MICRUSFRAME, DELTAFILL, and DELTAXSFT Microcoil Delivery Systems are intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. GALAXY G3 Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature. GALAXY G3 Mini Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
Device Story
Neurovascular embolization system; consists of platinum alloy microcoil attached to variable-stiffness Device Positioning Unit (DPU); includes introducer sheath, connecting cable, and Detachment Control Box (DCB). Used by physicians in clinical settings for endovascular procedures. Microcoil delivered via microcatheter to target site (aneurysm/malformation); radiopaque markers on DPU guide positioning. Thermo-mechanical detachment triggered by DCB via connecting cable; energy applied to detachment zone releases coil. Permanent obstruction of blood flow achieved via coil deployment. No changes to design, materials, or operating principles in this submission; modification limited to labeling/indications clarification.
Clinical Evidence
No clinical data required. Submission relies on previously cleared verification and validation data from K211344. Previous bench and in-vivo animal studies (porcine model) demonstrated reliable coil delivery, deployment, and detachment performance.
Technological Characteristics
Platinum alloy microcoils with polypropylene or absorbable polymer sutures; secondary shapes include spherical, complex, or helical. DPU features variable stiffness wire with radiopaque marker bands. Thermo-mechanical detachment via external DCB and connecting cable. No changes to materials or dimensions from predicate.
Indications for Use
Indicated for endovascular embolization of intracranial aneurysms, arteriovenous malformations, and arteriovenous fistulae; vascular occlusion of neurovascular blood vessels; and arterial/venous embolizations in peripheral vasculature. For adult patients (22+ years). Prescription use only.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
Optima Coil System (OptiBlock Line Extension) (K242582)
Submission Summary (Full Text)
{0}
**U.S. FOOD & DRUG**
ADMINISTRATION
August 14, 2026
Cerenovus, Inc.
Ryan Rivera
Senior Regulatory Affairs Program Lead
31 Technology Drive
Irvine, California 92618
Re: K262419
Trade/Device Name: Micrusframe (mfr100202); Micrusframe (mfr100253); Micrusframe (mfr100305);
Micrusframe (mfr100356); Micrusframe (mfr100407); Micrusframe (mfr100412);
Micrusframe (mfr100509); Micrusframe (mfr100517); Micrusframe (mfr100611);
Micrusframe (mfr100626); Micrusframe (mfr100713); Micrusframe (mfr100730);
Micrusframe (mfr100816); Micrusframe (mfr100829); Micrusframe (mfr140225);
Micrusframe (mfr140303); Micrusframe (mfr140304); Micrusframe (mfr140306);
Micrusframe (mfr140406); Micrusframe (mfr140408); Micrusframe (mfr140445);
Micrusframe (mfr140507); Micrusframe (mfr140512); Micrusframe (mfr140609);
Micrusframe (mfr140615); Micrusframe (mfr140711); Micrusframe (mfr140717);
Micrusframe (mfr140812); Micrusframe (mfr140820); Micrusframe (mfr140914);
Micrusframe (mfr140922); Micrusframe (mfr141016); Micrusframe (mfr141025);
Micrusframe (mfr141118); Micrusframe (mfr141127); Micrusframe (mfr141219);
Micrusframe (mfr141230); Micrusframe (mfr180510); Micrusframe (mfr180612);
Micrusframe (mfr180714); Micrusframe (mfr180813); Micrusframe (mfr180830);
Micrusframe (mfr180915); Micrusframe (mfr180933); Micrusframe (mfr181017);
Micrusframe (mfr181034); Micrusframe (mfr181118); Micrusframe (mfr181137);
Micrusframe (mfr181220); Micrusframe (mfr181240); Micrusframe (mfr181343);
Micrusframe (mfr181447); Micrusframe (mfr181550); Micrusframe (mfr181647);
Micrusframe (mfr181750); Micrusframe (mfr181846); Micrusframe (mfr181950);
Micrusframe (mfr182050); Deltafill (dlf100154); Deltafill (dlf100208)
Regulation Number: 21 CFR 882.5950
Regulation Name: Neurovascular Embolization Device
Regulatory Class: Class II
Product Code: HCG, KRD
Dated: July 14, 2026
Received: July 15, 2026
Dear Ryan Rivera:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262419 - Ryan Rivera
