Innovasis AxTiHA® Stand-Alone ALIF System;Innovasis TxTiHA® IBF System;Innovasis HAtetracell®-C Titanium Cervical IBF System

K262025 · INNO Holdings Inc. DBA Innovasis · OVD · Sep 11, 2026 · Orthopedic

Device Facts

Record IDK262025
Device NameInnovasis AxTiHA® Stand-Alone ALIF System;Innovasis TxTiHA® IBF System;Innovasis HAtetracell®-C Titanium Cervical IBF System
ApplicantINNO Holdings Inc. DBA Innovasis
Product CodeOVD · Orthopedic
Decision DateSep 11, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Innovasis AxTiHA Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The AxTiHA Stand-Alone interbody implants with a lordotic angle less than 20°, when used with all three internal fixation screws, do not require use of supplemental fixation. The interior of the AxTiHA implant is intended to be packed with autograft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft. The Innovasis® TxTiHA™ Transforaminal Lumbar IBF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via a transforaminal approach. This device is intended to be used with internal supplemental spinal fixation systems such as the Innovasis Excella® Spinal System. The interior of the TxTiHA implant is intended to be packed with autograft. The Innovasis HAtetracell™-C Titanium Cervical IBF System is indicated for cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. This device is to be used in patients who have had six weeks of non-operative treatment. The HAtetracell™-C device is intended to be used with supplemental fixation, such as the Innovasis Oryx® Cervical Plate System. The HAtetracell™-C device is intended to be used with autogenous and/or allogeneic bone graft comprised of cancellous, cortical, and/or corticocancellous bone and is to be implanted via an anterior approach.

Device Story

Intervertebral body fusion (IBF) implants for lumbar (ALIF/TLIF) and cervical (ACDF) spinal fusion; additive manufactured Ti-6Al-4V (ELI) titanium alloy; feature open-lattice Tetracell® technology for bone graft containment and through-growth; tapered leading edge for insertion; bi-convex profile; coated with 20-nanometer Promimic HA nano Surface® (hydroxyapatite) to enhance osseointegration. AxTiHA lumbar implants include integrated fixation screws and optional anti-backout clips. Used by surgeons in clinical settings to stabilize spinal segments, restore disc height, and facilitate fusion. Implants packed with autograft or allograft bone. Output is mechanical stabilization of the spinal segment. Benefits include reduced motion, pain relief, and promotion of bony fusion.

Clinical Evidence

No clinical data. Bench testing only; in vitro study confirmed nanoscale features and increased hydrophilicity of the HA coating compared to micro-rough HA surfaces.

