← Product Code [JHI](/productcode/JHI) · K262299

# Instant-View® hCG Combo II Cassette Test (K262299)

_Alfa Scientific Designs, Inc. · JHI · Sep 22, 2026 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K262299

## Device Facts

- **Applicant:** Alfa Scientific Designs, Inc.
- **Product Code:** [JHI](/productcode/JHI.md)
- **Decision Date:** Sep 22, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1155
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Instant-View® hCG Combo II Test is a chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine or serum. It is for use as an aid in the early detection of pregnancy. This test is for health care professional use only and not for self-testing.

## Device Story

Instant-View® hCG Combo II Cassette Test is a lateral flow chromatographic immunoassay; detects human chorionic gonadotropin (hCG) in urine or serum samples. Device consists of test strip with mouse monoclonal anti-hCG antibody-colloidal gold conjugate, nitrocellulose membrane with rabbit monoclonal anti-hCG α-subunit capture antibody (T line), and goat anti-mouse IgG (C line). Healthcare professionals apply specimen to sample well; hCG binds to gold conjugate; complex migrates to T line. Colored band develops at T line if hCG ≥ 20 mIU/mL (urine) or ≥ 10 mIU/mL (serum). C line confirms test validity. Used in clinical settings by professionals to aid pregnancy diagnosis. Results interpreted visually by provider to inform clinical decision-making regarding pregnancy status.

## Clinical Evidence

Bench testing only. Precision/reproducibility studies conducted across 3 sites with 6 operators using 3 lots confirmed performance at cut-offs. Analytical specificity evaluated against LH, FSH, TSH, and various exogenous compounds; no interference observed. Hook effect study confirmed no high-dose interference up to 2,000,000 mIU/mL. Method comparison study (n=250; 124 serum, 126 urine) against predicate device showed 100% agreement for both matrices.

## Technological Characteristics

Lateral flow sandwich immunochromatographic assay. Materials: nitrocellulose membrane, mouse monoclonal anti-hCG antibody, rabbit monoclonal anti-human CGA antibody, goat anti-mouse IgG, colloidal gold conjugate. Form factor: single cassette. Storage: 2-30°C. No electronic components, software, or connectivity.

## Regulatory Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

## Predicate Devices

- Alltest Pregnancy Rapid Combo Test Cassette ([K203272](/device/K203272.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

[LOGO]

FDA

U.S. FOOD & DRUG

ADMINISTRATION

### 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY

ASSAY ONLY

### I Background Information:

#### A 510(k) Number

K262299

#### B Applicant

Alfa Scientific Designs, Inc.

#### C Proprietary and Established Names

Instant-View® hCG Combo II Cassette Test

#### D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  JHI | II | 21 CFR § 862.1155 | CH – Clinical Chemistry  |

### II Submission/Device Overview:

#### A Purpose for Submission:

New device

#### B Measurand:

Human chorionic gonadotropin (hCG)

#### C Type of Test:

Qualitative chromatographic immunoassay

### III Intended Use/Indications for Use:

#### A Intended Use(s):

See Indications for Use below.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

# **B Indication(s) for Use:**

The Instant-View® hCG Combo II Test is a chromatographic immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in urine or serum. It is for use as an aid in the early detection of pregnancy. This test is for health care professional use only and not for self-testing.

# **C Special Conditions for Use Statement(s):**

Rx - For Prescription Use Only

# **D Special Instrument Requirements:**

N/A

# **IV Device/System Characteristics:**

# **A Device Description:**

The Instant-View® hCG Combo II Cassette Test is a professional-use only, qualitative assay for the detection of human chorionic gonadotropin (hCG) in urine (20 mIU/mL) or serum (10 mIU/mL) to aid in the early detection of pregnancy. The kit contains 25 test devices, each sealed together with a dropper pipette in a foil pouch, and 1 professional use package insert.

# **B Principle of Operation:**

This assay is a one-step lateral flow chromatographic immunoassay. The test strip in the device consists of a conjugate pad containing a mouse monoclonal anti-hCG antibody conjugated to colloidal gold, and a nitrocellulose membrane precoated with a rabbit monoclonal anti-human hCG α-subunit (CGA) antibody at the test line (T line) and goat anti-mouse IgG at the control line (C line).