Page 2
provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device
{2}
K262419 - Ryan Rivera
Page 3
Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeffrey J. Kustusch -S Digitally signed by Jeffrey J. Kustusch -S
For Sara S. Thompson, DVM
Assistant Director
DHT5A: Division of Neurosurgical,
Neurointerventional, and
Neurodiagnostic Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| MICRUSFRAME (MFR100202); MICRUSFRAME (MFR100253); MICRUSFRAME (MFR100305); MICRUSFRAME (MFR100356); MICRUSFRAME (MFR100407); MICRUSFRAME (MFR100412); MICRUSFRAME (MFR100509); MICRUSFRAME (MFR100517); MICRUSFRAME (MFR100611); MICRUSFRAME (MFR100626); MICRUSFRAME (MFR100713); MICRUSFRAME (MFR100730); MICRUSFRAME (MFR100816); MICRUSFRAME (MFR100829); MICRUSFRAME (MFR140225); MICRUSFRAME (MFR140303); MICRUSFRAME (MFR140304); MICRUSFRAME (MFR140306); MICRUSFRAME (MFR140406); MICRUSFRAME (MFR140408); MICRUSFRAME (MFR140445); MICRUSFRAME (MFR140507); MICRUSFRAME (MFR140512); MICRUSFRAME (MFR140609); MICRUSFRAME (MFR140615); MICRUSFRAME (MFR140711); MICRUSFRAME (MFR140717); MICRUSFRAME (MFR140812); MICRUSFRAME (MFR140820); MICRUSFRAME (MFR140914); MICRUSFRAME (MFR140922); MICRUSFRAME (MFR141016); MICRUSFRAME (MFR141025); MICRUSFRAME (MFR141118); MICRUSFRAME (MFR141127); MICRUSFRAME (MFR141219); MICRUSFRAME (MFR141230); MICRUSFRAME (MFR180510); MICRUSFRAME (MFR180612); MICRUSFRAME (MFR180714); MICRUSFRAME (MFR180813); MICRUSFRAME (MFR180830); MICRUSFRAME (MFR180915); MICRUSFRAME (MFR180933); MICRUSFRAME (MFR181017); | | |
{4}
MICRUSFRAME (MFR181034);
MICRUSFRAME (MFR181118);
MICRUSFRAME (MFR181137);
MICRUSFRAME (MFR181220);
MICRUSFRAME (MFR181240);
MICRUSFRAME (MFR181343);
MICRUSFRAME (MFR181447);
MICRUSFRAME (MFR181550);
MICRUSFRAME (MFR181647);
MICRUSFRAME (MFR181750);
MICRUSFRAME (MFR181846);
MICRUSFRAME (MFR181950);
MICRUSFRAME (MFR182050);
DELTAFILL (DLF100154);
DELTAFILL (DLF100208);
DELTAFILL (DLF100250);
DELTAFILL (DLF100258);
DELTAFILL (DLF100308);
DELTAFILL (DLF100310);
DELTAFILL (DLF100408);
DELTAFILL (DLF100410);
DELTAFILL (DLF100510);
DELTAFILL (DLF100515);
DELTAFILL (DLF100612);
DELTAFILL (DLF100616);
DELTAFILL (DLF100716);
DELTAFILL (DLF100720);
DELTAFILL (DLF100820);
DELTAFILL (DLF100925);
DELTAFILL (DLF180312);
DELTAFILL (DLF180415);
DELTAFILL (DLF180520);
DELTAFILL (DLF180625);
DELTAFILL (DLF180733);
DELTAFILL (DLF180835);
DELTAFILL (DLF180935);
DELTAFILL (DLF181040);
DELTAFILL (DLF181242);
DELTAFILL (DLF181445);
DELTAFILL (DLF181650);
DELTAFILL (DLF181855);
DELTAFILL (DLF182060);
DELTAFILL (DLF182260);
DELTAFILL (DLF182460);
DELTAXSFT (DLX100151);
DELTAXSFT (DLX100152);
DELTAXSFT (DLX100153);
DELTAXSFT (DLX100201);
DELTAXSFT (DLX100202);
DELTAXSFT (DLX100203);
DELTAXSFT (DLX100204);
DELTAXSFT (DLX100206);
{5}
DELTAXSFT (DLX100254);
DELTAXSFT (DLX100256);
DELTAXSFT (DLX100304);
DELTAXSFT (DLX100306);
DELTAXSFT (DLX100406);
DELTAXSFT (DLX100408);
GALAXY G3 (GLY120202);
GALAXY G3 (GLY120306);
GALAXY G3 (GLY120308);
GALAXY G3 (GLY120407);
GALAXY G3 (GLY120410);
GALAXY G3 (GLY120412);
GALAXY G3 (GLY120510);
GALAXY G3 (GLY120515);
GALAXY G3 (GLY120610);
GALAXY G3 (GLY120615);
GALAXY G3 (GLY120620);
GALAXY G3 (GLY120715);
GALAXY G3 (GLY120721);
GALAXY G3 (GLY120815);
GALAXY G3 (GLY120824);
GALAXY G3 (GLY120925);
GALAXY G3 (GLY121030);
GALAXY G3 (GLY121230);
GALAXY G3 (GLY122505);
GALAXY G3 (GLY122535);
GALAXY G3 (GLY123509);
GALAXY G3 (GLY123575);
GALAXY G3 MINI (GLM910010);
GALAXY G3 MINI (GLM910015);
GALAXY G3 MINI (GLM910020);
GALAXY G3 MINI (GLM910025);
GALAXY G3 MINI (GLM910030);
GALAXY G3 MINI (GLM910040);
GALAXY G3 MINI (GLM915020);
GALAXY G3 MINI (GLM915025);
GALAXY G3 MINI (GLM915030);
GALAXY G3 MINI (GLM915040);
GALAXY G3 MINI (GLM920030);
GALAXY G3 MINI (GLM920040);
GALAXY G3 MINI (GLM920060);
GALAXY G3 MINI (GLM925035);
GALAXY G3 MINI (GLM925045);
GALAXY G3 MINI (GLM925055);
GALAXY G3 MINI (GLM930040);
GALAXY G3 MINI (GLM930060);
GALAXY G3 MINI (GLM930080)
Please provide your Indications for Use below.
