Clinical expert evaluation: datasets included patients aged 12 years and older with permanent dentition
>1 (US-licensed dentists)
Indications for Use
DS Core CBCT Anatomy is a cloud-based AI/ML enabled software as a medical device. The device is available as a back end service via an API (Application Programming Interface). The device is intended to be used by dental professionals when reviewing Digital Impression (DI) scans and CBCT (DX) scans during standard of care diagnostic review and treatment planning. DS Core CBCT Anatomy analyses CBCT scans and Digital Impression (DI) scans to identify anatomical structures, propose a panoramic curve, segmentation of teeth, jaws, and the inferior alveolar nerve canal (IAN), as well as to perform image registration to support the review of CBCT dental images and DI scans.
Device Story
Cloud-based AI/ML SaMD; backend service via API; assists dental professionals in clinical review/treatment planning. Inputs: Digital Impression (DI/IOS) scans and CBCT (DX) scans. Processing: CNN-based algorithms identify anatomical structures, propose panoramic curves, segment teeth/jaws/IAN, perform image registration, compare DI scans, and identify specific tooth features/brackets/attachments. Output: Supplementary visualizations and structural information. Usage: Clinical setting; dental professional oversight. Benefit: Supports comprehensive interpretation of dental data; enhances workflow efficiency; provides automated anatomical identification/comparison without replacing clinician judgment.
Clinical Evidence
Standalone bench testing using generalizable datasets of patients aged 12+ with permanent dentition. DI Segmentation primary endpoints: tooth detection sensitivity 99.37% (95% CI: 98.97%, 99.70%), tooth numbering accuracy 99.37% (95% CI: 98.89%, 99.74%), mean Dice score 0.9806 (95% CI: 0.9791, 0.9820). DI Comparison primary endpoints: mandibular pass rate 85.86% (95% CI: 81.82%, 89.70%), maxillary pass rate 88.89% (95% CI: 86.06%, 91.32%). Subgroup analyses across sex, age, region, and scanner type confirmed performance consistency. All pre-specified success criteria met.
Technological Characteristics
Cloud-based software; API-delivered backend service. Utilizes locked convolutional neural network (CNN) models for image analysis, segmentation, and registration. Complies with 21 CFR 892.2050. No hardware components; software-only device.
Indications for Use
Indicated for dental professionals to assist in reviewing Digital Impression (DI) and CBCT (DX) scans for diagnostic review and treatment planning in patients aged 12 and older with permanent dentition.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 21, 2026
Dentsply Sirona
% Deepthi Paknikar
Senior Manager Regulatory Affairs
221 W. Philadelphia St.
YORK, PA 17401
Re: K262127
Trade/Device Name: DS Core CBCT Anatomy
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: QIH
Dated: June 24, 2026
Received: June 24, 2026
Dear Deepthi Paknikar:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K262127 - Deepthi Paknikar
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance).
{2}
K262127 - Deepthi Paknikar
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262127 | ? |
| Please provide the device trade name(s). | | ? |
| DS Core CBCT Anatomy | | |
| Please provide your Indications for Use below. | | ? |
| "DS Core CBCT Anatomy" is a cloud-based AI/ML enabled software as a medical device. The device is available as a back end service via an API (Application Programming Interface). The device is intended to be used by dental professionals when reviewing Digital Impression (DI) scans and CBCT (DX) scans during standard of care diagnostic review and treatment planning. DS Core CBCT Anatomy analyses CBCT scans and Digital Impression (DI) scans to identify anatomical structures, propose a panoramic curve, segmentation of teeth, jaws, and the inferior alveolar nerve canal (IAN), as well as to perform image registration to support the review of CBCT dental images and DI scans.This device is intended to be used with patients aged 12 and older with permanent dentition. | | |
| Please select the types of uses (select one or both, as applicable). | Prescription Use (21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | Neonates/Newborns (Birth to < 29 days old)Infants (29 days old to < 2 years old)Children (2 years old to < 12 years old)Adolescents (12 years old to < 22 years old)Adults (22 years old and greater) | ? |
{4}
510(k) #: K262127
# 510(k) Summary
Prepared on: 2026-08-21
Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Dentsply Sirona |
| --- | --- |
| Applicant Address | 221 West Philadelphia St. York PA 17401 United States |
| Applicant Contact Telephone | 6302011612 |
| Applicant Contact | Dr. Deepthi Paknikar |
| Applicant Contact Email | Deepthi.Paknikar@dentsplysirona.com |
| Correspondent Name | Dentsply Sirona |
| Correspondent Address | 221 West Philadelphia St. York PA 17401 United States |
| Correspondent Contact Telephone | 6302011612 |
