | Wire, Guide, Catheter, 21 CFR 870.1330 – Class II |
| Product Code: | MOF, DQX |
| Predicate Device: | 0.014” Willow Guidewire (K253168) |
| Reference Device: | Willow™ 18 Guidewire (K260130) |
| Reference Device: | Willow™ 24 Guidewire (K260537) |
| Reference Device: | Synchro 0.010” Guidewire (K032146) |
# Device Description:
Like the predicate device, the Willow 12 Guidewire is a single-use product with a shapeable tip, used to gain intravascular access to facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for neuro and peripheral diagnostic and interventional procedures. The wire can be torqued to facilitate navigation through the vasculature and has a distal outer diameter ≤0.012” (0.30mm) and a proximal outer diameter of <0.013” (0.33mm).
1
{5}
Arbor Endovascular, LLC
510(k) Premarket Notification
Willow® 12 Guidewire
# **Accessories:**
The Willow 12 Guidewire is provided with three (3) accessories: an Introducer, a Torque Device, and a Shaping Mandrel.
# **Indications for Use:**
The Indications for Use for the Willow 12 Guidewire are the same as for the predicate device and are as follows:
The Willow 12 Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.
# **Technological Characteristics and Product Feature Comparison:**
The Willow 12 Guidewire is substantially equivalent to the predicate device based on the following:
- Same indications for use
- Same fundamental materials and similar manufacturing process
- Same fundamental design and technology
- Same operating principles
- Same materials and processes for packaging
- Same sterilization method and process for devices
A comparison of the subject device with the predicate device is summarized in **Table 1** below.
**Table 1. Substantial Equivalence Comparison**
| Characteristic | Predicate Device 0.014" Willow Guidewire (K253168) | Subject Device Willow 12 Guidewire (K262424) |
| --- | --- | --- |
| **Classification Name** | Wire, Guide, Catheter, 21 CFR 870.1330, Class II | Same |
| **Product Code** | MOF, DQX | Same |
| **Review Panel** | Neurology | Same |
| **510(k) Submitter** | Arbor Endovascular, LLC 2345 Bering Drive San Jose, CA 95131 | Same |
2
{6}
Arbor Endovascular, LLC
510(k) Premarket Notification
Willow® 12 Guidewire
| Characteristic | Predicate Device 0.014" Willow Guidewire (K253168) | Subject Device Willow 12 Guidewire (K262424) |
| --- | --- | --- |
| Indications for Use | The 0.014" Willow Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures. | The Willow 12 Guidewire is intended for general intravascular use, including neurovascular and peripheral vasculatures. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neuro vasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures. |
| Device Description/ Principle of Operation | The 0.014" Willow Guidewire is a single-use product with a shapeable tip, used to gain intravascular access to facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for neuro and peripheral diagnostic and interventional procedures. The wire can be torqued to facilitate navigation through the vasculature. | The Willow 12 Guidewire is a single-use product with a shapeable tip, used to gain intravascular access to facilitate the positioning and exchange of interventional devices in small diameter, tortuous vasculature for neuro and peripheral diagnostic and interventional procedures. The wire can be torqued to facilitate navigation through the vasculature. |
| Target Population | Patients undergoing endovascular treatment, including in the neuro and peripheral vasculatures. | Same |
| Accessories | Guidewire Introducer, Torque Device, and Shaping Mandrel | Guidewire Introducer, Torque Device, and Shaping Mandrel |
| Core Wire | Commonly used metal alloy with PTFE coating on the proximal section | Same |
| Diameter | 0.014" | 0.012" Distal 0.013" Proximal |
| Core Wire Length | 215 cm Access Length 315 cm Exchange Length | 215 cm Access Length 350 cm Exchange Length |
| Guidewire Tip (slotted tube) | Nickel-Titanium Laser Cut Nitinol | Same |
| Radiopaque Coil | Platinum-Tungsten Alloy | Same |
| Radiopaque Marker | Platinum-Iridium Band | Platinum-Tungsten Alloy Coil |
| Adhesive | UV Curable Adhesive | Same |
| Primer | Parylene Dimer | Same |
| Hydrophilic Coating | Proprietary Hydrophilic Coating | Same |
| Guidewire Introducer | Included | Same |
| Shaping Mandrel | Included | Same |
| Torque Device | Included | Same |
| Dispenser Hoop | High Density Polyethylene (HDPE) | Same |
| Accessory Card | HDPE | Same |
| Sterile Pouch | Tyvek®/Nylon- Polyethylene | Same |
| Shipping Carton | Folding Box Board (FBB) | Same |
| Sterilization Method | 100% Ethylene Oxide | Same |
| How Supplied | Single Use/Sterile | Same |
3
{7}
Arbor Endovascular, LLC
510(k) Premarket Notification
Willow® 12 Guidewire
The difference between the devices, namely the smaller overall device diameter of the subject device, does not raise new questions of safety and effectiveness.