?
MICRUSFRAME, DELTAFILL, and DELTAXSFT Microcoil Delivery Systems are intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and are also intended for vascular occlusion of blood vessels
{6}
within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
GALAXY G3 Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
GALAXY G3 Mini Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| --- | --- | --- |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{7}
K262419
# 510(k) Summary
# I. Submitter
Cerenovus, Inc.
31 Technology
Drive
Irvine, CA 92618, USA
Tel: +1 714 876 5048
Contact Person: Ryan Rivera
Email: rriver82@ITS.JNJ.com
# II. Devices
| Table 1. Device | |
| --- | --- |
| Device Proprietary Name | MICRUSFRAME Microcoil Delivery System, DELTAFILL Microcoil Delivery System, DELTAXSFT Microcoil Delivery System, GALAXY G3 Microcoil Delivery System, GALAXY G3 XSFT Microcoil Delivery System, and GALAXY G3 Mini Microcoil Delivery System |
| Common or Usual Name | Device, Neurovascular Embolization & Vascular, for Promoting Embolization |
| Classification Name | Class II 21 CFR 882.5950 – Device, Neurovascular Embolization Class II 21 CFR 870.3300 – Device, Vascular, For Promoting Embolization |
| Regulatory Classification | II |
| Product Codes | HCG, KRD |
# III. Predicate Devices
The predicate devices are listed below in Table 2.
| Table 2: Predicate Devices | | | | |
| --- | --- | --- | --- | --- |
| Type | 510(k) Number | Date Cleared | Name | Manufacturer |
| Primary | K211344 | May 28, 2021 | MICRUSFRAME Microcoil Delivery System, DELTAFILL Microcoil Delivery System, DELTAXSFT Microcoil Delivery System, GALAXY G3 Microcoil Delivery System, GALAXY G3 XSFT Microcoil Delivery System, and GALAXY G3 Mini Microcoil Delivery System | Cerenovus, Inc. |
| Reference | K242582 | Sep 27, 2024 | Optima Coil System (OptiBlock Line Extension) | Balt USA |
| *510(k)s were previously held by Medos International SARL. The 510(k)s were transferred to Cerenovus, Inc., which is the current holder of the 510(k)s. | | | | |
Continued on next page
Page 1
{8}
K262419
# 510(k) Summary, Continued
# IV. Device Description
The MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3, and GALAXY G3 Mini Microcoil Delivery Systems consist of three components: a Microcoil System, a connecting cable, and a Detachment Control Box (DCB). Each component is sold separately.
As shown in Figure 1, the Microcoil System consists of a microcoil attached to a Device Positioning Unit (DPU).

The Microcoil System is packaged in an introducer sheath designed to protect the coil in the packaging dispenser and to provide support for introducing the coil into the microcatheter catheter. The microcoil is the implantable segment of the device and is detached from the Device Positioning Unit (DPU) using the Detachment Control System (Detachment Control Box and connecting cable).
For the MICRUSFRAME, DELTAFILL, DELTAXSFT, and GALAXY G3 microcoils:
- The microcoil is fabricated from a platinum alloy wire. The wire is wound into a primary coil which may contain either a polypropylene suture (SR) or an absorbable polymer suture and then formed into a secondary shape. The secondary shape may be spherical, complex, or helical.