Technological Characteristics

Materials: Ti-6Al-4V (ELI) per ASTM F3001; Ti-6Al-4V (ELI) per ASTM F136 (screws/clips). Surface: 20nm Promimic HA nano Surface® (hydroxyapatite). Manufacturing: Additive manufacturing. Form factor: Interbody cages with graft cavities and open-lattice structures. Sterilization: Sterile packaged.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous lumbar levels (L2-S1) or one cervical level (C2-T1). DDD defined as discogenic pain with radiographic confirmation. Lumbar patients require 6 months non-operative treatment; cervical patients require 6 weeks. Lumbar patients may have up to Grade 1 spondylolisthesis or retrolisthesis.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 11, 2026 INNO Holdings Inc. DBA Innovasis Michael Thomas Director of Regulatory Affairs 614 East 3900 South Salt Lake City, Utah 84107 Re: K262025 Trade/Device Name: Innovasis AxTiHA® Stand-Alone ALIF System;Innovasis TxTiHA® IBF System;Innovasis HAtetracell®-C Titanium Cervical IBF System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVD, MAX, ODP Dated: June 16, 2026 Received: June 16, 2026 Dear Michael Thomas: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K262025 - Michael Thomas Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K262025 - Michael Thomas Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # BRENT SHOWALTER -S Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} # Indications for Use Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K262025 Page 1 of 1 Please provide the device trade name(s). Innovasis AxTiHA® Stand-Alone ALIF System; Innovasis TxTiHA® IBF System; Innovasis HAtetracell®-C Titanium Cervical IBF System Please provide your Indications for Use below. The Innovasis AxTiHA Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved levels(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The AxTiHA Stand-Alone interbody implants with a lordotic angle less than 20°, when used with all three internal fixation screws, do not require use of supplemental fixation. The interior of the AxTiHA implant is intended to be packed with autograft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft. The Innovasis® TxTiHA™ Transforaminal Lumbar IBF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via a transforaminal approach. This device is intended to be used with internal supplemental spinal fixation systems such as the Innovasis Excella® Spinal System. The interior of the TxTiHA implant is intended to be packed with autograft. The Innovasis HAtetracell™-C Titanium Cervical IBF System is indicated for cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. This device is to be used in patients who have had six weeks of non-operative treatment. The HAtetracell™-C device is intended to be used with supplemental fixation, such as the Innovasis Oryx® Cervical Plate System. The HAtetracell™-C device is intended to be used with autogenous and/or allogeneic bone graft comprised of cancellous, cortical, and/or corticocancellous bone and is to be implanted via an anterior approach. Please select the types of uses (select one or both, as applicable). ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) Please select the age group(s) for which the device(s) is to be used. ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) ? ? ? ? ? {4} INNOVASIS INNOVATE / INVOLVE / INVENT # 510(k) Summary K262025 Page 1 of 4 # Contact Details Applicant Name: INNO Holdings Inc. DBA Innovasis Applicant Address: 614 E 3900 S Salt Lake City, UT 84107 United States Applicant Contact No.: (801) 261-2236 Applicant Contact: Mr. Michael Thomas Applicant Contact Email: mthomas@innovasis.com Date Prepared: September 10, 2026 # Device Name Device Trade Name: Innovasis AxTiHA® Stand-Alone ALIF System Innovasis TxTiHA® IBF System Innovasis HAtetracell®-C Titanium Cervical IBF System Common Name: Intervertebral body fusion device Regulation Number: 888.3080 Classification Name & Innovasis AxTiHA® Stand-Alone ALIF System Product Codes: OVD – Intervertebral Fusion Device with Integrated Fixation, Lumbar Innovasis TxTiHA® IBF System MAX – intervertebral Fusion Device with Bone Graft, Lumbar Innovasis HAtetracell®-C Titanium Cervical IBF System ODP – Intervertebral Fusion Device with Bone Graft, Cervical Legally Marketed Predicate Devices | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | --- | --- | --- | | K201614 | Innovasis AxTiHA® Stand-Alone ALIF System | OVD | | K201614 | Innovasis TxTiHA® IBF System | MAX | | K231899 | Innovasis HAtetracell®-C Titanium Cervical IBF System | ODP | # Device Description Summary The AxTiHA system is for Anterior Lumbar Interbody Fusion (ALIF). The IBF implants are an additive manufactured device comprised of Ti-6Al-4V (ELI) per ASTM F3001 with a Promimic HA nano Surface® (Hydroxyapatite (HA)) coating and are available in multiple size options to facilitate a more precise anatomical fit. The IBF implants have a tapered leading edge which aids in implant insertion due to limited anatomical space, feature a bi-convex profile to match the anatomy, and include anti-migration features to ensure implant stability during the fusion process. The large graft cavity and open geometric Tetracell® Technology structure provide increased volume for autograft loading and bone through-growth. The AxTiHA IBF devices include integrated fixation by way of three converging bone screws and optional screw anti-backout locking clips manufactured from Ti-6Al-4V (ELI) per ASTM F136. Reusable instruments to support the ALIF surgery are provided with the implants in sterilization trays. Page 1 of 4 {5} K262025 Page 2 of 4 INNOVASIS INNOVATE / INVOLVE / INVENT The TxTiHA system is an intervertebral body fusion device with associated instrumentation for use in Transforaminal Lumbar Interbody Fusion (TLIF) surgeries. The implants are an additive manufactured device made from the titanium alloy Ti-6Al-4V (ELI) per ASTM F3001 with a Promimic HA nano Surface® (Hydroxyapatite (HA)) coating and are available in multiple size options to facilitate a more precise anatomical fit. The implants have a tapered leading edge which aids in implant insertion due to limited anatomical space, feature a bi-convex profile to match the anatomy and include