When an adequate amount of specimen is applied to the sample well of the device, hCG in the specimen binds to the anti-hCG antibody gold conjugate in the conjugate pad, forming a complex that migrates along the membrane strip. If the hCG level is at or above 20 mIU/mL in urine or 10 mIU/mL in serum, the complex will bind to the capture antibody pre-coated on the T line, developing a colored band. If the specimen contains no hCG or if the level below the detection limit, the T line will not develop.

The goat anti-mouse antibody pre-coated on the C line will always bind to the gold-antibody conjugate, forms a colored line regardless of the presence or absence of hCG.

# **V Substantial Equivalence Information:**

# **A Predicate Device Name(s):**

Alltest Pregnancy Rapid Combo Test Cassette

# **B Predicate 510(k) Number(s):**

K203272

# **C Comparison with Predicate(s):**

K262299 - Page 2 of 8

{2}

|  Device & Predicate Device(s): | K262299 | K203272  |
| --- | --- | --- |
|  Device Trade Name | Instant-View® hCG Combo II Cassette Test | Alltest Pregnancy Rapid Combo Test Cassette  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | Rapid qualitative detection of hCG to aid in the early detection of pregnancy | Same  |
|  Principle of Measurement | Lateral flow Sandwich Immunochromatographic Assay | Same  |
|  Cut-off values | Urine: 20 mIU/mL Serum: 10 mIU/mL | Same  |
|  General Device Characteristic Differences |  |   |
|  Read time | Urine: 5 minutes Serum: 5 minutes | Urine: 3 minutes Serum: 5 minutes  |

## VI Standards/Guidance Documents Referenced:

None referenced.

## VII Performance Characteristics (if/when applicable):

### A Analytical Performance:

#### 1. Precision/Reproducibility:

Negative female serum or urine specimens were spiked with varying hCG (commercially available and traceable to the 6th WHO international Standard) concentrations. The spiked samples were measured in replicates of 6 each day for 5 days using 3 different lots at 3 different testing sites representative of the intended use environment. Tests were performed by 6 different operators for each sample concentration. The cut-off values of urine and serum are 20 mIU/mL hCG and 10 mIU/mL and, respectively. Results are shown in the following tables.

Urine

|  hCG Concentration (mIU/mL) | Site 1 Lot 3 |   | Site 2 Lot 1 |   | Site 3 Lot 2 |   | Total result |   | % Negative | % Positive  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  - | + | - | + | - | + | - | +  |   |   |
|  0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100 | 0  |
|  5 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100 | 0  |

K262299 - Page 3 of 8

{3}

|  hCG Concentration (mIU/mL) | Site 1 Lot 3 |   | Site 2 Lot 1 |   | Site 3 Lot 2 |   | Total result |   | % Negative | % Positive  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  - | + | - | + | - | + | - | +  |   |   |
|  10 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100 | 0  |
|  15 | 26 | 4 | 27 | 3 | 27 | 3 | 80 | 10 | 88.9 | 11.1  |
|  17.5 | 2 | 28 | 2 | 28 | 3 | 27 | 7 | 83 | 7.8 | 92.2  |
|  20 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100  |
|  25 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100  |
|  30 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100  |
|  50 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100  |

Serum

|  hCG Concentration (mIU/mL) | Site 1 Lot 3 |   | Site 2 Lot 1 |   | Site 3 Lot 2 |   | Total result |   | % Negative | % Positive  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  - | + | - | + | - | + | - | +  |   |   |
|  0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100 | 0  |
|  2 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100 | 0  |
|  5 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100 | 0  |
|  7 | 4 | 26 | 3 | 27 | 4 | 26 | 11 | 79 | 12.2 | 87.8  |
|  8.5 | 3 | 27 | 2 | 28 | 2 | 28 | 7 | 83 | 7.8 | 92.2  |
|  10 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100  |
|  12 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100  |
|  15 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100  |
|  25 | 0 | 30 | 0 | 30 | 0 | 30 | 0 | 90 | 0 | 100  |

# 2. Linearity:

Not applicable. This is a qualitative test.

