Use for virtually all bonding needs in restorative dentistry for dental patients of all ages. MPa Max adhesive bonds to the following materials:Dentin and enamelPorcelain (hydrofluoric acid and silane required)ZirconiaMetalComposite
Device Story
MPa Max adhesive is a bottle-delivered, 7.5% filled, ethanol-based dental bonding resin containing 0.2% chlorhexidine. Used by dental clinicians in an etch-and-rinse technique; dentin and enamel are etched with phosphoric acid, rinsed, and dried prior to adhesive application. The ethanol solvent facilitates resin penetration into the prepared tooth structure. Upon light-curing with high-intensity LEDs, the adhesive forms a polymerized layer providing chemical bonding and micromechanical interlocking between the tooth structure and restorative materials (porcelain, zirconia, metal, composite). The cured adhesive film thickness is approximately 2 µm. The device is intended for use in clinical dental settings to improve long-term bond strengths and facilitate restorative procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing. Shear bond strength testing performed per ISO 29022:2013 on dentin, enamel, porcelain, metal, composite, and zirconia (n=7 per substrate); all samples met the 25 MPa acceptance criterion. Packaging light-protection testing confirmed adhesive stability after light exposure.
Technological Characteristics
7.5% filled resin with ethyl alcohol solvent; contains 0.2% chlorhexidine. Light-cure mechanism. Bottle delivery. Cured film thickness ~2 µm. Non-sterile. Bench testing per ISO 29022:2013.
Indications for Use
Indicated for restorative dentistry bonding needs in patients of all ages. Bonds to dentin, enamel, porcelain, zirconia, metal, and composite.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 24, 2026
Den-Mat Holdings, LLC
Brianna Alvarez
Regulatory Affairs Specialist
1017 W. Central Ave.
Lompoc, California 93436
Re: K262252
Trade/Device Name: MPa Max adhesive
Regulation Number: 21 CFR 872.3200
Regulation Name: Resin Tooth Bonding Agent
Regulatory Class: Class II
Product Code: KLE
Dated: July 1, 2026
Received: July 2, 2026
Dear Brianna Alvarez:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262252 - Brianna Alvarez
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262252 - Brianna Alvarez
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**MICHAEL E. ADJODHA -S**
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262252 | ? |
| Please provide the device trade name(s). | | ? |
| MPa Max adhesive | | |
| Please provide your Indications for Use below. | | ? |
| Use for virtually all bonding needs in restorative dentistry for dental patients of all ages. MPa Max adhesive bonds to the following materials:Dentin and enamelPorcelain (hydrofluoric acid and silane required)ZirconiaMetalComposite | | |
| Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | Neonates/Newborns (Birth to < 29 days old)Infants (29 days old to < 2 years old)Children (2 years old to < 12 years old)Adolescents (12 years old to < 22 years old)Adults (22 years old and greater) | ? |
{4}
Clinician's Choice®
Clinician's Choice is a division of Den-Mat Holdings, LLC
Traditional 510K Submission
MPa Max Adhesive
### K262252 - 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR 807.92.
| Submitter Information | 21 CFR 807.92(a)(1) |
| --- | --- |
| Submitter: | Den-Mat Holdings, LLC 1017 W. Central Avenue Lompoc, CA 93436 USA Phone: 805-346-3700 Establishment Registration Number: 2018957 |
| --- | --- |
| Contact Person: | Brianna Alvarez Regulatory Affairs Specialist Phone: 805-346-3700 Email: regulatory@denmat.com |
| --- | --- |
| Date Prepared: | September 8, 2026 |
| --- | --- |
| Device Information | 21 CFR 807.92(a)(2) |
| --- | --- |
| Trade Name: | MPa™ Max adhesive |
| --- | --- |
| Common Name: | Dental tooth bonding resin |
| Regulation Number: | 21 CFR 872.3200 |
| Regulation Name: | Agent, Tooth Bonding, Resin |
| Classification: | Class II |
| Product Code: | KLE |
| Review Panel: | Dental |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) |
| --- | --- |
| Device Trade Name: | Peak™ Universal Bond |
| --- | --- |
| 510(k) Number: | K100752 (Primary Predicate) |
| Product Code: | KLE |
| Device Trade Name: | Peak™ Universal Bond |
| --- | --- |
| 510(k) Number: | K240743 (Secondary Predicate) |
| Product Code: | KLE |
Den-Mat Holdings, LLC dba Clinician's Choice
1017 W Central Ave., Lompoc, CA 93436, USA
Page 1 of 4
P: 805-346-3700
www.clinicianschoice.com
{5}
Clinician's Choice®
Clinician's Choice is a division of Den-Mat Holdings, LLC
Traditional 510K Submission
MPa Max Adhesive
# Device Description
21 CFR 807.92(a)(4)
MPa Max adhesive is a bottle delivered bonding resin. It can be used with total-etch techniques. It is 7.5% filled with an ethyl alcohol solvent carrier and will cure with most high intensity lights including LEDs. MPa Max adhesive contains 0.2% chlorhexidine which may ensure long term bond strengths. The resin is intended for bonding restorative materials to dentin, enamel, porcelain, zirconia, metal, and composite.
MPa Max is used with an etch-and-rinse bonding technique. Dentin and enamel are etched with phosphoric acid, rinsed with water, and prepared in accordance with the instructions for use before application of the adhesive. Phosphoric-acid etching removes or modifies the smear layer and demineralizes the superficial tooth structure to facilitate penetration of the adhesive into the prepared substrate.
The ethanol-based formulation facilitates penetration of the adhesive resin into the prepared dentin surface. Upon light curing, the adhesive forms a polymerized bonding layer between the prepared tooth structure and restorative material through chemical bonding and micromechanical interlocking.