anti-migration features to ensure implant stability during the fusion process. The large graft cavity and open geometric Tetracell® Technology structure provide increased volume for autograft loading and bone through-growth. Implants are supplied sterile. Reusable instruments to support the TLIF surgery are provided with the implants in sterilization trays. The HAtetracell-C Titanium Cervical IBF System is an intervertebral body fusion (IBF) device with associated instrumentation, used with bone graft material, intended to stabilize a cervical spinal segment to promote fusion which restricts motion and decreases pain. The HAtetracell-C Titanium Cervical IBF System is implanted via an Anterior Cervical Discectomy and Fusion (ACDF) surgical approach. The HAtetracell-C implants are an additive manufactured device made from titanium alloy Ti-6Al-4V ELI per ASTM F3001 and are coated with a hydroxyapatite Promimic HA nano Surface®. The trabecular structures on the endplate contact surfaces feature porous macro-, micro-, and nanofeatures that are designed to mimic cortical and cancellous bone to help facilitate intervertebral fusion. The HAtetracell-C implant features a tapered nose to aid in insertion and protect the porous layer during insertion, a graft cavity to provide volume for bone graft, and the open lattice structure is designed to reduce the radiographic signature. The implant is available in multiple size options to match vertebral anatomy and is designed to restore height in the cervical spinal column during the fusion process. Implants are supplied sterile. Reusable instruments to support the ACDF surgery are provided with the implants in sterilization trays. AxTiHA interbody cages, TxTiHA interbody cages, and HAtetracell-C interbody cages are coated with a 20-nanometer thin Promimic HA nano Surface® implant surface modification composed of crystalline hydroxyapatite (HA) particles, which have the same shape, composition, and structure as HA found in human bone. This surface treatment increases implant anchoring by facilitating osseointegration and enhancing early bone growth. All subject interbody cages are only available sterile packaged. ### Intended Use/Indications for Use The Innovasis AxTiHA Stand-Alone ALIF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to a Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via an anterior (ALIF) approach. Hyperlordotic implants (those with a lordotic angle greater than or equal to 20°) are indicated for use with a supplemental spinal fixation system such as the Innovasis® Excella® Spinal System. The AxTiHA Stand-Alone interbody implants with a lordotic angle less than 20°, when used with all three internal fixation screws, do not require use of supplemental fixation. The interior of the AxTiHA implant is intended to be packed with autograft or allogenic bone graft composed of cancellous and/or corticocancellous bone graft. Page 2 of 4 {6} K262025 Page 3 of 4 INNOVASIS INNOVATE / INVOLVE / INVENT The Innovasis® TxTiHA™ Transforaminal Lumbar IBF System is an intervertebral body fusion device for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbar spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of non-operative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These implants are used to facilitate fusion in the lumbar spine and are placed via a transforaminal approach. This device is intended to be used with internal supplemental spinal fixation systems such as the Innovasis Excella® Spinal System. The interior of the TxTiHA implant is intended to be packed with autograft. The Innovasis HAtetracell™-C Titanium Cervical IBF System is indicated for cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level from C2-T1. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. This device is to be used in patients who have had six weeks of non-operative treatment. The HAtetracell™-C device is intended to be used with supplemental fixation, such as the Innovasis Oryx® Cervical Plate System. The HAtetracell™-C device is intended to be used with autogenous and/or allogeneic bone graft comprised of cancellous, cortical, and/or corticocancellous bone and is to be implanted via an anterior approach. ### Indications for Use Comparison The subject AxTiHA Stand-Alone ALIF System and TxTiHA IBF System have identical Indications for Use as the predicate (K201614). The subject HAtetracell-C Titanium Cervical IBF System has identical Indications for Use as the predicate (K231899). The subject devices are indicated for the same patient population, condition, and mechanism of action as the predicate devices. ### Technological Comparison The technological design features of the subject implants were compared to the predicates in intended use, indications for use, design, function and technology and it was demonstrated that they are substantially equivalent. ### Non-Clinical and/or Clinical Tests Summary & Conclusions In consideration of the FDA's Guidance for Industry: Considering Whether an FDA-Regulated Product Involves the Application of Nanotechnology, in vitro study results demonstrate that the nanoscale features of the Innovasis AxTiHA Stand-Alone ALIF System, TxTiHA IBF System and HAtetracell-C Titanium Cervical IBF System implants are engineered and uniformly present on the implant surface and exhibit increased hydrophilicity as represented by contact angle when compared to micro-rough HA coated titanium surfaces. The subject devices are identical to the legally marketed predicate in intended use, design, materials, and technological characteristics, including coating properties. No modifications have been made that could affect safety or effectiveness; therefore, no new questions of safety or effectiveness are raised, and additional performance testing is not necessary to support a determination of substantial equivalence. Clinical testing was not applicable. Page 3 of 4 {7} K262025 Page 4 of 4 ![img-0.jpeg](img-0.jpeg) Substantial equivalence is supported by device comparison demonstrating same intended use and technological characteristics as the predicate, with no new questions of safety or effectiveness; therefore, performance is equivalent. Page 4 of 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...