# 3. Analytical Specificity/Interference:

Cross-reactivity: The sponsor performed a study to evaluate potential cross-reactivity from LH (Luteinizing hormone), FSH (Follicle-stimulating hormone), and TSH (Thyroid-stimulating hormone). Negative and positive samples (10 and 20 mIU/mL hCG for urine, 5 and 10 mIU/mL hCG for serum) were spiked with various concentrations of LH, FSH, and TSH. Samples were tested using 3 different lots by 3 operators. No interference was observed for these samples for the device at LH concentrations up to 500 mIU/mL, FSH concentrations up to 1000 mIU/mL, and TSH concentrations up to 1000 μIU/mL.

Interference: The sponsor performed a study to evaluate potential interference from exogenous compounds. Each interferent was spiked into both hCG negative (10 mIU/mL for urine, 5 mIU/mL for serum) and hCG positive (20 mIU/mL for urine, 10 mIU/mL for serum) samples. Samples were tested using 3 device lots.

K262299 - Page 4 of 8

{4}

The results showed that no interference was observed from substances at the following concentrations for both negative and positive hCG urine and serum samples. Results are shown in the following table.

|  Interferent | Concentration  |
| --- | --- |
|  Acetaminophen | 20 mg/dL  |
|  Acetoacetic Acid | 2,000 mg/dL  |
|  Acetylsalicylic Acid | 20 mg/dL  |
|  Albumin | 2,000 mg/dL  |
|  Ampicillin | 20 mg/dL  |
|  Ascorbic Acid | 20 mg/dL  |
|  Atropine | 20 mg/dL  |
|  B-hydroxybutyrate | 2,000 mg/dL  |
|  Benzoylecgonine | 10 mg/dL  |
|  Bilirubin (urine) | 2 mg/dL  |
|  Bilirubin (serum) | 40 mg/dL  |
|  Caffeine | 20 mg/dL  |
|  Cannabinol | 10 mg/dL  |
|  Codeine | 10 mg/dL  |
|  EDTA | 80 mg/dL  |
|  Ephedrine | 20 mg/dL  |
|  Estriol-17-beta | 1.4 mg/dL  |
|  Ethanol | 1%  |
|  Gentisic Acid | 20 mg/dL  |
|  Glucose | 2,000 mg/dL  |
|  Hemoglobin | 2,000 mg/dL  |
|  High-density lipoprotein (serum only) | 70 mg/dL  |
|  Ketone | 20 mg/dL  |
|  Methadone | 10 mg/dL  |
|  Methanol | 10%  |
|  Phenothiazine | 20 mg/dL  |
|  Phenylpropanolamine | 20 mg/dL  |
|  Pregnanediol | 1.5 mg/dL  |
|  Salicylic Acid | 20 mg/dL  |
|  Tetracycline | 20 mg/dL  |
|  Thiophene | 20 mg/dL  |
|  Total cholesterol (serum only) | 250 mg/dL  |
|  Triglycerides (serum only) | 1,200 mg/dL  |

# Effect of hCG beta-core fragment

To evaluate potential interference by hCG β-core fragment, negative and positive samples (10 and 20 mIU/mL hCG in urine, 5 and 10 mIU/mL hCG in serum;) were spiked with various concentrations of hCG β -core fragment (0 to 2x10⁶ pmol/L). These samples were

K262299 - Page 5 of 8

{5}

tested by 3 different operators using 3 different device lots. No interference from β-core fragment was observed at concentrations up to 2.0 x 10⁶ pmol/L for both urine and serum samples.

# Effect of urine pH

To evaluate potential interference from changes in urine pH, negative (i.e., 10 mIU/mL hCG) and positive (i.e., 20 mIU/mL hCG) urine samples were tested at pH 4 to 9 by 3 different operators using 3 different device lots. The data shows that there is no interference from pH ranging from 4 to 9 on tested urine samples.

# Effect of urine specific gravity

To evaluate potential interference from changes in specific gravity, negative (i.e., 10 mIU/mL hCG) and positive (i.e., 20 mIU/mL hCG) urine samples were tested at density values ranging from 1.000 to 1.035 by 3 different operators using 3 different device lots. The results indicate that the varying of urine density from 1.000 to 1.035 does not impact the performance of the test device.

# Hook effect study

To evaluate high dose hook effect, negative urine and serum samples were spiked with hCG concentrations ranging from 500 to 2,000,000 mIU/mL and were tested in replicates of 5 by 3 different operators using 3 different device lots. The results showed no hook effect up to 2,000,000 mIU/mL for both urine and serum samples tested.

