K261950 · Health Line International Corporation · DYB · Jun 22, 2026 · Cardiovascular
Device Facts
Record ID
K261950
Device Name
Health Line Tearaway Introducer
Applicant
Health Line International Corporation
Product Code
DYB · Cardiovascular
Decision Date
Jun 22, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1340
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Health Line Tearaway Introducer is used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters, into the vasculature.
Device Story
Health Line Tearaway Introducer is a sterile, single-use vascular access device consisting of an outer tearaway sheath and an inner dilator. The device is used by clinicians for percutaneous introduction of diagnostic or therapeutic devices (e.g., catheters) into the vasculature using the Seldinger technique. The outer sheath features a winged hub that allows the cannula to be split longitudinally for removal from around an indwelling device. The inner dilator, which includes a female Luer hub for guidewire insertion, features a thermally tapered distal tip for atraumatic vessel entry and a snap-fit locking mechanism to secure it to the sheath hub during insertion. The device facilitates safe vascular access and efficient sheath removal without requiring excessive force at the insertion site.
Clinical Evidence
Bench testing only. Testing included hub break force, sheath peel force, dilator break/insertion force, guidewire compatibility, dimensional analysis, and simulation use. Biocompatibility testing (ISO 10993-1) included cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogen, hemocompatibility, and genotoxicity. Sterilization and shelf-life validated per ISO 11135 and ISO 11607-1.
Indicated for percutaneous introduction of diagnostic or therapeutic devices, such as catheters, into the vasculature in adult patients.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
Tearaway Introducer Sheath (K153533) by Galt Medical Corp.
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 22, 2026
Health Line International Corporation
% Prithul Bom
Most Responsible Person
Regulatory Technology Services, LLC
1000 Westgate Dr. Suite #510k
Saint Paul, Minnesota 55114
Re: K261950
Trade/Device Name: Health Line Tearaway Introducer
Regulation Number: 21 CFR 870.1340
Regulation Name: Catheter introducer
Regulatory Class: Class II
Product Code: DYB
Dated: June 10, 2026
Received: June 10, 2026
Dear Prithul Bom:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261950 - Prithul Bom
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K261950 - Prithul Bom
Page 3
Sincerely,
FINN E.
DONALDSON -S
Digitally signed by
FINN E. DONALDSON -S
Date: 2026.06.22
10:10:42 -04'00'
Finn Donaldson
Acting Assistant Director
DHT2C: Division of Coronary and
Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K261950
Device Name
Health Line Tearaway Introducer
Indications for Use (Describe)
The Health Line Tearaway Introducer is used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters, into the vasculature.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Health Line International Corporation
510(k) Premarket Notification Submission:*Health Line Tearaway Introducer*
K261950
Page 1 of 4
# **510(k) SUMMARY**
(21 CFR 807.92)
| **SUBMITTER** | Name: | Health Line International Corporation |
| --- | --- | --- |
| | Address: | 260 North Ace Yeager Court, Unit D Salt Lake City, Utah 84116 USA |
| | FDA Registration: | 3010882065 |
| | Contact Name: | Aaron G. Faulkner VP QA and RA |
| | Telephone: | 801-773-7798 Ext. 109 |
| | Email: | agfaulkner@hlic.net |
| | Date Prepared: | November 26, 2025 |
| **SUBJECT DEVICE** | Name: | ***Health Line Tearaway Introducer*** |
| | Regulation Name: | Catheter Introducer |
| | Classification Name: | Introducer, Catheter |
| | Classification Panel: | Cardiovascular |
| | Regulatory Class: | Class II |
| | Product Code: | DYB |
| | Regulation Number: | 21 CFR 870.1340 |
| **PREDICATE DEVICE** | Name: | ***Tearaway Introducer Sheath (K153533) by Galt Medical Corp.*** |
| | Regulation Name: | Catheter Introducer |
| | Classification Name: | Introducer, Catheter |
| | Classification Panel: | Cardiovascular |
| | Regulatory Class: | Class II |
| | Product Code: | DYB |
| | Regulation Number: | 21 CFR 870.1340 |
| **DEVICE DESCRIPTION** | The *Health Line Tearaway Introducer* is a sterile, single-use vascular access device composed of two main components: an outer tearaway sheath and an inner dilator. | |
| | The outer tearaway sheath is manufactured as a one-piece molded hub with integral wings, attached to a cylindrical peelable cannula. The design allows the sheath to be split longitudinally along its entire length when minimal separating force is applied at the winged hub. This controlled tear mechanism enables safe and efficient removal of the sheath from around an indwelling catheter or other intravascular devices without exerting excessive force on the insertion site. | |
| | The inner dilator consists of a rigid cylindrical cannula integrally molded with a proximal female Luer hub to accommodate guidewire insertion. The distal end of the dilator is thermally tapered to facilitate atraumatic vessel entry. The dilator is designed to be longer than the tearaway sheath to provide optimal guidance during insertion. A snap-fit locking mechanism secures the dilator hub to the tearaway sheath hub, preventing relative movement during vascular access. This locking mechanism may be disengaged by applying a controlled rotational motion, permitting separation of the components when required. | |
| | The *Health Line Tearaway Introducer* is provided in different sizes which is indicated in the front face of the dilator hub. | |
{5}
Health Line International Corporation
510(k) Premarket Notification Submission:*Health Line Tearaway Introducer*
K261950^{}[] Page 2 of 4
# **INTENDED USE**
The *Health Line Tearaway Introducer* is used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters, into the vasculature.
