Missing tooth structure in anterior and posterior teeth. Areas of application: 1. Metal free indirect restorations: full crown, inlay, onlay, laminated veneer 2. Implant supported crown
Device Story
Hybrid ceramic CAD/CAM block; used for fabrication of fixed all-ceramic dental restorations; milled using dental CAD/CAM system; finished and polished by dental professional; bonded to tooth structure or attached to implant abutment; restores anatomical form and function of missing or defective tooth structure; used in clinical dental settings.
Clinical Evidence
No clinical data. Bench testing only. Evaluated per ISO 10993-1:2018 and ISO 7405:2018 for cytotoxicity, skin sensitization, and irritation. Physical/mechanical testing performed per ISO 10477, ISO 4049, ISO 9917-1, ADA No. 27, and JIS T 6517, including flexural strength, hardness, water sorption, solubility, color stability, compressive strength, elastic modulus, and radio-opacity.
Technological Characteristics
Hybrid ceramic CAD/CAM block. Composition: Barium glass, methacrylate polymer, silicon dioxide, pigments. Form factor: Pre-cured block for milling. Sterilization: Non-sterile. Storage: 4–25°C. Shelf-life: 5 years. Mechanical/physical properties verified via ISO/ADA/JIS standards.
Indications for Use
Indicated for patients with missing tooth structure in anterior and posterior teeth requiring metal-free indirect restorations (full crown, inlay, onlay, laminated veneer) or implant-supported crowns.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 10, 2026
GC America, Inc.
Futoshi Fusejima
Director of PE and Regulatory Affairs
3737 W. 127th St.
Alsip, Illinois 60803
Re: K262202
Trade/Device Name: Cerasmart Forte
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth shade resin material
Regulatory Class: Class II
Product Code: EBF
Dated: June 29, 2026
Received: June 29, 2026
Dear Futoshi Fusejima:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262202 - Futoshi Fusejima
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262202 - Futoshi Fusejima
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K262202
Device Name
Cerasmart Forte
Indications for Use (Describe)
Missing tooth structure in anterior and posterior teeth
Areas of application:
1. Metal free indirect restorations: full crown, inlay, onlay, laminated veneer
2. Implant supported crown
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
K262202
## 510(k) Summary
CERASMART FORTE
Submitter: GC America Inc.
3737 W. 127th Street
Alsip, IL 60803
Contact Person: Futoshi Fusejima
Title: Director of PE and Regulatory Affairs
Phone: (708)926-3050
Alternative Contact: Tamiko Scott
Title: Regulatory Affairs Specialist
Date Prepared: June 26, 202
| Trade/Device Name | CERASMART FORTE |
| --- | --- |
| Common or Usual Name | Tooth Shade Resin Material |
| Regulatory Class | Class II |
| Product Code | EBF |
| Regulation Number | 21 CFR 872.3690 |
Predicate Devices:
| Product | Applicant | 510(k) No. | Code No. | Predicate/reference device | Decision DateUni |
| --- | --- | --- | --- | --- | --- |
| CERASMART | GC America Inc. | K133824 | EBF | Primary Predicate | 03/14/2014 |
| GRADIA CORE | GC America Inc. | K082171 | EBF | Reference | 10/30/2008 |
| Nmf004a (G-ænial Universal Injectable) | GC America Inc. | K173500 | EBF | Reference | 03/8/2018 |
{5}
**Device Description:**
CERASMART FORTE is a hybrid ceramic for fixed all-ceramic dental restorations. The material is indicated for use of :
Missing tooth structure in anterior and posterior teeth.
Areas of application:
1. Metal free indirect restorations: full crown, inlay, onlay, laminated veneer.
2. Implant supported crown.
For professional use only.
**Intended Use/Indications for Use:**
Missing tooth structure in anterior and posterior teeth.
Areas of application:
1. Metal free indirect restorations: full crown, inlay, onlay, laminated veneer
2. Implant supported crown
**Summary of Technological Characteristics:**
CERASMART FORTE is a hybrid ceramic CAD/CAM block intended for the fabrication of fixed all-ceramic dental restorations, including metal-free indirect restorations such as full crowns, inlays, onlays, and laminate veneers, as well as implant-supported crowns. The block is mechanically milled using a dental CAD/CAM system to fabricate the intended restoration. After milling, the restoration is finished and polished, and then bonded to the tooth structure or attached to an implant abutment using conventional dental procedures. The device functions as a passive restorative material by replacing or restoring missing or defective tooth structure and restoring the anatomical form and function of the tooth.
Applicant device and predicate devices are substantially equivalent in the following points.
- Pre-cured composite block
- Indications for Use; that is the material to use for indirect prosthetic restorations (crowns, inlays, and veneers) including crowns on implants
- Provided in pre-cured block
- Method of use
- Performance evaluation
Therefore, the new and predicate devices are the same in function, and similar in composition and intended use. This supports that the compatibility and safety of the applicant device are substantially equivalent to the predicate devices.