| Correspondent Contact | Dr. Deepthi Paknikar |
| Correspondent Contact Email | Deepthi.Paknikar@dentsplysirona.com |
Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | DS Core CBCT Anatomy |
| --- | --- |
| Common Name | Medical image management and processing system |
| Classification Name | Medical image management and processing system |
| Regulation Number | 892.2050 |
| Product Code(s) | QIH |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K260785 | DS Core CBCT Anatomy | QIH |
Device Description Summary
21 CFR 807.92(a)(4)
"DS Core CBCT Anatomy" is a cloud-based, AI/ML-enabled software as a medical device (SaMD) that operates as a backend service accessible through an Application Programming Interface (API). The software is designed to support dental professionals during routine diagnostic review and treatment planning by analyzing Digital Impression (DI) scans (also referred to as "IOS" intra oral scans) and cone-beam computed tomography (CBCT) (DX) scans. Using automated algorithms, DS Core CBCT Anatomy identifies key anatomical structures, proposes a panoramic curve, segments teeth, jaws, and the inferior alveolar nerve canal (IAN), and performs image registration to assist in comprehensive interpretation of CBCT and DI data. The device provides automatic comparison of two different DI Scans. The device is intended to integrate seamlessly into clinical workflows, providing supplementary visualizations and structural information while preserving clinician oversight and decision-making. This device is intended to be used with patients aged 12 years and older who have permanent dentition.
{5}
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
"DS Core CBCT Anatomy" is a cloud-based AI/ML enabled software as a medical device. The device is available as a back end service via an API (Application Programming Interface). The device is intended to be used by dental professionals when reviewing Digital Impression (DI) scans and CBCT (DX) scans during standard of care diagnostic review and treatment planning. DS Core CBCT Anatomy analyses CBCT scans and Digital Impression (DI) scans to identify anatomical structures, propose a panoramic curve, segmentation of teeth, jaws, and the inferior alveolar nerve canal (IAN), as well as to perform image registration to support the review of CBCT dental images and DI scans.
This device is intended to be used with patients aged 12 and older with permanent dentition.
# Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use of the Subject Device is identical to that of the predicate (DS Core CBCT Anatomy, K260785). Both devices are intended to assist dental professionals in reviewing CBCT (DX) and Digital Impression (DI/IOS) scans for clinical assessment, communication, and treatment planning during standard-of-care diagnostic review. Like the predicate, the Subject Device performs automated analysis of CBCT and DI data to identify anatomical structures, propose a panoramic curve, segment the teeth, jaws, and inferior alveolar nerve canal (IAN), and perform image registration, for patients aged 12 and older with permanent dentition. The Subject Device adds two additional algorithms:
1. DI Scan Comparison - enables clinicians to view and compare two intraoral scans from a patient and observe any potential differences.
2. DI Segmentation – identifies and marks specific features in a DI scan:
• A specific tooth (with its associated tooth number)
• Brackets
• Attachments
This addition does not alter the fundamental intended use, which remains an assistive software function that provides supplementary anatomical information under clinician oversight while retaining full clinical autonomy. Consistent with the predicate, the Subject Device does not provide diagnostic conclusions and does not direct treatment. Therefore, the additional capabilities do not constitute a new intended use.
# Technological Comparison
21 CFR 807.92(a)(6)
The Subject Device and the predicate share the same fundamental technological principle: automated, machine-learning-based image analysis using computer-vision neural network (CNN) algorithms to identify and display dental anatomy for clinical use. Both are cloud-based AI/ML-enabled software as a medical device, delivered as a back-end service via an API, and both fall under the same regulation (892.2050), product code (QIH), and device class (II). The devices are technologically equivalent in that they process CBCT and DI scans and apply CNN-based algorithms for detection, segmentation, and registration. The only difference is that the Subject Device includes two additional algorithms, one for DI Scan Comparison and one for DI Segmentation. These algorithms rely on the same fundamental technological principles as the predicate and do not introduce any new principles of operation. Like the predicate, the Subject Device is software-based, uses machine learning algorithms, and produces results intended to support, rather than replace, clinician review. Accordingly, the devices are considered technologically comparable.
# Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The device uses several locked, machine-learning-based convolutional neural network (CNN) models that were independently evaluated through standalone testing. These models support anatomical segmentation (for DI and CBCT "DX" scans), keypoint detection (tooth and root tip detection), and automatic scan orientation (ASO) with their performance assessed against expert ground truth created by U.S. licensed dentists. The DI/DX Matching algorithm incorporates ML components, such as the Keypoint & Automatic Scan Orientation (ASO) model, and was evaluated through clinical expert evaluation of the output registration. The DI Comparison algorithm also incorporates an ML component, DI Segmentation, and was evaluated through a clinical expert evaluation of the output registration (scan comparison). Other than the DI Segmentation and DI Comparison algorithms, all others were cleared in the previous version of the device and were not retested for this subject device submission.
The device uses locked, machine learning-based convolutional neural network (CNN) models to assist dental professionals in reviewing intraoral (Digital Impression, or DI) scan data. These models do not adapt or change once deployed. The performance of the previously cleared algorithms under K260785 remains unchanged. The two new algorithms for the subject device are addressed here: DI Segmentation, which is an ML-based CNN algorithm evaluated against a consensus ground truth from U.S.-licensed dentists in a standalone study, and DI Comparison, which compares two different DI scans. DI Comparison is not itself ML-enabled, but its pipeline incorporates the DI Segmentation model, and it was evaluated through expert clinical evaluation by U.S.-licensed, experienced dentists who reviewed the final comparison output for clinical acceptability.
Standalone testing used generalizable datasets with expert ground truth. The datasets included patients aged 12 years and older with permanent dentition and reflected diverse demographics and imaging characteristics. All algorithms were tested against predefined
{6}
acceptance criteria, and all evaluations met their prespecified success criteria.
# Primary endpoint results
DI Segmentation met all three co-primary endpoints. Tooth detection sensitivity was 99.37%, with a 95% confidence interval of (98.97%, 99.70%). Tooth numbering classification accuracy was likewise 99.37%, with a 95% confidence interval of (98.89%, 99.74%). The tooth segmentation Dice score had a mean of 0.9806 (about 98%), with a 95% confidence interval of (0.9791, 0.9820). All lower bounds met or exceeded their thresholds.
DI Comparison met its primary acceptance criteria for both jaws. The mandibular (lower) full-arch pass rate was 85.86%, with a 95% confidence interval of (81.82%, 89.70%), and the maxillary (upper) pass rate was 88.89%, with a 95% confidence interval of (86.06%, 91.32%); the acceptance criterion was met or exceeded in each case.
# Subgroup performance
Subgroup analyses supported generalizability across patient and imaging factors.
For DI Segmentation, tooth detection and numbering remained high across subgroups. By sex, sensitivity was 100% (female), about 99% (male), and 99% (unknown), with tooth numbering accuracy near 99–100% across the same groups. Across patient age bands (from 11–20 through 71+), sensitivity stayed roughly 99–100% and numbering accuracy roughly 98–100%. By U.S. region, sensitivity ranged from about 99% to 100%, same for tooth numbering accuracy. By arch type, lower-arch and upper-arch sensitivities were about 99% each, with numbering accuracy of roughly 99% (lower) and 100% (upper). Across scanner types, sensitivity was about 99% for Primescan, 100% for Primescan 2, and 100% for the unknown group. Tooth-level Dice scores were consistently around 0.98, including roughly 0.98 across age bands, 0.98–0.99 by sex, about 0.98 across U.S. regions, 0.98 for both arches, and 0.97–0.99 across scanner types.
For DI Comparison, the sextant pass rates were 95.35% (lower anterior), 92.32% (upper anterior), 96.58% (lower posterior right), 97.65% (upper posterior right), and 97.90% for both lower and upper posterior left. At the overall jaw level, subgroup pass rates by sex were about 84% (lower) and 89% (upper) for female, 88% and 89% for male, and 89% and 89% for the unknown group. Across age bands the lower-arch rates ranged from roughly 82% to 88% and the upper-arch rates ranged roughly 87–93%. Across U.S. regions, lower-arch rates ranged from about 81% (South) to 88% (West) and upper-arch rates from about 85% (Northeast) to 93% (Midwest and West).
All performance evaluations met their pre specified success criteria.
Results collectively demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.