# **Risk Assessment:**
Risk assessment of the Willow 12 Guidewire has been conducted in accordance with ISO 14971:2019 to show that no new risks were identified compared to the commercially available predicate device. Results of non-clinical testing have demonstrated the subject device is substantially equivalent to the predicate device.
# **Bench Performance Testing:**
The results of the bench testing conducted on the subject device demonstrate that it performs as intended and meets design specifications. A summary of the bench performance testing is provided in **Table 2**.
**Table 2: Bench Testing Summary**
| Test | Test Method Summary | Conclusions |
| --- | --- | --- |
| Coating Lubricity | Hydrophilic coating lubricity was assessed after multiple pull cycles through silicone pads. | Acceptance criteria were met. |
| Coating Durability | Coating durability was assessed after repeating multiple pull cycles through silicone pads. | Acceptance criteria were met. |
| Coating Integrity | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Coating integrity inspected pre- and post-simulated use tracking with comparison to predicate and reference devices. | Coating integrity was reported and considered acceptable. |
| Corrosion Resistance | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). | Acceptance criteria were met. |
| Dimensional and Visual Inspection | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Verified key dimensions of the guidewire. | Acceptance criteria were met. |
| Flex Fatigue | Subjected the guidewire to multiple flexure cycles around cylindrical pins. | Acceptance criteria were met. |
| Fracture | Subjected the guidewire to multiple wrappings around cylinder and visually inspected for signs of fracture. | Acceptance criteria were met. |
| Kink Resistance | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Wrapped the guidewire around test fixture with clinically relevant radii. | Acceptance criteria were met. |
| Particulate Characterization | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Counted particulates of various size ranges after tracking through a tortuous simulated use model, with comparison to the reference device. | Particulate counts were reported and considered acceptable. |
| Radiopacity | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Subject and reference guidewires | All Willow Guidewires demonstrated acceptable radiopacity. |
4
{8}
Arbor Endovascular, LLC
510(k) Premarket Notification
Willow® 12 Guidewire
| Test | Test Method Summary | Conclusions |
| --- | --- | --- |
| | were evaluated under fluoroscopy. | The radiopacity of the subject device was comparable to that of the reference device. |
| Simulated Use | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Guidewires were tested for use with a microcatheter, guidewire introducer, and torque device while navigating to target locations in a tortuous simulated use model. | Acceptance criteria were met. |
| Tensile Strength | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Measured the force required to break at each bond, including distal tip. | Acceptance criteria were met. |
| Tip Flexibility | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Measured the force required to deflect the guidewire tip when held at 5 mm, 10 mm and 20 mm gauge lengths. | Acceptance criteria were met. |
| Tip Shapeability | Shaped the guidewire tip for a total of three (3) times per labeling. | Acceptance criteria were met. |
| Tip Shape Retention | Measured the tip shape retention after tracking the guidewire through a tortuous simulated use model. | Acceptance criteria were met. |
| Torqueability | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Tracked the guidewire through a tortuous simulated use model and evaluated the torque response. | Acceptance criteria were met. |
| Torque Strength | Testing completed per FDA guidance document “*Coronary, Peripheral, and Neurovascular Guidewires – Performance Tests and Recommended Labeling*” (2019). Recorded the number of times the proximal end of the wire can be rotated until it exceeds its maximum rotations and fails. | Acceptance criteria were met. |
### **Sterilization and Shelf-Life Testing:**
The device and its accessories are sterilized by 100% Ethylene Oxide and have been adopted into the existing validated sterilization process in accordance with the principles of ISO 11135:2014 “Sterilization of health-care products - Ethylene oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices” and AAMI TIR28:2016(R)2024 “Product adoption and process equivalence for ethylene oxide sterilization”. A sterility assurance level (SAL) of 10⁻⁶ has been demonstrated. The performance testing completed supports a 1-year shelf-life.
### **Performance Data – Animal, Clinical:**
No animal or clinical studies were conducted as the indications for use and the fundamental scientific technology are the same as that of the predicate. Substantial equivalence of the subject device has been established to the predicate device through the results of non-clinical performance testing.
5
{9}
Arbor Endovascular, LLC
510(k) Premarket Notification
Willow® 12 Guidewire
### **Biocompatibility Testing:**
No new biocompatibility testing was performed - prior biocompatibility evaluations were used to support the biological safety of the subject device. The biocompatibility testing was conducted using a larger guidewire (Willow 24 Guidewire, reference device) with the same materials, manufacturing, and geometry present in the Willow 12 Guidewire in accordance with FDA guidance document, “*Use of International Standard ISO 10993-1, “Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process”*,” and demonstrates that the guidewire and accessories meet biological safety requirements for externally communicating medical devices with direct circulating blood contact for ≤ 24 hours. This statement is supported by the standards assessment and the testing conducted by Arbor Endovascular.
### **Conclusion:**
The subject device is substantially equivalent to the predicate device with regards to device design, materials, intended use, and patient population. The conclusions drawn from the risk assessment and non-clinical testing conducted demonstrate that the subject device performs as intended and is substantially equivalent to the predicate device.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.