← Product Code [HRS](/productcode/HRS) · K262098

# EnHAnce PEEK Bunion System (K262098)

_Nvision Biomedical Technologies · HRS · Sep 18, 2026 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K262098

## Device Facts

- **Applicant:** Nvision Biomedical Technologies
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Sep 18, 2026
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The enHAnce PEEK Bunion System is intended for fixation of osteotomies and corrective procedures of the hallux and associated disorders, such as hallux valgus.

## Device Story

EnHAnce PEEK Bunion System is a single-use, permanently implanted bone fixation device for hallux valgus correction; consists of a plate and pins manufactured from HA Enhanced PEEK. Device provides intramedullary fixation for the first metatarsal. Used in surgical settings by orthopedic surgeons; implanted during bunion osteotomy procedures. Subject device adds a minimally invasive surgery (MIS) configuration to the existing system. Output is physical stabilization of bone segments; aids in bone healing and deformity correction. Benefits include mechanical fixation of osteotomies. Device provided non-sterile for facility steam sterilization or terminally gamma-sterilized.

## Clinical Evidence

Bench testing only. Engineering analysis compared mechanical performance characteristics of the subject MIS PEEK Bunion implant against the predicate. Results demonstrated that the subject device does not create a new global worst-case and maintains equivalent mechanical performance and load resistance in vivo.

## Technological Characteristics

Material: HA Enhanced PEEK (ASTM F2026). Principle: Intramedullary bone fixation via plate and pins. Form factor: Cannulated plate and pins. Sterilization: Steam sterilization (healthcare facility) or terminal gamma radiation. Connectivity: None. Software: None.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- enHAnce PEEK Bunion System ([K253718](/device/K253718.md))
- K-Wire ([K241014](/device/K241014.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** U.S. FOOD & DRUG
ADMINISTRATION

September 18, 2026

Nvision Biomedical Technologies
Marisa Zink
In-House Counsel
12449 Silicon Dr.
San Antonio, Texas 78249

Re: K262098

Trade/Device Name: EnHAnce PEEK Bunion System

Regulation Number: 21 CFR 888.3030

Regulation Name: Single/multiple component metallic bone fixation appliances and accessories

Regulatory Class: Class II

Product Code: HRS

Dated: June 22, 2026

Received: June 22, 2026

Dear Marisa Zink:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K262098 - Marisa Zink

Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K262098 - Marisa Zink

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

**Thomas Mcnamara -S**

For: Christopher Ferreira, M.S.

Assistant Director

DHT6C: Division of Restorative,

Repair, and Trauma Devices

OHT6: Office of Orthopedic Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K262098 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  EnHAnce PEEK Bunion System  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The enHAnce PEEK Bunion System is intended for fixation of osteotomies and corrective procedures of the hallux and associated disorders, such as hallux valgus.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

510(k) #: K262098

# 510(k) Summary

Prepared on: 2026-09-13

Contact Details

21 CFR 807.92(a)(1)

|  Applicant Name | Nvision Biomedical Technologies  |
| --- | --- |
|  Applicant Address | 12449 Silicon Drive San Antonio TX 78249 United States  |
|  Applicant Contact Telephone | 2108706261  |
|  Applicant Contact | Mrs. Marisa Zink  |
|  Applicant Contact Email | marisazink@nvisionbiomed.com  |

Device Name

21 CFR 807.92(a)(2)

|  Device Trade Name | EnHAnce PEEK Bunion System  |
| --- | --- |
|  Common Name | Single/multiple component metallic bone fixation appliances and accessories  |
|  Classification Name | Plate, Fixation, Bone  |
|  Regulation Number | 888.3030  |
|  Product Code(s) | HRS  |

Legally Marketed Predicate Devices

21 CFR 807.92(a)(3)

|  Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code  |
| --- | --- | --- |
|  K253718 | enHAnce PEEK Bunion System | HRS  |
|  K241014 | K-Wire | HTY  |