The following difference may be noted between applicant device and predicate device.
- The substances used in the applicant device differ from those used in the predicate device.
Although the chemical composition is not identical, both the applicant and predicate device contain methacrylate based monomers, initiators, inorganic fillers, and pigments,
{6}
which are commonly used in dental composite resin materials including the reference devices; therefore, the two devices are considered comparable in material characteristics. The applicant device does not introduce any new indications, as all intended uses of the applicant device are included within the predicate device CERASMART.
Furthermore, the biological safety of this product was evaluated in accordance with ISO 10993-1:2018, Use of International Standard ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and ISO 7405:2018 and following the flowchart in "ISO 10993-1:2018 Figure 1 - Summary of the systematic approach to a biological evaluation of medical devices as part of a risk management process".
Q-sub Q260748 was conducted June 15th, 2026, and this biological evaluation was conducted based on the FDA suggestions at Q260748.
In conclusion, the applicant device is substantially equivalent to the predicate device in technological principle.
### Performance Data:
The biological safety of this product was evaluated in accordance with ISO 10993-1:2018, Use of International Standard ISO10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and ISO 7405:2018 and following the flowchart in "ISO 10993-1:2018 Figure 1 - Summary of the systematic approach to a biological evaluation of medical devices as part of a risk management process".
In accordance with ISO 10993-1:2018 "Annex A Endpoints to be addressed in a biological risk assessment", the following are contact and contact duration of this product with the body.
| Category | : Externally communicating medical device |
| --- | --- |
| Contact | : Tissue/bone/dentin |
| Contact duration | : Long term (> 30 d) |
- ISO 10993-5 In Vitro Cytotoxicity
- ISO 10993-10 Skin Sensitization
- ISO 10993-23 Skin Irritation
Additional bench testing based on the test with reference to ISO 10477:2020, ISO 4049:2019, ISO9917-1:2007, ADA specification No. 27-1993, JIS T 6517:2011, and the FDA guidance documents entitled "Dental Composite Resin Devices - Premarket Notification [510(k)] Submissions - Guidance for Industry and FDA Staff".
- Appearance
- Shade
- Surface finish
{7}
- Hardness
- Flexural strength
- Water sorption
- Solubility
- Color stability
- Compressive strength
- Elastic modulus
- Radio-opacity
{8}
CERASMART FORTE
Substantial Equivalence Chart
| | Applicant device | Predicate device | Reference device 1 | Reference device 2 | Rational |
| --- | --- | --- | --- | --- | --- |
| Trade name | CERASMART FORTE | CERASMART K133824 | GRADIA CORE K082171 | Nmf004a (G-ænial Universal Injectable) K173500 | |
| Manufacturer | GC Corporation | GC Corporation | GC Corporation | GC Corporation | |
| Product category | EBF | EBF | EBF | EBF | Applicant device is identical to predicate device and reference devices. |
| Indications for Use | Missing tooth structure in anterior and posterior teeth. Areas of application: 1. Metal free indirect restorations: full crown, inlay, onlay, laminated veneer 2. Implant supported crown | The product is indicated for inlays, onlays, veneers and full crown restorations, including crowns on implants. | GRADIA CORE is intended to be used for restoration, core build-up and post cementation. | 1. Liner or base 2. Blocking out undercuts 3. Repair of (in) direct aesthetic restorations, temporary crown & bridge, defect margins when margins are in enamel 4. Sealing hypersensitive areas 5. Fissure sealant 6. Direct restorative for Class I, II, III, IV, V cavities 7. Core build-up | Applicant device is equivalent to predicate device. |
| Device description | CERASMART FORTE is a hybrid ceramic for fixed all-ceramic dental restorations. | CERASMART is a pre-cured composite block for milling CAD/CAM indirect restorations. The milled device is used for the restorations of both anterior and posterior teeth. | GRADIA CORE cartridge is a two-paste (base and catalyst) type, dual-cured composite resin filled in double syringe to provide auto-mixing system. GRADIA CORE cartridge paste | NMF004A is a light cured nano-filled radiopaque composite resin filled in a syringe. The device is used for the restorations of both anterior and posterior teeth. The material is available in 30 shades. | Applicant device is substantially equivalent to predicate device |
{9}
| | | | comprises four shades, Universal, Dentin, White and Blue. | | |
| --- | --- | --- | --- | --- | --- |
| Material Type | Pre-cured composite block | Pre-cured composite block | Dual-cured composite resin | Light-cured composite resin | Applicant device is equivalent to predicate device. |
