K262073 · Dürr Dental SE · PEM · Sep 17, 2026 · General, Plastic Surgery
Device Facts
Record ID
K262073
Device Name
Hygienic Protective Bite Cover (2207-010-50)
Applicant
Dürr Dental SE
Product Code
PEM · General, Plastic Surgery
Decision Date
Sep 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4370
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Hygienic protective covers serve as hygienic barrier between dental instruments, equipment, accessories, and patient during use. The device is non-sterile and intended for single patient use only.
Device Story
Disposable polyethylene sleeve; functions as passive hygienic barrier; placed over dental bite blocks during extraoral radiological examinations (e.g., OPG, CBCT). Prevents direct contact between bite block and patient oral cavity; reduces cross-contamination risk. Used in dental clinics; operated by dental staff. Single-use; non-sterile. Benefits patient by providing physical barrier against potential contaminants on reusable dental equipment.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via biocompatibility (ISO 10993-1, -5, -10) and physical property testing (ASTM F1670, F1671, D882, F1342, D1004).
Indicated for use as a hygienic barrier between dental instruments/equipment and patients (children, adolescents, and adults) during dental diagnostic procedures to reduce cross-contamination risk.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 17, 2026
Dürr Dental SE
Corinna Gordillo
Regulatory Affairs Manager
Höpfigheimer St. 17
Bietigheim-Bissingen, BW 74321
Germany
Re: K262073
Trade/Device Name: Hygienic Protective Bite Cover (2207-010-50)
Regulation Number: 21 CFR 878.4370
Regulation Name: Surgical Drape And Drape Accessories
Regulatory Class: Class II
Product Code: PEM,
Dated: June 19, 2026
Received: June 22, 2026
Dear Corinna Gordillo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K262073 - Corinna Gordillo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K262073 - Corinna Gordillo
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

For Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K262073 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| Hygienic Protective Bite Cover (2207-010-50) | | | |
| Please provide your Indications for Use below. | | | ? |
| Hygienic protective covers serve as hygienic barrier between dental instruments, equipment, accessories, and patient during use. The device is non-sterile and intended for single patient use only. | | | |
| Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
DÜRR DENTAL
# 510(k) SUMMARY
(In accordance with 21 CFR 807.92)
K262073
# Hygienic Protective Bite Cover
DÜRR DENTAL SE
Höpfigheimer Str. 17
74321 Bietigheim-Bissingen
Germany
1. Date Summary Prepared
May 29, 2026
2. Submitter's Information
• Company: DÜRR DENTAL SE
• Role: Manufacturer
• Owner/Operator No.: 10047925
• Establishment Reg. No.: 3015509619
• Address: Höpfigheimer Str. 17
74321 Bietigheim-Bissingen
Germany
• Website: www.duerrdental.com
• Phone: +49 7142 70 5-0
• Responsible Person: Oliver Lange
Director Quality Management
oliver.lange@duerrdental.com
+49 7142 705-190
• Submission Correspondent: Corinna Gordillo
Regulatory Affairs Manager
corinna.gordillo@duerrdental.com
3. Type of Submission
• 510(k): Traditional 510(k)
• Purpose of this submission: Initial Submission
1
{5}
DÜRR DENTAL
#### 4. Subject Device
• Trade/Device Name: Hygienic Protective Bite Cover
• Manufacturer: DÜRR DENTAL SE
• Classification Name: Surgical Drape and Drape Accessories
• Regulation Number: 21 CFR 878.4370
• Regulatory Class: Class II
• Product Code: PEM
#### 5. Predicate Device
• Trade/Device Name: Dental Barrier and Sleeves
• Manufacturer: Yueying Medical Instruments (Langfang) Co., Ltd
• 510(k) Number: K250606
• Classification Name: Surgical Drape and Drape Accessories
• Regulation Number: 21 CFR 878.4370
• Regulatory Class: Class II
• Product Code: PEM
#### 6. Indications for Use
Hygienic protective covers serve as hygienic barrier between dental instruments, equipment, accessories, and patient during use. The device is non-sterile and intended for single patient use only.
#### 7. Device Description
The hygienic protective bite cover is a disposable polyethylene cover designed to function as a hygienic barrier between patient and dental devices or device components. It is intended to prevent direct contact and reduce the risk of contamination during dental diagnostic procedures. The cover is supplied non-sterile and is intended for single use only.
| Product Family | Trade Name | Reference Number (REF) | Application |
| --- | --- | --- | --- |
| VistaVox / VistaPano | Hygienic Protective Bite Cover | 2207-010-50 | Cover for bite block |
Prior to dental extraoral radiological examinations (like OPG or CBCT), the hygienic protective bite cover is placed over the bite block and remains in position for the duration of the examination. It serves as a disposable hygienic barrier, preventing direct contact between the bite block and the patient's oral cavity.
The use of the hygienic protective bite cover is recommended but not mandatory to reduce the risk of cross-contamination, as the bite block is positioned intraorally. The bite block may be used with or without hygienic protective bite cover, as it is designed to be reusable and can be reprocessed in accordance with validated cleaning, sterilization and disinfection procedures (as described in the instruction for use).