Device Description Summary

21 CFR 807.92(a)(4)

The enHAnce PEEK Bunion System is a single-use bone fixation device intended to be permanently implanted. The system consists of a plate and pins that provide fixation for the first metatarsal in the correction of a bunion. The plate and pins are both manufactured from HA Enhanced PEEK (ASTM F2026). Both the enHAnce PEEK Bunion Plate cleared in K253718 and the subject MIS PEEK Bunion Plate are designed to allow for intramedullary insertion of the stem portion. While the originally cleared plate has pin receiving holes, the subject device does not. The originally cleared plate and the subject MIS plate are both cannulated and provided in one size and configuration that can be used across all patient anatomy, including left and right feet.

Intended Use/Indications for Use

21 CFR 807.92(a)(5)

The enHAnce PEEK Bunion System is intended for fixation of osteotomies and corrective procedures of the hallux and associated disorders, such as hallux valgus.

Indications for Use Comparison

21 CFR 807.92(a)(5)

Nvision believes that the subject enHAnce MIS PEEK Bunion System is substantially equivalent to Nvision's own predicate device (K253718), and the fundamental scientific technology remains unchanged. The system modifications are not due to any complaint, nor are they based on a corrective fix to the device. Rather, the primary reason for these updates is to make the system more accessible for surgeons who are performing the procedure differently. The modification is:

{5}

1. Addition of MIS Bunion Implant: add a bunion implant configuration for minimally invasive bunion surgery to the already existing system consisting of a bunion implant not intended to be used for minimally invasive surgery in the previously cleared enHAnce PEEK Bunion System.

The enHAnce device is provided non-sterile in a sterilization case for subsequent steam sterilization at a healthcare facility prior to use (as cleared in K253718). Alternately, the enHAnce device can be terminally gamma-sterilized in a sealed sterile barrier package. The enHAnce devices may only undergo one method of sterilization and are not intended to be reprocessed.

## Technological Comparison

21 CFR 807.92(a)(6)

The subject enHAnce implants maintain similar features as the previously cleared enHAnce device, including the same material of construction (PEEK HA), plate feature, stemmed portion of implant, etc. In addition, the MIS PEEK Bunion implant does not impose a new worst-case concerning size compared to the previously cleared implant, thus not affecting its mechanical strength or performance. Furthermore, the subject implant is used with some of the same instruments and new instruments very similar to the original enHAnce instruments. The changes in the instrumentation were made to accommodate the new MIS PEEK Bunion Implant, as it follows its own technique based on the surgeon's preference.

## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)

Engineering Analysis

Engineering analysis of the device characteristics demonstrated equivalence to the predicate devices with respect to mechanical performance. Key parameters for the worst-case subject enHAnce MIS PEEK Bunion, using the same criteria as the initial submission, were compared to those of the enHAnce PEEK Bunion to verify that the subject device system does not create a new global worst case and thus remained substantially equivalent to the predicates.

A Design Verification Summary is presented in Table 18.1. Verification results demonstrated that the performance characteristics of the worst-case subject enHAnce MIS PEEK Bunion are at least equivalent to those of the previously cited predicate. Both devices have equivalent dimensions and features, however variations in several parameters result in the predicates having reduced mechanical properties relative to the enHAnce PEEK Bunion System. Overall, the subject and predicate devices would have at least equivalent mechanical performance and would equivalently resist loads in vivo.

Furthermore, the subject and predicate devices are manufactured equivalently using conventional machining, cleaning, and surface treatments. They are also all equivalently sterilized, which does not raise questions of equivalence because the performance of Enhanced PEEK HA is not negatively impacted by steam sterilization or gamma radiation.

Conclusion: Engineering Analysis demonstrates that the subject enHAnce MIS PEEK Bunion does not create a new worst-case relative to the predicate systems and thus confirms substantial equivalence with respect to mechanical performance.

---

**Source:** [https://fda-staging.innolitics.com/device/K262098](https://fda-staging.innolitics.com/device/K262098)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).