# **INDICATIONS FOR USE**
The *Health Line Tearaway Introducer* is used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters, into the vasculature.
**Subject Device:** *Health Line Tearaway Introducer*
**Predicate Device:** *Tearaway Introducer Sheath (K153533)*
# **TECHNOLOGICAL CHARACTERISTICS - COMPARISON**
| ATTRIBUTES | SUBJECT DEVICE | PREDICATE DEVICE |
| --- | --- | --- |
| Intended use | The *Health Line Tearaway Introducer* is used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters, into the vasculature. | These introducers are used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters and pacing leads, into the vasculature. |
| Indications for Use | The *Health Line Tearaway Introducer* is used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters, into the vasculature. | These introducers are used for the percutaneous introduction of diagnostic or therapeutic devices, such as catheters and pacing leads, into the vasculature. |
| Intended duration | Transient (≤ 24 Hours) | Transient (≤ 24 Hours) |
| Prescription device | Yes | Yes |
| Insertion sites | Percutaneous | Percutaneous |
| Insertion technique | Seldinger (Over the guidewire) | Seldinger (Over the guidewire) |
| Intended population | Adults | Adults |
| Size Configurations | 3.5Fr - 6.5Fr | 3Fr - 16Fr |
| Length Configurations | 5, 7, 10cm | 5 - 30cm |
| Design | The Health Line Tearaway Introducer Sheath assembly consists of an outer peelable sheath and a dilator. The tearaway sheath has a winged hub to initiate the tear in the sheath. | The Galt Tearaway Introducer Sheath assembly consists of an outer peelable sheath and a dilator. The tearaway sheath has a winged hub to initiate the tear in the sheath. |
| Assembly | Tearaway Sheath/Dilator | Tearaway Sheath/Dilator |
| Dilator lock | Yes | Yes |
| Insertion markings | None | None |
| Dilator tip design | Symmetrical soft tapered tip | Symmetrical soft tapered tip |
| Shape design | Sheath hub: Winged Sheath cannula: Cylindrical Dilator hub: Round Dilator cannula: Cylindrical | Sheath hub: Winged Sheath cannula: Cylindrical Dilator hub: Round Dilator cannula: Cylindrical |
| Color | Sheath: Color coded winged hub, gray cannula Dilator: White hub, white cannula | Sheath: Color coded winged hub, gray cannula Dilator: White hub, green cannula |
| Winged hub material | Polypropylene | Polypropylene |
| Sheath material | PTFE | PTFE |
| Dilator hub material | Polyurethane | PEBAX |
| Dilator cannula material | PEBAX | PEBAX |
{6}
Health Line International Corporation
510(k) Premarket Notification Submission: Health Line Tearaway Introducer
K261950
Page 3 of 4
| Size identification | Marked in dilator hub | Marked in dilator hub |
| --- | --- | --- |
| Sterilization method | Ethylene Oxide | Ethylene Oxide |
| Method of use | Single use, disposable | Single use, disposable |
| Shelf life | 3 years | 4 years |
| Sterile barrier material | Tyvek | Tyvek |
| Conical fittings | ISO 80369-7 Compatible | ISO 80369-7 Compatible |
| Non-Pyrogenic | Yes | Yes |
| Made with latex rubber | No | No |
| Made with DEHP | No | No |
The subject device has similar technological characteristics as compared to the predicate device. Differences, if any, are not critical to the intended use of the subject device (as discussed in the section corresponding to Substantial Equivalence Comparison) and do not raise new questions regarding safety and effectiveness.
The Health Line Tearaway Introducer followed verification and validation activities in accordance with Design Controls as per 21 CFR Section 820.30.
The following bench testing was conducted in the subject device to demonstrate substantial equivalence to the predicate device and verify conformity to international standards, FDA-recognized consensus standards and in-house requirements:
1. Hub Break Force
2. Sheath Peel Force
3. Dilator Break Force
4. Dilator Insertion Force
5. Guidewire Compatibility
6. Dimensional
7. Simulation Use
**SAFETY AND PERFORMANCE TESTING**
Functional testing was conducted in accordance with ISO 11135 and ISO 11607-1 to evaluate sterilization and shelf life of the finished subject device.
The following biocompatibility testing was conducted in accordance with ISO 10993-1 and based on a biological evaluation plan (BEP) and report (BER) to evaluate the subject device on:
1. Cytotoxicity
2. Sensitization
3. Irritation or Intracutaneous Reactivity
4. Acute Systemic Toxicity
5. Material Mediated Pyrogen
6. Hemocompatibility
7. Genotoxicity
Results of the biocompatibility testing support the determination of substantial equivalence.
**SUMMARY OF SUBSTANTIAL EQUIVALENCE**
Functional and biological comparison confirms that the technological characteristics between the subject and predicate devices are substantial equivalent, including:
- Principle of Operation
- Design
- Key Features
{7}
Health Line International Corporation
510(k) Premarket Notification Submission: Health Line Tearaway Introducer
K261950
Page 4 of 4
- Material Composition
- Sterilization
- Biocompatibility
- Performance Testing
CONCLUSION
In accordance with FDA 21 CFR Section 807.92, and based on the indications for use, technological characteristics, biocompatibility and safety and performance testing, the subject Health Line Tearaway Introducer met the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation, and indications for use to the currently marketed Tearaway Introducer Sheath (K153533) predicate device without raising any new questions regarding safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.