# 4. Detection Limit and Assay Reportable Range

This assay is intended to detect hCG at levels equal to or above 20 mIU/mL in urine or 10 mIU/mL in serum. See section VII. A.1. for precision of the device near the cutoff and section VII. A.6 for verification of the device's performance at the cut-off.

# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

# Traceability

The Instant-View® hCG Combo II Cassette Test is traceable to the WHO 6th International Standard (IS) for human Chorionic Gonadotropin (hCG) 18/244.

# Stability

To investigate stability of the device, accelerated stability testing, shelf-life determination, transport simulation testing, and sample stability testing were performed, each by 3 different operators. Based on the results from these studies, the sponsor claims a 24-month shelf life for the device and sample stability for at least 3 days when stored at 2-30°C.

# 6. Assay Cut-Off:

To verify the cut-off value of the Instant-View® hCG Combo II Cassette Test, negative urine and serum specimens were spiked with varying hCG (commercially available and traceable to the 6th WHO International Standard) concentrations. The spiked samples were randomly labeled and tested using 3 different lots of the device by 3 different

K262299 - Page 6 of 8

{6}

operators. Twenty replicates were tested for each sample concentration and each device lot. Results are shown in the following tables.

Urine

|  hCG Concentration (mIU/mL) | Lot 1 |   | Lot 2 |   | Lot 3 |   | Total result |   | % Negative | % Positive  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  - | + | - | + | - | + | - | +  |   |   |
|  0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0  |
|  5 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0  |
|  10 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0  |
|  15 | 19 | 1 | 18 | 2 | 17 | 3 | 54 | 6 | 90 | 10  |
|  17.5 | 1 | 19 | 1 | 19 | 2 | 18 | 4 | 56 | 6.7 | 93.3  |
|  20 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100  |
|  25 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100  |
|  30 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100  |
|  50 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100  |

Serum

|  hCG Concentration (mIU/mL) | Lot 1 |   | Lot 2 |   | Lot 3 |   | Total result |   | % Negative | % Positive  |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
|   |  - | + | - | + | - | + | - | +  |   |   |
|  0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0  |
|  2 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0  |
|  5 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100 | 0  |
|  7 | 3 | 17 | 2 | 18 | 2 | 18 | 7 | 53 | 12 | 88  |
|  8.5 | 2 | 18 | 1 | 19 | 1 | 19 | 4 | 56 | 6.7 | 93.3  |
|  10 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100  |
|  12 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100  |
|  15 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100  |
|  25 | 0 | 20 | 0 | 20 | 0 | 20 | 0 | 60 | 0 | 100  |

### B Comparison Studies:

#### 1. Method Comparison with Predicate Device:

A total of 124 serum samples and 126 urine were collected from 250 women with ages ranging from 18 to 49 years, who presented to 3 clinical sites for pregnancy testing. All subjects were suspected to be pregnant and all subjects confirmed to be pregnant were in the early stages of pregnancy (less than 5 weeks). The samples were masked and randomized prior to testing by 3 different health professionals using the candidate device

K262299 - Page 7 of 8

{7}

and the predicate device (Alltest Pregnancy Rapid Combo Test Cassette). Three different lots of the proposed device were used in the study for both urine and serum samples. A summary of the results is presented in the tables below:

Urine

|  New Device | Cleared device | + | -  |
| --- | --- | --- | --- |
|   |  + | 49 | 0  |
|   |  - | 0 | 77  |

Serum

|  New Device | Cleared device | + | -  |
| --- | --- | --- | --- |
|   |  + | 54 | 0  |
|   |  - | 0 | 70  |

The data show that the agreement of the Instant-View® hCG Combo II Cassette Test with the predicate device was 100% for both urine and serum samples.

#### 2. Matrix Comparison:

Not applicable; see sections VII. A.1. and VII. B.1. for precision and method comparison studies conducted with urine and serum samples. Performance with both matrices is described in the performance sections above.

### C Clinical Studies:

1. Clinical Sensitivity: Not applicable.
2. Clinical Specificity: Not applicable.
3. Clinical Cut-Off: Not applicable.
4. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable.

### D Expected Values/Reference Range:

Not applicable.

### VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

### IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K262299 - Page 8 of 8

---

**Source:** [https://fda-staging.innolitics.com/device/K262299](https://fda-staging.innolitics.com/device/K262299)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).