2
{6}
DÜRR DENTAL
### 8. Substantial Equivalence Comparison
Shown below are the technological characteristics of the subject device compared with the predicate device.
| Device | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Manufacturer | DÜRR DENTAL SE | Yueying Medical Instruments (Langfang) Co., Ltd | / |
| 510(k) Number | - | K250606 | / |
| Product Name | Hygienic Protective Bite Cover | Dental Barrier and Sleeves | / |
| Classification | Class II Device (21 CFR 878.4370) | Class II Device (21 CFR 878.4370) | Same |
| Product Code | PEM | PEM | Same |
| Device Reference No. | 2207-010-50 | Refer to K250606 | / |
| Indications for Use | Hygienic protective covers serve as hygienic barrier between dental instruments, equipment, accessories, and patient during use. The device is non-sterile and intended for single patient use only. | Dental Barrier and Sleeves are intended to be used as a disposable barrier for dental instruments and equipment. This device is non-sterile and intended for single patient use only | The intended use of both product types is equivalent. |
| Device Description | The hygienic protective bite cover is a polyethylene protective sleeve designed to be placed over the bite block during dental radiological examinations. The cover is supplied non-sterile, prepackaged and is intended for single use only. | Dental Barrier and Sleeves consist of various sizes and shapes of polyethylene covers which are positioned on various small hand-held dental instruments such as hand pieces, curing lights, air/water syringes and similar hand instruments. In other forms, they are used to cover various devices such as dental chairs, dental instrument trays, x-ray heads, and others. The products are sold non-sterile, prepackaged, and are disposable, single use only. | The intended use of both product types is equivalent. The predicate device also includes covers for intraoral use, such as bite block covers, as does the subject device. |
| Sterility | Non-sterile | Non-sterile | Same |
| Labeling | Single-use device | Single-use device | Same |
3
{7}
DÜRR DENTAL
| Device | Subject Device | Predicate Device | Comparison |
| --- | --- | --- | --- |
| Technology/Function | Passive dental barrier/sleeve used to cover equipment. | Passive dental barrier/sleeve used to cover equipment. | Same fundamental barrier technology. |
| Material | Polyethylene film | Polyethylene film | Equivalent material class. |
| Material Composition | LDPE XSF-68.3 | LLDPE (80%), LDPE (20%) | Equivalent material class. |
| Performance Properties | - Synthetic Blood Penetration test (ASTM F1670) - Viral Penetration test (ASTM F1671) | - Synthetic Blood Penetration test (ASTM F1670) - Viral Penetration test (ASTM F1671) | Same |
| Biocompatibility | - Non-cytotoxic - Non-sensitizing - Non-irritating | - Non-cytotoxic - Non-sensitizing - Non-irritating | Equivalent. All devices meet the ISO 10993 requirements for contact duration and exposure. |
| Dimensions | One size only. | Multiple barrier sleeve sizes available. Difference in dimensions is due to the size of instruments and equipment they cover. | Differences are dimensional only. Not safety-critical. |
| FDA recognized Standards | ASTM F1670 ASTM F1671 ASTM D882 ASTM F1342 ASTM D1004 ISO 10993-1 ISO 10993-5 ISO 10993-10 | ASTM F1670 ASTM F1671 ASTM D882 ASTM F1342 ASTM D1004 ISO 10993-5 ISO 10993-10 ISO 10993-23 | Equivalent |
### Substantial Equivalence Discussion
The subject device and the predicate device share the same intended use. Both are single-use devices and are equivalent with respect to materials and performance characteristics. Compared to the subject device, the predicate device also includes intraoral accessories such as bite block covers. Subject and predicate device share the same fundamental barrier technology, and the material class is equivalent. Minor specification differences or differences in dimensions between the subject and predicate devices do not introduce new questions of safety or effectiveness and are deemed not safety-critical.
4
{8}
DÜRR
DENTAL
## 9. Non-Clinical Testing
### 9.1 Biocompatibility Testing
Non-clinical performance testing demonstrated that the subject device met all predefined acceptance criteria for the applicable standard tests performed. The testing strategy, rationales and test reports are documented in the attached “*Biological Risk Assessment*” document. Refer to:
| Topic | Standard ID | Submission Attachment |
| --- | --- | --- |
| **Biological Risk Assessment** | ISO 10993-1 | 08_Biological Risk Assessment.pdf |
| **Cytotoxicity** | ISO 10993-5 | 08_Biological Risk Assessment.pdf |
| **Sensitization** | ISO 10993-10 | 08_Biological Risk Assessment.pdf |
| **Irritation** | ISO 10993-10 | 08_Biological Risk Assessment.pdf |
### 9.2 Performance Testing
Non-clinical performance testing demonstrated that the material of the subject device meets the predefined acceptance criteria for the applicable standard tests. The covers are made of a transparent polyethylene material that has been demonstrated to be safe and effective through more than 20 years of use. The material was successfully cleared as part of the approval of another product line from our subsidiary (see K190949, ScanX Barrier Envelopes).
A rationale supporting the relevance and applicability of these test reports to the subject device is provided in the attached “*Performance Testing Assessment*” document. Refer to:
| Topic | Standard ID | Submission Attachment |
| --- | --- | --- |
| **Performance Testing Assessment** | / | 09_Performance Testing Assessment.pdf |
| **Tear Resistance** | ASTM D1004 | 09_Performance Testing Assessment.pdf |
| **Tensile Strength** | ASTM D882 | 09_Performance Testing Assessment.pdf |
| **Synthetic Blood Penetration** | ASTM F1670 | 09_Performance Testing Assessment.pdf |
| **Viral Penetration** | ASTM F1671 | 09_Performance Testing Assessment.pdf |
| **Puncture Resistance** | ASTM F1342 | 09_Performance Testing Assessment.pdf |
5
{9}
DÜRR^{}[] DENTAL
## **10. Shelf-Life**
Not applicable. The device is supplied non-sterile and intended for single-use; therefore, no shelf-life evaluation is required.
## **11. Clinical Data**
Not applicable. No clinical studies are required to support the safety or performance of the device.
## **12. Conclusion**
Based on the comparison and analysis above, the subject device is as safe, as effective, and performs as well as the legally marketed predicate devices.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.