ImageGrid StreamVue is a device that receives medical images and data from various imaging sources. Images and data can be stored, communicated, and displayed within the system or across computer networks at distributed locations. Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians as an adjunctive to standard radiology practices for diagnosis. Typical users of this system are trained professional physicians, radiologists, nurses, medical technicians, and assistants.
Device Story
Software-only medical image viewing application; part of ImageGrid product family. Receives medical images from upstream PACS; performs server-side image rendering; streams rendered viewport images to client application for display/interaction. Eliminates local storage/persistence of DICOM objects on client. Used in clinical environments by radiologists, clinicians, and medical staff. Does not perform image acquisition, reconstruction, or analysis; no AI/ML algorithms. Output is visual representation of medical images for diagnostic review. Benefits include efficient image delivery and viewing without requiring full local dataset storage.
Clinical Evidence
Bench testing only. Comparative diagnostic image quality evaluation performed using representative DICOM studies across CR, CT, MR, US, MG, and fluoroscopy. Results demonstrated comparable diagnostic image quality across all modalities. System-level software verification and validation, including regression testing, confirmed successful retrieval, server-side rendering, streaming, and display functionality. All test cases passed.
Technological Characteristics
Software-only medical image management and processing system. Server-side rendering architecture with client-side streaming. Supports DICOM standard (NEMA PS 3.1–3.20 2024e). Developed per IEC 62304 (software lifecycle) and ISO 14971 (risk management). No hardware components. Connectivity via computer networks.
Indications for Use
Indicated for trained healthcare professionals (physicians, radiologists, nurses, technicians, assistants) to receive, store, communicate, and display medical images for diagnostic viewing. Applicable to various imaging modalities. Contraindicated for primary diagnosis of mammography using lossy compressed or digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 23, 2026
Candelis, Inc.
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Dr. Suite #510k
Saint Paul, Minnesota 55114
Re: K262143
Trade/Device Name: ImageGrid StreamVue
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical image management and processing system
Regulatory Class: Class II
Product Code: LLZ
Dated: June 25, 2026
Received: June 25, 2026
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K262143 - Prithul Bom
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262143 - Prithul Bom
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jessica Lamb, Ph.D.
Assistant Director, Imaging Software Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262143 | ? |
| Please provide the device trade name(s). | | ? |
| ImageGrid StreamVue | | |
| Please provide your Indications for Use below. | | ? |
| ImageGrid StreamVue is a device that receives medical images and data from various imaging sources. Images and data can be stored, communicated, and displayed within the system or across computer networks at distributed locations. Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians as an adjunctive to standard radiology practices for diagnosis. Typical users of this system are trained professional physicians, radiologists, nurses, medical technicians, and assistants. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) #: K262143
# 510(k) Summary
Prepared on: 2026-06-16
Contact Details
21 CFR 807.92(a)(1)
Applicant Name
Candelis, Inc.
Applicant Address
9903 South Santa Monica Blvd #2101 Beverly Hills CA 90212 United States
Applicant Contact Telephone
714-788-8152
Applicant Contact
Mr. Josh Baker
Applicant Contact Email
josh@jbshelp.com
Correspondent Name
Josh Baker Solutions
Correspondent Address
32565 B Golden Lantern #1048 Dana Point CA 92629 United States
Correspondent Contact Telephone
714-788-8152
Correspondent Contact
Mr. Josh Baker
Correspondent Contact Email
josh@jbshelp.com
Device Name
21 CFR 807.92(a)(2)
Device Trade Name
ImageGrid StreamVue
Common Name
Medical image management and processing system
Classification Name
System, Image Processing, Radiological
Regulation Number
892.2050
Product Code(s)
LLZ
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
Predicate #
Predicate Trade Name (Primary Predicate is listed first)
Product Code
K092949
ImageGrid Radiology Viewer System
LLZ
Device Description Summary
21 CFR 807.92(a)(4)
ImageGrid StreamVue is a software-only medical device that functions as a diagnostic image viewing application within the ImageGrid product family. The device provides authorized users with the ability to view medical images that are stored and managed by upstream image management systems.