| Delivery form | Pre-cured block | Pre-cured block | Paste-paste formulation in a double-chamber syringe | Paste in a syringe | Applicant device is equivalent to predicate device. |
| Chemical composition | Barium glass, methacrylate polymer, silicon dioxide, pigment | Barium glass, methacrylate polymer, silicon dioxide, pigment | Paste A: Fluoro-alumino-silicate glass, Di methacrylate, silicon dioxide, initiator, stabilizer, pigment, Paste B: Fluoro-alumino-silicate glass, Di methacrylate, silicon dioxide, initiator, stabilizer, pigment, | Barium glass, di methacrylate, initiator, pigment, silicon dioxide, stabilizer | Applicant device and predicate device are considered equivalent in that they are composed of inorganic fillers, methacrylate polymers, stabilizers, and pigments. |
| Sterilization | Not a sterile medical device. | Not a sterile medical device. | Not a sterile medical device. | Not a sterile medical device. | The applicant device is the same as predicate device and reference devices. |
| Storage temperature | 4 – 25°C | 4 – 25°C | 2 – 8°C | 4 – 25°C | The applicant device is the same as predicate device. |
| Shelf-Life | 5 years | 5 years | 2 years | 3 years | The applicant device is the same as predicate device. |
| Available shades | LT: B1, A1, A2, A3, A3.5, A4 HT: B1, A1, A2, A3, A3.5 | A1 HT, A2 HT, A3 HT, A3.5 HT, A4 HT, B1 HT, B2 HT, B3 HT, B4 HT, C1 HT, C2 HT, C3 HT, | Universal, Dentin, White, Blue | A1, A2, A3, A3.5, A4, A5, B1, B2, B3, B4, C1, C2, C3, C4, | Applicant device shades are included in the predicate device shades. |
{10}
| | | C4 HT, D2 HT, D3 HT, D4 HT, A1 LT, A2 LT, A3 LT, A3.5 LT, A4 LT, B1 LT, B2 LT, B3 LT, B4 LT, C1 LT, C2 LT, C3 LT, C4 LT, D2 LT, D3 LT, D4 LT, BL | | D2, D3, D4, E1, E2, E3, U AO1, AO2, AO3, AO3.5, AO4, AO5, WO, BW, XBW | |
| --- | --- | --- | --- | --- | --- |
| Direction for use | PREPARATION DESIGN MILLING FINISHING AND POLISHING CEMENTATION ADDITIONAL APPLICATION AND REPAIR CHARACTERIZATION | I. MILLING PROCEDURE II. ADDITIONAL APPLICATION AND REPAIR | 1. Loading cartridge into dispenser 2. Cavity preparation 3. Application of G-Premio BOND and G-Premio BOND DCA or GRADIA CORE SELF-ETCHING BOND A/B 4. Post cementation and core build-up 5. Contouring 6. Additional build-up | 1. Preparations 2. Shade Selection 3. Cavity Preparation 4. Bonding Treatment 5. Placement of G-aenial Universal Injectable 6. Light curing 7. Finishing and polishing | Differences are limited to wording and level of detail; method of use is the same as predicate device. |
| Biocompatibility | Conformed to ISO-10993-1 and ISO 7405 | Conformed to ISO 10993-1 | Conformed to ISO-10993-1 | Conformed to ISO-10993-1 | |
| Performance Testing | Reference to: FDA guidance document "Dental Composite Resin Devices, ISO 10477, ISO 4049, ISO 9917-1 and JIS T 6517 • Appearance • Shade • Surface finish • Hardness • Flexural strength • Water sorption • Solubility • Color stability • Compressive strength • Elastic modulus | ISO 10477 • Surface finish • Flexural strength • Water sorption • Solubility • Shade consistency • Color stability | ISO 4049 Type 2, Class 3 • Working time(sec.)(at 23°C) • Setting time(min.)(at 37°C) • Flexural strength • Water sorption • Solubility • Radio-opacity | ISO 4049 Type 1, Class 2, Group 1 • Stability to light • Depth of cure • Flexural strength • Water sorption • Solubility • Shade of restorative materials • Color stability after irradiation and water sorption • Radio-opacity | |
{11}
| | Radio-opacity | | | | |
| --- | --- | --- | --- | --- | --- |
| Appearance | Conformed | Conformed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
| Shade | Conformed | Conformed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
| Surface finish | Conformed | Conformed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
| Hardness | Passed | Passed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
| Flexural strength | Passed | Passed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
| Water sorption | Passed | Passed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
| Solubility | Passed | Passed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
| Colour stability | Conformed | Conformed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
| Compressive strength | Passed | Passed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
{12}
| Elastic modulus | Passed | Passed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
| --- | --- | --- | --- | --- | --- |
| Radio-opacity | Passed | Passed | N/A | N/A | Applicant device is substantially equivalent to predicate device. |
{13}
In all instances, the CERASMART FORTE functioned as intended, and the outcomes were as expected.
# **Conclusions:**
Based on similarities in indications for use, technology, safety and effectiveness, the applicant device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.