For the GALAXY G3 Mini microcoils:
- The microcoil is fabricated from a platinum alloy wire. The wire is wound into a primary coil which contains a polypropylene suture (SR) and then formed into a secondary shape. The secondary shape is complex.
For all Microcoil Delivery Systems:
- The DPU is a variable stiffness wire and has a radiopaque marker band located three (3) cm from its distal end. The Device Positioning Unit includes five (5) fluoro saver markers on the proximal section of the shaft. The markers are intended to indicate when the tip of the microcoil is approaching the tip of the microcatheter. When the distal-most marker reaches the proximal end of the Rotating Hemostatic Valve (RHV) on the microcatheter, the tip of the coil is approaching the tip of the microcatheter and fluoroscopy should be used to guide further coil insertion.
- The introducer sheath has three main components: an introducer tip, a translucent introducer body, and a re-sheathing tool.
Continued on next page
Page 2
{9}
K262419
# 510(k) Summary, Continued
IV. Device Description, continued
The ENPOWER Detachment Control Box (DCB) provides the energy necessary to allow for a thermo-mechanical detachment of the microcoil from the DPU. The connecting cable delivers the energy necessary to detach the embolic coil from the Microcoil System's detachment zone. The connecting cable is connected between the Microcoil System's hub connector on the DPU and the output connector on the DCB.
- The connecting cables may be one of two types: one with a remote detach button (the ENPOWER Control Cable) catalog no. ECB000182-00, or one without a detach button (standard connecting cable) catalog no. CCB00157-00.
- The ENPOWER Detachment Control Box, catalog no. DCB2000500, works with the ENPOWER Control Cable and with the standard connecting cable.
V. Indications for Use
MICRUSFRAME, DELTAFILL, and DELTAXSFT Microcoil Delivery Systems are intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The GALAXY G3 Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
The GALAXY G3 Mini Microcoil Delivery System is intended for endovascular embolization of intracranial aneurysms, other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation and for arterial and venous embolizations in the peripheral vasculature.
VI. Comparison of Technological Characteristic with Predicate Device
There are no differences in technological characteristics between the MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3, and GALAXY G3 Mini Microcoil Delivery Systems and the predicate device (K211344).
Continued on next page
Page 3
{10}
K262419
# 510(k) Summary, Continued
VI.
Comparison of Technological Characteristic with Predicate Device
| Table 3-1. Technological Characteristics of the Predicate and Proposed Device | | | |
| --- | --- | --- | --- |
| Description | Predicate Device: MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3, GALAXY G3 XSFT, and GALAXY G3 Mini Microcoil Delivery System (K211344) | Reference Device: Optima Coil System (OptiBlock Line Extension) (K242582) | This Submission: MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3, and GALAXY G3 Mini Microcoil Delivery System |
| Indications for Use | MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3, and GALAXY G3 Mini Microcoil Delivery Systems | The Optima Coil System | MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3, and GALAXY G3 Mini Microcoil Delivery Systems |
| | are intended for endovascular embolization of intracranial aneurysms, and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and | is intended for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae | are intended for endovascular embolization of intracranial aneurysms, and other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae, and |
| | are also intended for | is also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation | are also intended for vascular occlusion of blood vessels within the neurovascular system to permanently obstruct blood flow to an aneurysm or other vascular malformation |
| | | and for | and for |
| | arterial and venous embolizations in the peripheral vasculature. | arterial and venous embolizations in the peripheral vasculature. | arterial and venous embolizations in the peripheral vasculature. |
Continued on next page
Page 4
{11}
K262419
# 510(k) Summary, Continued
VI.