ImageGrid StreamVue is a modified version of the ImageGrid Radiology Viewer System previously cleared under FDA 510(k) number K092949, with no changes to intended use or indications for use.
The subject device uses image streaming technology to deliver medical images to the end user for viewing. Image rendering and interaction occur without requiring full local storage of the image dataset on the client system. This change replaces the prior method of locally loading complete image data while maintaining the same user-facing diagnostic viewing functionality.
The device is intended to be used as part of a medical image management workflow and does not acquire medical images directly. ImageGrid StreamVue does not perform image acquisition, image reconstruction, or image analysis. No computer-aided detection,
{5}
computer-aided diagnosis, or artificial intelligence algorithms are introduced or modified.
ImageGrid StreamVue is implemented as a software application that interfaces with existing ImageGrid infrastructure to request, stream, and display medical images. The device supports interactive image viewing functions consistent with the previously cleared device. The introduction of streaming technology represents a change in data delivery and rendering architecture only. There are no changes to the device's intended use, indications for use, clinical functionality, or diagnostic purpose.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
ImageGrid StreamVue is a device that receives medical images and data from various imaging sources. Images and data can be stored, communicated, and displayed within the system or across computer networks at distributed locations. Only pre-processed DICOM for presentation images can be interpreted for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor that offers at least 5 Mpixel resolution and meets other technical specifications reviewed and accepted by FDA. Diagnosis is not performed by the software but by Radiologists, Clinicians and referring Physicians as an adjunctive to standard radiology practices for diagnosis. Typical users of this system are trained professional physicians, radiologists, nurses, medical technicians, and assistants.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
There are no changes to the device's indications for use.
## Technological Comparison
21 CFR 807.92(a)(6)
ImageGrid StreamVue and the predicate device, ImageGrid Radiology Viewer System (K092949), are both software-only medical image viewing systems classified under 21 CFR 892.2050, Product Code LLZ. Both devices are intended to receive, retrieve, and display medical images from an upstream PACS for diagnostic viewing by trained healthcare professionals. Both devices support standard diagnostic image viewing workflows and use the same fundamental scientific technology of medical image management and display.
The primary technological difference between ImageGrid StreamVue and the predicate device is the image delivery architecture. The predicate device downloads image data to the client workstation and performs local image decoding. ImageGrid StreamVue performs image rendering on the server side and streams rendered viewport images to the client application for display and interaction. This architecture eliminates storage or persistence of DICOM image objects on the client system.
This difference is limited to the method of image delivery. It does not change the device's intended use, indications for use, user population, diagnostic functionality, supported image viewing purpose, or fundamental scientific technology. The subject device remains a software-based medical image viewing system used by trained healthcare professionals as an adjunct to standard radiology practices.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Non-clinical performance testing was conducted to support a determination of substantial equivalence between ImageGrid StreamVue and the predicate device, ImageGrid Radiology Viewer System (K092949).
Testing included comparative diagnostic image quality evaluation, system-level software verification and validation, regression testing, and validation activities. Comparative image quality testing was performed using representative DICOM studies across CR, CT, MR, US, MG, and fluoroscopy. The reviewer evaluated diagnostic quality, artifacts, and functional limitations. The results showed diagnostic image quality for all reviewed modalities, with comparable image quality for mammography, ultrasound, fluoroscopy, CT, and MR, and superior/faster performance for X-ray/CR.
System-level software testing evaluated retrieval of medical images from the upstream PACS, server-side image rendering and streaming to the client, display and interaction with streamed medical images, standard diagnostic viewing functions, and system behavior under expected operating conditions. All planned system-level test cases passed, and any observations or anomalies were evaluated and determined not to impact safety or effectiveness.
Software verification and validation activities included requirements verification testing, system-level testing, image quality testing, and regression testing. These activities were traceable to software requirements and identified risks through the Traceability Matrix.
Testing and risk management activities reference the FDA-recognized standards declared in this submission, including NEMA PS 3.1–3.20 2024e (DICOM), IEC 62304, IEC 60812, and ISO 14971.
{6}
Not Applicable
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.