MPa Max has a viscosity designed to facilitate application and coating of the prepared bonding surfaces. The cured adhesive has a film thickness of approximately 2 µm.
# Intended Use / Indications for Use
21 CFR 807.92(a)(5)
Use for virtually all bonding needs in restorative dentistry for dental patients of all ages.
MPa Max adhesive bonds to the following materials:
- Dentin and enamel
- Porcelain (hydrofluoric acid and silane required)
- Zirconia
- Metal
- Composite
# Comparison of Technological Characteristics
21 CFR 807.92(a)(6)
MPa Max adhesive is substantially equivalent to the predicate devices Peak Universal Bond in terms of design, chemical composition, mode of action, principle of operation, manufacturing, packaging, and labeling. While the indications for use for the subject and primary predicate devices are substantially equivalent, the subject device can also bond to Zirconia. The secondary predicate device is included specifically to support this additional material compatibility. Any differences between the subject and predicate devices are considered minor and do not raise new questions concerning safety and effectiveness.
Den-Mat Holdings, LLC dba Clinician's Choice
1017 W Central Ave., Lompoc, CA 93436, USA
Page 2 of 4
P: 805-346-3700
www.clinicianschoice.com
{6}
Clinician's Choice®
Clinician's Choice is a division of Den-Mat Holdings, LLC
Traditional 510K Submission
MPa Max Adhesive
| | Subject Device | Primary Predicate Device | Secondary Predicate Device | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | MPa Max Adhesive | Peak Universal Bond | Peak Universal Bond | N/A |
| Manufacturer | Den-Mat Holdings, LLC | Ultradent Products, Inc. | Ultradent Products, Inc. | N/A |
| 510(k) number | K262252 | K100752 | K240743 | N/A |
| Classification Name | Agent, Tooth Bonding, Resin | Agent, Tooth Bonding, Resin | Agent, Tooth Bonding, Resin | Identical |
| Regulation Number | 21 CFR 872.3200 | 21 CFR 872.3200 | 21 CFR 872.3200 | Identical |
| Product Code | KLE | KLE | KLE | Identical |
| Indications for Use | Use for virtually all bonding needs in restorative dentistry for dental patients of all ages. MPa Max adhesive bonds to the following materials: • Dentin and enamel • Porcelain (hydrofluoric acid and silane required) • Zirconia • Metal • Composite | Use for virtually all etching and bonding needs in restorative dentistry for dental patients of all ages. Peak Universal Bond bonds to the following materials: • Dentin and enamel • Porcelain • Metal • Composite | Use for virtually all etching and bonding needs in restorative dentistry for patients of all ages. Peak Universal Bond is used for desensitizing/sealing of tooth structure. Peak Universal Bond adhesive bonds to the following materials: • Dentin and Enamel • Porcelain, Zirconia • Metal • Composite | While the indications for use for the subject and primary predicate devices are substantially equivalent, the subject device can also bond to zirconia. The secondary predicate device is included specifically to support this additional material compatibility. |
| Delivery System | Bottle | Bottle | Bottle | Identical |
| Sterile | No | No | No | Identical |
| Curing Mechanism | Light-cure | Light-cure | Light-cure | Identical |
Den-Mat Holdings, LLC dba Clinician's Choice
1017 W Central Ave., Lompoc, CA 93436, USA
P: 805-346-3700
www.clinicianschoice.com
Page 3 of 4
{7}
Clinician's Choice®
Clinician's Choice is a division of Den-Mat Holdings, LLC
Traditional 510K Submission
MPa Max Adhesive
Performance Data
21 CFR 807.92(b)
### 21 CFR 807.92(b)(1) – Non-Clinical Testing
Non-clinical performance testing was conducted on representative samples of MPa Max adhesive to evaluate shear bond strength and the light-protection performance of the primary packaging.
### Shear Bond Strength
Shear bond strength testing was conducted in accordance with ISO 29022:2013, Dentistry — Adhesion — Notched-edge shear bond strength test. Seven samples per substrate were evaluated for dentin, enamel, porcelain, metal, composite, and zirconia. All samples met the predefined acceptance criterion of a minimum shear bond strength of 25 MPa.
### Light-Protection Performance of Primary Packaging
An in-house test method was used to evaluate the ability of the primary packaging bottle to protect MPa Max adhesive from light exposure. Two bottles containing MPa Max adhesive were tested. Each bottle was first exposed to a high-intensity light source for 5 minutes, after which adhesive was expressed from the bottle and evaluated for its ability to cure. The same bottles were then exposed to a standard light source for an additional 10 minutes, after which the adhesive was again expressed and evaluated for its ability to cure. Both bottles met the predefined acceptance criterion following each light-exposure condition, demonstrating that the primary packaging provided adequate light protection under the specified test conditions.
The results of the non-clinical performance testing demonstrate that MPa Max adhesive met the predefined acceptance criteria and support the substantial equivalence determination.
### 21 CFR 807.92(b)(2) – Clinical Testing
Clinical testing was not required to demonstrate substantial equivalence to the predicate devices. The substantial equivalence of the MPa Max adhesive is supported by non-clinical performance testing and technological characteristics.
Conclusion
Based on the indications for use, technological characteristics, and the summary of data submitted, Den-Mat Holdings, LLC has determined that the subject device does not raise new questions of safety and effectiveness compared to the predicate devices. Therefore, the proposed subject device is substantially equivalent to the legally marketed predicate devices.
Den-Mat Holdings, LLC dba Clinician's Choice
1017 W Central Ave., Lompoc, CA 93436, USA
P: 805-346-3700
www.clinicianschoice.com
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.