Comparison of Technological Characteristic with Predicate Device
| Table 11-1. Technological Characteristics of the Predicate, Reference and Proposed Device, continued | | | |
| --- | --- | --- | --- |
| Description | Predicate Device: MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3, GALAXY G3 XSFT, and GALAXY G3 Mini Microcoil Delivery System (K211344) | Reference Device: Optima Coil System (OptiBlock Line Extension) (K242582) | This Submission: MICRUSFRAME, DELTAFILL, DELTAXSFT, GALAXY G3, and GALAXY G3 Mini Microcoil Delivery System |
| Microcoil | | | |
| Microcoil Stretch-Resistant | PGA= Polyglycolic Acid Suture PP= Polypropylene Suture | .0022" Polyolefin Engage thread (Annealed) .0018" Polyolefin Engage thread (Annealed) | Same as Predicate |
| | Galaxy G3 Mini Microcoil PP= Polypropylene Suture | | |
| Primary Coil Wind Outer Diameter (OD) | 0.009" – 0.015" | 0.010" - 0.014" | Same as Predicate |
| | Galaxy G3 Mini Microcoil 0.009" | | |
| Secondary Shape OD Ranges | 1.5mm – 24mm | 1mm-14mm | Same as Predicate |
| | Galaxy G3 Mini Microcoil 1.0mm – 3.0mm | | |
| Microcoil Length Ranges | 1cm - 60cm | 3 cm – 65 cm | Same as Predicate |
| | Galaxy G3 Mini Microcoil 1cm - 8cm | | |
| Delivery System | | | |
| Delivery System Type | Wire Shaft with radiopaque marker | Body coil laser welded to hypotube | Same as Predicate |
| Mechanism of Detachment | Connection to Microcoil System: Uses Connecting Cable or EnPOWER Control Cable | Thermal via detachment controller | Same as Predicate |
| | Detachment: Thermo-Mechanical System uses the ENPOWER Detachment Control Box (DCB) with ENPOWER Control Cable or Connecting Cable | | |
Continued on next page
Page 5
{12}
K262419
# 510(k) Summary, Continued
VII. Non-Clinical Data Performance Data
No additional design verification testing is required because the proposed modification is limited to the addition of wording in the Indications for Use and does not involve any change to the device design, materials, technological characteristics, operating principles, delivery system, detachment mechanism, or performance requirements. The previously submitted verification testing, which was reviewed and cleared via K211344 remains applicable because the evaluated device configurations are representative of the currently marketed Spectra embolic coil systems and the subject change does not alter the way the device is delivered, deployed, or detached.
No additional validation testing was required because the change is limited to the addition of wording in the Indications for Use. The previous validation testing remains applicable and continues to support the clarified wording, as the change does not impact the device design, technological characteristics, principles of operation, clinical use, or user requirements.
| Table 5: Validation Testing | | |
| --- | --- | --- |
| Test | Test Method Summary | Results |
| Usability: Animal Study Report | The purpose of this testing is to assess the effect of resistance heater position in the Device Position Unit of Spectra with respect to the microcatheter tip and coil mass density on the reliability of detachment of the Coil detachment system. An in vivo model with its complex nature of hemodynamics and physiologic tortuosity, allows the assessment of the acute performance of the test article to deliver an embolic coil to the target vessel. | Pass Samples passed the established acceptance criteria |
| DPU3 Design Validation In-Vivo Study Report | The study was performed to assess the efficacy of the test sample's coil insertion, radiopacity, re-sheathing and to deliver an embolic coil to the target vessel in a porcine model. There were no aneurysms created for this study the embolic coils were delivered to parent vessel target site that were suitable in diameter that allowed for multiple coil deployment while maintaining high blood flow rate | Pass Samples passed the established acceptance criteria |
| Spectra in-vivo evaluation Report | The purpose of this report is to document the in-vivo evaluation for friction in Spectra DeltaXSFT coils. | Pass Samples passed the established acceptance criteria |
| Spectra Usability Study In-Vivo | The purpose of this in-vivo usability evaluation was to assess the deliverability, deployment behaviour, microcatheter stability, and friction performance of representative Spectra-family embolic coils when delivered to suitable parent-vessel target sites. The evaluation supports the proposed indication by demonstrating that the coil systems can be delivered and deployed under physiologically relevant vessel conditions without reliance on aneurysm creation. | Pass Samples passed the established acceptance criteria |
Continued on next page
Page 6
{13}
K262419
# 510(k) Summary, Continued
# Animal Testing
N/A – No animal studies were required as appropriate verification and validation of the modifications were achieved based on the similarities of the proposed device to the predicate device, and from results of bench testing.
# Shelf Life Testing
N/A – Changes did not impact the shelf-life of the product.
# Biocompatibility Testing
N/A – Changes did not impact biocompatibility.
# Sterilization
N/A – Changes did not impact sterilization.
# VIII. Clinical Performance Data
Clinical studies were not required because the proposed change consists only of additional wording clarification to the Indications for Use. The previously cleared verification and validation evidence remains applicable, as the modification does not alter the device design, intended use, technological characteristics, or raise new questions of safety or effectiveness.
# IX. Conclusion
The modifications made to the additional wording to the Indications for Use does not alter the design, intended use, and technological characteristics of the proposed and predicate devices. The risk assessment of the proposed subject and the predicate device remains the same and previous verification /validation testing raised no new questions regarding the safety and effectiveness of the devices. Therefore, the modified Indications for Use are substantially equivalent to their predicate devices.
End of 510(k) Summary
Page 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.